K Number
K090824

Validate with FDA (Live)

Date Cleared
2009-06-02

(68 days)

Product Code
Regulation Number
866.3120
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in PreservCyt® Solution using an aliquot that is removed prior to processing for additional gynecological testing. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

The BD Viper System, when used with the BD ProbeTec amplified nucleic assay(s), is intended for the in vitro detection of targeted organisms from specimens as identified in the assay-specific reagent package insert(s).

Device Description

The BD ProbeTec CT Q* Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe (8, 9). The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of C. trachomatis DNA is determined by calculating the peak fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

In addition to the fluorescent probe used to detect amplified C. trachomatis target DNA, a second labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the C. trachomatisspecific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and C. trachomatis-specific signals to report results as positive, negative, or EC failure.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Acceptance Criteria and Device Performance

The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined set of thresholds. Instead, it presents clinical performance characteristics (sensitivity and specificity) that are implicitly the targets for the device to be considered effective. The reported device performance is then compared against these implied expectations.

MetricAcceptance Criteria (Implied)Reported Device Performance (BD ProbeTec CT Q* Assay)
SensitivityHigh sensitivity for C. trachomatis detectionSymptomatic: 96.7% (95% C.I. 88.7% - 99.6%) Asymptomatic: 91.9% (95% C.I. 83.2% - 97.0%)Total: 94.1% (95% C.I. 88.7% - 97.4%)
SpecificityHigh specificity for C. trachomatis detectionSymptomatic: 99.8% (95% C.I. 99.2% - 100.0%) Asymptomatic: 99.8% (95% C.I. 99.4% - 100.0%)Total: 99.8% (95% C.I. 99.5% - 100.0%)
LOD (Analytical Sensitivity)≤ 30 CT EB per mL in PreservCyt specimens (or ≤ 1 IFU per mL)≤ 30 CT EB per mL in PreservCyt specimens (corresponds to ≤ 1 IFU per mL)
Detection of Serovars> 95% proportion positive at 15 EB per mLDetected 16 CT serovars with > 95% proportion positive at 15 EB per mL
Interfering SubstancesNo interference from common vaginal products, blood, etc.No interference from most listed substances; Glacial Acetic Acid + Blood (≤5%/1% V/V) may cause EC failures or false negatives.

Study Details

1. Sample sized used for the test set and the data provenance:

  • Sample Size: 2071 female subjects were evaluated in the clinical study.
  • Data Provenance: The data was collected from prospective clinical sites in North America. Specifically, specimens were collected from female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at eleven geographically diverse clinical sites in North America.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not specify the number of experts or their qualifications for establishing the ground truth. It refers to a "patient infected status (PIS) algorithm" which relied on results from three reference methods.

3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • The adjudication method was based on a "patient infected status (PIS) algorithm".
    • 2+1: At least two positive reference results were required to establish a subject as PIS-positive.
    • 2+1: At least two negative reference results were required to establish a subject as PIS-negative.
    • This implies a 2-out-of-3 consensus approach using results from the three reference endocervical swabs (BD ProbeTec ET CT/GC/AC assay, BD ProbeTec CT Q* assay, and another commercially available NAAT).

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study evaluates the performance of an in vitro diagnostic (IVD) device (DNA assay) and not an AI-assisted diagnostic system where human readers would be involved in interpreting AI output.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Yes, this was a standalone study in the context of an IVD device. The BD ProbeTec CT Q* Amplified DNA Assay, when used with the BD Viper System, automatically processes and reports results (positive, negative, or EC failure) based on fluorescence measurements and an automated algorithm. There is no human interpretation step being evaluated as part of results generation within the device's operational workflow.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth was established by expert consensus based on multiple reference methods. Specifically, it was determined by a Patient Infected Status (PIS) algorithm which required at least two concordant results (positive or negative) from three different commercially available NAATs on endocervical swab specimens.

7. The sample size for the training set:

  • The document does not specify a training set size. This is typical for a traditional IVD assay development, where "training" in the machine learning sense isn't applicable. The assay's parameters would have been optimized during earlier development (e.g., analytical performance characteristics like LOD, specificity for serovars, interfering substances), but a distinct "training set" as understood in AI/ML is not part of this documentation.

8. How the ground truth for the training set was established:

  • As no "training set" in the AI/ML sense is explicitly mentioned, the method for establishing ground truth for such a set is not applicable/not provided. The analytical performance characteristics were established using controlled laboratory experiments (e.g., seeding known concentrations of C. trachomatis serovars).

