Search Results
Found 2 results
510(k) Data Aggregation
(113 days)
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support.
Here's a breakdown of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The document primarily focuses on analytical and clinical performance without explicitly stating numerical acceptance criteria prior to presenting the results. However, we can infer implied acceptance criteria from the reported performance, especially for linearity and specificity (inhibition). For clinical sensitivity and specificity, the provided confidence intervals suggest a critical range.
Performance Metric | Acceptance Criteria (Implied / Inferred) | Reported Device Performance |
---|---|---|
Precision | Within-run and total precision (CV%) generally low for various allergen concentrations. (No explicit numerical target provided.) | Within-Run CV: 3.28% - 5.42% |
Total CV: 4.91% - 6.94% | ||
Linearity (Slope) | Slope of regression (observed vs. expected) close to 1.00. (No explicit numerical range provided, but 95% CI is given.) | Slopes: 0.997 - 1.011 (All 95% CIs include 1.00) |
Linearity (Intercept) | Intercept of regression (observed vs. expected) close to 0.00. (No explicit numerical range provided, but 95% CI is given.) | Intercepts: -0.322 - 0.425 (All 95% CIs include 0.00) |
Specificity (Inhibition) | Target % inhibition of 50% met by relevant inhibitor extract in a concentration-dependent fashion. | All allergens met the 50% inhibition target, showing concentration-dependent inhibition. Lowest reported inhibition at 0.008 mg/mL was 48.28% (Walnut) and 48.95% (Egg). |
Clinical Sensitivity | No explicit numerical target provided, but 95% confidence interval presented. | 54.4% (95% CI: 48% - 60%) |
Clinical Specificity | No explicit numerical target provided, but 95% confidence interval presented. | 96.9% (95% CI: 96% - 98%) |
Overall Agreement | No explicit numerical target provided, but 95% confidence interval presented. | 87.3% (95% CI: 85% - 89%) |
2. Sample Sizes Used for the Test Set and Data Provenance
- Precision: For each of the seven allergens, three positive samples and one negative sample were tested. Each sample was assayed in two aliquots per day for 20 different days, totaling 40 measurements per sample.
- Linearity: For each allergen, two samples were diluted into 5 levels (undiluted + 4 dilutions). The table shows N=12 for each allergen, implying 2 samples * 5 dilutions + 2 undiluted samples.
- Specificity (Inhibition): A single serum sample or pool of sera was used for each allergen.
- Clinical Performance:
- Total Patients: 1,157 individuals.
- Clinical Data: 261 positive, 896 normal.
- Data Provenance: The document does not explicitly state the country of origin. It samples were described as "samples from non-atopic individuals and samples from atopic patients with case histories of suspected clinical reactions." This indicates that the data is retrospective clinical data for the "clinical data" ground truth, likely collected from patients presenting with symptoms.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- The document does not specify the number of experts or their qualifications.
- For clinical performance, the ground truth is referred to as "clinical data" which includes "case histories of suspected clinical reactions to the specific allergen or allergy group" and "documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." This suggests that the clinical diagnosis was established by medical professionals, but not explicitly stated as a panel of "experts" for the study itself.
4. Adjudication Method for the Test Set
- The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The "clinical data" appears to be the established reference.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This study evaluates an in vitro diagnostic (IVD) assay (IMMULITE® 2000 3gAllergy™ Specific IgE Assay), which is a laboratory test, not an AI system that assists human readers. Therefore, the concept of "human readers improve with AI vs. without AI assistance" does not apply.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, the performance presented for the IMMULITE® 2000 3gAllergy™ Specific IgE Assay is standalone (algorithm only). This device is an automated in vitro diagnostic test that quantitatively measures allergen-specific IgE. Its performance is evaluated independently of human interpretation of the assay results, though human clinicians will interpret a patient's overall clinical picture, including the assay results.
7. The Type of Ground Truth Used
- Precision and Linearity: Internal quality control materials or spiked samples with known concentrations.
- Specificity (Inhibition): Known relevant inhibitor extracts and negative control samples.
- Clinical Performance: "Clinical data" described as "case histories of suspected clinical reactions to the specific allergen or allergy group" and "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity." For the egg allergen, comparison studies were done against previously cleared individual egg constituents (egg white and egg yolk allergens). This is best characterized as clinical documentation/diagnosis.
8. The Sample Size for the Training Set
- The document does not mention a training set. This is not applicable as the device is a chemiluminescent immunoassay, not a machine learning algorithm that requires a training set. The "assay" itself is the device, and its development involves chemical and biological optimization, not algorithm training.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(121 days)
For in vitro diagnostic use with the IMMULITE® 2000 Analyzer - for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.
IMMULITE® 2000 3gAllergy™ Specific IgE is a solid-phase, two-step, chemiluminescent immunoassay that exploits liquid phase kinetics in a bead format. It represents a significant advance over conventional methods relying on allergens attached to a solid-phase support, such as a paper disk. The allergens are covalently bound to a soluble polymer matrix, which in turn is labeled with a ligand. The use of an amino acid co-polymer amplifies the amount of allergen that the matrix can support. Incubation Cycles: 2 × 30 minutes.
1. Table of Acceptance Criteria and Reported Device Performance:
The document describes several performance criteria for the IMMULITE® 2000 3gAllergy™ Specific IgE Assay, including precision, linearity, and specificity. Clinical performance is also assessed.
