K Number
K053446

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-07-25

(225 days)

Product Code
Regulation Number
866.3120
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The APTIMA Assay for Chlamydia trachomatis is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) in clinician-collected endocervical, vaginal and male urethral swab specimens, patient-collected vaginal swab specimens, and female and male urine specimens. The assay is also intended for use with the testing of gynecological specimens collected in the PreservCyt® Solution and processed with the Cytyc ThinPrep® 2000 System. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

Device Description

Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE APTIMA Assay for Chlamydia trachomatis to include PreservCyt liquid Pap specimens (collected and processed by the Cytyc ThinPrep 2000 Processor) as acceptable testing specimens. The ancillary kit formulated for this specific application is the commercially available GEN-PROBE APTIMA Specimen Transfer Kit. The components of the APTIMA Specimen Transfer Kit include: (1) a transport tube containing transport media with a penetrable cap and (2) specific instructions for use regarding decontamination and specimen processing procedures. The APTIMA Specimen Transfer Kit may only be used in conjunction with GEN-PROBE APTIMA Assays for the detection of Chlamydia trachomatis and/or Neisseria gonorrhoeae.

AI/ML Overview

The GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis, specifically for its expanded indication with ThinPrep Specimens, underwent a clinical study to evaluate its performance.

1. Table of Acceptance Criteria and Reported Device Performance:

The document describes the performance in terms of sensitivity and specificity. While explicit acceptance criteria are not called out as "acceptance criteria," the clinical study results are presented, suggesting that these performance metrics were deemed acceptable for the device's expanded use.

MetricAcceptance Criteria (Implied)Reported Device Performance (Overall)
Sensitivity(Not explicitly stated, but high sensitivity desired for diagnostic tests)95.6% (86/90)
Specificity(Not explicitly stated, but high specificity desired for diagnostic tests)98.8% (1539/1557)

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: 1,647 female subjects were evaluated in the clinical study.
    • 1,288 asymptomatic subjects
    • 359 symptomatic subjects
  • Data Provenance: The study was a prospective multi-center clinical study. The data was collected from OB/GYN, family planning, public health, women’s, and STD clinics. The country of origin is not explicitly stated, but the submission to the FDA and use of brand names like Cytyc ThinPrep 2000 System suggest a US-centric study or at least a study adhering to US regulatory standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • The document does not mention the use of experts to establish the ground truth in the traditional sense of human readers adjudicating images or cases.
  • Instead, the ground truth (patient infected status) was established using a patient infected status algorithm. This algorithm relied on the results of two reference Nucleic Acid Amplification Tests (NAATs) on endocervical swab specimens.

4. Adjudication Method for the Test Set:

  • Adjudication Method: The patient infected status was determined by an algorithm that used two reference NAATs (APTIMA Combo 2 Assay and APTIMA CT Assay) on endocervical swab specimens.
    • An infected patient status required both reference NAATs to be positive.
    • A non-infected patient status required at least one reference NAAT to be negative.
    • This can be considered a form of "2-out-of-2" positive for infection, and "1-out-of-2" negative for non-infection, rather than human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No MRMC comparative effectiveness study was done. This study focuses on the standalone performance of the APTIMA Assay for Chlamydia trachomatis when using PreservCyt liquid Pap specimens, not a comparison involving human readers with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance evaluation was done. The clinical study directly assessed the performance of the APTIMA CT Assay on PreservCyt liquid Pap specimens against the defined patient infected status algorithm. This represents the algorithm's performance without direct human interpretation of the assay results, other than performing the lab test.

7. The Type of Ground Truth Used:

  • Algorithm Consensus (from reference NAATs): The ground truth for the clinical study was established by a patient infected status algorithm based on the results of two highly sensitive and specific reference Nucleic Acid Amplification Tests (NAATs) – the APTIMA Combo 2 Assay and the APTIMA CT Assay – performed on endocervical swab specimens. This is a common and robust method for establishing ground truth for infectious disease diagnostics.

8. The Sample Size for the Training Set:

  • The document does not explicitly state the sample size for a training set. Diagnostic assays like the APTIMA Assay are developed through extensive research and analytical validation (limit of detection, analytical specificity, interference, recovery studies) before large-scale clinical validation. The details provided primarily cover the validation study for the expanded indication. It is typical for the assay to have been developed and optimized using various samples, but a specific "training set" in the machine learning sense is not described.

