K Number
K090301

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-04-14

(432 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BinaxNOW® PBP2a Test is a qualitative, in vitro immunochromatographic assay for the rapid detection of penicillin-binding protein 2a (PBP2a) present in methicillinresistant Staphylococcus aureus (MRSA). The test is performed directly on blood culture samples positive for S. aureus.

The BinaxNOW® PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. Subculturing positive blood cultures is necessary to recover organisms for susceptibility testing or epidemiological typing.

Device Description

The BinaxNOW® PBP2a Test is a rapid immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect the PBP2a protein directly from blood cultures which have been identified as being positive for S. aureus. These antibodies and a control antibody are immobilized onto a test strip as two distinct lines and combined with other reagents/pads. This test strip is mounted inside a cardboard, book-shaped hinged test device.

Specimens are aliquots from blood cultures which have been identified as positive for Staphylococcus aureus. After the sample is prepared, it is added to the sample pad at the top of the test strip and the device is closed. Results are read at 10 minutes.

AI/ML Overview

Acceptance Criteria and Study for BinaxNOW® PBP2a Test

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the BinaxNOW® PBP2a Test, as derived from the provided document, are presented in the table below, alongside the reported device performance.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Clinical Performance (Positive Agreement)
Cefoxitin (30 µg) disc diffusionHigh (e.g., >95%)96.9% (62/64) (89.3 - 99.1% CI)
Oxacillin (1 µg) disc diffusionHigh (e.g., >95%)96.5% (55/57) (88.1 - 99.0% CI)
Automated Antimicrobial SusceptibilityHigh (e.g., >95%)97.6% (41/42) (87.7 - 99.6% CI)
Clinical Performance (Negative Agreement)
Cefoxitin (30 µg) disc diffusionHigh (e.g., >99%)100.0% (67/67) (94.6 - 100.0% CI)
Oxacillin (1 µg) disc diffusionHigh (e.g., >99%)100.0% (58/58) (93.8 - 100.0% CI)
Automated Antimicrobial SusceptibilityHigh (e.g., >99%)100.0% (29/29) (88.3 - 100.0% CI)
Overall Clinical PerformanceHigh (e.g., >95% for positive, >99% for negative)97.1% positive agreement, 100.0% negative agreement (for overall 199 samples)
Analytical Reactivity (MRSA strains)All tested strains positiveAll listed MRSA strains (NARSA and ATCC) tested positive.
Analytical Specificity (MSSA strains)All tested strains negativeAll listed MSSA strains tested negative.
Analytical Specificity (Other Staphylococcal strains)All tested strains negative (except for expected cross-reactivity)All tested strains negative except Staphylococcus sciuri.
Analytical Specificity (Non-Staphylococcal strains)All tested strains negative (except for expected cross-reactivity)All tested strains negative except Cryptococcus neoformans.
Interfering SubstancesNo interferenceAll 20 listed substances produced appropriate results.
Analytical Sensitivity (Limit of Detection)Specific CFU/mL value expected2.5 x 10^7 cells/mL (turbidity 0.03) / 2.36 x 10^7 CFU/mL (from ATCC BAA44)
Reproducibility100% agreement expected100% (599/599) agreement with expected results. No significant differences.

Note: The document does not explicitly state numerical acceptance criteria. The "Implied Acceptance Criteria" are inferred from the demonstrated performance and the context of a 510(k) submission, where high agreement with established methods is required for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Clinical Performance Test Set: 199 S. aureus samples.
  • Data Provenance: Multi-center clinical study conducted in 2008-09 at four geographically diverse hospital laboratories within the United States. The study was prospective in nature, as samples were "evaluated in the BinaxNOW® PBP2a Test and compared to standard methods."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states that the ground truth for the clinical performance was established by "standard methods used routinely by the laboratories: Cefoxitin (30 µg) disc diffusion, Oxacillin (1 µg) disc diffusion, and automated Minimum Inhibitory Concentration (MIC) Systems." It also mentions "Individual samples were evaluated by multiple laboratory methods, and in all cases there was 100% agreement between the reference methods."

While specific "experts" for establishing ground truth are not explicitly named or quantified, the ground truth was determined by multiple, routinely used laboratory methods performed by qualified laboratory personnel within the four hospital laboratories. The qualifications of these individuals would typically include clinical microbiologists or medical technologists with experience in performing and interpreting these standard susceptibility tests. The document implies that the "experts" were the laboratory staff routinely performing these validated reference methods.

