K Number
K081128

Validate with FDA (Live)

Device Name
MOLAR IMPLANT
Manufacturer
Date Cleared
2008-11-07

(200 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nobel Biocare's Molar Implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The Molar Implants are indicated for single or multiple unit restorations in the molar region. The Molar Implants can be used in splinted or non-splinted applications. The Molar Implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

Device Description

Nobel Biocare's Molar Implants are threaded, root-form dental implants intended for use in the upper and lower jaw arches to support prosthetic devices, such as an artificial tooth, in order to restore esthetics and chewing function to partially or fully edentulous patients.

The Molar Implants are similar in design to the NobelSpeedy, Groovy, TiUnite, and NobelActive Internal Connection Implants. Like these implants, the Molar Implants are manufactured from commercially pure titanium. The implants utilize the TiUnite surface treatment and have the Groovy groove design feature.

AI/ML Overview

The provided text describes a 510(k) summary for a dental implant, not a medical device with acceptance criteria for performance against a specific condition or disease, or a study to prove such performance. Therefore, I cannot extract the requested information like expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.

The submission focuses on establishing substantial equivalence to previously cleared devices based on material, design, and intended use. The "acceptance criteria" in this context would primarily relate to meeting regulatory standards and demonstrating similarity to predicate devices, rather than performance metrics for diagnostic or therapeutic accuracy.

Therefore, I am unable to provide a response to the prompt as medical devices without AI functionality and performance criteria are not included in this document.

{0}------------------------------------------------

K081128

510(k) Summary of Safety and Effectiveness 1.4

Submitted by:Herbert CraneDirector, Global Regulatory AffairsNOV - 7 2008
Address:Nobel Biocare USA LLC22715 Savi Ranch ParkwayYorba Linda, CA 92887
Telephone:(714) 282-4800, ext. 5074
Facsimile:(714) 282-9023
Date of Submission:April 18, 2008
Classification Name:Endosseous Dental Implant (21 CFR 872.3640)
Trade or Proprietaryor Model Name:Molar Implant
Legally Marketed Device(s):TiUnite Implants (K050705)NobelSpeedy Implants (K050406)Groovy Implants (K050258)NobelActive Internal Connection Implant (K071370)

Device Description:

Nobel Biocare's Molar Implants are threaded, root-form dental implants intended for use in the upper and lower jaw arches to support prosthetic devices, such as an artificial tooth, in order to restore esthetics and chewing function to partially or fully edentulous patients.

The Molar Implants are similar in design to the NobelSpeedy, Groovy, TiUnite, and NobelActive Internal Connection Implants. Like these implants, the Molar Implants are manufactured from commercially pure titanium. The implants utilize the TiUnite surface treatment and have the Groovy groove design feature.

Indications for Use:

Nobel Biocare's Molar Implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The Molar Implants are indicated for single or multiple unit restorations in the molar region. The Molar Implants can be used in splinted or non-splinted applications. The Molar Implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

{1}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV - 7 2008

Nobel Biocare AB C/O Ms. Thuong Son Nguyen Regulatory Affairs Specialist Nobel Biocare USA, LLC 22715 Savi Ranch Parkway Yorba Linda, California 92887

Re: K081128

Trade/Device Name: Molar Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: П Product Code: DZE Dated: October 3, 2008 Received: October 6, 2008

Dear Ms. Nguyen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Ms. Nguyen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric' s (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Suite H. Michaud, MD

FOR DR. CHILL Chiu S. Lin. Ph. D Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

KO81128 510(k) Number (if known):

Device Name: Molar Implant

Indications For Use:

Nobel Biocare's Molar Implants are endosseous dental implants intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as an artificial tooth, in order to restore patient esthetics and chewing function. The Molar Implants are indicated for single or multiple unit restorations in the molar region. The Molar Implants can be used in splinted or non-splications. The Molar Implants may be placed immediately and put into immediate function provided that initial stability requirements detailed in the manual are satisfied.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

My Geth for MSR

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 1

510(k) Number:

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.