9. How the Ground Truth for the Training Set was Established:

  • As a specific "training set" and its ground truth establishment are not detailed in the provided document, this information cannot be extracted. The document focuses on the validation of an existing assay for an expanded specimen type. The ground truth for the extensive analytical studies would likely have been established using well-characterized culture isolates and spiked samples with known concentrations of C. trachomatis as described in the Limit of Detection, Analytical Specificity, and Recovery sections (e.g., using IFU/assay as a measure for CT organisms).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo and text for GEN-PROBE INCORPORATED. The logo is a stylized black shape on the left. Below the company name is the text "APTIMA Assay for Chlamydia trac."

K053446

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis

IJUL 2 5 2006

General Information

Submitted By:Company Contact:
Gen-Probe IncorporatedBrian J. Shea, RAC
10210 Genetic Center DriveRegulatory Affairs Associate
San Diego, California 92121
phone: (858) 410-8000phone: (858) 410-8535
fax: (858) 410-8622fax: (858) 410-8622
e-mail: brians@gen-probe.com
Trade Name:GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis
Common or Usual Name:rRNA target-amplified nucleic acid probe test for the in vitrodiagnostic detection of Chlamydia trachomatis
Classification Name:DNA Probe, Nucleic Acid Amplification, Chlamydia
Classification Code:Medical Specialty: Microbiology
Product Code: MKZ
Registration Number: CFR 866.3120
Device Class: 1
Description: Reagents used to identify chlamydia directlyfrom clinical specimens or cultured isolates derived fromclinical specimens. The identification aids in the diagnosis ofdisease caused by bacteria belonging to the genus Chlamydiaand provides epidemiological information on these diseases.Chlamydia are the causative agents of psittacosis (a form ofpneumonia), lymphogranuloma venereum (a venereal disease),and trachoma (a chronic disease of the eye and eyelid).

Substantially Equivalent Device:

GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black and white graphic of a stylized fist. The fist is clenched and angled slightly to the right. The design is simple and bold, with clean lines and a focus on the shape of the fist. The background is plain white, which makes the black fist stand out.

PTIMA Assay for Chlamydia trachomtias - Expanded Indication: ThinPrep Specimens

Device Description

Clearance of this premarket notification extends the clinical performance claims of the commercially available GEN-PROBE APTIMA Assay for Chlamydia trachomatis to include PreservCyt liquid Pap specimens (collected and processed by the Cytyc ThinPrep 2000 Processor) as acceptable testing specimens. The ancillary kit formulated for this specific application is the commercially available GEN-PROBE APTIMA Specimen Transfer Kit. The components of the APTIMA Specimen Transfer Kit include: (1) a transport tube containing transport media with a penetrable cap and (2) specific instructions for use regarding decontamination and specimen processing procedures. The APTIMA Specimen Transfer Kit may only be used in conjunction with GEN-PROBE APTIMA Assays for the detection of Chlamydia trachomatis and/or Neisseria gonorrhoeae.

Intended Use

APTIMA Assay for Chlamydia trachomatis package insert:

The APTIMA Assay for Chlamydia trachomatis is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) in clinician-collected endocervical, vaginal and male urethral swab specimens, patient-collected vaginal swab specimens, and female and male urine specimens. The assay is also intended for use with the testing of gynecological specimens collected in the PreservCyt® Solution and processed with the Cytyc ThinPrep® 2000 System. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

1 Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

{2}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Ancillary Kit package insert:

The GEN-PROBE APTIMA Specimen Transfer Kit is only for use with GEN-PROBE APTIMA Assays for the detection of Chlamydia trachomatis and/or Neisseria gonorrhoeae. The GEN-PROBE APTIMA Specimen Transfer Kit allows for APTIMA Assay testing of gynecological specimens collected and processed by the Cytyc ThinPrep 2000 Processor according to the instructions provided.

APTIMA Assay for Chlamydia trachomatis

A complete description of the APTIMA Assay for Chlamydia trachomatis is provided in the commercialized package insert.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a black and white graphic. The graphic appears to be a stylized symbol or logo. It is composed of geometric shapes, primarily angled lines and curves, forming an abstract design. The symbol is solid black against a white background, creating a high contrast image.

