K Number
K091778

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-08-31

(75 days)

Product Code
Regulation Number
870.3450
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

"bypass or reconstruction of peripheral arterial blood vessels".
The Vascutek Cobrahood™ vascular grafts are still intended for prescription use only.

Device Description

Vascuteks ePTFE vascular grafts were originally cleared by FDA in 1999 (K992832: unsealed and K993667: gelatin sealed). In 2004, Vascutek received clearance for the application of an external ePTFE wrap to the original FDA cleared ePTFE vascular graft range (K041528). As a result of development in line with design control procedures, it became possible to flare the end of a standard wrapped ePTFE graft to create the Cobrahood™ design as shown in Figures 1a and 1b of this submission. The intended use is, "bypass or reconstruction of peripheral arterial blood vessels".

All Vascutek ePTFE grafts are manufactured from virgin expanded polytetrafluoroethylene resin, including the ePTFE external wrap. Cobrahood™ ePTFE vascular grafts are not sealed, i.e. do not have an external gelatin coating. The grafts are available with or without an external support constructed of PTFE. This is identical to the support present on the predicate devices, cleared by FDA in 2005 (K041528).

All Vascutek ePTFE vascular grafts are supplied sterile. The method of sterilization is Ethylene Oxide. A shelf-life of five years has been established. All related parameters remain unchanged to the predicate grafts.

The Cobrahood™ ePTFE vascular graft range is detailed fully in Table 2 of this submission. Manufacturing drawings are provided in Attachment I.

AI/ML Overview

The document provided is a 510(k) summary for the Vascutek Cobrahood™ ePTFE Vascular Grafts and the FDA's clearance letter. This type of regulatory submission focuses on demonstrating substantial equivalence to a predicate device, rather than explicit acceptance criteria and performance against those criteria in a typical clinical study.

Therefore, the information regarding acceptance criteria and a study proving the device meets those criteria will be interpreted in the context of a 510(k) submission, specifically focusing on bench testing for substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Bench Testing for Substantial Equivalence)Reported Device Performance
Mechanical/Physical Properties: Tests to ensure Cobrahood™ grafts meet established standards for ePTFE vascular grafts and are comparable to predicate devices. (Specific tests or quantitative criteria are not detailed in the summary, but implied by "physical testing").All samples tested passed the relevant acceptance criteria. Results demonstrated that the Cobrahood™ ePTFE vascular grafts are substantially equivalent to the predicate devices (Vascutek MAXIFLO™ Wrap Grafts and IMPRA Inc. Distaflo Bypass Grafts).
Material Composition: Use of virgin expanded polytetrafluoroethylene resin for both graft and external wrap.The device uses the same materials of construction (virgin expanded polytetrafluoroethylene resin, including the ePTFE external wrap) as the predicate devices.
Sterilization Method: Ethylene Oxide.The device is supplied sterile via Ethylene Oxide. Method unchanged from predicate.
Shelf-Life: 5 years.A shelf-life of five years has been established, and all related parameters remain unchanged to the predicate grafts.
Intended Use: Bypass or reconstruction of peripheral arterial blood vessels.The intended use is "bypass or reconstruction of peripheral arterial blood vessels," identical to the predicate.
Technological Characteristics: No changes to materials, sterilization, packaging, or shelf life compared to predicates.The device has the same technological characteristics as the predicates; no changes to materials of construction, sterilization, packaging, or shelf life.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The document mentions "All samples tested" for bench testing, but does not provide the number of Cobrahood™ grafts or predicate grafts used.
  • Data Provenance: The bench testing was conducted by Vascutek Ltd. in Scotland. It is a retrospective comparison against previously cleared predicate devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable in the context of this 510(k) submission. The "ground truth" for demonstrating substantial equivalence against predicate devices is established through engineering and performance specifications, and comparison of physical characteristics and bench test results, rather than expert consensus on clinical outcomes or images. The "ground truth" is effectively the established performance and characteristics of the predicate devices.

