K Number
K090286

Validate with FDA (Live)

Date Cleared
2009-03-23

(46 days)

Product Code
Regulation Number
870.4450
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is indicated for use to provide hemostasis of the femoral vascular access site during and following catheterization or cannulation procedures.

Device Description

The PressureMate™ Compression Assist Handle mates with the SuperComfort™ Discs (K040615) to provide an alternative to the use of mechanical clamping systems or direct hand holding pressure to obtain hemostasis following fernoral vascular catheterization procedures.

The PressureMate™ handle itself is an aluminum handle with stainless steel stem. The stem ends in a female connection which mates to the male connector located on the SuperComfort Discs. The handle can not be used without a Disc.

The PressureMate™ comes in two designs: i) symmetrical, with both ends symmetrically tapering down at each end and centrally located stem or ii) asymmetrical, with one end tapered smaller than the other end and the stem slightly off center. The handle designs, round with tapered ends, enable better fit to an operator's hand.

Use of the handle and disc by a medical practitioner avoids prolonged direct contact with bodily fluids, and alleviates bio-mechanical stress which may occur during traditional direct digital compression of the femoral artery post-cardiac catheterization.

AI/ML Overview

The provided text describes the PressureMate™ Compression Assist Handle, a device designed to assist with hemostasis following femoral vascular catheterization procedures. Here's an analysis of the acceptance criteria and the study information available:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Withstand holding force of 35 poundsThe device is designed and has been tested to withstand a holding force of 35 pounds.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" in the context of clinical or performance testing on subjects/patients. The reported performance refers to a device design and material strength test.

  • Test Set Description: The test involved evaluating the device's ability to withstand a specific mechanical force.
  • Sample Size: Not explicitly stated for this mechanical test, but implies multiple units were tested to ensure consistency.
  • Data Provenance: Not applicable as it's a mechanical test, not clinical data from patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. The "ground truth" for the mechanical test is the physical capacity of the device to withstand 35 pounds, which is a manufacturing specification, not an expert assessment.

4. Adjudication Method for the Test Set

Not applicable. This was a mechanical engineering test, not a clinical study requiring expert adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a physical handle for manual compression, not an AI-assisted diagnostic tool or an imaging device that would typically involve MRMC studies.

6. Standalone (i.e. algorithm only without human-in-the loop performance) Study

No, a standalone study focusing on an algorithm's performance was not done. This device is a manual assist tool, not an algorithm or software.

7. Type of Ground Truth Used

The "ground truth" for the reported performance criterion (withstanding 35 pounds) is a physical/mechanical specification and measurement. It's a measure of the device's structural integrity and strength, rather than a clinical outcome or expert judgment.

8. Sample Size for the Training Set

Not applicable. As this device is a physical, manual compression assist handle and not an AI/software product, there is no "training set" in the context of machine learning or algorithms.

9. How the Ground Truth for the Training Set Was Established

Not applicable for the same reason as above. There is no training set for this type of medical device.

{0}------------------------------------------------

KØ9Ø286

MAR 2 3 2009

31 December 2008

510(k) Summary

Company Contact Information

Advanced Vascular Dynamics 1910 NW 23td Place, Portland, Oregon 97210 Herbert J. Semler, President, Advanced Vascular Dynamics Phone: (800) 525-2555 Fax: (503)223-8585

Name of Device

Trade name - PressureMate™ Compression Assist Handle Common name -- Femoral Access Compression Device Procode/Classification name - DXC/Clamp, Vascular Regulation Number - 870.4450

Predicate Device(s)

Compass Compression Assist Handle (K053398) SuperComfort™ Discs (K040615)

Indication for Use

The device is indicated for use to provide hemostasis of the femoral vascular access site during and following catheterization or cannulation procedures.

Device Description

The PressureMate™ Compression Assist Handle mates with the SuperComfort™ Discs (K040615) to provide an alternative to the use of mechanical clamping systems or direct hand holding pressure to obtain hemostasis following fernoral vascular catheterization procedures.

The PressureMate™ handle itself is an aluminum handle with stainless steel stem. The stem ends in a female connection which mates to the male connector located on the SuperComfort Discs. The handle can not be used without a Disc.

The PressureMate™ comes in two designs: i) symmetrical, with both ends symmetrically tapering down at each end and centrally located stem or ii) asymmetrical, with one end tapered smaller than the other end and the stem slightly off center. The handle designs, round with tapered ends, enable better fit to an operator's hand.

Use of the handle and disc by a medical practitioner avoids prolonged direct contact with bodily fluids, and alleviates bio-mechanical stress which may occur during traditional direct digital compression of the femoral artery post-cardiac catheterization.

Performance Data

The device is designed and has been tested to withstand a holding force of 35 pounds.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. The text is in all capital letters.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 3 2009

Advanced Vascular Dynamics c/o Mr. Matthew Semler 1910 NW 23rd Place Portland, OR 97210

Re: K090286

PressureMate™ Manual Femoral Access Compression Device Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Compression Device Regulatory Class: Class II Product Code: DXC Dated: January 30, 2009 Received: February 5, 2009

Dear Mr. Semler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Matthew Semler

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Qimia D: Vulumen

Bram D. Zuckerman, M.D. Division Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known): K090286

Device Name: PressureMate™ Manual Femoral Access Compression Device_

Indications for Use:

This device is indicated for use to provide hemostasis of the femoral vascular access site during and following catheterization or cannulation procedures.

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

una R. Valner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number_ko90286_

§ 870.4450 Vascular clamp.

(a)
Identification. A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.(b)
Classification. Class II (performance standards).