GEN-PROBE INCORPORATED PTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Summary of Non-Clinical (Analytical Laboratory) Performance Data

Limit of Detection (Analytical Sensitivity)

C. trachomatis analytical sensitivity (limit of detection) was determined by directly comparing dilutions of CT organisms in cell culture and in the APTIMA CT assay. The analytical sensitivity claim for the assay is one Inclusion-Forming Unit (IFU) per assay (7.25 IFU/swab, 5 IFU/mL urine, and 9.75 IFU/mL PreservCyt Solution liquid Pap) for all 15 CT serovars. However, dilutions of less than one IFU/assay of all serovars tested positive.

Analytical Specificity

A total of 154 culture isolates were evaluated using the APTIMA CT Assay. These isolates included 86 organisms that may be isolated from the urogenital tract and 68 additional organisms that represent a phylogenetic cross-section of organisms. The tested organisms included bacteria, fungi, yeast, parasites and viruses. All organisms except C. psittaci, C. pneumoniae, U. urealyticum and the viruses were tested at 1.0x100 cells/assay in Kova-trol/Urine Transport Media and 60 organisms were tested in Swab Transport Media. The Chlamydia and Neisseria organisms were tested in the PreservCyt Solution media. C. psittaci VR601 was tested at 8x104 cells/assay and C. psittaci VR125 was tested at 1x105 cells/assay. C. pneumoniaewas tested at 4x103 cells/assay and U. urealyticum was tested at 6.7x106 cells/assay. The viruses were tested as follows: (a) herpes simplex virus I: 2.5x104 TCID50/assay, (b) herpes simplex virus II: 6.0x104 TCID50/assay, (c) human papillomavirus 16: 2.9x106 DNA copies/assay and (d) cytomegalovirus: 4.8x10 cells/assay. The list of organisms tested is shown in Table 1.

{4}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Table 1 - APTIMA CT Assay Analytical SpecificityORGANISMORGANISMORGANISM
Achromobacter xerosisEscherichia coliNeisseria mucosa (3)
Acinetobacter calcoaceticusFlavobacterium meningosepticumNeisseria sicca (3)
Acinetobacter IwoffiFusobacterium nucleatumNeisseria subflava (14)
Actinomyces israeliiGardnerella vaginalisNeisseria perflava
Actinomyces pyogenesGemella haemolysansNeisseria polysaccharea
Aerococcus viridansHaemophilus ducreyiParacoccus denitrificans
Aeromonas hydrophilaHaemophilus influenzaePeptostreptococcusanaerobius
Agrobacterium radiobacterHerpes simplex virus IPeptostreptococcusproductus
Alcaligenes faecalisHerpes simplex virus IIPlesiomonas shigelloides
Bacillus subtilisHuman papilloma virus 16Propionibacterium acnes
Bacteriodes fragilisKingella dentrificansProteus mirabilis
Bacteriodes ureolyticusKingella kingaeProteus vulgaris
Bifidobacterium adolescentisKlebsiella oxytocaProvidencia stuartii
Bifidobacterium breviKlebsiella pneumoniaePseudomonas aeruginosa
Branhamella catarrhalisLactobacillus acidophilusPseudomonas fluorescens
Brevibacterium linensLactobacillus brevisPseudomonas putida
Campylobacter jejuniLactobacillus jensoniiRahnella aquatilis
Candida albicansLactobacillus lactisRhodospirillum rubrum
Candida glabrataLegionella pneumophila (2)Saccharomyces cerevisiae
Candida parapsilosisLeuconostoc paramensenteroidesSalmonella minnesota
Candida tropicalisListeria monocytogenesSalmonella typhimurium
Chlamydia pneumoniaeMicrococcus luteusSerratia marcescens
Chlamydia psittaci (2)Moraxella lacunataStaphylococcussaprophyticus
Chromobacterium violaceumMoraxella osloensisStaphylococcus aureus
Citrobacter freundiiMorganella morganiiStaphylococcusepidermidis
Clostridium perfringensMycobacterium smegmatisStreptococcus agalactiae
Corynebacterium genitaliumMycoplasma genitaliumStreptococcus bovis
Corynebacterium xerosisMycoplasma hominisStreptococcus mitis
Cryptococcus neoformansN. meningitidis Serogroup AStreptococcus mutans
CytomegalovirusN. meningitidis Serogroup BStreptococcus pneumoniae
Deinococcus radioduransN. meningitidis Serogroup C (4)Streptococcus pyogenes
Derxia gummosaN. meningitidis Serogroup DStreptococcus salivarius
Eikenella corrodensN. meningitidis Serogroup YStreptococcus sanguis
Enterobacter aerogenesN. meningitidis Serogroup W135Streptomyces griseinus
Enterobacter cloacaeNeisseria cinerea (4)Trichomonas vaginalis
Entercoccus aviumNeisseria dentrificansUreaplasma urealyticum
Entercoccus faecalisNeisseria elongata (3)Vibrio parahaemolyticus
Entercoccus faeciumNeisseria flavaYersinia enterocolitica
Erwinia herbicolaNeisseria flavescens (2)
Erysipelothrix rhusiopathiaeNeisseria lactamica (9)

