Search Results
Found 2 results
510(k) Data Aggregation
(432 days)
The BinaxNOW® PBP2a Test is a qualitative, in vitro immunochromatographic assay for the rapid detection of penicillin-binding protein 2a (PBP2a) present in methicillinresistant Staphylococcus aureus (MRSA). The test is performed directly on blood culture samples positive for S. aureus.
The BinaxNOW® PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. Subculturing positive blood cultures is necessary to recover organisms for susceptibility testing or epidemiological typing.
The BinaxNOW® PBP2a Test is a rapid immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect the PBP2a protein directly from blood cultures which have been identified as being positive for S. aureus. These antibodies and a control antibody are immobilized onto a test strip as two distinct lines and combined with other reagents/pads. This test strip is mounted inside a cardboard, book-shaped hinged test device.
Specimens are aliquots from blood cultures which have been identified as positive for Staphylococcus aureus. After the sample is prepared, it is added to the sample pad at the top of the test strip and the device is closed. Results are read at 10 minutes.
Acceptance Criteria and Study for BinaxNOW® PBP2a Test
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the BinaxNOW® PBP2a Test, as derived from the provided document, are presented in the table below, alongside the reported device performance.
Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Clinical Performance (Positive Agreement) | ||
Cefoxitin (30 µg) disc diffusion | High (e.g., >95%) | 96.9% (62/64) (89.3 - 99.1% CI) |
Oxacillin (1 µg) disc diffusion | High (e.g., >95%) | 96.5% (55/57) (88.1 - 99.0% CI) |
Automated Antimicrobial Susceptibility | High (e.g., >95%) | 97.6% (41/42) (87.7 - 99.6% CI) |
Clinical Performance (Negative Agreement) | ||
Cefoxitin (30 µg) disc diffusion | High (e.g., >99%) | 100.0% (67/67) (94.6 - 100.0% CI) |
Oxacillin (1 µg) disc diffusion | High (e.g., >99%) | 100.0% (58/58) (93.8 - 100.0% CI) |
Automated Antimicrobial Susceptibility | High (e.g., >99%) | 100.0% (29/29) (88.3 - 100.0% CI) |
Overall Clinical Performance | High (e.g., >95% for positive, >99% for negative) | 97.1% positive agreement, 100.0% negative agreement (for overall 199 samples) |
Analytical Reactivity (MRSA strains) | All tested strains positive | All listed MRSA strains (NARSA and ATCC) tested positive. |
Analytical Specificity (MSSA strains) | All tested strains negative | All listed MSSA strains tested negative. |
Analytical Specificity (Other Staphylococcal strains) | All tested strains negative (except for expected cross-reactivity) | All tested strains negative except Staphylococcus sciuri. |
Analytical Specificity (Non-Staphylococcal strains) | All tested strains negative (except for expected cross-reactivity) | All tested strains negative except Cryptococcus neoformans. |
Interfering Substances | No interference | All 20 listed substances produced appropriate results. |
Analytical Sensitivity (Limit of Detection) | Specific CFU/mL value expected | 2.5 x 10^7 cells/mL (turbidity 0.03) / 2.36 x 10^7 CFU/mL (from ATCC BAA44) |
Reproducibility | 100% agreement expected | 100% (599/599) agreement with expected results. No significant differences. |
Note: The document does not explicitly state numerical acceptance criteria. The "Implied Acceptance Criteria" are inferred from the demonstrated performance and the context of a 510(k) submission, where high agreement with established methods is required for substantial equivalence.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Clinical Performance Test Set: 199 S. aureus samples.
- Data Provenance: Multi-center clinical study conducted in 2008-09 at four geographically diverse hospital laboratories within the United States. The study was prospective in nature, as samples were "evaluated in the BinaxNOW® PBP2a Test and compared to standard methods."
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document states that the ground truth for the clinical performance was established by "standard methods used routinely by the laboratories: Cefoxitin (30 µg) disc diffusion, Oxacillin (1 µg) disc diffusion, and automated Minimum Inhibitory Concentration (MIC) Systems." It also mentions "Individual samples were evaluated by multiple laboratory methods, and in all cases there was 100% agreement between the reference methods."
While specific "experts" for establishing ground truth are not explicitly named or quantified, the ground truth was determined by multiple, routinely used laboratory methods performed by qualified laboratory personnel within the four hospital laboratories. The qualifications of these individuals would typically include clinical microbiologists or medical technologists with experience in performing and interpreting these standard susceptibility tests. The document implies that the "experts" were the laboratory staff routinely performing these validated reference methods.
4. Adjudication Method for the Test Set
The document notes that "Individual samples were evaluated by multiple laboratory methods, and in all cases there was 100% agreement between the reference methods." This indicates that if there were any discrepancies between results from the Cefoxitin disc diffusion, Oxacillin disc diffusion, and automated MIC systems, they were resolved or did not occur. The fact that only three clinical samples (1.5%) produced "discrepant results" overall (compared to the BinaxNOW test) suggests that there was an established reference standard, and these three were just discordant with the device, not necessarily with the reference methods themselves. No specific formal adjudication method (e.g., 2+1, 3+1) for resolving conflicts between the reference methods is described, as 100% agreement among them was reported. Discrepancies between the device and the reference methods were simply noted.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study focused on the diagnostic accuracy of the device against standard laboratory methods, not on how human readers' performance improved with or without AI assistance. The BinaxNOW® PBP2a Test is a rapid immunochromatographic assay, meaning it is a diagnostic test kit that provides a direct result, not an AI system designed to assist human readers.
6. Standalone Performance
Yes, a standalone performance study was done. The entire clinical performance study section describes the algorithm's (the BinaxNOW® PBP2a Test's) performance "alone" against established reference methods. The results presented in the table are the direct output of the BinaxNOW® PBP2a Test without human-in-the-loop intervention in the interpretation of the enzymatic reaction as positive or negative. While a human reads the test, the test itself provides a direct biochemical result that is interpreted.
