K Number
K083384

Validate with FDA (Live)

Date Cleared
2010-01-05

(414 days)

Product Code
Regulation Number
870.1200
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Biosensors flow-directed Thermodilution catheters are designed for use in critical care patients to measure cardiac output, right atrium, pulmonary artery and pulmonary capillary wedge pressures; continuously monitor pulmonary artery temperature, sample blood, intravenously administer drugs and solution and measure cardiac output via the cardiac output computers which interface with 14 kilo-ohms catheters. Different models are available for use in pediatric and adult patients.

The tips of the catheters are mounted with a latex balloon which when inflated, protects the heart tissues from the product's tips during insertion; utilizes blood flow to direct the catheter tip through the right ventricle into the pulmonary artery.

Device Description

Blosensors flow-directed Thermodilution catheter is fabricated from extruded tubing, connected to a hub carrying 4 or 5 lumens. The total length of catheter is either of 90 or 110 cm. The extensions include a proximal, a distal, a thermistor and an inflation extension, where the distal end of the inflation extension ends with a tip mounted with a balloon.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details based on the provided text, formatted as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
FunctionalityISO 10555-1Meets ISO 10555-1 standard
BiocompatibilityISO 10993-1 Part 1Biocompatible and safe
USP 27 <88>Biocompatible and safe

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size for the test set of the Thermodilution Catheter models with PU catheter tubing. However, it indicates "performance tests were performed on the Thermodilution Catheter models with PU catheter tubing". The provenance of the data is not specified (e.g., country of origin, retrospective or prospective), but given the applicant is Biosensors International Pte Ltd based in Singapore, the testing would likely have originated from or been managed by them.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document describes performance testing against standards (ISO 10555-1, ISO 10993-1, USP 27 <88>), not a study requiring expert-established ground truth in the context of diagnostic accuracy. Therefore, information regarding the number and qualifications of experts for ground truth establishment is not applicable to this type of performance assessment.

4. Adjudication Method for the Test Set

Not applicable as it's not a study involving human adjudication of results. The performance was assessed against established technical and biological standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document describes engineering and biocompatibility testing for a medical device (thermodilution catheter), not a diagnostic algorithm where human readers would improve with AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is not an AI-based diagnostic algorithm but a physical medical device. The "standalone" performance refers to the device's inherent functionality and biocompatibility as tested against standards.

7. Type of Ground Truth Used

The "ground truth" for the device's performance was established by recognized international and national standards:

  • ISO 10555-1: For functionality and integrity of intravascular catheters.
  • ISO 10993-1 Part 1: Evaluation & Testing: For biocompatibility.
  • USP 27 <88> Biological Reactivity Tests, In Vivo: For biocompatibility.

8. Sample Size for the Training Set

Not applicable. This is a medical device, not a machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for Biosensors International. The logo consists of a stylized symbol above the company name. The symbol is made up of three horizontal lines that are connected by vertical lines, forming a shape that resembles the letter 'B'. The text 'BIOSENSORS INTERNATIONAL' is written in a bold, sans-serif font below the symbol.

SPECIAL 510K

Paga

510K SUMMARY [As required by section 807.92(c)]

JAN - 5 2010

Revised on 2 December 2009

1. Applicant Information

Company Address:Biosensors International Pte Ltd21 Kallang Avenue #07-165/171Singapore 339412
Registration Number:8043983
Contact Person:Moses Tham
Telephone:(65) 6411 0506
Facsimile:(65) 6298 6242

2. Device Identification

Trade Name:Thermodilution Catheter and Accessories
Classification Name:Diagnostic Intravascular Catheter

3. Predicate Devices

Thermodilution Catheter (K911710)

4. Device Description

Blosensors flow-directed Thermodilution catheter is fabricated from extruded tubing, connected to a hub carrying 4 or 5 lumens. The total length of catheter is either of 90 or 110 cm. The extensions include a proximal, a distal, a thermistor and an inflation extension, where the distal end of the inflation extension ends with a tip mounted with a balloon.

કે. Intended Use

Blosensors flow-directed Thermodilution catheter is designed for use in the critical care patients to measure right atrium, pulmonary artery and pulmonary capillary wedge pressure; continuously monitor pulmonary artery temperature, sample blood, intravenously administer drugs and solution and measure cardiac output via the cardiac output computer which interfaces with 14 kilo-ohms catheters. Different models are available for use in pediatric and adult patients.

The tips of the catheters are mounted with a latex balloon which when inflated, protects the heart tissues from the product's tips during insertion; utilizes blood flow to direct the catheter tip through the right ventricle into the pulmonary artery.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo for Biosensors International. The logo consists of a stylized letter "M" stacked on top of the words "BIOSENSORS INTERNATIONAL". The letter "M" is made up of three horizontal bars connected by vertical lines. The text is in a bold, sans-serif font.