A PTIMA CT Assay Analytical Specificity Table

(n) = number of strains tested

All organisms tested produced a negative result in the APTIMA CT Assay.

{5}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Interference Studies

The following interfering substances were individually spiked into swab, PreservCyt liquid Pap and/or urine specimens: 10% blood, contraceptive jelly, spermicide, moisturizer, hemorrhoidal anesthetic, body oil, powder, anti-fungal cream, vaginal lubricants, femininc spray and leukocytes (1x106 cells/mL). The following interfering substances were individually spiked into urine specimens: 30% blood, urine analytes, protein, glucose, ketones, bilirubin, nitrate, urobilinogen, pH 4 (acidic), pH 9 (alkaline), leukocytes (1x106 cells/mL), cellular debris, vitamins, minerals, acetaminophen, aspirin and ibuprofen. All were tested for potential assay interference in the absence and presence of CT at the estimated rRNA equivalent of 1 cell/assay (5 fg/assay). The rRNA equivalents were calculated based on the genome size and estimated DNA:RNA ratio/cell of each organism. No interference was observed with any of the tested substances. No inhibitors of amplification were observed in the APTIMA CT Assay.

Recovery

Escherichia coli, Gardnerella vaginalis, Lactobacillus acidophilus, Bacteroides ureolyticus, and Staphylococcus epidermidis (1 x 10° cells/assay) were added to samples containing the rRNA equivalent of approximately one CT IFU (5 fg). These additions did not interfere with the amplification and detection of CT rRNA using the APTIMA CT Assay.

Liquid Pap Specimen Stability Studies

Data to support the recommended shipping and storage conditions for PreservCyt Solution liquid Pap samples were generated with negative processed and unprocessed liquid Pap samples. For the unprocessed samples, four pools of PreservCyt Solution samples were tested after being stored in the Cytyc PreservCyt Solution vial. Each specimen pool was spiked with 1-10 IFU CT/assay, held at 2°C, 10°C, and 30°C, then tested at baseline and on days 5, 7, 8, 14, 18, 21, 25 and 36 depending on the storage

{6}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

temperature. All of the spiked samples were positive for CT at all times and temperatures.

For the processed samples, four pools of PreservCyt Solution samples were used to determine processed specimen stability at 2℃ to 30℃. Each negative sample pool was spiked with 1-10 IFU CT/assay, then tested at baseline. Prior to processing, the PreservCyt Solution samples were stored at 30℃ for seven (7) days to simulate the time lapse between sample collection, Pap processing and shipment to a microbiology testing lab. After seven days at 30°C, 1 mL aliquots of each pool were transferred to an APTIMA Specimen Transfer Tube and tested at baseline before being placed at 2°C, 10°C, and 30°C. The processed samples were then tested for 17 days stored at 30°C and 36 days stored at 2°C to 10°C. All of the spiked samples were positive for CT at all times and temperatures.

Data to support longer storage conditions were generated from four pools of negative processed PreservCyt Solution samples tested at below freezing temperatures. Each pool was spiked with 1-10 IFU CT/assay, then tested at baseline. Each pool was first placed at 30°C for 14 days and then stored at -20°C or -70°C over the course of 106 days. All of the spiked samples were positive for CT at all times and temperatures.

