K Number
K091538
Date Cleared
2010-03-25

(303 days)

Product Code
Regulation Number
868.5870
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Wet Nose Technologies (WNT) Pressure Release Valve is intended for use with 'positive pressure' breathing gas delivery systems (e.g., High Flow Nasal Cannula or Bubble CPAP type systems) for pediatric/infants patients utilizing continuous flow systems with flow rates greater than 0 L/min and up to, and including 15 L/min. The device is intended for use when a disposable, low pressure (20 cm H20 +/- 4 cm H2O activation), audible, non-adjustable pressure relief valve is needed to be placed upstream of the patient in-line with the circuit to protect the patient from excessive inspiratory pressure in event of a downstream occlusion or an increase of system pressure above the relief pressure. Intended to alert the healthcare provider(s) of a need to take corrective action to reduce system pressure, the device emits an audible sound when circuit pressure is above relief pressure.

Device Description

The Pressure Release Valve (PRV) is a custom valve designed specifically for pediatric/infant use within a Continuous Positive Airway Pressure (CPAP), High/Heated Flow Nasal Cannual (HFNC) and other pressure systems. This valve is a safety feature designed to limit the system pressure of the circuit to pressures below the PRV relief pressure range. The device is intended for use with flow rates greater than 0 L/min up to, and including, 15 L/min. The device is placed upstream of the patient in-line with the circuit to protect the patient from excessive inspiratory pressure in event of a downstream occlusion or an increase of system pressure above the relief pressure. Thus, this device may be used to regulate the maximum pressure achievable within a pressure system. Opening the PRV will activate an audible sound to alert Healthcare professionals over pressurization and the possible need for corrective action. The activation pressure 20 cm H2O +/- 4 cm H2O is based on the relief pressure at 8 L/min is 16 is 20 cm H2O and at 15 L/min is 24 cm H2O. The valve reacts instantaneously to an occlusion and automatically resets upon release of the occlusion. The device is disposable, single-patient use and is prescription only.

The Pressure Release Valve connects with the pressure tubing by way of a Wet Nose Technologies 'T6' Adapter.

The Pressure Release Valve and the 'T6' Adapter are sterilized by gamma irradiation and supplied sterile in bag pouch. Fifty bag pouches are packaged in a dispenser box. Each bag pouch and dispenser box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.

AI/ML Overview

The Wet Nose Technologies Pressure Release Valve is designed as a safety feature for pediatric/infant breathing gas delivery systems to prevent excessive inspiratory pressure. The acceptance criteria and performance are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Activation Pressure
General activation level20 cm H2O
Activation pressure at 8 L/min20 cm H2O
Activation pressure at 15 L/min24 cm H2O
Tolerance for Activation Pressure+/- 4 cm H2O
Audible AlertDevice emits an audible sound when circuit pressure is above relief pressure.
Response TimeThe valve reacts instantaneously to an occlusion.
Reset MechanismAutomatically resets upon release of the occlusion.
Disposable/Single-Patient UseDevice is disposable, single-patient use.
Flow Rate CapabilityGreater than 0 L/min up to, and including, 15 L/min.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" in terms of patient data or clinical trials. Instead, performance testing was conducted on the device itself.

  • Sample Size: Not explicitly stated as a number of devices, but implies testing of the manufactured product.
  • Data Provenance: The testing was nonclinical and conducted by the manufacturer, Wet Nose Technologies, LLC. The studies are prospective in nature, as they involve testing the device against its specifications. There is no indication of country of origin of patient data, as no clinical data was used.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. Ground truth for this device's performance was established through engineering specifications and objective measurements against those specifications, rather than expert interpretation of patient data.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in clinical imaging studies or diagnostic accuracy assessments involving human interpretation. For this device, performance was measured objectively.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. An MRMC comparative effectiveness study was not done. The device is a mechanical safety valve, and its performance is evaluated against engineering specifications, not human diagnostic performance. Therefore, there is no effect size of human readers improving with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, in a sense. The performance testing conducted (e.g., change in relief pressure with flow rate, leak testing, response time) assessed the device's functional characteristics independently. This is analogous to a "standalone" evaluation of the device's mechanical and operational performance.

7. The Type of Ground Truth Used

The ground truth used was the engineering specifications and established physical principles for a pressure relief valve. Performance was measured against these predefined thresholds (e.g., activation pressure, response time).

8. The Sample Size for the Training Set

Not applicable. This device is a mechanical safety valve, not an AI or machine learning algorithm requiring a "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

iC0915388

510(k) Summary (As required by 21 CFR 807.92(a))

MAR 2 5 2010

Application Date:3/15/2010
------------------------------

A. Submitter Information

Wet Nose Technologies, LLC3750 2nd AvenueLos Angeles, CA 90018
Establishment Number:3007521506
Phone Number:949-481-5713
Fax Number:949-481-5745
Contact:Jim Barley
Trade Name:Wet Nose Technologies Pressure ReleaseValve

B. Device Information

Trade/Proprietary Name:Wet Nose Technologies PressureRelease Valve
Common name of device:Pressure Relief Valve
Classification Name:Nonrebreathing Valve
Product Code:CBP
Regulatory Class:II
Classification Number:868.5870
Reason for 510(k):New Device

C. Predicate Device:

Fisher & Paykel BC110Pressure Manifold
Predicate 510(k) #:K040366
Predicate product code:CBP