{0}------------------------------------------------

Image of BD logo and text

BD

Helping all people
live healthy lives

BD ProbeTec™ Chlamydia trachomatis (CT) ( Amplified DNA Assay

ApplicantBD Diagnostic Systems7 Loveton CircleSparks, MD 21152
Establishment Registration No.1119779
Contact PersonSaba Modjarradtel. 410-316-4685fax. 410-316-4499saba_modjarrad@bd.com
Summary DateMarch 25, 2009
Proprietary NameBD ProbeTec TM Chlamydia trachomatis (CT) Q x Amplified DNA Assay
Generic NameDNA probe, nucleic acid amplification, Chlamydia
ClassificationClass I
Classification NameChlamydia serological reagents
Regulation Number866.3120
Product CodeMKZ
Predicate DevicesBD ProbeTec TM Chlamydia trachomatis (CT) Q x Amplified DNA Assay (K081824),Gen-Probe APTIMA Assay for Chlamydia trachomatis (ACT) (K053446)

Device Description

The BD ProbeTec CT Q* Amplified DNA Assay is based on the simultaneous amplification and detection of target DNA using amplification primers and a fluorescently-labeled detector probe (8, 9). The reagents for SDA are dried in two separate disposable microwells: the Priming Microwell contains the amplification primers, fluorescently-labeled detector probe, nucleotides and other reagents necessary for amplification, while the Amplification Microwell contains the two enzymes (a DNA polymerase and a restriction endonuclease) that are required for SDA. The BD Viper™ System pipettes a portion of the purified DNA solution from each Extraction Tube into a Priming Microwell to rehydrate the contents. After a brief incubation, the reaction mixture is transferred to a corresponding, pre-warmed Amplification Microwell which is sealed to prevent contamination and then incubated in one of the two thermally-controlled fluorescent readers. The presence or absence of C. trachomatis DNA is determined by calculating the peak

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the BD logo. The logo consists of a stylized sun-like graphic on the left and the letters "BD" on the right. Below the letters, there is a tagline that reads "Helping all people live healthy lives."

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

fluorescence (Maximum Relative Fluorescent Units (MaxRFU)) over the course of the amplification process and by comparing this measurement to a predetermined threshold value.

In addition to the fluorescent probe used to detect amplified C. trachomatis target DNA, a second labeled oligonucleotide is incorporated in each reaction. The Extraction Control (EC) oligonucleotide is labeled with a different dye than that used for detection of the C. trachomatisspecific target and is used to confirm the validity of the extraction process. The EC is dried in the Extraction Tubes and is rehydrated upon addition of the specimen and extraction reagents. At the end of the extraction process, the EC fluorescence is monitored by the BD Viper System and an automated algorithm is applied to both the EC and C. trachomatis-specific signals to report results as positive, negative, or EC failure.

Intended Use

The BD ProbeTecT™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in PreservCvt® Solution using an aliquot that is removed prior to processing for additional gynecological testing. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

Summary and Principles of Operation

When used with the BD Viper System, the BD ProbeTec CT Q* Amplified DNA Assay involves automated extraction of DNA from clinical specimens through the chemical lysis of cells, followed by binding of DNA to para-magnetic particles, washing of the bound nucleic acid and elution in an amplification-compatible buffer. When present, C. trachomatis DNA is then detected by Strand Displacement Amplification (SDA) of a specific target sequence in the presence of a fluorescently labeled detector probe.

Analytical Performance Characteristics

Limit of Detection (Analytical Sensitivity)

The Limits of Detection (LODs) for the CT Q* Assay with C. trachomatis seroyar H in PreservCyt specimens when extracted on the BD Viper System were determined to be ≤ 30 CT EB per mL. A correlation of EB to IFU suggests that the CT Q* assay LODs with serovar H in PreservCyt specimens correspond to ≤ 1 IFU per mL. The CT Q* Assay on the BD Viper System in extracted mode was able to detect 16 CT serovars with > 95% proportion positive at a concentration of 15 EB per mL in clean diluted PreservCyt Solution.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of a stylized sun-like graphic with a smiling face incorporated into the design, followed by the letters "BD" in bold font. Below the letters, the company's slogan reads, "Helping all people live healthy lives" in a smaller, italicized font.

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

Interfering Substances

Potential interfering substances which may be encountered in PreseryCyt specimens were extracted from diluted PreservCyt matrix in the absence and presence of CT target (90 CT EBs/mL) and tested with the BD ProbeTec CT Q* Amplified DNA Assay on the BD Viper System. Results are summarized in Table 2.