4. Adjudication Method for the Test Set

The document notes that "Individual samples were evaluated by multiple laboratory methods, and in all cases there was 100% agreement between the reference methods." This indicates that if there were any discrepancies between results from the Cefoxitin disc diffusion, Oxacillin disc diffusion, and automated MIC systems, they were resolved or did not occur. The fact that only three clinical samples (1.5%) produced "discrepant results" overall (compared to the BinaxNOW test) suggests that there was an established reference standard, and these three were just discordant with the device, not necessarily with the reference methods themselves. No specific formal adjudication method (e.g., 2+1, 3+1) for resolving conflicts between the reference methods is described, as 100% agreement among them was reported. Discrepancies between the device and the reference methods were simply noted.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focused on the diagnostic accuracy of the device against standard laboratory methods, not on how human readers' performance improved with or without AI assistance. The BinaxNOW® PBP2a Test is a rapid immunochromatographic assay, meaning it is a diagnostic test kit that provides a direct result, not an AI system designed to assist human readers.

6. Standalone Performance

Yes, a standalone performance study was done. The entire clinical performance study section describes the algorithm's (the BinaxNOW® PBP2a Test's) performance "alone" against established reference methods. The results presented in the table are the direct output of the BinaxNOW® PBP2a Test without human-in-the-loop intervention in the interpretation of the enzymatic reaction as positive or negative. While a human reads the test, the test itself provides a direct biochemical result that is interpreted.

7. Type of Ground Truth Used

The type of ground truth used for the clinical performance study was expert consensus (of established laboratory methods). Specifically, it relied on the results from:

  • Cefoxitin (30 µg) disc diffusion
  • Oxacillin (1 µg) disc diffusion
  • Automated Minimum Inhibitory Concentration (MIC) Systems

The document states there was "100% agreement between the reference methods" for individual samples, indicating these methods collectively formed the definitive truth.

8. Sample Size for the Training Set

The document does not specify a sample size for a "training set." The BinaxNOW® PBP2a Test is described as an immunochromatographic assay, which is a chemical reaction-based test, not a machine learning or AI model that typically requires a separate training set. The various analytical studies (reactivity, specificity, sensitivity, interfering substances) might be considered part of the development and "training" (calibration/validation) of the assay itself, but these are based on known reference strains and substances rather than patient data used in a typical machine learning training set.

9. How the Ground Truth for the Training Set Was Established

As noted above, there is no explicit "training set" in the context of a machine learning model. However, if we consider the development and validation of the assay, the "ground truth" for analytical studies was established using known, well-characterized bacterial strains from sources like the American Type Culture Collection (ATCC) and the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA). For example, MRSA strains from these collections were expected to test positive, and MSSA, other Staphylococcal, and non-Staphylococcal strains were expected to test negative. These strains have established classifications regarding their methicillin resistance.

{0}------------------------------------------------

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K090301

SUBMITTER

Binax, Inc. 10 Southgate Road Scarborough, Maine 04074 (207) 730-5733 (Office) (207) 730-5717 (FAX) Establishment Registration Number: 1221359

APR 1 | 2010

CONTACT PERSON

Suzanne M. Vogel suzanne.vogel@invmed.com (email)

DATE PREPARED April 12th, 2010

TRADE NAME BinaxNOW® PBP2a Test

COMMON NAME

BinaxNOW® PBP2a Test, BinaxNOW® PBP2a, Binax NOW® PBP2a Test, Binax NOW® PBP2a, NOW® PBP2a Test, NOW® PBP2a

CLASSIFICATION NAME

System, Test. Genotypic Detection, Resistant Markers, Staphylococcus Colonies (MYI) (per 21 CFR 866.1640)

PREDICATE DEVICES

Mueller Hinton Agar w/4% NaCl w/Antibiotics (Remel) K850291 PBP22 Latex Agglutination Test (Oxoid) K011710

DEVICE DESCRIPTION

The BinaxNOW® PBP2a Test is a rapid immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect the PBP2a protein directly from blood cultures which have been identified as being positive for S. aureus. These antibodies and a control antibody are immobilized onto a test strip as two distinct lines and combined with other reagents/pads. This test strip is mounted inside a cardboard, book-shaped hinged test device.

