K Number
K090762

Validate with FDA (Live)

Device Name
LEDA
Manufacturer
Date Cleared
2009-08-28

(158 days)

Product Code
Regulation Number
878.4810
Age Range
All
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LEDA System with the Applicator LEDA SCR 585 yellow is indicated for the treatment of periorbital wrinkles and rhytides.

The LEDA System with the Applicator LEDA SCR 635 red is indicated for the treatment of superficial, benign vascular and pigmented lesions.

The LEDA System with the Applicators LEDA EPI 808 and LEDA EPI 980 is indicated for hair removal and permanent hair reduction defined as the stable, long-term reduction in hair counts at 6, 9, or 12 months following a treatment regime.

Device Description

The system is composed of a table-top base unit that includes the power supply, software and cooling system. Different applicators containing the light sources can be plugged into the base unit. The device is controlled by a touch-screen graphic user interface. There are three kinds of applicators each of which is available with several wavelengths. For details see the following section.

The LEDA is a device designed for dermatological use. Depending on the applicator selected wavelength in the range of 550-1,000 nm are produced either by Light Emitting Diodes (LED) or Diode Lasers. For each applicator and thus each wavelength there is a specific set of medical indications.

AI/ML Overview

The provided text is a 510(k) Summary for a medical device called LEDA. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study proving the device meets specific acceptance criteria with quantifiable performance metrics.

However, I can extract information related to the device's indications for use and the basis for its substantial equivalence claim.

Here's an analysis based on the provided text, structured to address your questions where possible:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not contain a table of acceptance criteria with quantifiable performance metrics that a study aimed to prove. Instead, the submission relies on demonstrating substantial equivalence to already cleared predicate devices. The "performance" being reported is the similarity in technological characteristics and intended use.

Therefore, I cannot create such a table directly from this document. The "criteria" are implicitly met if the device's technology and indications are sufficiently similar to approved predicates, implying comparable safety and efficacy.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a specific clinical study with a test set of patients or data. The submission relies on a comparison of device specifications and indications for use with established predicate devices.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

As there is no clinical test set described, there is no information about experts used to establish ground truth. The submission hinges on the regulatory agency's assessment of substantial equivalence based on technical specifications and established indications of predicate devices.

4. Adjudication Method for the Test Set

Not applicable, as no clinical test set is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or reported in this 510(k) summary. The submission's focus is on demonstrating substantial equivalence through technological comparison rather than a human-in-the-loop performance study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This refers to a device that is inherently an algorithm-only device, such as AI-based diagnostic software. The LEDA device is a physical light-based therapeutic device. Therefore, the concept of a "standalone" algorithm-only performance test is not applicable in this context. The performance is inherent to the physical device's operation.

7. The Type of Ground Truth Used

The "ground truth" implicitly used in this 510(k) submission is the established safety and effectiveness of the predicate devices based on their prior regulatory clearance. The LEDA device aims to demonstrate that it is sufficiently similar to these predicates that it shares their established safety and efficacy profiles. There is no new ground truth established via pathology, outcomes data, or expert consensus specifically for the LEDA device in this document.

8. The Sample Size for the Training Set

Not applicable, as this is a traditional medical device submission for a light-based therapy device, not an AI/machine learning device that would require a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set mentioned or implied.

{0}------------------------------------------------

90762

Section 5

AUG 2 8 2009

510(k) Summary

This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.87 and 807.92. Summary preparation date 04-10-08 [21 CFR 807.92(a)(1)].

Contact Information [21 CFR 807.92(a)(1)] A.

Quantel Derma GmbH

Am Wolfsmantel 46

D-91056 Erlangen

Tel: +49 9131 94088-40

Fax: +49 9131 94088-99

Contact person: Dr. Dietmar Fischer, Director R&D

Device Trade Name, Common Name, Classification [21 CFR B. 807.92(a)(2)]

Trade Name: LEDA

Device Common Name: Light-based Therapy Device for Dermatology

Classification Name: Laser Instrument, Surgical Powered (per 21 CFR 878.4810)

Product Code: GEX

Panel: Dermatology and Plastic Surgery

Device Classification: Class II

{1}------------------------------------------------

090762

Predicate Devices [21 CFR 807.92(a)(3)] C.

The LEDA device uses similar technology and has equivalent physical output characteristics as the following predicate devices:

  • For LEDA applicator SCR 585 yellow: .
    • GentleWaves LED Light Photomodulation Device (K031425) manufactured by BioScience L.L.C.
  • For LEDA applicator SCR 635 red: .
    • O Omnilux Revive (K030426) manufactured by Photo Therapeutics Inc.
    • VersaClear (K051259) manufactured by TheraLight Inc. ಂ

Summary: The LEDA LED applicators, SCR 585 yellow and SCR 635 red, are technically nearly identical ith the predicated devices. They all use LED technology and the physical output parameters (selectable power and dose) are nearly identical.

Power range: 4 mW/cm2 up to 120 mW/cm2.

Dose range: 0.1 J/cm2 up to 100 J/cm2.

  • For LEDA applicators EPI 808 and EPI 980: .
    • LightSheer (K973324, K982940, K001746) manufactured ം by Lumenis (formerly Star Medical / Coherent Star)
    • Quanta System's Diode Medical Laser Family (K072034) o manufactured by Quanta System SpA
    • MYDON (K040384) manufactured by Quantel Derma O GmbH (formerly Wavelight)

Summary: The LEDA LED applicators EPI 808 and EPI 980 use laser diode technology. Predicated devices are LightSheer (for 808 nm) and the Quanta System's Diode Medical Laser Family (for 808 nm and 980 nm). The device MYDON (a solid state laser) is an additional predicate for the LEDA EPI 980. The medical effect of a hair removal laser depends on the fluence and pulse duration. The technological differences do not raise new types of safety and efficacy issues because physical output parameters of all devices are absolutely comparable.