7. Type of Ground Truth Used
The type of ground truth used for the clinical performance study was expert consensus (of established laboratory methods). Specifically, it relied on the results from:
- Cefoxitin (30 µg) disc diffusion
- Oxacillin (1 µg) disc diffusion
- Automated Minimum Inhibitory Concentration (MIC) Systems
The document states there was "100% agreement between the reference methods" for individual samples, indicating these methods collectively formed the definitive truth.
8. Sample Size for the Training Set
The document does not specify a sample size for a "training set." The BinaxNOW® PBP2a Test is described as an immunochromatographic assay, which is a chemical reaction-based test, not a machine learning or AI model that typically requires a separate training set. The various analytical studies (reactivity, specificity, sensitivity, interfering substances) might be considered part of the development and "training" (calibration/validation) of the assay itself, but these are based on known reference strains and substances rather than patient data used in a typical machine learning training set.
9. How the Ground Truth for the Training Set Was Established
As noted above, there is no explicit "training set" in the context of a machine learning model. However, if we consider the development and validation of the assay, the "ground truth" for analytical studies was established using known, well-characterized bacterial strains from sources like the American Type Culture Collection (ATCC) and the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA). For example, MRSA strains from these collections were expected to test positive, and MSSA, other Staphylococcal, and non-Staphylococcal strains were expected to test negative. These strains have established classifications regarding their methicillin resistance.
Ask a specific question about this device
(252 days)
The Clearview® Exact PBP2a Test is a qualitative, in vitro immunochromatographic assay for the detection of penicillin-binding protein 2a (PBP2a) in isolates identified as Staphylococcus aureus, as an aid in detecting methicillin-resistant Staphylococcus aureus (MRSA). The Clearview® Exact PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections.
The Clearview® Exact PBP2a Test is a rapid immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect the PBP2a protein directly from bacterial isolates. These antibodies and a control antibody are immobilized onto a nitrocellulose membrane as two distinct lines and combined with a sample pad, a blue conjugate pad, and an absorption pad to form a test strip.
Isolates are sampled directly from the culture plate and eluted into an assay tube containing Reagent 1. Reagent 2 is then added and the dipstick is placed in the assay tube. Results are read visually at 5 minutes.
The Clearview® Exact PBP2a Test is a rapid immunochromatographic assay for detecting penicillin-binding protein 2a (PBP2a) in Staphylococcus aureus isolates, aiding in the detection of methicillin-resistant Staphylococcus aureus (MRSA).
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria. However, it reports sensitivity and specificity performance values for the device compared to a reference method. We can infer that the reported performance values were considered acceptable for regulatory clearance.
Performance Metric | Reported Device Performance (Tryptic Soy Agar with 5% sheep blood) | Reported Device Performance (Columbia Agar with 5% sheep blood) | Reported Device Performance (Mueller Hinton with 1 µg oxacillin induction) |
---|---|---|---|
Sensitivity | 98.1% (95.2-99.3% CI) | 99.0% (96.6-99.7% CI) | 99.5% (97.4-99.9% CI) |
Specificity | 98.8% (96.5-99.6% CI) | 98.8% (96.5-99.6% CI) | 98.8% (96.5-99.6% CI) |
2. Sample Size and Data Provenance for the Test Set
- Sample Size: A total of 457 S. aureus samples were evaluated in the clinical performance study.
- Data Provenance: The study was a multicenter clinical study conducted in 2009 at three geographically-diverse laboratories. The analytical performance section also mentions bacterial strains obtained from the Network on Antimicrobial Resistance in Staphylococcus aureus (NARSA), American Type Culture Collection (ATCC), and a collection from the Department of Infectious Disease Epidemiology of the Imperial College in London, England. This indicates a mix of strains from reference collections and clinical isolates, and at least some data provenance from England in addition to the diverse US laboratories. The study appears to be retrospective in the sense that existing S. aureus samples were evaluated by the new device.
3. Number of Experts and Qualifications for Ground Truth
The document does not mention the use of experts to establish ground truth for the test set.
4. Adjudication Method for the Test Set
The document does not mention an adjudication method for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study compares the device's performance to a standard method (cefoxitin disk diffusion), not to human readers' performance with and without AI assistance.
6. Standalone Performance Study
Yes, a standalone study was done. The clinical performance study directly evaluated the performance of the Clearview® Exact PBP2a Test against the reference method without human intervention in the interpretation of the device's results, as it is a visual read test that relies on the device's output. The reproducibility study also tested the device in a standalone manner.
7. Type of Ground Truth Used
The ground truth used for the clinical performance study was cefoxitin (30 ug) disk diffusion, interpreted according to CLSI (Clinical and Laboratory Standards Institute) standards. This is a recognized phenotypic method for determining methicillin resistance in S. aureus.
8. Sample Size for the Training Set
The document does not explicitly mention a dedicated "training set" or its size for the development of the Clearview® Exact PBP2a Test. The analytical performance section mentions that 162 MRSA strains and 112 MSSA strains were tested for analytical reactivity and specificity, which might represent samples used during later stages of development or validation, but it's not explicitly labeled as a training set.
9. How Ground Truth for the Training Set Was Established
Since a distinct training set is not explicitly defined, the method for establishing its ground truth is not detailed. However, for the strains used in analytical performance (162 MRSA and 112 MSSA), it is implied that their methicillin-resistant/sensitive status was known, likely established through standard microbiological identification and susceptibility testing methods (e.g., CLSI guidelines, reference lab testing) given their origin from reputable collections (NARSA, ATCC, Imperial College).
Ask a specific question about this device
Page 1 of 1