4. Adjudication Method for the Test Set

Not applicable. This is not a study involving human interpretation or clinical adjudication. Substantial equivalence is determined by comparing design, materials, and performance data from bench tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a vascular graft, not an AI-powered diagnostic or interpretive tool. No MRMC study was conducted.

6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

Not applicable. This device is a physical medical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the bench testing was primarily the engineering specifications and established performance characteristics of the predicate devices, against which the Cobrahood™ graft's physical and performance data were compared. The objective was to show that the new design (flared outflow) did not alter the safety and effectiveness, and that the performance remained unchanged, thereby achieving substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI model requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the number 5091778 at the top. Below that is the word "VASCUTEK" in a stylized font, with a logo to the left of the word. Underneath that is the word "TERUMO".

VASCUTEK L a TERUMO Company Newmains Avenue Renfrewshire Scotland

Tel: (+44) (0)141 812-5555 Fax: (+44) (0)141 812 7650 www.vascutek.com

510(k) SUMMARY

VASCUTEK COBRAHOOD™ ePTFE VASCULAR GRAFTS

Date prepared: 12th of June 2009

Prepared in accordance with CFR21 Part 807, Sec. 807.92; Content and format of a 510(k) summary.

Page 1 of 3

Common/Usual Name : Prosthesis, vascular graft

Vascutek Cobrahood™ ePTFE vascular grafts Proprietary Name(s) :

Classification Name : Prosthesis, vascular graft, of 6mm and greater diameter

The Food and Drug Administration has classified these devices Classification : as Class II devices under classification 21 CFR 870.3450. Classification, DSY.

Applicant and : 510(k) submitter/ contact -

Karen Kelso Regulatory Affairs Manager Vascutek Ltd. Newmains Avenue Inchinnan Industrial Estate Renfrewshire, Scotland PA4 9RR, United Kingdom

+44 (141) 814 5527 Phone: +44 (141) 814 3431 Fax:

AUG 8 1 2009

{1}------------------------------------------------

Device Predicates:

Predicate Device NamesFDA Clearance DateOriginal 510(k)
Vascutek MAXIFLOTM Wrap Grafts
ePTFE wrapped ePTFE vascular grafts *14th September 2004K041528
NOTE: *PTFE external support cleared under K043552 on 5th January 2005 forwrapped grafts (K041528)
IMPRA Inc. Distaflo Bypass Grafts
ePTFE wrapped ePTFE vascular graftswith flared distal end24th November 1998K983861

Device Description:

Vascuteks ePTFE vascular grafts were originally cleared by FDA in 1999 (K992832: unsealed and K993667: gelatin sealed). In 2004, Vascutek received clearance for the application of an external ePTFE wrap to the original FDA cleared ePTFE vascular graft range (K041528). As a result of development in line with design control procedures, it became possible to flare the end of a standard wrapped ePTFE graft to create the Cobrahood™ design as shown in Figures 1a and 1b of this submission. The intended use is, "bypass or reconstruction of peripheral arterial blood vessels".

All Vascutek ePTFE grafts are manufactured from virgin expanded polytetrafluoroethylene resin, including the ePTFE external wrap. Cobrahood™ ePTFE vascular grafts are not sealed, i.e. do not have an external gelatin coating. The grafts are available with or without an external support constructed of PTFE. This is identical to the support present on the predicate devices, cleared by FDA in 2005 (K041528).

All Vascutek ePTFE vascular grafts are supplied sterile. The method of sterilization is Ethylene Oxide. A shelf-life of five years has been established. All related parameters remain unchanged to the predicate grafts.

The Cobrahood™ ePTFE vascular graft range is detailed fully in Table 2 of this submission. Manufacturing drawings are provided in Attachment I.

