K Number
K091920

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-09-28

(90 days)

Product Code
Regulation Number
872.3630
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement and screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the following manufacturers' implant systems: Nobel Replace® Straight Groovy 3.5mm, Nobel Replace® Select Tapered and Straight 3.5mm and NobelSpeedyTM 3.5mm implants.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Device Description

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. Zirconia may have a variation in shade. The subject abutments are indicated for cemented restorations. The Atlantis™ Abutments in Zirconia for Nobel Replace Implants and abutment screws are made from biocompatible yttria-stabilized tetragonal zirconia polycrystals (Y-TZP) and meets ISO Standards 6872 & 13356 and may be produced in any composition complying with the values specified by these standards.. The abutment screws are made from titanium grade Ti-6A1-V ELI and meets ASTM Standard F-136. The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with Nobel Replace® Straight Groovy 3.5mm. Nobel Replace® Select Tapered and Straight 3.5mm, and NobelSpeedy™ 3.5mm implants.

AI/ML Overview

The provided document is a 510(k) summary for a dental abutment, not a study report for an AI/ML device. Therefore, it does not contain information on acceptance criteria, reported device performance, study types (e.g., MRMC, standalone), ground truth establishment, or sample sizes related to AI/ML device development and evaluation.

The document discusses the substantial equivalence of the "Atlantis™ Abutment in Zirconia for Nobel Replace Implant" to predicate devices based on intended use, material, design, and performance, typical for a traditional medical device submission.

As a result, I cannot populate the requested table and answer the study-related questions from the provided text. The questions are designed for AI/ML device evaluation studies, which are not detailed in this 510(k) submission.

{0}------------------------------------------------

Ko91920

Pre-market Notification Page - 000009 - - - - - - - - - - - - - - -

-1

VII. SECTION 10 - 510(K) SUMMARY

This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR Part 807.92.

1. Applicant's Name and Address

Astra Tech Inc. 590 Lincoln Street Waltham, Massachusetts 02451 Telephone Number: Fax Number: Contact Person:

781-810-6462 781-810-6719 Franklin Uyleman Manager of Quality and Regulatory Affairs

2. Name of Device

Atlantis™ Abutment in Zirconia for Nobel Replace Trade Name: Implant ´ Common Name: Endosseous dental implant abutment Endosseous dental implant abutment Classification Name: 21 CFR 872.3630 Product code NHA

Legally Marketed Device to which Equivalence is claimed (Predicate Device) 3.

ManufacturerDevice510(k)Number
Astra Tech Inc.(formerly AtlantisComponents Inc.)Atlantis™ Abutment in Zirconia forNobel Replace ImplantK062277
Nobel BiocareReplace TiUniteK023113
Astra Tech Inc.(formerly AtlantisComponents Inc.)Atlantis Abutment in ZirconiaK052070

{1}------------------------------------------------

Pre-market Page - 000010 - - - - - - - - - - - - - - - - -

4. Description of the Device

The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. Zirconia may have a variation in shade. The subject abutments are indicated for cemented restorations. The Atlantis™ Abutments in Zirconia for Nobel Replace Implants and abutment screws are made from biocompatible yttria-stabilized tetragonal zirconia polycrystals (Y-TZP) and meets ISO Standards 6872 & 13356 and may be produced in any composition complying with the values specified by these standards.. The abutment screws are made from titanium grade Ti-6A1-V ELI and meets ASTM Standard F-136. The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with Nobel Replace® Straight Groovy 3.5mm. Nobel Replace® Select Tapered and Straight 3.5mm, and NobelSpeedy™ 3.5mm implants.

5. Intended Use of the Device

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely endentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement or screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the following manufacturers' implant systems: Nobel Replace® Straight Groovy 3.5mm, Nobel Replace® Select Tapered and Straight 3.5mm and NobelSpeedy TM 3.5mm implants.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e.30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Basis for Substantial Equivalence 6.

The Atlantis™ Abutments in Zirconia for Nobel Replace Implants are substantially equivalent in intended use, material, design and performance to the Atlantis Abutments in Zirconia cleared under K052070. Atlantis Abutments in Zirconia for Nobel Replace cleared under K062277 and Nobel Replace cleared under K023113.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three bars above it, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle symbol. The image is in black and white and appears to be a scan or photocopy.

SEP 2 8 2009

Food and Drug Administration 10903 New Hampshire Avenne Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Astra Tech, Incorporated C/O Ms. Betsy A. Brown Consültant B.A. Brown & Associates 8944 Tamaroa Terrace Skokie, Illinois 60076

Re: K091920

Trade/Device Name: Atlantis™ Abutment in Zirconia for Nobel Replace Implant Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 15, 2009 Received: June 30, 2009

Dear Ms. Brown:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Brown

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies, You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Pre-market Notification Page - 000227

Indications for Use

510(k) Number (if Known) Ko91920

Device Name: Atlantis ™ Abutment in Zirconia for Nobel Replace Implant

Indication for Use:

The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement and screw retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.

This device is compatible with the following manufacturers' implant systems: Nobel Replace® Straight Groovy 3.5mm, Nobel Replace® Select Tapered and Straight 3.5mm and NobelSpeedyTM 3.5mm implants.

Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.

Highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.

Prescription Use x (Part 21 CFR 801 SubpartD)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kein Haley fa WSR

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K091920

§ 872.3630 Endosseous dental implant abutment.

(a)
Identification. An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)