Table 2 Interfering Substances

InterpretationPreservCyt
No Interference ObservedBlood (≤ 1%)Seminal FluidMucusOver The Counter vaginal products and contraceptivesHemorrhoidal creamPrescription vaginal treatmentsLeukocytes (1x106 cells/mL)1x106 cells/mL Neisseria gonorrhoeae
May cause extraction control(EC) failuresGlacial Acetic Acid + Blood (≤5%/1% V/V)
May cause False NegativeresultsGlacial Acetic Acid + Blood (≤5%/1% V/V)

Clinical Performance Characteristics

Endocervical swab specimens and PreservCyt specimens were collected from 2079 compliant female subjects attending family planning, OB/GYN, and sexually transmitted disease clinics at eleven geographically diverse clinical sites in North America. Subjects were classified as symptomatic if they reported symptoms such as dysuria, coital pain/difficulty/bleeding, abnormal vaginal discharge, or pelvic/uterine/adnexal pain. Subjects were classified as asymptomatic if they did not report symptoms. Five subjects were excluded due to an undetermined patient infected status. Three subjects did not have a PreservCyt specimen result. There were 2071 subjects evaluated.

Three randomized endocervical swab specimens and a PreservCyt specimen were collected from each female subject. The three reference endocervical swabs were tested with the BD ProbeTec ET CT/GC/AC assay, the BD ProbeTec CT Q* assay, and another commercially available NAAT (Nucleic Acid Amplification Test). Sensitivity and specificity for PreservCyt specimens were calculated by comparing results to a patient infected status (PIS) algorithm. The designation of positive or negative PIS was based on the endocervical swab specimen results

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the BD logo. On the left is an image of a person with their arms raised, with rays of light emanating from behind them. To the right of the image are the letters "BD" in bold font. Below the letters is the text "Helping all people live healthy lives".

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

from the three reference methods. At least two positive reference results were required to establish a subject as PIS-positive. At least two negative reference results were required to establish a subject as PIS-negative. Sensitivity and specificity by symptomatic status are presented in Table 4.

The distribution of cervical sampling devices used in the clinical study according to clinical collection site is summarized in Table 3.

Table 3 Summary of Cervical Sampling Devices Used in the PreservCyt Specimen Clinical Study

ПРИЗИОНСЕНИЕ ПРИЗИОНИЕ ПРОДОВЕННИЕ ПРИЗЕРИНИЕ В ИЗВЕННИЦИИ ПОДОСТОВСКОГО ПРОДОВИЕ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДОВОДИЛИ ПОДО1 - - - - - - - - - - - - - - -1. 2 2 2 2 2 2 2 12Career States of Canadian Career StationATAYAY YOUI SEE SIL BI EE EE EEE BEE EE LE LE LE LE LE LE CLA C L C . L . L . C . C . C . CCompany of The Property of Children Company of Children Children Children Children Children Children Children Children Children Children Children Children Children Children CCONSULEMENT CONSULTION COLLECTION CONSULTERS CONSULTERS CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTER CONSULTI2000O DUPO MOUNT ENT FRE HERLANA TRANSA
1a province as any arman and and sure andСтава В серетенсовского массиии са седать в в с в полностия с.should with the " contine for shoulder " would seller belief includes been a shoState Clouding2000 0000Concession of CattleCareer and College ofANN AN AN E LANDER38.581Concession. Career of the programCompany Call Call of.
Device11 8 0000 BLA 18 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A SUNCHILL CREASE & AND AND A BUSINESS Aver Children Aven Children

Table 4 CT Q+ Assay Performance for PreservCyt Specimens Compared to Patient Infected Status (by symptomatic status)

Performance Compared to Patient Infected Status
SymptomaticnSensitivity95% C.I.Specificity95% C.I.PPV%NPV%Error Initial/Final
A134791.9%(68/74)(83.2% -97.0%)99.8%(1271/1273)(99.4% -100.0%)96.4%99.5%1/0
S72496.7%(59/61)(88.7% -99.6%)99.8% (662/663)(99.2% -100.0%)97.8%99.7%0/0
Total207194.1%(127/135)(88.7% -97.4%)99.8%(1933/1936)(99.5% -100.0%)97.0%99.6%1/0

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a stylized sun-like symbol with a human figure in the center, and the letters "BD" in bold font. Below the letters, there is a tagline that reads "Helping all people live healthy lives."

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

A total of 2071 CT Q* Assay results from PreservCyt specimens was evaluated from cleven geographically diverse clinical sites. A frequency distribution of the initial MaxRFU values for the CT Q* assay is shown in Figure A.

Image /page/4/Figure/4 description: The image shows the title of a figure. The title is "Figure A Frequency Distribution of MaxRFU for the CT Q* Assay (PreservCyt Specimens)". The title indicates that the figure will show the frequency distribution of MaxRFU for the CT Q* Assay.

Image /page/4/Figure/5 description: The image is a bar graph showing frequency on the y-axis and MaxRFU on the x-axis. The first bar, representing 0-49 MaxRFU, has a frequency of 1930. The bar representing >=800 MaxRFU has a frequency of 128, while the other bars have frequencies of 0 or 1.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the BD logo. The logo consists of a stylized sun-like graphic on the left and the letters "BD" in bold font on the right. Below the letters, there is a tagline that reads "Helping all people live healthy lives" in a smaller font.