Performance Characteristic | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Precision | Undefined explicit acceptance criteria in the provided text. Implied acceptable %CV values. | Within-Run %CV: Ranged from 3.01% to 5.24% for positive samples across all allergens tested. Total %CV: Ranged from 4.27% to 7.03% for positive samples across all allergens tested. |
Linearity | Slopes close to 1.00 and intercepts close to 0 for regression equation. 95% CI for slope should ideally include 1.00, and for intercept should ideally include 0. | Red Maple: Slope = 0.999 (95% CI: 0.961-1.037), Intercept = 0.059 (95% CI: -0.124-0.242) White Hickory: Slope = 1.004 (95% CI: 0.981-1.028), Intercept = 0.062 (95% CI: -0.122-0.246) Red Cedar: Slope = 1.003 (95% CI: 0.987-1.019), Intercept = -0.007 (95% CI: -0.081-0.067) Sweet Gum: Slope = 0.997 (95% CI: 0.965-1.029), Intercept = 0.306 (95% CI: 0.051-0.561) Common Sagebrush: Slope = 0.999 (95% CI: 0.970-1.028), Intercept = 0.615 (95% CI: -0.400-1.270) Wingscale: Slope = 0.998 (95% CI: 0.981-1.015), Intercept = -0.098 (95% CI: -0.252-0.056) All slopes are very close to 1.00, and intercepts are close to 0, with 95% CIs generally encompassing these ideal values. |
Specificity (Inhibition) | Target % inhibition of 50% met by relevant inhibitor extract in a concentration-dependent fashion. | The inhibition study demonstrated that the allergens tested are inhibited by the relevant inhibitor extract in a concentration dependent fashion. The target % inhibition of 50% was met for all six allergens. Specific inhibition percentages at different concentrations are provided in tables for each allergen, showing gradual increases in inhibition with increasing inhibitor concentration. For example, for Red Maple, inhibition reaches 96.83% at 0.2 mg/mL and 100% at 5 mg/mL. |
Clinical Performance | Results compare well with clinical documentation of presence or absence of signs, symptoms, and other diagnostic evidence of allergen sensitivity. | Specificity: 95.4% (Based on 775 Negative results out of 812 Clinical Normal cases) Agreement: 88.3% (Based on 196 Positive + 775 Negative results out of 1100 Total cases) |
2. Sample size used for the test set and the data provenance:
- Precision Studies: Three positive samples and one negative sample were tested for each allergen lot. The testing involved two aliquots of each test sample in two runs per day on 20 different days, following CLSI document EP5-A2. The exact number of individual measurements can be calculated from these parameters (3 positive samples * 2 aliquots/sample * 2 runs/day * 20 days = 240 measurements for positive samples per lot, plus negative control measurements).
- Linearity Studies: For each allergen, two samples were diluted into 5 levels. These 12 samples (2 original + 10 diluted) were tested.
- Specificity (Inhibition) Studies: A single serum sample or pool of sera was used for each allergen. For each inhibition experiment, 70uL of undiluted and 4 levels of 5-fold serially diluted inhibitor extract were mixed with 250uL of sample or pool.
- Clinical Studies: A total of 1,100 samples were used. This included 288 samples from atopic patients with case histories of suspected clinical reactions to the specific allergen or allergy group, and 812 samples from non-atopic individuals.
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the clinical studies describe "samples from non-atopic individuals" and "samples from atopic patients with case histories of suspected clinical reactions to the specific allergen or allergy group," which implies a prospective or at least carefully selected retrospective collection for diagnostic relevance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
The document does not mention the use of experts to establish a "ground truth" for the test set in the traditional sense of image interpretation or complex diagnostic decision-making.
- For the Precision, Linearity, and Specificity (Inhibition) studies, the "ground truth" is based on the inherent biochemical properties of the assay and the measured values, not expert interpretation.
- For the Clinical Studies, "clinical data" was used as the reference. This "clinical data" is described as "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity" and "case histories of suspected clinical reactions." This implies that the ground truth was established by medical professionals (e.g., allergists, physicians) based on patient history, symptoms, and other diagnostic findings; however, the specific number and qualifications of these experts are not provided in this document.
4. Adjudication method for the test set:
Not applicable. The types of studies described (analytical performance and comparison to clinical documentation) do not involve adjudication methods like those used for expert consensus in medical image analysis.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI versus without AI assistance:
Not applicable. This document describes an in vitro diagnostic assay, not an AI-assisted diagnostic tool that would involve human readers.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Yes, the studies presented are all standalone performance evaluations of the IMMULITE® 2000 3gAllergy™ Specific IgE Assay itself (the "device" being the assay system), without human-in-the-loop interaction for interpretation beyond standard laboratory procedures. The device provides a quantitative measurement of allergen-specific IgE.
7. The type of ground truth used:
- Precision, Linearity, Specificity: Ground truth is based on physical and chemical measurement principles and established scientific methods for evaluating assay performance (e.g., known concentrations for linearity, defined inhibition protocols for specificity).
- Clinical Studies: The ground truth used was "clinical data," consisting of "clinical documentation of presence or absence of signs, symptoms and other diagnostic evidence of allergen sensitivity" and "case histories of suspected clinical reactions" to specific allergens. This can be categorized as outcomes data and expert clinical assessment.
8. The sample size for the training set:
The document describes performance evaluation studies and does not refer to a "training set" in the context of machine learning or AI development. The studies are for validating the assay's performance against established clinical and analytical benchmarks.
9. How the ground truth for the training set was established:
Not applicable, as there is no mention of a "training set" in the provided document. The studies focus on analytical validation and clinical concordance of the assay itself.
Ask a specific question about this device
Page 1 of 1