K Number
K082676

Validate with FDA (Live)

Date Cleared
2008-12-18

(94 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.

Device Description

The product is a portable suction device that may promote wound healing when used with accessory wound sealing kits.

AI/ML Overview

The provided 510(k) summary for the Genadyne A4 Wound Vacuum System (K082676) describes a substantial equivalence determination to a predicate device, the BlueSky VISTA™ Wound Vacuum System (K061367). This type of submission does not typically involve extensive clinical studies with acceptance criteria as one might find for novel devices or PMAs. Instead, it focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device.

Here's an analysis of the available information regarding acceptance criteria and the supporting study, based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) generally compares the new device's technical specifications and intended use against the predicate device to argue substantial equivalence, rather than setting and meeting specific performance acceptance criteria for a new clinical outcome. The acceptance criteria essentially boil down to aligning with or being acceptably similar to the predicate device’s established characteristics.

Acceptance Criterion (Implicit)Reported Device Performance (Genadyne A4)Comparison to Predicate Device (BlueSky VISTA™)
Suction Capacity (Functional equivalence)5 liters per minutePredicate: 8 liters per minute (Difference noted, but not explicitly stated as non-equivalent for safety/effectiveness)
Max Vacuum (Functional equivalence)230 mmHgPredicate: 200 mmHg (Difference noted, but not explicitly stated as non-equivalent for safety/effectiveness)
Power Requirements (Technical equivalence)24 VDC, 1APredicate: 100-240 V AC, 60 Hz (Difference noted, but not explicitly stated as non-equivalent for safety/effectiveness)
Battery Type (Component equivalence)Ni-MHPredicate: Ni-MH (Equivalent)
Dimensions and Weight (Physical characteristics)200 x 180 x 80 mm / 1.36 KgPredicate: 260 x 250 x 106 mm / 1.9 Kg (Genadyne A4 is lighter; noted as a difference that "does not affect safety and effectiveness")
Canisters (Accessory compatibility)800 ml disposable with hydrophobic shut off filterPredicate: 250 ml or 800 ml disposable (Similar capacity, added filter in Genadyne A4)
Reusable (Operational characteristic)YesPredicate: Yes (Equivalent)
Sterile (Operational characteristic)Non SterilePredicate: Non Sterile (Equivalent)
Accessory Kit Components (Functional equivalence)Non-adherent gauze, Anti-Microbial gauze, Transparent film dressing, Silicone flat drainPredicate: Non-adherent gauze, Anti-Microbial gauze, Transparent film dressing, Silicone flat drain (Equivalent)
Indications for Use (Clinical equivalence)"indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris."Predicate: "indicated for patients who would benefit from a suction device particularly as the device may promote wound healing." (Stated as identical)
Contraindications (Safety equivalence)Necrotic tissue, Untreated osteomyelitis, Malignancy (with exception to enhance quality of life), Untreated malnutrition, Exposed arteries, veins, or organsPredicate: Necrotic tissue, Untreated osteomyelitis, Malignancy (with exception to enhance quality of life), Untreated malnutrition, Exposed arteries, veins, or organs (Equivalent)
Precautions (Safety equivalence)Patients receiving anticoagulant therapy, suffering from difficult hemostasis, untreated for malnutrition, non-complaint or combativePredicate: Patients receiving anticoagulant therapy, suffering from difficult hemostasis, untreated for malnutrition, non-complaint or combative (Equivalent)
Compliance (Electrical Safety)UL 60601-1, CAN/CSA C22.2 No. 601-1-M90, IEC 60601-1-2, FCC part 15 Class B, EN 55011, IEC 61000-4-2, IEC 61000-4-3Predicate: UL 60601-1, IEC 0601-1-2, CAN/CSA C22.2 No. 601.1 (Similar standards, Genadyne lists additional EMC/EMI tests)
Storage/Transport and Operation Conditions (Environmental Tolerance)Similar temperature and humidity rangesSimilar temperature and humidity ranges (Differences noted, but within acceptable operating ranges for device type)