Specimens are aliquots from blood cultures which have been identified as positive for Staphylococcus aureus. After the sample is prepared, it is added to the sample pad at the top of the test strip and the device is closed. Results are read at 10 minutes.

Binax NOW® PBP2a Test 510(k) Notification K090301 Rev. 04/12/10

{1}------------------------------------------------

INTENDED USE

The BinaxNOW® PBP2a Test is a qualitative, in vitro immunochromatographic assay for the rapid detection of penicillin-binding protein 2a (PBP2a) present in methicillinresistant Staphylococcus aureus (MRSA). The test is performed directly on blood culture samples positive for S. aureus.

The BinaxNOW® PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. Subculturing positive blood cultures is necessary to recover organisms for susceptibility testing or epidemiological typing.

Performance Characteristics

Clinical Performance

The clinical performance of the BinaxNOW® PBP2a Test was established in a multi-center clinical study conducted in 2008-09 at four geographically diverse hospital laboratories within the ાટે.

A total of 199 S. aureus samples were evaluated in the BinaxNOW® PBP2a Test and compared to standard methods used routinely by the laboratories: Cefoxitin (30 ug) disc diffusion, Oxacillin (1 ug) disc diffusion, and automated Minimum Inhibitory Concentration (MIC) Systems. Individual samples were evaluated by multiple laboratory methods, and in all cases there was 100% agreement between the reference methods. Among the clinical samples tested, , only three clinical samples (3/199 or 1.5%) produced discrepant results. Overall, the BinaxNOW PBP2a assay identified 97.1% of the specimens positive for MRSA and 100.0% of the specimens negative for MRSA.

The table below presents BinaxNOW® PBP2a Test performance by reference method. Because each sample was tested on more than one reference method, there are more observations in this table (n=317) than the total number of samples (n=199).

Reference MethodPositiveAgreement(95% CI)NegativeAgreement(95% CI)
Cefoxitin (30 µg) disc diffusion96.9% (62/64)(89.3 - 99.1%)100.0% (67/67)(94.6 - 100.0%)
Oxacillin (1 µg) disc diffusion96.5% (55/57)(88.1 - 99.0%)100.0% (58/58)(93.8 - 100.0%)
Automated AntimicrobialSusceptibility Test System97.6% (41/42)(87.7 - 99.6%)100.0% (29/29)(88.3 - 100.0%)

BinaxNOW® PBP2a Test Performance vs. Reference Methods

Expected Values

In the clinical evaluation of the BinaxNOW® PBP2a Test conducted in 2008-09 at four geographically diverse hospital laboratories within the US, the overall expected rate of PBP2a (MRSA) was 51.3% (102/199), and among the four site populations the expected positive rate ranged from 48.3% to 61.5%.

Binax NOW® PBP2a Test 510(k) Notification K090301 Rev. 04/12/10

{2}------------------------------------------------

Analytical Reactivity

The human pathogenic Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA) and American Type Culture Collection (ATCC) methicillin-resistant Staphylococcus aureus (MRSA) strains listed below tested positive in the BinaxNOW® PBP2a Test as expected.

Methicillin-Resistant Staphylococcus aureus (MRSA)

