K Number
K992832

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-11-19

(88 days)

Product Code
Regulation Number
870.3450
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sulzer Vascutek ePTFE grafts are intended for the creation of subcutaneous arteriovenous conduits for blood access, bypass, or reconstruction of occluded or diseased arterial blood vessels.

Device Description

The device is composed of polytetrafluoroethylene, which has been fabricated in tubular form and expanded to impart porosity to the structure.

AI/ML Overview

The provided text is a 510(k) summary for the Sulzer Vascutek ePTFE Vascular Prosthesis, asserting its substantial equivalence to an existing device rather than presenting a study to prove its inherent performance against specific acceptance criteria. Therefore, most of the requested information regarding acceptance criteria, study design, and ground truth cannot be extracted directly from this document.

However, I can extract the information that is present and indicate what is not available.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Substantial equivalence to Impra commercial grafts.In vitro testing: Sulzer Vascutek ePTFE graft line shown to be substantially equivalent to Impra commercial grafts.
Substantial equivalence to Impra standard wall grafts (for standard wall grafts).Animal studies: Sulzer Vascutek standard wall grafts shown to be equivalent to Impra standard wall grafts.
Biocompatibility and non-toxicity.Biomaterial testing: Sulzer Vascutek ePTFE grafts demonstrated to be biocompatible and non-toxic.
Mechanical properties (strength, suture retention, handling) substantially equivalent to currently marketed products.Mechanical properties: Stated as substantially equivalent to products currently in commercial distribution. (No specific values provided).

2. Sample size used for the test set and the data provenance

  • Sample size for test set: Not explicitly stated. The document refers to "in vitro testing," "animal studies," and "biomaterial testing" without specifying the number of grafts or animals used in each.
  • Data provenance: Not explicitly stated. The testing was conducted by Sulzer Vascutek Inc. or on their behalf, but the location or whether the data was retrospective or prospective is not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. The document describes engineering and biological testing against a predicate device, not a diagnostic device requiring expert interpretation for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. The methods described are laboratory and animal testing, not human expert adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an AI/diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is not an AI/diagnostic device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The "ground truth" here is the performance of the predicate device (Impra commercial grafts and Impra standard wall grafts) as established through:
    • In vitro measurements: For evaluating mechanical properties and overall equivalence.
    • Animal study results: For assessing in-vivo equivalence.
    • Biomaterial test standards: For biocompatibility and non-toxicity.

8. The sample size for the training set

  • Not applicable. This is not a machine learning or AI device.

9. How the ground truth for the training set was established

  • Not applicable. This is not a machine learning or AI device.

Summary of what the document indicates instead:

This 510(k) submission primarily relies on demonstrating substantial equivalence to an already legally marketed predicate device (Impra's ePTFE vascular grafts) rather than proving performance against independent acceptance criteria. The "study" mentioned involves various forms of testing (in vitro, animal, biomaterial) to compare the Sulzer Vascutek ePTFE grafts to the predicate device. The goal is to show that the new device performs similarly and is as safe and effective as the predicate.

{0}------------------------------------------------

NOV 1 9 1999

Sulzer Carbomedics Inc.

1300 East Anderson Lane Austin, Texas 78752-1793

Phone (512) 435-3200 FAX (512) 435-3350 WATS (800) 648-1579 (US and Canada)

510(k) SUMMARY SULZER VASCUTEK ePTFE VASCULAR PROSTHESIS

The Sulzer Vascutek expanded polytetrafluoroethylene (ePTFE) graft line is substantially equivalent to devices in commercial distribution by Impra, a division of C. R. Bard. All devices referenced under this application are 6mm in diameter or greater and are used to treat diseased or occluded systemic arteries and for A/V access. In addition, the application covers externally supported graft designs.

The device is composed of polytetrafluoroethylene, which has been fabricated in tubular form and expanded to impart porosity to the structure. Similar devices have been used clinically since the 1970's with few reported complications or material failures. ePTFE typically is used as a synthetic conduit to replace natural vessels or as a shunt for AV/dialysis access. The mechanical properties of Sulzer Vascutek ePTFE such as strength, suture retention, and handling are substantially equivalent to products currently in commercial distribution.

Sulzer Vascutek ePTFE grafts are intended for the creation of subcutaneous conduits for blood access, bypass, or reconstruction of occluded or diseased arterial blood vessels. Typical applications for standard wall grafts include systemic vascular repair, primarily for axillo-femoral bypass and femoral-popliteal reconstruction. Typical applications for thin wall grafts include systemic vascular repair, but not for axillofemoral/bifemoral bypass reconstruction.

In vitro testing conducted on the Sulzer Vascutek ePTFE graft line shows it to be substantially equivalent to Impra commercial grafts. Results from animal studies demonstrate that the Sulzer Vascutek standard wall grafts to be equivalent to Impra standard wall grafts. Results from biomaterial testing demonstrate that the Sulzer Vascutek ePTFE grafts are biocompatible and non-toxic.

In summary, all testing demonstrates that the Sulzer Vascutek ePTFE graft line to be substantially equivalent to the grafts in commercial distribution by Impra, Division of C.R. Bard for the reconstruction and bypass of diseased or occluded systemic blood vessels and construction of subcutaneous a-v conduits for blood access.

Common name of the Device:Vascular Graft
Trade name of Proprietary Name:Sulzer Vascutek ePTFE Vascular Prosthesis
Submitter and Contact Person:Edward E. NewtonRegulatory Affairs Manager1300 E. Anderson Lane, Austin, TX 78752Phone: (512) 435-3407 Fax: (512) 435-3350
Submission Submitted on:August 20, 1999

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping, wavy lines, which are meant to represent the essence of human health and services.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 9 1999

Mr. Edward E. Newton Requlatory Affairs Manager Sulzer Carbomedics, Inc. 1300 East Anderson Lane Austin, TX 78752-1793

K992832 Re : Sulzer Vascutek ePTFE Vascular Prosthesis Regulatory Class: II (Two) Product Code: 74 DSY August 20, 1999 Dated: Received: August 23, 1999

Dear Mr. Newton:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of The general controls provisions of the Act include the Act. requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you miqht have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Mr. Edward E. Newton

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding described in your sio!, promains a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance Also, please note the regulation entitled, at (301) 594-4639. "Misbranding by reference to premarket notification" (21CFR Other general information on your responsibilities under 807.97) . . the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html."

Sincerely yours,

for
Celia M. Witten, Ph.D., M.D.
& exper Ce

Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1 _

510(k) Number (if known): __ K992832

Device Name: _ Sulzer Vascutek ePTFE Vascular Prosthesis

Indications For Use: Sulzer Vascutek ePTFE grafts are intended for the creation of subcutaneous arteriovenous conduits for blood access, bypass, or reconstruction of occluded or diseased arterial blood vessels.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bree B. Bempere Qe

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K 992832 510(k) Number _

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 870.3450 Vascular graft prosthesis.

(a)
Identification. A vascular graft prosthesis is an implanted device intended to repair, replace, or bypass sections of native or artificial vessels, excluding coronary or cerebral vasculature, and to provide vascular access. It is commonly constructed of materials such as polyethylene terephthalate and polytetrafluoroethylene, and it may be coated with a biological coating, such as albumin or collagen, or a synthetic coating, such as silicone. The graft structure itself is not made of materials of animal origin, including human umbilical cords.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance Document for Vascular Prostheses 510(k) Submissions.”