The study proving the device meets these (implicit) acceptance criteria is a bench testing and comparative analysis study. The 510(k) states: "Performance testing has been completed to demonstrate the safe and effective use of the Genadyne A4 Wound Vacuum System for the intended use." and "Performance testing and device comparison demonstrates that the subject device is substantially equivalent to the predicate device, and is safe and effective for the intended use." This is primarily a comparison of technical specifications and claimed indications for use.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly mentioned in terms of patient numbers or specific datasets. For a mechanical device like a wound vacuum system, "test set" typically refers to the physical devices undergoing various engineering and performance tests, rather than a clinical patient cohort. The testing appears to be focused on the device itself and its components (e.g., suction capacity, maximum vacuum, battery life, compliance with electrical safety standards, kit component performance).
  • Data Provenance: The nature of the testing (bench testing, comparison of specifications) indicates that the "data" would be generated from laboratory tests of the device itself. There's no mention of country of origin for any clinical data, as this is not a clinical study in the traditional sense, but a demonstration of substantial equivalence based on technical and performance characteristics.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable for this type of submission. "Ground truth" in this context refers to established engineering standards, existing scientific understanding of wound therapy mechanisms, and the performance of the predicate device. No expert panel was involved in creating a "ground truth" for a test set of data in a clinical trial model.

4. Adjudication Method for the Test Set

Not applicable. As noted above, this is a technical equivalence submission, not a clinical study requiring adjudication of outcomes or diagnoses.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device for wound therapy (a physical pump), not an imaging or diagnostic AI device that involves human "readers" or AI assistance in interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithmic or AI device. It's a pump that provides negative pressure. The "standalone" performance here refers to the device's ability to meet its technical specifications (e.g., suction rate, maximum vacuum) independent of a human operator, which would have been covered by the performance testing mentioned.

7. The Type of Ground Truth Used

The "ground truth" for this substantial equivalence submission is primarily:

  • Predicate Device Performance: The established safety and effectiveness profile and technical specifications of the BlueSky VISTA™ Wound Vacuum System (K061367).
  • Engineering Standards: Compliance with recognized international and national standards for medical devices (e.g., UL 60601-1, IEC 60601-1-2, FCC part 15 Class B, EN 55011, IEC 61000-4-x series).
  • Scientific Principles: The understanding of how negative pressure wound therapy works and the role of suction in removing exudates, infectious material, and tissue debris.

8. The Sample Size for the Training Set

Not applicable. There is no AI/algorithm being trained described in this 510(k) submission.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no AI/algorithm or training set.

{0}------------------------------------------------

K082676 page 1/8

510k Summary

General Information

    1. Applicant:
      Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 (t) 516.487.8787 (f) 516.487.7878 www.genadyne.com

DEC 1 8 2008

  • Mr. Chien-Ming GOH (Andrew) 2. Contact Person: Vice President Genadyne Biotechnologies Inc. 65 Watermill Lane, Great Neck, NY 11021 Tel: 516-487-8787 Fax: 516-487-7878 Andrew@gcnadyne.com

3. Trade/Proprietary Name Including Model Number of Device:

Genadyne A4 Wound Vacuum System

4. Common Name or Classification Name (21 CFR Part 807.87) of Device:

Powered Suction Pump (21 CFR 878.4780, Product Code BTA)

5. Class in which Device has been placed:

Class II

6. Reason for Premarket Notification:

Introduction of a new device that is substantially equivalent to a legally marketed device.

7. Identification of Legally Marketed Device Which We Claim Substantial Equivalence (Predicate Device):

BlueSky VISTA"M Wound Vacuum System, K061367 (Aug 10, 2006)

{1}------------------------------------------------

K082676 Page 2/8

8. Compliance with Requirements of the Federal FD&C Act:

The General and Rostorative Device Panel (DGRD) has classified this device as Class II, 21 CFR 878.4780

Product Code: BTA

9. Kit Certification and Information:

This device kit has been tested and is substantially equivalent to the kit of the predicate device.

Please refer to Attachment E for the performance testing on the dressing kit for both the Genadyne A4 Wound Vacuum System and the predicate device. Please do refer to Attachment F for the kit information and material data safety sheets for each component of the dressing kit.

10. Description of the Device

The product is a portable suction device that may promote wound healing when used with accessory wound sealing kits.

11. Intended use of the Device

The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.

12. Substantial Equivalence

In establishing substantial equivalence to the predicate device, Genadyne Biotechnologies evaluated the indications for use, material, technology, product snecifications, and energy requirements of the system. Performance testing has been completed to demonstrate the safe and cffective use of the Genadyne A4 Wound Vacuum System for the intended use.

13. Summary of Safety and Effectiveness

Performance testing and device comparison demonstrates that the subject device is substantially equivalent to the predicate device, and is safe and effective for the intended use.