Precision

PreservCyt specimen within-laboratory precision with the ACT Assay was determined by spiking PreservCyt vials with 20 CT IFU per vial (0.1 IFU per reaction) and 100 CT IFU per vial (0.5 IFU per reaction). Vials containing 1,000 CT IFU per vial (5 IFU per reaction) and unspiked PreservCyt vials were tested as positive and negative controls. Ten vials spiked at each IFU level and ten unspiked vials were divided between two operators. The operators vortexed the vials and then transferred 14 aliquots (1.0 mL each) per vial into 14 APTIMA Transfer Tubes as per the APTIMA Specimen Transfer Kit package insert. The operators were blinded to the samples' titers. Each of the resulting Pap-STM samples was tested once in the ACT Assay. A total of five runs were

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the text "GEN-PROBE INCORPORATED" in a bold, sans-serif font. The text is arranged horizontally, with each word clearly legible. The word "GEN-PROBE" is hyphenated. A black arrow is pointing to the text.

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

performed over a five-day period for 140 results at each IFU level. The results are summarized in Table 2.

{8}------------------------------------------------

PTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimen N-PROBE INCORPORAT

Table 2: APTIMA CT Assay Within-Laboratory Precision Data for PreservCyt Using a 4-Member Precision Panel Containing
to 1000 IFU/20 mL of CT Cells.

PanelIFU/20mlIFU!్టర్లి మైనMeanRUWithin-OperatorBetween-DaOperator0etween-Total
Member PreservCyrxntlAgreedAgrit.×1000------------------------------------------------------------------------------------------------------------------------------------------------------------------------------x1000SD(%)CVx1000SD(%CV(x1000SD(%)CVx1000SD(%CV
07i1401401006501.7734.811.30.0546.98 4THE ARREAL COLLECTION OF CONSULT OF CHEARTH Aત્વન્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામના લોકોનો મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામCommender Company College of Children Artists of Article14.1
01000.514038a8 e6337.71054.76.600.0947.214.9417.6ને22.4
()100011 )140140------------------------------------------------------------------------------------------------------------------------------------------------------------------------------1006521.9a0013.9247 .3.8393.910215.7------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
11401401001.208N/A0NA0.4N/A0.9N/A
discordant results wore one nonative saand the starter of the starting the starting the starting the starting the starting the starting the state the state the state the state the state the state the state the staるクリリックなどParties on 1

Ingredients:

Organic Rye Flour, Water, Sea Salt.

secoder readlity result and 1 a quirce lesult
ote: Yariallity iss me lection may be which can occur if the variability do to hose factors is very snal. When his cocus wit
he

{9}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

PreservCyt Liquid Pap Specimen Clinical Study Results

A prospective multi-center clinical study was conducted to evaluate the use of the PreservCyt Solution (a component of the ThinPrep 2000 System) as an alternative medium for gynecological specimens for the detection of CT by the APTIMA CT Assay. One thousand six hundred forty-seven (1,647) symptomatic and asymptomatic female subjects attending OB/GYN, family planning, public health, women's, and STD clinics were evaluated in the clinical study. Of the 1,647 evaluable subjects, 1,288 were asymptomatic subjects and 359 were symptomatic subjects. Subjects were enrolled from sites with CT prevalence that ranged from 2.8% to 14.0%.

Two specimens were collected from each eligible subject: one PreservCyt Solution liquid Pap specimen and one endocervical swab specimen. PreservCyt Solution liquid Pap specimens were collected with the spatula/cyto-brush or a broom-like brush cervical sampling device. The distribution of cervical sampling devices is summarized in Table 3 by specimen collection site and overall.

PreservCyt Solution liquid Pap specimens were processed in accordance with the ThinPrep 2000 Processor Operator's Manual and APTIMA Specimen Transfer Kit Package Insert. After processing the PreservCyt Solution liquid Pap specimen with the ThinPrep 2000 Processor, the specimen was transferred into the APTIMA Specimen Transfer Kit for testing with the APTIMA CT Assay.

Sensitivity and specificity of the APTIMA CT Assay in PreservCvt Solution liguid Pap specimens were calculated by comparing results to a patient infected status algorithm. The algorithm included APTIMA Combo 2 Assay and APTIMA CT Assay results in endocervical swab specimens. Both reference NAATs were required to be positive to establish an infected patient status. At least one reference NAAT was required to be negative to establish a noninfected patient status. Table 4 summarizes the frequency of test outcomes for the two reference NAATs.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows a logo with a stylized letter 'G' on the left and the word 'G' followed by 'A' on the right. The 'G' on the left is a geometric shape with sharp angles, resembling a stylized letter. The 'G' and 'A' on the right are in a simple, sans-serif font, with 'G' being larger than 'A'.