Methicillin-Resistant Staphylococcus aureus (MRSA)
BacteriumBacterium
Staphylococcus aureus ATCC 33591Staphylococcus aureus NRS643(USA300)
Staphylococcus aureus ATCC 33592Staphylococcus aureus NRS647(USA300)
Staphylococcus aureus ATCC 43300Staphylococcus aureus NRS657(USA300)
Staphylococcus aureus ATCC 49476Staphylococcus aureus NRS658(USA100)
Staphylococcus aureus ATCC 51153Staphylococcus aureus NRS659(USA300)
Staphylococcus aureus ATCC 700698Staphylococcus aureus NRS660(USA100)
Staphylococcus aureus ATCC 700699Staphylococcus aureus NRS661(USA100)
Staphylococcus aureus ATCC 700789Staphylococcus aureus NRS667(USA300)
Staphylococcus aureus ATCC BAA1026Staphylococcus aureus NRS670(USA100)
Staphylococcus aureus ATCC BAA38Staphylococcus aureus NRS671(USA100)
Staphylococcus aureus ATCC BAA39Staphylococcus aureus NRS672(USA100)
Staphylococcus aureus ATCC BAA41Staphylococcus aureus NRS673(USA100)
Staphylococcus aureus ATCC BAA43Staphylococcus aureus NRS674(USA100)
Staphylococcus aureus ATCC BAA44Staphylococcus aureus NRS679(USA100)
Staphylococcus aureus NRS123(USA 400)Staphylococcus aureus NRS687(USA300)
Staphylococcus aureus NRS172Staphylococcus aureus NRS688(USA300)
Staphylococcus aureus NRS192Staphylococcus aureus NRS693(USA300)
Staphylococcus aureus NRS193Staphylococcus aureus NRS694(USA300)
Staphylococcus aureus NRS194Staphylococcus aureus NRS697(USA100)
Staphylococcus aureus NRS22(USA 600)Staphylococcus aureus NRS699(USA100)
Staphylococcus aureus NRS241Staphylococcus aureus NRS710(USA100)
Staphylococcus aureus NRS245Staphylococcus aureus NRS711(USA100)
Staphylococcus aureus NRS248Staphylococcus aureus NRS716(USA300)
Staphylococcus aureus NRS249Staphylococcus aureus NRS717(USA100)
Staphylococcus aureus NRS382(USA 100)Staphylococcus aureus NRS721(USA100)
Staphylococcus aureus NRS383(USA 200)Staphylococcus aureus NRS725(USA300)
Staphylococcus aureus NRS384(USA 300)Staphylococcus aureus NRS732(USA300)
Staphylococcus aureus NRS385(USA 500)Staphylococcus aureus NRS733(USA300)
Staphylococcus aureus NRS386(USA 700)Staphylococcus aureus NRS736(USA300)
Staphylococcus aureus NRS387(USA 800)Staphylococcus aureus NRS739(USA300)

Analytical Specificity (Cross-Reactivity)

To determine the analytical specificity of the BinaxNOW® PBP2a Test, methicillin-susceptible Staphylococcus aureus (MSSA), Staphylococcal strains (other than S. aureus) and non-Staphylococcal strains were tested. All strains tested negative in the BinaxNOW® test except Cryptococcus neoformans and Staphylococcus sciuri.

Binax NOW® PBP2a Test 510(k) Notification K090301 Rev. 04/12/10

{3}------------------------------------------------

Methicillin-Sensitive Staphylococcus aureus (MSSA)

BacteriumBacterium
Staphylococcus aureus ATCC 33862Staphylococcus aureus NRS167
Staphylococcus aureus ATCC 13150Staphylococcus aureus NRS168
Staphylococcus aureus ATCC 11632Staphylococcus aureus NRS169
Staphylococcus aureus ATCC 14776Staphylococcus aureus NRS170
Staphylococcus aureus ATCC 6538PStaphylococcus aureus NRS171
Staphylococcus aureus ATCC 29213Staphylococcus aureus NRS173
Staphylococcus aureus ATCC BAA977Staphylococcus aureus NRS174
Staphylococcus aureus NRS164Staphylococcus aureus NRS175
Staphylococcus aureus NRS165Staphylococcus aureus NRS176
Staphylococcus aureus NRS166Staphylococcus aureus NRS177
Staphylococcus aureus ATCC 9144Staphylococcus aureus Lafferty
Staphylococcus aureus ATCC 51740Staphylococcus aureus ATCC 31153
Staphylococcus aureus ATCC 29737Staphylococcus aureus ATCC 12600
Staphylococcus aureus ATCC 15564Staphylococcus aureus ATCC14993
Staphylococcus aureus ATCC 14775Staphylococcus aureus ATCC 33862
Staphylococcus aureus ATCC 25923 .Staphylococcus aureus subsp. Anaerobius ATCC 35844

Staphylococcal Strains (other than S. aureus)

