K Number
K090705

Validate with FDA (Live)

Date Cleared
2009-10-09

(206 days)

Product Code
Regulation Number
888.3560
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ortho Development Balanced Knee® System is indicated for patients suffering from severe knee pain and disability. Specific indications include femoral, tibial and patellar replacement due to:

  1. Loss of joint configuration and joint function.
  2. Osteoarthritis of the knee joint.
  3. Rheumatoid arthritis of the knee joint.
  4. Post-traumatic arthritis of the knee joint.
  5. Valgus, varus, or flexion deformities.
  6. Revision procedures where other treatments or devices have failed.

The Balanced Knee® System is indicated in the salvage of previously failed surgical attempts where femoral bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.

Device Description

The Balanced Knee System Ultracongruent Tibial Insert may be used in conjunction with the Balanced Knee System CR femorals, standard and modular tibial trays, tibial augments, stems, and patellae. The Ultracongruent Tibial Insert has an anterior lip and tighter anterior curvature to stabilize the knee in the anterior direction and are snap fitted into the tibial trays and locked in place by mating features, the circumferential rim of the tray, and tabs on the anterior and posterior sides. The Ultraconqruent Tibial Insert is available in seven sizes to fit the majority of patients encountered. The sizes match each size of the Balanced Knee System tibial trays cleared under Premarket Notifications K994370, K020383, and K031201. The insert is symmetric, not left/right specific. It is supplied in thicknesses of 7, 8, 9, 10, 11, 12, 13. 14. 16. 18. and 20mm.

Materials: The device is manufactured from Compression Molded Ultra High Molecular Weight Polyethylene (UHMWPE, ASTM F-648).

Function: The system functions to provide restoration of function as a replacement for diseased and arthritic knees.

AI/ML Overview

The provided text does not contain specific acceptance criteria or detailed study results that would allow for the construction of a table with reported device performance or a comprehensive description of a study as requested. The document is a 510(k) summary for a medical device (Balanced Knee System® Ultracongruent Tibial Insert) and focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a detailed performance study with acceptance criteria.

The key information regarding performance testing is as follows:

  • Performance Testing: "The Food and Drug Administration have established no performance standards applicable to the Ultracongruent Tibial Insert, however biomechanical testing and analysis of the device was performed with results included as part of the submission. Clinical data and conclusion were not needed for this device."
  • Basis for substantial equivalency: "Testing met all acceptance criteria and verifies that the performance of the Ultracongruent Tibial Insert is substantially equivalent in design, function, material and intended use to the predicate devices."

Based on the available information, I can only provide a limited response:

Acceptance Criteria and Study Details for K090705: Balanced Knee System® Ultracongruent Tibial Insert

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly state specific numerical acceptance criteria or detail the reported device performance. It only states that "Testing met all acceptance criteria." Therefore, a table detailing these cannot be created from the provided text. The submission relied on biomechanical testing and analysis to demonstrate substantial equivalence, rather than specific performance standards established by the FDA for this device type.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions "biomechanical testing and analysis," but does not specify the sample size used for any test set or the data provenance (e.g., country of origin, retrospective/prospective).

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable as the submission states that "Clinical data and conclusion were not needed for this device." This implies that the evaluation for substantial equivalence did not rely on human expert interpretation of clinical outcomes or images to establish a 'ground truth' in the way a diagnostic AI device would. The evaluation focused on engineering and material characteristics.

4. Adjudication Method

This information is not applicable for the same reasons as point 3.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC comparative effectiveness study was not conducted as this is a knee implant device, not a diagnostic imaging or AI-assisted interpretation device that would involve human readers.

6. Standalone (Algorithm Only) Performance Study

A standalone performance study of an algorithm was not conducted as this is a physical medical device (knee implant), not a software algorithm.

7. Type of Ground Truth Used

The "ground truth" for the biomechanical testing would have been based on established engineering and material science principles and physical measurements/simulations of the device's performance against defined parameters (e.g., strength, durability, wear resistance) compared to predicate devices. It was not clinical, pathological, or outcomes data in the traditional sense for a diagnostic device.

8. Sample Size for the Training Set

This is not applicable as there is no mention of an algorithm or machine learning model that would require a training set.

9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no mention of an algorithm or machine learning model.

{0}------------------------------------------------

K090705

OCT ~ 9 2009

Image /page/0/Picture/2 description: The image shows the logo for "Ortho Development". The logo consists of a circular graphic to the left of the text. The graphic contains three curved lines that appear to be orbiting a central point. Below the words "Ortho Development" is the phrase "engineering simplicity" in a smaller font.