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Table 5 shows the sensitivities and specificities of the APTIMA CT Assay by symptom status and overall. Overall sensitivity was 95.6% (86/90). In symptomatic and asymptomatic subjects, sensitivities were 96.7% (29/30) and 95.0% (57/60), respectively. Overall specificity was 98.8% (1539/1557). In symptomatic and asymptomatic subjects, specificities were 98.8% (325/329) and 98.9% (1214/1228), respectively.

Table 6 shows the sensitivities and specificities of the APTIMA CT Assay by specimen collection site and overall. Sensitivities ranged from 92.9% to 100%. Specificities ranged from 96.5% to 100%.

{11}------------------------------------------------

GEN-PROBE INCORPORATED
APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimen

able 3: Distribution of Cervical Sampling Device Used for PreservCyt Liquid Pap Specimen

Cervical Sampling Device UsedClinical Collection Site123456Total
Spatula/Cytobrush0124475287573641307
Broom-Type Device1000002400340

'able 4: PreservCyt Solution Liquid Pap Specimen Analysis for Patient Infected Statu

Endocervical SwabSymptom StatusAPTIMA CTPreservCytSolution result+/++/--/+-/-Sensitivity (%)(95% CI)Specificity (%)(95% CI)
Patient Infected StatusAPTIMA COMBO2 AssayAPTIMA CTAssaySymptomaticAsymptomaticTotalSymptomaticPositive2901396.7 (29/30)(82.8 – 99.9)98.8 (325/329)(96.9 – 99.7)
InfectedPositivePositive3060SymptomaticNegative133319
Non-InfectedNegativeNegative3221214SymptomaticTotal3034322
Non-InfectedNegativePositive412AsymptomaticPositive57011395.0 (57/60)(86.1 – 99.0)98.9 (1214/1228)(98.1 – 99.4)
Non-InfectedPositiveNegative32AsymptomaticNegative32111201
Total3591288AsymptomaticTotal602121214
AllPositive86021695.6 (86/90)(89.0 – 98.8)98.8 (1539/1557)(98.2 – 99.3)
AllNegative45141520
AllTotal905161536

{12}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimen N-PROBE INCORPORATE

'able 5: Sensitivity and Specificity of the APTIMA CT Assay Relative to Patient Infected Status by Symptom Stat nd Overall for PreservCyt Solution Liquid Pap Specimen

.

বিভ্যাশ
বিভাগ
বিভাগ
বিভাগ
বিভাগ বিভাগ
বিভাগ বিভাগ = = Noservical swab speciment in the APTIMA COMEO 2 Assaylbegives wab spainer resuli in beFTMA (min PTMA PTIMA PTMA
= Negative endocervical swab specimentes (PTIMA COMBO 2 As 1

{13}------------------------------------------------

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimen EN-PROBE INCORPORATE

'able 6: Sensitivity, Specificity and Predictive Values of the APTIMA CT Assay Relative to Patient Infected Status Clinical Site and Overall for PreservCyt Solution Liquid Pap Specimen

SiteAPTIMA CTPreservCytSolutionResult+++/--+--Prev(%)Sensitivity(%) (95% C.I.)Specificity (%(95% C.I.)PPV (%)NPV (%)
1Positive1401214.0100 (14/14)(76.8 – 100)96.5 (83/86)(90.1 – 99.3)82.4100
Negative00083
Total140185
2Positive40003.2100 (4/4)(39.8 – 100)100 (120/120)(97.0 – 100)100100
Negative002118
Total402118
3Positive290066.593.5 (29/31)(78.6 – 99.2)98.6 (438/444)(97.1 – 99.5)82.999.5
Negative202436
Total3102442
4Positive80042.8100 (8/8)(63.1 – 100)98.6 (275/279)(96.4 – 99.6)66.7100
Negative031271
Total831275
5Positive130034.792.9 (13/14)(66.1 – 99.8)98.9 (280/283)(96.9 – 99.8)81.399.6
Negative114275
Total1414278
6Positive180115.294.7 (18/19)(74.0 – 99.9)99.4 (343/345)(97.9 – 99.9)90.099.7
Negative115337
Total1916338
AllPositive8602165.595.6 (86/90)(89.0 – 98.8)98.8 (1539/1557)(98.2 – 99.3)82.799.7
Negative45141520
Total905161536

Page 14

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains a stylized logo. The logo features a bold, abstract shape that resembles a stylized letter or symbol. The symbol is composed of thick lines and angles, creating a sense of movement and dynamism. The logo is simple, modern, and visually striking.