BacteriumATCC#BacteriumATCC#
Staphylococcus simulans27851Staphylococcus equorum43958
Staphylococcus warneri49454Staphylococcus lentus700403
Staphylococcus lugdunensis43809Staphylococcus hyicus11249
Staphylococcus sciuri29601Staphylococcus carnosus51365
Staphylococcus saprophyticus15305Staphylococcus capitis35661
Staphylococcus schleiferi43808Staphylococcus arlettae43957
Staphylococcus haemolyticus29970Staphylococcus piscifermentans51136
Staphylococcus kloosii43959Staphylococcus hominis27844
Staphylococcus cohnii29972Staphylococcus caprae51548
Staphylococcus xylosus49148Staphylococcus pasteuri51128
Staphylococcus succinus700337Staphylococcus chromogenes43764
Staphylococcus vitulinus51162Staphylococcus lutrae700373
Staphylococcus pulvereri51698Staphylococcus muscae49910
Staphylococcus intermedius29663Staphylococcus felis49168
Staphylococcus gallinarum700401Staphylococcus auricularis33753
Staphylococcus epidermidis14990Staphylococcus delphini49171
Staphylococcus epidermidis12228Staphylococcus saccharolyticus14953
Staphylococcus epidermidis35984Staphylococcus schleiferi subspcoagulans49545
Staphylococcus fleurettiiBAA-274Staphylococcuspseudintermedius49444
BacteriumATCC#BacteriumATCC#
Acinetobacter calcoaceticus51432Macrococcus caseolyticus35662
Aerococcus viridans10400Macrococcus equipercicus51831
Aeromonas hydrophila35654Micrococcus luteus27141
Bacillus cereus11778Moraxella catarrhalis25238
Bacillus subtilis6633Morganella morganii25830-T
Bacteroides fragilis23745Neisseria gonorrhoeae49226
Beta Strep Group F12392Neisseria meningitidis, serogroup A13077
Burkholderia cepacia25416-TNeisseria sicca9913
Candida parapsilosis90018Parvimonas micra (formerlyPeptostreptococcus micros)33270
Candida krusei14243Pasturella multocida51687
Cellulomonas turbata (formerlyOerskovia)25835Pediococcus acidilactici12697
Citrobacter freundii8090Peptostreptococcus anaerobius27337
Citrobacter koseri25408Planococcus citreus14404
Clostridium perfringens3624Proteus mirabilis7002
Corynebacterium xerosus7711Proteus vulgaris33420
Corynebacterium amycolatum49368Providencia stuartii49809
Corynebacterium diphtheriae13812Pseudomonas aeruginosa15442
Corynebacterium glutamicum13869Pseudomonas fluorescens49271
Corynebacterium jeikeium43734Pseudomonas putida49128
Corynebacteriumpseudodiphtheriticum10700-TRhodococcus equi10146
Corynebacterium urealyticum43042Rothis mucilaginosa (Stomatococcus)25296
Cryptococcus neoformans14116Salmonella adelaide10718
Escherichia coli (ESBL producer)ClinicalisolateSerratia marcescens13880
Enterobacter aerogenes35029Stenotrophomonas maltophilia12637-T
Enterobacter cloacae49141Streptococcus agalactiae, group B13813
Enterococcus avium49465Streptococcus anginosis (milleri)33397
Enterococcus casseliflavus12817Streptococcus dysgalactiae, group C(strain C74)12388
Enterococcus durans49135Streptococcus dysgalactiae, group G(strain Lf D166B)12394
Enterococcus faecalis49474Streptococcus intermedius (milleri)27335
Enterococcus faecium12952Streptococcus mitis49456
Enterococcus gallinarum49608Streptococcus mutans25175
Enterococcus hirae10541Streptococcus pasteurianus (bovis)49133
Enterococcus mundtii43187Streptococcus pneumoniae33400
Enterococcus raffinosus49464Streptococcus pneumoniae49136
Escherichia coli10798Streptococcus pneumoniaeSSI-1
Finegoldia magna (formerlyPeptostreptococcus magnus)15794Streptococcus pneumoniaeSSI-10A
Gemella bergeri700627Streptococcus pneumoniae6301
Haemophilus influenzae49247Streptococcus pneumoniae33938
Haemophilus parainfluenzae33392-TStreptococcus pneumoniae49619
Klebsiella oxytoca49131Streptococcus pneumoniae51937
Klebsiella pneumoniae49472Streptococcus pneumoniaeSSI-14
Klebsiella pneumoniae (ESBL prodand KPC pos)ClinicalisolateStreptococcus pneumoniaeSSI-7F
Kocuria kristinaeBAA752Streptococcus pneumoniae51938
Kytococcus chroeter13884Streptococcus pyogenes, group A12384
Lactobacillus casei393Streptococcus salivarius13419
Lactococcus garvieae49157YeastATCC#
Leuconostoc mesenteroides10877Candida albicans60193
Listeria monocytogenes, serotype 4b19115Candida glabrata66032
Candida tropicalis750

Non- Staphylococcal Strains

Binax NOW® PBP2a Test
510(k) Notification K090301 Rev. 04/12/10

Confidential

/

{4}------------------------------------------------

Binax NOW® PBP2a Test
5 IO(k) Notification K090301
Rev. 04/12/10

.