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

Reproducibility

A reproducibility study of the BD Viper System using the BD ProbeTec CT Q* Assay was conducted for Liquid Based Cytology (LBC) specimens at three clinical sites on one BD Viper System per site. A panel of simulated specimens comprising CT and GC organisms seeded into LBC Specimen Dilution Tubes containing LBC medium was tested with the BD ProbeTec CT Q Assay. Uninoculated LBC Specimen Dilution Tubes containing LBC medium were used for the CT negative samples. Nine replicates of each panel member were tested every day for five days on each BD Viper System. The data are summarized in Table S.

Two additional target levels were included in the panels to characterize the reproducibility of test results (i.e., proportion positive or negative) at target levels below the analytical Limit of Detection (LOD) of the BD ProbeTec CT Q* Assay. These additional specimens comprised CT and GC organisms seeded into LBC Specimen Dilution Tubes containing LBC medium at dilutions of 1:10 and 1:100 of the respective analytical LODs of each analyte. These levels were selected to fall within the dynamic range of the analytical LOD curves for the BD ProbeTec CT Q and GC Q assays. Nine replicates of each panel member were tested every day for five days across the three BD Viper Systems. The data are summarized in Table 6.

Table 5 Summary of Reproducibility Data for LBC Specimens on the BD Viper System for the CT Q* Assay

CTEBs/mLGCCells/mL%Correct95% CIMeanMaxRFUWithin RunBetween RunsWithin SiteBetweenSite
00100.0%(135/135)(97.3% -100.0%)1.304.66357.640.8565.290.2015.12
300100.0%(135/135)(97.3% -100.0%)2021.95225.9411.1716.580.8221.521.06
0100100.0%(135/135)(97.3% -100.0%)1.353.63268.970.000.000.8764.48
30250100.0%(135/135)(97.3% -100.0%)2028.41155.457.669.930.490.000.00
75100100.0%(135/135)(97.3% -100.0%)1964.40170.918.7044.372.268.700.44

Page 6 of 7

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for BD, a medical technology company. The logo consists of a stylized sunburst graphic to the left, followed by the letters "BD" in bold, sans-serif font. Below the letters, the company's tagline reads "Helping all people live healthy lives" in a smaller, more delicate font.

BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

Table 6 Characterization of System Reproducibility at Target Levels below the Analytical Limit of Detection for the CT Q* Assay for LBC Specimens

DilutionofAnalyticalLOD% Positive95% CI(Positive)MaxRFUMean(Positive)% Negative95% CI(Negative)MaxRFUMean(Negative)
1:1050.4 (68/135)(41.6 - 59.1)1935.949.6 (67/135)(40.9 - 58.4)11.5
1:1007.4 (10/135)(3.6 - 13.2)1835.792.6 (125/135)(86.8 - 96.4)9.4

Conclusions

The analytical and clinical study results for the BD ProbeTec Chlamydia trachomatis (CT) Q* Amplified DNA Assay with PreservCyt specimens support the determination of substantial equivalence in accordance with the intended use as stated in the product labeling.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's commitment to health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Saba Modiarrad Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

JUN - 2 2009

Re: K090824

Trade/Device Name: BD Probetec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

Regulation Number: 21 CFR 866.3120 Regulation Name: Chlamydia serological reagents Regulatory Class: Class I Product Code: MKZ Dated: March 25, 2009 Received: March 26, 2009

Dear Ms. Modjarrad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indicrions for use stated in the enclosure) to legally marketed predicate devices market the interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmonts, or to devices that have been reclassified in accordance with the provisions of the Federallients, ou o and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, l'usting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Dov. In a

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing prectice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{8}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrio's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attaym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

1090824 510(k) Number (if known):

Device Name: BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay

Indications For Use:

The BD ProbeTec™ Chlamydia trachomatis (CT) Q* Amplified DNA Assay, when tested with the BD Viper™ System in Extracted Mode, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in PreservCyt® Solution using an aliquot that is removed prior to processing for additional gynecological testing. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

The BD Viper System, when used with the BD ProbeTec amplified nucleic assay(s), is intended for the in vitro detection of targeted organisms from specimens as identified in the assay-specific reagent package insert(s).

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off
---------------------

Office of In Vitro Diagnostic Device
Evaluation and Safety510(k) K090824

Page 1 of 1

BD Diagnostic Systems Becton, Dickinson and Company

Page ix

§ 866.3120 Chlamydia serological reagents.

(a)
Identification. Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusChlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).(b)
Classification. Class I (general controls).