K Number
K082400

Validate with FDA (Live)

Date Cleared
2009-02-03

(167 days)

Product Code
Regulation Number
876.5665
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Culligan Soft Water Service Company Deionizer and Carbon Exchange tanks are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62. These tanks remove chlorine, chloramines, total organic carbons and total dissolved solids from water used to dilute dialysis concentrate to form dialysate, in reprocessing of hemodialyzers, as supply water to the dialysis machines and equipment rinse and disinfection. Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether. These tanks are components of a larger water treatment system employing adequate pre-treatment and post-treatment sections. Culligan's tanks are not to be used alone.

Device Description

The Culligan Soft Water Service Company, Deionizer and Carbon Exchange Service for Hemodialysis includes carbon filtration for the removal of chlorine and chloramines and deionizer (DI) exchange tanks for the removal of contaminants from the water through an ion exchange process to provide AAMI quality water for hemodialysis applications.

AI/ML Overview

This document is a 510(k) summary for the Culligan Soft Water Service Company Deionizer and Carbon Exchange tanks for hemodialysis, not a study report. It states that the device is substantially equivalent to a predicate device (Ameriwater Dialysis Deionizer Exchange Service, K991519) and defines its intended use and general specifications.

Therefore, the requested information regarding acceptance criteria, study data, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness is not available within the provided text. This document focuses on the regulatory submission process and establishing substantial equivalence rather than presenting performance study results.

To fulfill the request, a hypothetical example based on the type of device (water purification for hemodialysis) would typically involve:

1. Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance Criteria (e.g., based on ANSI/AAMI RD62)Reported Device Performance
Chlorine Removal< 0.1 mg/L(e.g., < 0.05 mg/L)
Chloramine Removal< 0.1 mg/L(e.g., < 0.05 mg/L)
Total Organic Carbon (TOC) Removal< 2.0 mg/L(e.g., < 1.0 mg/L)
Total Dissolved Solids (TDS) Removal< 10 ppm(e.g., < 5 ppm)
Flow Rate (at specified pressure)(e.g., 5 GPM)(e.g., 5.2 GPM)
Capacity (Volume processed before exhaustion)(e.g., 10,000 gallons)(e.g., 10,500 gallons)

2. Sample Size and Data Provenance for Test Set:

  • Sample Size: This would refer to the number of water samples tested under various conditions and the number of exchange tanks evaluated. For example, "10 tanks were tested over a period of 6 months, with daily water samples analyzed for each parameter."
  • Data Provenance: "Prospective testing conducted in a laboratory setting using water sources simulating typical municipal water supplies in the United States."

3. Number and Qualifications of Experts for Ground Truth:

  • For water quality, "ground truth" is typically established by chemical analysis using validated laboratory methods rather than expert consensus. Therefore, expert qualifications would refer to the chemists or laboratory technicians performing the analyses.
  • Number of Experts: (e.g., 3 laboratory technicians)
  • Qualifications: (e.g., Certified Analytical Chemists with 5+ years of experience in water quality analysis, specifically trained in AAMI RD62 standards).

4. Adjudication Method:

  • Not applicable for objective chemical measurements. Results are quantitative.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Not applicable as this is a water purification device, not an imaging or diagnostic AI system that interacts with human readers.

6. Standalone Performance:

  • Yes, the performance of the device (tanks) is inherently standalone. It is measured by its ability to purify water to AAMI standards without human intervention in the purification process itself. The "study" would involve measuring the water quality output from the tanks under various operational conditions.

7. Type of Ground Truth Used:

  • Objective Chemical Analysis: Water samples were analyzed using standard laboratory techniques (e.g., spectrophotometry for chlorine/chloramines, conductivity meters for TDS, TOC analyzers) to determine the concentration of specified contaminants. These analytical results serve as the ground truth.

8. Sample Size for Training Set:

  • Not applicable in the traditional sense of machine learning. The "training" for such a device involves engineering design, material selection, and optimization of the resin beds, which is based on established principles of water chemistry and filtration, not a data-driven training set for an algorithm.

9. How Ground Truth for Training Set was Established:

  • Again, not directly applicable. The performance characteristics and design parameters of the resins and tank configurations would be based on scientific literature, material specifications from resin manufacturers, pilot studies, and engineering calculations to achieve the desired removal efficiencies. This is a design and testing process, not "ground truth establishment" in the AI or diagnostic context.

{0}------------------------------------------------

2/3/09

Subject: 510 (K) SUMMARY

510 (K) Number:K082400
Contact Name:Mike Jablonski, General ManagerJoe Baumtrog, Project Coordinator
Proprietary Name:Culligan Soft Water Service Company, Deionizer, and Carbon ExchangeService for Hemodialysis
Common Name:DI and Carbon Exchange Tanks for Hemodialysis
Classification Name:Carbon and Deionized Dialysis Exchange Tanks for Hemodialysis
Classification:Class II Medical Device under 21 CFR 876.5665Panel: GastroenterologyProduct Code:FIR

Intended Use: The Culligan Soft Water Service Company Deionizer and Carbon Exchange tanks are primary or temporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62. These tanks remove chlorine, chloramines, total organic carbons and total dissolved solids from water used to dilute dialysis concentrate to form dialysate, in reprocessing of hemodialyzers, as supply water to the dialysis machines and equipment rinse and disinfection. Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether. These tanks are components of a larger water treatment system employing adequate pre-treatment and post-treatment sections. Culligan's tanks are not to be used alone.

Device Description: The Culligan Soft Water Service Company, Deionizer and Carbon Exchange Service for Hemodialysis includes carbon filtration for the removal of chlorine and chloramines and deionizer (DI) exchange tanks for the removal of contaminants from the water through an ion exchange process to provide AAMI quality water for hemodialysis applications.

Statement of Substantial Equivalence: The Culligan Soft Water Service Company, Deionizer and Carbon Exchange Service for Hemodialysis is substantially equivalent in intended use, function, and technology to the Ameriwater Dialysis Deionizer Exchange Service (K991519).

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three stripes representing the three branches of government. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB -3 2009

Mr. Mike Jablonski General Manager Culligan Soft Water Service Company 6030 Culligan Way MINNETONKA MN 55345-5970 -

Re: K082400

Trade/Device Name: Carbon and Deionized Dialysis Exchange Tanks for Hemodialysis

Regulation Number: 21 CFR §876.5665

Regulation Name: Water purification system for hemodialysis

Regulatory Class: II

Product Code: FIP

Dated: January 28, 2009

Received: January 30, 2009

Dear Mr. Jablonski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act. or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours.

Janine M. Morris

anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number:K082400
Device Name:Carbon and Deionized Dialysis Exchange Tanks for Hemodialysis

Culligan Tank Model Numbers:

Carbon TanksMixed-Bed TanksDuo-Bed Tanks
DRM50UDM50DC9
DRM9UDM9DA9
DRM4UDM4DC4

Indications For Use: The Culligan Soft Water Service Company Deionizer and Carbon Exchange tanks are primary or icmporary devices to provide water for hemodialysis applications per the requirements of ANSI/AAMI RD62. These tanks remove chloramines, total organic carbons and total dissolved solids from water used to dilute dialysis concentrate to form dialysate, in reprocessing of hemodialyzers, as supply water to the dialysis machines and equipment rinse and disinfection. Upon exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or with new resin altogether. These tanks are components of a larger water treatment system employing adequate pretreatment and post-treatment sections. Culligan's tanks are not to be used alone.

X Prescription Use (Pait 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

DA4

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helen Lew

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

3.0

§ 876.5665 Water purification system for hemodialysis.

(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.