{1}------------------------------------------------

D. Device Description

The Pressure Release Valve (PRV) is a custom valve designed specifically for pediatric/infant use within a Continuous Positive Airway Pressure (CPAP), High/Heated Flow Nasal Cannual (HFNC) and other pressure systems. This valve is a safety feature designed to limit the system pressure of the circuit to pressures below the PRV relief pressure range. The device is intended for use with flow rates greater than 0 L/min up to, and including, 15 L/min. The device is placed upstream of the patient in-line with the circuit to protect the patient from excessive inspiratory pressure in event of a downstream occlusion or an increase of system pressure above the relief pressure. Thus, this device may be used to regulate the maximum pressure achievable within a pressure system. Opening the PRV will activate an audible sound to alert Healthcare professionals over pressurization and the possible need for corrective action. The activation pressure 20 cm H2O +/- 4 cm H2O is based on the relief pressure at 8 L/min is 16 is 20 cm H2O and at 15 L/min is 24 cm H2O. The valve reacts instantaneously to an occlusion and automatically resets upon release of the occlusion. The device is disposable, single-patient use and is prescription only.

The Pressure Release Valve connects with the pressure tubing by way of a Wet Nose Technologies 'T6' Adapter.

The Pressure Release Valve and the 'T6' Adapter are sterilized by gamma irradiation and supplied sterile in bag pouch. Fifty bag pouches are packaged in a dispenser box. Each bag pouch and dispenser box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.

  • Statement of Indications for Use E.
    The Wet Nose Technologies (WNT) Pressure Release Valve is intended for use with 'positive pressure' breathing gas delivery systems (e.g., High Flow Nasal Cannula or Bubble CPAP type systems) for pediatric/infants patients utilizing continuous flow systems with flow rates greater than 0 L/min and up to, and including 15 L/min.

{2}------------------------------------------------

The device is intended for use when a disposable, low pressure (20 cm H20 +/- 4 cm H2O activation), audible, non-adjustable pressure relief valve is needed to be placed upstream of the patient in-line with the circuit to protect the patient from excessive inspiratory pressure in event of a downstream occlusion or an increase of system pressure above the relief pressure. Intended to alert the healthcare provider(s) of a need to take corrective action to reduce system pressure, the device emits an audible sound when circuit pressure is above relief pressure.

  • Comparison of Required Technological Characteristics: F.
    Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Wet Nose Technologies Pressure Release Valve and the cited predicate device.

  • Discussion of Nonclinical Tests: G.
    The intended use of the Wet Nose Technologies Pressure Release Valve is identical to that of the cited predicate device. Any differences in technological characteristics were insignificant and do not raise new issues of safety or effectiveness.

Performance testing consisted of the following:

    1. Change in relief pressure with flow rate
    1. Leak Testing
    1. Response time to an occlusion

In addition, testing for compliance to the applicable sections of the following voluntary standards was performed:

    1. ANSI/AAMI/ISO 11137 Sterilization of health care products --Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
    1. ISO 11607;2003 Packaging for terminally sterilized medical devices
    1. ISO 10993-4:2006 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood

{3}------------------------------------------------

  • H. Discussion of Clinical Tests:
    None submitted

  • Conclusions Demonstrating Safety, Effectiveness and Performance: I.
    The WNT Pressure Relief Valves has a pressure activation level of 20 cm H20. The Instructions for Use provide the pressure activation levels by flow rate for the device.

The design of the Pressure Release Valve and the T6 Adapter minimize the chance of user error. The device has been tested and found to meet all product specifications and requirements. Accelerated aging was used to verify the performance of the product over the life of the device.

Instructions for Use detail how to use the devices and the conditions of use. Product labeling clearly shows that the device is for single patient use only.

The WNT Pressure Release Valve and T6 Adapter have been found to be safe and effective for their intended use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features an abstract eagle design with three stylized wing strokes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Jim Barley Director of Regulatory Affairs/Quality Assurance Wet Nose Technologies, LLC 3750 2nd Avenue Los Angeles, California 90018

MAR 2 5 2010

Re: K091538

Trade/Device Name: Wet Nose Technologies Pressure Release Valve Regulation Number: 21CFR 868.5870 Regulation Name: Nonrebreathing Valve Regulatory Class: II Product Code: CBP Dated: March 15, 2010 Received: March 19, 2010

Dear Mr. Barley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Mr. Barley

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reprting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/com 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/ default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications For Use

510(k) Number (if known):

Device Name: Wet Nose Technologies (WNT) Pressure Release Valve

Indications For Use:

The Wet Nose Technologies (WNT) Pressure Release Valve is intended for use with 'positive pressure' breathing gas delivery systems (e.g., High Flow Nasal Cannula or Bubble CPAP type systems) for pediatric/infants patients utilizing continuous flow systems with flow rates greater than 0 L/min and up to, and including 15 L/min. The device is intended for use when a disposable, low pressure (20 cm H20 +/- 4 cm H2O activation), audible, non-adjustable pressure relief valve is needed to be placed upstream of the patient in-line with the circuit to protect the patient from excessive inspiratory pressure in event of a downstream occlusion or an increase of system pressure above the relief pressure. Intended to alert the healthcare provider(s) of a need to take corrective action to reduce system pressure, the device emits an audible sound when circuit pressure is above relief pressure.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

L. Schultheis

(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 868.5870 Nonrebreathing valve.

(a)
Identification. A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.(b)
Classification. Class II (performance standards).