Fluence range: 6 - 60 J/cm2 (LEDA EPI 808)

9 - 90 J/cm2 (LEDA EPI 980)

15 - 100 J/cm2 (MYDON)

10 - 60 J/cm2 (LightSheer)

{2}------------------------------------------------

K090762

Pulse duration range: 6-60 ms (LEDA EPI 808 and EPI 980) 10 - 90 ms (MYDON)

5 - 100 ms (LightSheer)

Device Description [21 CFR 807.92(a)(4)] D.

The system is composed of a table-top base unit that includes the power supply, software and cooling system. Different applicators containing the light sources can be plugged into the base unit. The device is controlled by a touch-screen graphic user interface. There are three kinds of applicators each of which is available with several wavelengths. For details see the following section.

The LEDA is a device designed for dermatological use. Depending on the applicator selected wavelength in the range of 550-1,000 nm are produced either by Light Emitting Diodes (LED) or Diode Lasers. For each applicator and thus each wavelength there is a specific set of medical indications.

Device Specifications [21 CFR 807.92(a)(6)] E.

The LEDA emits light of different wavelength depending on the selected applicator. Depending on the applicator a subset of the following parameters can be adjusted by the user: pulse duration, power density, fluence, repetition rate, irradiation dose, pause between irradiations and pulse on / off times.

There are two different kinds of applicators:

  • . SCR applicators: These are screen (SCR) applicators with LED light sources that are mounted on a holder. The patient is placed 14 cm in front of the screen without direct contact to the device. An area of about 16 cm times 10 cm is irradiated. SCR applicators are available with the wavelengths 635 nm (red) and 585 nm (yellow). The user selects dose, power density, and pulse on / off times. After pressing the start button on the display of the base unit irradiation starts until the entered dose is reached or the user switches off the radiation.
  • . EPI applicators: These are handheld applicators intended for epilation (EPI) treatments. They contain a diode laser light source and a scanner that moves the treatment spot over skin. EPI applicators are available with wavelengths 808 nm or 980 nm. The

б

{3}------------------------------------------------

Ko 90762

scanned area is 1.2 cm times 5.0 cm and can be reduced to 1.2 cm times 1.2 cm by another distance holder. The user selects fluence, pulse duration, and pause between scans. When pressing a foot switch the scanning starts until the whole area has been scanned or the user releases the foot switch.

Indications for Use [21 CFR 807.92(a)(5)] F.

510(k) Number (if known): K090762

Device Name: LEDA

Indications for Use:

The LEDA System with the Applicator LEDA SCR 585 yellow is indicated for the treatment of periorbital wrinkles and rhytides.

The LEDA System with the Applicator LEDA SCR 635 red is indicated for the treatment of superficial, benign vascular and pigmented lesions.

The LEDA System with the Applicators LEDA EPI 808 and LEDA EPI 980 is indicated for hair removal and permanent hair reduction defined as the stable, long-term reduction in hair counts at 6, 9, or 12 months following a treatment regime.

Conclusion [21 CFR 807.92(b)(3)] G.

.

Technologically, the LEDA was found to be substantially equivalent to the following the currently cleared devices depending on the LEDA applicator:

  • For LEDA applicator SCR 585 yellow: .
    • GentleWaves LED Light Photomodulation Device o (K031425) manufactured by BioScience L.L.C.
  • For LEDA applicator SCR 635 red: .
    • o Omnilux Revive (K030426) manufactured by Photo Therapeutics Inc.
    • VersaClear (K051259) manufactured by TheraLight Inc. o
  • For LEDA applicators EPI 808 and EPI 980: ♥
    • LightSheer (K973324, K982940, K001746) manufactured o by Lumenis (formerly Star Medical / Coherent Star)

7

{4}------------------------------------------------

090762

  • Quanta System's Diode Medical Laser Family (K072034) manufactured by Quanta System SpA
  • MYDON (K040384) manufactured by Quantel Derma o GmbH (formerly Wavelight)

Thus, the risks and benefits for the LEDA are comparable to the predicate devices.

The indications for use are exactly the same as the previously cleared systems which are listed above.

We believe that there are no new questions of safety or efficacy raised by the introduction of the LEDA.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Quantel Derma GmbH % Underwriters Laboratories, Inc. Mr. Casey Conry 1285 Walt Whitman Road Melville, New York 11747

AUG 2 8 2009

Re: K090762

Trade/Device Name: LEDA Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX, ONF

Dated: August 7, 2009

Received: August 11, 2009

Dear Mr. Conry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2 - Mr. Casey Conry

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Mishranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Neil R.P. Ogden for
Mark N. Miller

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section 4

Indications for Use

510(k) Number (if known): K090762

Device Name: LEDA

Indications for Use:

The LEDA System with the Applicator LEDA SCR 585 yellow is indicated for the treatment of periorbital wrinkles and rhytides.

The LEDA System with the Applicator LEDA SCR 635 red is indicated for the treatment of superficial, benign vascular, and pigmented lesions.

The LEDA System with the Applicators LEDA EPI 808 and LEDA EPI 980 is indicated for hair removal and permanent hair reduction defined as the stable, long-term reduction in hair counts at 6, 9, or 12 months following a treatment regime.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND / OR .

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil R.P. Ogden Sorman

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090762

3

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.