{2}------------------------------------------------

K082676 Page 3/5

14. Comparison to Predicate Device

Table of Comparison to Legally Marketed Device:

Comparative Information
New DevicePredicate Device
CompanyGenadyne BiotechnologiesBlueSky Medical Group, Inc.
BlueSky VISTATM Wound VacuumSystem
Device NameA4 Wound Vacuum SystemSystem
510 (K) NumberK082676K061367
Technical Data
Suction Capacity5 liters per minute8 liters per minute
Max Vacuum230 mmHg200 mmHg
Power Requirements24 VDC, 1A100-240 V AC, 60 Hz
Battery TypeNi-MHNi-MH
Dimensions andWeight200 x 180 x 80 mm / 1.36 Kg260 x 250 x 106 mm / 1.9 Kg
Accessories
Canisters800 ml disposable canister witha build-in hydrophobic shut offfilter for overflow protection250 ml or 800 ml disposablecanister
ReusableYesYes
SterileNon SterileNon Sterile
Accessory Kit
A4 Wound Sealing Kit which isbased on the teachings of Dr.Mark Chariker and Dr.Katherine Jeter1, which includes:BlueSky Medical Wound SealingKits, which includes:
- Non-adherent gauze- Non-adherent gauze
- Anti-Microbial gauze- Anti-Microbial gauze
- Transparent film dressing- Transparent film dressing
- Silicone flat drain- Silicone flat drain
Indications for Use
The Genadyne A4 WoundVacuum System is indicatedfor use in patients whowould benefit from negativepressure wound therapyparticularly as the devicemay promote wound healingby the removal of excessexudates, infectious materialand tissue debris.The BlueSky VISTATM WoundVacuum System is indicated forpatients who would benefit from asuction device particularly as thedevice may promote wound healing.
Contraindications
The Genadyne A4 iscontraindicated in the presenceof :The VISTATM is contraindicated inthe presence of:
Necrotic tissueNecrotic tissue
-Untreated osteomyelitisUntreated osteomyelitis
-Malignancy (with exception toenhance quality of life)Malignancy (with exception toenhance quality of life)
Untreated malnutritionUntreated malnutrition
Exposed arteries, veins, ororgansExposed arteries, veins, or organs
Precautions
Precautions should be taken forPrecautions should be taken for
patients who are or may be:patients who are or may be:
Receiving anticoagulanttherapyReceiving anticoagulant therapy
Suffering from difficulthemostasisSuffering from difficult hemostasis
Untreated for malnutritionUntreated for malnutrition
Non-complaint or combativeNon-complaint or combative
Compliance
UL 60601-1UL 60601-1
CAN/CSA C22.2 No. 601-1-M90IEC 0601-1-2
CAN/CSA C22.2 No. 601.1
Storage / Transport
-18°C to +43°C (0°F to 110°F)-30 to +50° C (-22 to 122° F)
Relative Humidity 10% to 95 %5 to 90% Humidity, non condensing
pressurepressure
Operation18°C to 34°C (65°F to 94°F)+5 to +35° C (41 to 95° F)
Relative Humidity 10% to 95 %20 to 80 % Humidity, non condensing
700 - 1060 mbar Atmosphericpressure700 - 1060 mbar Atmosphericpressure
Testing
IEC 60601-1-2-
FCC part 15 Class B-
EN 55011-
IEC 61000-4-2-
IEC 61000-4-3-

1 Chariker ME, Jeter KF, Tintle TE. Effective management of incisional and cutaneous fistulae with closed suction wound drainage. Contemporary Surg. 1989;(34):59-63.

{3}------------------------------------------------

K082676 page 4/8

:

{4}------------------------------------------------

K082676 Page 5/6

15. Comparative Performance Evaluation:

The FDA Decision Tree for substantial equivalence was followed and the steps involved have been considered. The rationale for each step is discussed below. For reviewer convenience, the numbering system used by FDA on the decision tree has been followed by Gcnadyne in their process for the substantial equivalence determination rationale.

{5}------------------------------------------------

K082676 page
6/45

510(k) “SUBSTANTIAL EQUIVALENCE”