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS:

Submitter:

ﺎ ﻓ

Ortho Development Corporation 12187 South Business Park Drive Draper, Utah 84020

Contact Person:

Johanne Young Manger, Quality Assurance and Regulatory/Clinical Affairs (801) 619-3450 Telephone: Fax: (801) 619-8950

Date Prepared:

October 5, 2009

Name of the Device

Trade Name:Balanced Knee System® Ultracongruent Tibial Insert
Common Name:Knee Joint Prosthesis
Classification Name:Prosthesis, Knee patellofemorotibial, Semi-constrained
Cemented, Polymer/Metal/Polymer (888.3560)
Product Code:JWH

Predicate or legally marketed devices which are substantially equivalent:

  • Ortho Development Corporation Balanced Knee System CR or PS Insert � (K994770)
  • Depuy Sigma Deep Dish Insert (K033272) .
  • Depuy LCS Deep Dish Bearing (P830055) .
  • Encore Medical Foundation Ultracongruent Insert (K923277) .
  • Hayes Medical Concensus Ultracongruent Insert (K001456) .
  • OMNI Life Science Apex Knee Ultracongruent Insert (K001456) .
  • Smith & Nephew Genesis II Dished Insert (K951987) .
  • Smith & Nephew (Plus Orthopedics) TC-PLUS Solution Ultracongruent Insert . (Коореве)
  • Zimmer Insall-Bernstein PSCK Tibial Insert (K872379) .
  • Zimmer (Intermedics) Natural Knee II Ultracongruent and PS Insert . (K912663, K936159)

{1}------------------------------------------------

Description of the device:

The Balanced Knee System Ultracongruent Tibial Insert may be used in conjunction with the Balanced Knee System CR femorals, standard and modular tibial trays, tibial augments, stems, and patellae. The Ultracongruent Tibial Insert has an anterior lip and tighter anterior curvature to stabilize the knee in the anterior direction and are snap fitted into the tibial trays and locked in place by mating features, the circumferential rim of the tray, and tabs on the anterior and posterior sides. The Ultraconqruent Tibial Insert is available in seven sizes to fit the majority of patients encountered. The sizes match each size of the Balanced Knee System tibial trays cleared under Premarket Notifications K994370, K020383, and K031201. The insert is symmetric, not left/right specific. It is supplied in thicknesses of 7, 8, 9, 10, 11, 12, 13. 14. 16. 18. and 20mm.

Materials: The device is manufactured from Compression Molded Ultra High Molecular Weight Polyethylene (UHMWPE, ASTM F-648). Function:

The system functions to provide restoration of function as a replacement for diseased and arthritic knees.

Intended Use:

The Ortho Development Balanced Knee™ System is indicated for patients suffering from severe knee pain and disability. Specific indications include femoral, tibial and patellar replacement due to:

    1. Loss of joint configuration and joint function.
    1. Osteoarthritis of the knee joint.
    1. Rheumatoid arthritis of the knee joint.
    1. Post-traumatic arthritis of the knee joint.
    1. Valgus, varus, or flexion deformities.
    1. Revision procedures where other treatments or devices have failed.

The Balanced Knee® System is indicated in the salvage of previously failed surgical attempts where femoral bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.

Performance Testing:

The Food and Drug Administration have established no performance standards applicable to the Ultracongruent Tibial Insert, however biomechanical testing and analysis of the device was performed with results included as part of the submission. Clinical data and conclusion were not needed for this device.

Basis for substantial equivalency:

There are no significant differences between the components of the Ortho Development Ultracongruent Tibial Insert and other commercially available tibial inserts listed in the substantial equivalency section currently being marketed, which would adversely affect the use of the product. Testing met all acceptance criteria and verifies that the performance of the Ultracongruent Tibial Insert is substantially equivalent in design, function, material and intended use to the predicate devices.

INTENDED USE/INDICATIONS

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of several curved lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Ortho Development Corporation % Ms. Johanne Young Manager, Quality Assurance & Regulatory/Clinical Affairs 12187 South Business Park Drive Draper, Utah 84020

OCT - 9 2009

Re: K090705

Trade Name: Balanced Knee System Ultracongruent Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH

Dated: August 20, 2009 Received: August 21, 2009

Dear Ms. Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Johanne Young

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

fel

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4 Indications for Use Statement

Indications for Use of Device Form

1090705

510(k) Number (if known):

Device Name: Balanced Knee® System Ultracongruent Tibial Insert

Indications for Use

The Ortho Development Balanced Knee® System is indicated for patients suffering from severe knee pain and disability. Specific indications include femoral, tibial and patellar replacement due to:

    1. Loss of joint configuration and joint function.
    1. Osteoarthritis of the knee joint.
    1. Rheumatoid arthritis of the knee joint.
    1. Post-traumatic arthritis of the knee joint.
    1. Valgus, varus, or flexion deformities.
    1. Revision procedures where other treatments or devices have failed.

The Balanced Knee® System is indicated in the salvage of previously failed surgical attempts where femoral bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral (M/L) stability.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart C)

(Division Sign-Off) (Division of Surgical, Orthopedic, Divisionative Devices

510(k) Number

N/A