6. Summary of Technological Characteristics of Device in relation to Predicate Devices(s)

Biosensors Thermodilution catheter product family included in 510K (K911710) is currently licensed for heparin-coated them of addin and intelled in 5700 (KB11710) is currently
goolication involves the including of additions of additions to of the application involves the inclusion of additional models of Themodilution Catheler where the catheter is designed with the non-coated PU catheter tubing.

Comparison was made for the above models against those that have already been licensed in K911710. The additional cather models have the same intended use and design construction with equivalent components as the predicated catheter models which models with the new mated with a ballon. The safety and efficiency of the subject device models with the new material type have been assessed and discussed further in Section 7 below.

7. Assessment of Performance Data used to justify Substantial Equivalence

Test Summary, In-vitro

Models with non-coated PU catheter tubing

Inclusion of the category of models is to provide an alternative to physician who prefers a different tautness of the catheter tubing.

Functionality performance tests were performed on the Thermodilution Catheter models with PU catheter tubing to evaluate the performance and integrity of the catheter in accordance to ISO 1055-1. Based on the per results which are of the acceptance cifleria, the design construction of the catheter with PU tubing meets the ISO10555-1 standard.

Test Summary, Biocompatibility

Biocompatibility tests were performed in accordance to ISO10993-1 Part 1: Evaluation & Testing and USP 27 <88> Biological Reactivity Tests, In Vivo and was proven that the themodilution catheter, with PU catheter tubing, is blocompatible and safe for its intended usa.

8. Conclusion:

The Functionality Tests and Blocompatibility tests have demonstrated that the Biosensors Themodilution catheter that is substantially equivaled the predicated devices is safe and effective for its intended use.

{2}------------------------------------------------

PREMARKET NOTIFICATIONTHERMODILUTION CATHETER
BIOSENSORS INTERNATIONALSPECIAL 510KPage3 of 3

510K Substantial Equivalence Decision-Making Tree

Image /page/2/Figure/2 description: This image is a flowchart titled "510(k) "Substantial Equivalence" Decision-Making Process (Overview)". The flowchart outlines the steps for determining if a new device is substantially equivalent to a marketed device. The process begins with comparing the new device to a marketed device and proceeds through a series of questions about intended use, technological characteristics, and performance equivalence, ultimately leading to a determination of substantial equivalence or a need for more information.

Decision to substantial Equivalence is based on 510(k) Mamorandum #486-3. 510(k). Questions in the decision path are evaluated as follow:

Q1. Does New Device have same Intended use?

Yes. The intended use of the naw range of models is the same as the models currently listed.

Q2. Does New Device have Technological Characteristics that raise new types of Safety or effectiveness questions?

No. New models are of the same design construction is the models currently listed.

Q3. Does descriptive or Performance information demonstrate equivalence?

Yes. The product description and the intended use on the instruction for Use between the subject models and predicated models demonstrated the essential equivalence. The physical and performance characteristics evaluated in Table 2 in the Section 10 - Description to the Modification to the device also showed the equivalence subject models and predicated models.

Conclusion:

As according to the decision path, the new models of Thermodiation Catheter can be submitted using substantial Equivalence.

{3}------------------------------------------------

Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles facing to the right. The figure is abstract and appears to be in motion.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

MAR 2 3 2010

Biosensors International Pte., Ltd. c/o Mr. Moses Tham Quality Assurance Manager 21 Kallang Avenue #07-165/171 Singapore, 339412

Re: K083384

Trade/Device Name: Thermodilution Catheter and Accessories Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II (two) Product Code: DQO Dated: December 2, 2009 Received: December 7, 2009

Dear Mr. Tham:

This letter corrects our substantially equivalent letter of January 5, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Moses Tham

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): J. K083384

THERMODILUTION CATHETER AND ACCESSORIES Device Name: . _ _ .

· Indications for Use:

Biosensors flow-directed Thermodilution catheters are designed for use in critical care patients to measure cardiac output, right atrium, pulmonary artery and pulmonary capillary wedge pressures; continuously monitor pulmonary artery temperature, sample blood, intravenously administer drugs and solution and measure cardiac output via the cardiac output computers which interface with 14 kilo-ohms catheters. Different models are available for use in pediatric and adult patients.

The tips of the catheters are mounted with a latex balloon which when inflated, protects the heart tissues from the product's tips during insertion; utilizes blood flow to direct the catheter tip through the right ventricle into the pulmonary artery.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

W.M.F.

(Division Sign-Off) (Division of Cardiovascular Devices

510(k) Number.

Page \ of |

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).