K Number
K041801

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-09-08

(68 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Expedium MIS Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous, posterior approach with MIS instrumentation, the Expedium MIS System screw components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Device Description

The Expedium MIS System consists of cannulated polyaxial pedicle screws and 5.5mm rods in various lengths. The Expedium MIS System also contains Class 1 manual surgical instruments to aid in the percutaneous approach, and cases that are considered exempt from premarket notification.

AI/ML Overview

The provided text is a 510(k) summary for the DePuy Spine, Inc. Expedium™ MIS Spine System. It describes the device, its intended use, materials, and states that "Performance data were submitted to characterize the notched rod with the cannulated polyaxial screws."

However, this document does not contain specific acceptance criteria, reported device performance (numerical or otherwise), details about study designs (sample sizes, provenance, ground truth establishment, expert qualifications, adjudication methods), or information regarding MRMC studies or standalone algorithm performance.

Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets them based solely on the provided text. The document only mentions that performance data were submitted but does not present them.

{0}------------------------------------------------

SEP - 8 2004

  • X. 510(k) Summary
    K0411801
    page 1 of 2
SUBMITTER:DePuy Spine, Inc.
325 Paramount Drive
Raynham, MA 02780

Jennifer Mooney CONTACT PERSON:

June 30, 2004 DATE PREPARED:

CLASSIFICATION NAME: Appliance, Fixation, Spinal Interlaminal Orthosis, Spinal Pedicle Fixation

Expedium™ MIS Spine System PROPRIETARY NAME:

Expedium™ Spine System (K033901), CD Horizon® PREDICATE DEVICES: Spinal System (K032033 and K032265)

The Expedium MIS System consists of cannulated DEVICE DESCRIPTION: polyaxial pedicle screws and 5.5mm rods in various lenaths.

The Expedium MIS System also contains Class 1 manual surgical instruments to aid in the percutaneous approach, and cases that are considered exempt from premarket notification.

The Expedium MIS Spine System is intended for INTENDED USE: noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous, posterior approach with MIS instrumentation, the Expedium MIS System screw components are intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and

{1}------------------------------------------------

K041801
page 2 of 2

radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Manufactured from ASTM F-136 implant grade MATERIALS: titanium alloy.

PERFORMANCE Performance data were submitted to characterize the DATA: notched rod with the cannulated polyaxial screws.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three curved lines representing its wings and body. The eagle is facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP = 8 2004

Ms. Jennifer Mooney Regulatory Affairs Associate Depuy Acromed, Inc. 325 Paramount Drive Raynham, Massachusetts 02767-0350

Re: K041801

Trade/Device Name: Expedium™ MIS Spine System Frace/Doviet Number: 21 CFR 888.3070, 21 CFR 888.3050, 21 CFR 888.3060 Regulation Name: Pedicle screw spinal system, Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis Regulatory Class: III Product Code: NKB, MNI, MNH, KWP, KWQ Dated: July 1, 2004 Received: July 2, 2004

Dear Ms. Mooney:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreation of the enactment date of the Medical Device Amendments, or to conniner or prior to may 20, 2011 - 11:11 accordance with the provisions of the Federal Food, Drug, de neces that have been require approval of a premarket approval application (PMA). and Cosmetter For (110) that as asvice, subject to the general controls provisions of the Act. The Tournaly, therefore, maines of the Act include requirements for annual registration, listing of general controls pro ractice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elassified (600 as controls. Existing major regulations affecting your device can may or sabject to back areas Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dr Hab Intact a and regulations administered by other Federal agencies. You must or any I odolar statutes and states and station but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Ms. Jennifer Mooney

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin mailioning of substantial equivalence of your device to a legally premits that in the PDA mians of cation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for Jour as 1) 594-4692. Also, please note the regulation entitled, p Colliact the Office of Companies market notification" (21CFR Part 807.97). You may obtain Misolanding by reference to promanton in the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark N. Milliken

Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use IV.

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Expedium MIS Spine System Device Name:

Indications For Use:

The Expedium MIS Spine System is intended for noncervical The Expedium MIO Opine Oyetting for the following indications: pedicle fixation and nomported as back pain of discogenic degenerative also discuss of the disc confirmed by history and ongin with degenoration of en of clickly trauma (i.e., fracture or radiographio studios); spen: ; curvatures (i.e., scoliosis, kyphosis, ulsiocation); Spinal Stonocio; Jan thereosis; and failed previous fusion in skeletally mature patients.

When used in a percutaneous, posterior approach with MIS Instrumentation, the Expedium MIS System screw components are intentation, the Expodian Mis Uyearam Mis Uynand nonpedicle fixation for interior for nonomations: degenerative disc disease (defined as the following manogenic origin with degeneration of the disc back pain of diosege and radiographic studies); spondylolisthesis; conlinned by nietery arr dislocation); spinal stenosis; curvatures tradina (i.o., maolaro of alend/or lordosis); tumor; pseudoarthrosis; and failed previous fusion in skeletally mature patients.

Mark N. Milkeron

Division of General, Restorative, and Neurological Devices

510(k) Number K041801

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR Over-The-Counter Use: _____________________________________________________________________________________________________________________________________________________ Prescription Use: _ (Per 21 CFR 801.109)

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.