Intended Use:

The subject of this 510(k) is to request FDA clearance for the Cobrahood™ ePTFE vascular grafts, which have the following indication for use;

"bypass or reconstruction of peripheral arterial blood vessels".

The Vascutek Cobrahood 10 vascular grafts are still intended for prescription use only.

{2}------------------------------------------------

The difference in design of the Vascutek Cobrahood™ vascular grafts, i.e. the presence of the flared outflow is not critical to the intended therapeutic, diagnostic, prosthelic or surgical use of the devices. The difference also does not affect the safety and effectiveness of the device when used as labelled.

Principles of Operation and Technology:

The performance of the Vascutek Cobrahood™ ePTFE vascular grafts remains unchanged. The grafts are still used in surgery in the same manner as standard ePTFE vascular grafts.

Vascutek Cobrahood MePTFE vascular grafts are intended only for use in the hospital operating room by suitably qualified surgeons.

Vascutek Cobrahood M ePTFE vascular grafts are provided sterile by Ethylene Oxide for single use only.

The devices have the same technological characteristics as the predicates i.e. there are no changes to the materials of construction, sterilization, packaging or shelf life.

Performance: The performance of the Vascutek Cobrahood ™ ePTFE vascular grafts remains unchanged. The grafts are still used in surgery in the same manner as standard ePTFE vascular grafts.

Bench Testing: For verification purposes, physical testing was undertaken on Cobrahood™ grafts and both predicate grafts, MAXIFLO™ Wrap and Distatio. All samples tested passed the relevant acceptance criteria. Results have demonstrated that the Cobrahood 10 ePTFE vascular grafts are substantially equivalent to the predicate devices.

Clinical Information; No clinical testing has been undertaken on the Cobrahood™ ePTFE vascular grafts, which have been shown to be substantially equivalent to the predicate devices. However, a review of the literature in relation to clinical use of the predicate Distaflo grafts is provided, which has been available commercially in the USA since 1998. It is also demonstrated that clinical use of comparable techniques is an established concept.

Conclusion: In conclusion, the Vascutek Cobrahood™ ePTFE vascular grafts are substantially equivalent to the predicate devices in commercial distribution and results of non-clinical tests and supporting information from the literature demonstrates that the devices continue to be safe, effective and perform as well as the predicate devices.

Substantial Equivalence:

The Vascutek Cobrahood™ ePTFE vascular grafts are substantially equivalent in overall intended use, design and materials, performance, principles of operation and technology to the predicate devices. This is shown comprehensively in Table 2, Section 10 of this pre-market notification.

There are no other design changes to the Vascutek Cobrahood™ ePTFE vascular grafts other than the presence of the flared graft portion.

Harsh Kothari

Signature

.............................................................................................................................................................................. Date

्र के

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's logo. The logo is a stylized depiction of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

AUG 3 1 2009

Vascutek Ltd. c/o Dr. Karen Kelso Regulatory Affairs Manager Newmains Avenue Inchinnan Renfrewshire, PA4 9PR Scotland

Re: K091778

Vascutek Cobrahood™ ePTFE Vascular Grafts Regulation Number: 21 CFR 870.3450 Regulation Name: Vascular Graft Prosthesis Regulatory Class: Class II Product Code: DSY Dated: June 12, 2009 Received: June 17, 2009

Dear Dr. Kelso:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Dr. Karen Kelso

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Dana R. Wachner

Image /page/4/Picture/8 description: The image shows a close-up of a handwritten signature or symbol. The lines are dark and bold, creating a flowing, abstract shape. The overall impression is that of a stylized mark, possibly representing initials or a unique identifier.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

K091778

Device Name:

Vascutek Cobrahood™ cPTFE vascular grafts

Indications For Use:

Bypass or reconstruction of peripheral arterial blood vessels

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

una R. lehner

Page \ of I

(Division Sign-Off) (Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number Kog) 7 4 g

§ 870.3450 Vascular graft prosthesis.

(a)
Identification. A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”