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Prevalence

The prevalence of CT in patient populations depends on risk factors such as age, gender, the presence of symptoms, the type of clinic, and the test method. A summary of the prevalence of CT, for PreservCyt specimens, is shown in Table 7.

Table 7: Prevalence of C. trachomatis by Clinical Site and Overall as Determined by APTIMA CT Assay Results Using PreservCyt Solution Liquid Pap Specimens

Site% (#positive / #tested)
117.0(17/100)
743.2(4/124)
ﻟﻠﻘﺐ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ7.4(35/475)
44.2(12/287)
55.4(16/297)
్రీ5.5(20/364)
All6.3(104/1647)

Conclusions from Non-Clinical and Clinical Data

The non-clinical and clinical study results support the use of PreservCyt liquid Pap specimens collected and processed by the Cytyc ThinPrep 2000 Processor in the currently marketed GEN-PROBE APTIMA Assay for Chlamydia trachomatis. The currently marketed GEN-PROBE APTIMA Specimen Transfer Kit provides necessary materials and instructions to allow for the testing of PreservCyt liquid Pap specimens in the APTIMA CT Assay.

The results of the clinical study demonstrate reasonable evidence that when the APTIMA CT Assay and the APTIMA Specimen Transfer Kit are labeled as proposed, the APTIMA CT Assay continues to be safe and effective for its stated intended use.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows a logo. The logo is a stylized letter 'S' or a symbol resembling two interlocking 'C' shapes, presented in a bold, black color. The symbol is positioned to the left of some text, which is partially visible on the right side of the image.

APTIMA Assay for Chlamydia trachomtias – Expanded Indication: ThinPrep Specimens

Contraindications and Cautions

There are no contraindications or cautions.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design, with three stylized wing segments. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 25 2006

Brian Shea, RAC Regulatory Affairs Associate Gen-Probe Incorporated 10210 Genetic Center Drive San Diego, CA 92121

Re: K053446

Trade/Device Name: GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis Regulation Number: 21 CFR 866.3120 Regulation Name: Chlamydia serological reagents Regulatory Class: Class I Product Code: MKZ Dated: June 8, 2006 Received: June 9, 2006

Dear Mr. Shea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Spccial Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{17}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to Iogal|y marketed predicate device results in a classification for your device and thus, perroits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Sally, artm

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{18}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: _GEN-PROBE® APTIMA® Assay for Chlamydia trachomatis

Indications For Use:

The APTIMA Assay for Chlamydia trachomatis is a target amplification nucleic acid probe test that utilizes target capture for the in vitro qualitative detection of ribosomal RNA (rRNA) from Chlamydia trachomatis (CT) in clinician-collected endocervical, vaginal and male urethral swab specimens, patient-collected vaginal swab specimens. and female and male urine specimens. The assay is also intended for use with the testing of gynecological specimens collected in the PreservCyt® Solution and processed with the Cytyc ThinPrep® 2000 System. The assay may be used to test specimens from symptomatic and asymptomatic individuals to aid in the diagnosis of chlamydial urogenital disease.

1 Patient-collected vaginal swab specimens are an option for screening women when a pelvic exam is not otherwise indicated. The vaginal swab specimen collection kit is not for home use.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie Poole

Division Sign-Off

Division Sign-Off

Office of In Titre Diagnostic Device Evaluation and Surery

1001 K053446

§ 866.3120 Chlamydia serological reagents.

(a)
Identification. Chlamydia serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to chlamydia in serum. Additionally, some of these reagents consist of chlamydia antisera conjugated with a fluorescent dye used to identify chlamydia directly from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genusChlamydia and provides epidemiological information on these diseases. Chlamydia are the causative agents of psittacosis (a form of pneumonia), lymphogranuloma venereum (a venereal disease), and trachoma (a chronic disease of the eye and eyelid).(b)
Classification. Class I (general controls).