.

Confidential

Page 5 of 7

{5}------------------------------------------------

Interfering Substances:

The 20 potentially interfering substances listed below produced appropriate results in the BinaxNOW® PBP2a test.

Anti-Inflammatory DrugsTestConcentrationEndogenous Blood ComponentsTestConcentration
Acetaminophen1324 µmol/LHemoglobin2 g/L
Acetylsalicyclic acid3.62 mmol/LTriglyceride sera37 mmol/L
Ibuprofen2425 µmol/LConjugated bilirubin342 µmol/L
AntibioticsTestConcentrationUnconjugated bilirubin342 µmol/L
Amoxicillin206 µmol/Lγ- globulin120g/L
Cephalexin337 µmol/LAnti-coagulantTestConcentration
Chloramphenicol155 µmol/LSodium Polyanetholesulfonate (SPS)1%
Ciprofloxacin30.2 µmol/L
Erythromycin81.6 µmol/L
Gentamicin21 µmol/L
Tetracycline34 µmol/L
Sulfisoxazole1.12 mmol/L
Sulfamethoxazole1.58 mmol/L
Trimethoprim138 µmol/L
Vancomycin69 µmol/L

Analytical Sensitivity:

The analytical limit of detection of the BinaxNOW® PBP2a Test in ATCC strain BAA44 at a turbidity level of 0.03 is 2.5 x 10' cells/mL, and the equivalent concentration in CFU/mL is 2.36 x 10'.

495 4 77LA LA BELL MALE MARK L. MILL AN A BELL AN A BELANDING MARK, CAL BAAN A & AN ENGLAND, AND AN AND CHARACT AND ANDREA CARD CHANAL
------1 - 1 - 4 - 1 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10 - 10

Binax NOW® PBP2a Test 510(k) Notification K090301 Rev. 04/12/10

Confidential

.

{6}------------------------------------------------

$3.33x10^7$20/20100
$2.5x10^7$19/2095
$4.94x10^6$13/2065
$2.19x10^6$3/2015
Whole Blood0/200

Reproducibility Study:

A study of the BinaxNOW® PBP2a Test was conducted at 3 separate sites using panels of blind coded specimens containing negative and positive samples. Participants tested each sample twice on 5 different days. There was 100% (599/599) agreement with expected test results, with no significant differences within run (replicates tested by one operator), between run (5 different days), between sites (3 sites), or between operators (6 operators).

Signed
Suzanne M. Vogel, MPH

Suzanne M. Vogel, MPH Clinical Affairs Binax, Inc.

Date 4/12/10

Binax NOW® PBP2a Test 5 l 0(k) Notification K090301 Rev. 04/12/10

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing human services, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The text is in all caps and is black.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center-WO66-G609 Silver Spring, MD 20993-0002

Binax. Inc c/o Ms. Suzanne Vogel Department of Clinical Affairs Inverness Medical 10 Southgate Rd. Scarborough, Maine 04074

APR 1 4 2010

Re: K090301

Trade/Device Name: BinaxNOW® PBP2a Test Regulation Number: 21 CFR& 866.1640 Regulation Name: Methicillin Resistant Staphylococcus aureus (MRSA) Regulatory Class: Class II Product Code: MYI Dated: March 29, 2010 Received: March 31, 2010

Dear Ms. Vogel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{8}------------------------------------------------

Page 2 - Suzanne Vogel

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket" notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Sally attym

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K090301

Device Name: BinaxNOW® PBP2a Test

Indications for Use:

The BinaxNOW® PBP2a Test is a qualitative, in vitro immunochromatographic assay for the rapid detection of penicillin-binding protein 2a (PBP2a) present in methicillinresistant Staphylococcus aureus (MRSA). The test is performed directly on blood culture samples positive for S. aureus.

The BinaxNOW® PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. Subculturing positive blood cultures is necessary to recover organisms for susceptibility testing or epidemiological typing.

Prescription Use 2 (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Freddie W. Poole

Vision Sign-Off

Signon Sign-Off

Binax NOW® PBP2a Test 510(k) Notification K090301 Rev. 3/29/10

Confidential

Office of In Viterage againstic Device Evaluation and Safety

$10(k) K09030/

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).