DECISION-MAKING PROCESS New Device is Compared to Marketed Device * 2 l Descriptive Information Does New Device Have Same NO Do the Differences Alter the Intended Not Substantially about New or Marketed Indication Statement? Therapeutic/Diagnostic/etc. Effect YES Equivalent Determination Device Requested as Needed (in Deciding, May Consider Impact on AES Safety and Effectiveness)?** New Device Has Same Intended NO Use and May be "Substantially Equivalent" O New Device Has New Intended {Jse ﺮ ﻟﻠ 4 Does New Device Have Same ර Technological Characteristics, NO Could the New Do the New Characteristics e.g. Design, Materials, etc. ? Characteristics YES Affect Safety or · Raise New Types of Safety YES Effectiveness? or Effectiveness Questions? 5 NO Are the Descriptive NO Characteristics Precise Enough NO to Ensure Equivalence? -8 7 NO Are Performance Data Do Accepted Scientific Available to Assess Equivalence? YES Methods Exist for NO Assessing Effects of the New Characteristics? YES YES ହୁ Performance Are Performance Data Available NO Data Required To Assess Effects of New Characteristics? *** YES 9 0 Performance Data Demonstrate Performance Data Demonstrate Equivalence? Equivalence? YES YES NO NO "Substantially Equivalent" へ To べ Determination To

  • 510(k) Submissions compare new devices to marketed devices. FDA requests additional information if the relationship between marketed and "predicate" (pre-Amendments or reclassified post-Amendments) devices is unclear.

本本 This decision is normally based on descriptive information alone, but limited testing information is sometimes required.

*** Data maybe in the 510(k), other 510(k)s, the Center's classification files, or the literature

{6}------------------------------------------------

K0552676 Page 7/8

16. Discussion of Substantial Equivalence:

Note 1: Does new device have same indication statement as the predicate device(s)?

Yes. The Genadyne A4 Wound Vacuum System and the BlueSky VISTATM Wound Vacuum System are both indicated for patients who would benefit from a suction device particularly as the device may promote wound healing.

Note 3: Does the device have the same technological characteristics, e.g., design, materials, etc. ?

Yes. The Genadyne A4 Wound Vacuum System is design with a suction pump, can be externally and internally powered and has a collection canister, same as the BlueSky VISTA™ Wound Vacuum System, which also features a suction pump, a collection canister and can be powered externally and internally.

Note 5: Are the descriptive characteristics precise enough to ensure equivalence?

Yes. The characteristics of the Genadyne A4 Wound Vacuum System and the BlueSky VISTATM Wound Vacuum System are precise enough to ensure equivalence, based on the table in item 14 in the 510K Summary.

17. Discussion of Similarities and Differences

Device Similarities

Indication for use

The indication for use is identical for the Genadyne A4 Wound Vacuum System and the predicate device.

Configuration

All devices are sold non-sterile and are intended to be reusable. Devices are compatible with off-the-shelf accessories, such as disposable 800ml canisters.

Basic Product Function

The Genadyne A4 Wound Vacuum System and the predicate device have the same product function of generating a vacuum to provide general use suction and collection of liquids into an off-the-shelf canister reservoir which may promote wound healing for patient who would benefit from it.

{7}------------------------------------------------

K082676 Pages/s

Device Differences

In comparison to the predicate devices, the Genadyne A4 Powered Wound Vacuum System has several differences which do not affect the device safety and effectiveness of the Genadyne A4 Wound Vacuum System. These differences between the Genadyne A4 Wound Vacuum System and the predicate device are described in further detail below.

BlueSky VISTA134 Wound Vacuum System

The differences between the Genadyne A4 Wound Vacuum System and the BlueSky VISTA™ Wound Vacuum System are that the Genadyne A4 Wound Vacuum System is lighter in weight and has a longer battery life. In all other aspects, the Genadyne A4 Wound Vacuum System and the predicate device is substantially equivalent.

18. Conclusions:

Genadyne believes the Genadyne A4 Wound Vacuum System is substantially equivalent to the predicate device.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The eagle is positioned to the right of a circular emblem containing the words "DEPARTMENT OF HEALTH + HUMAN SERVICES USA" arranged around the perimeter.

Public Health Service

APR - 7 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Genadyne Biotechnologies, Inc. % Mr. Chien-Ming Goh 65 Watermill Lane Great Neck, New York 11021

Re: K082676

Trade/Device Name: Genadyne A4 Wound Vacuum System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: II Product Code: OMP Dated: November 12, 2008 Received: November 14, 2008

Dear Mr. Goh:

This letter corrects our substantially equivalent letter of December 18, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not

{9}------------------------------------------------

limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

K 062676

Indications for Use

510(k) Number (if known):

Device Name: Genadyne A4 Wound Vacuum System

Indications For Use:

The Genadyne A4 Wound Vacuum System is indicated for use in patients who would benefit from negative pressure wound therapy particularly as the device may promote wound healing by the removal of excess exudates, infectious material and tissue debris.

Prescription Use X (Per 21 CFR 801 Subpart D)

OR

Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
--------------------------------------------------------

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K082676

Page 1 of 1

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.