K Number
K080370

Validate with FDA (Live)

Date Cleared
2008-07-31

(170 days)

Regulation Number
862.1120
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ABL80 FLEX CO-OX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose and oximetry in whole blood. The ABL80 FLEX CO-OX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order:

pH, pO2, and pCO2 : pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Potassium (cK'): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa"): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCI): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The following six parameters will be introduced with the ABL80 FLEX CO-OX:

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sOx oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus reduced hemoglobin.

FO₂Hb: oxyhemoglobin as a fraction of total hemoglobin.

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoglobin as a fraction of total hemoglobin.

Device Description

The ABL80 FLEX CO-OX syste:n consists òf a modular analy.cer incorporating a user interface module with a large color touch screen interfacing to analyzer electronic and fluidis modules. The user interface module contains the analyzer CP1J and all of the required electronic interfaces for external communication and data stor:ge.

Sensors that measure pH, pCO2, pO2, cCa2', cCI, cK* cNa* and cGlu are contained in a cassette that also contains the sample inlet. This cassette attaches to the front of the analyzer.

The oximetry module measures ctHb, sO2, FO2Hb, F MetHb and FHHb. This module consists of a spectrometer, an ultrasonic hemolyzer and thermostatic components integrated into the analyzer.

The system also includes a reagent cartridge for the calibration and automatic quality control of the sensor and oximetry system. The calibration and quality control reagents are packaged in sealed foil pouches.

The analyzer and consumables incorporate "smart chip" technology for unique identification and lot specific calibration and quality control data.

AI/ML Overview

The provided document describes the ABL80 FLEX CO-OX system, an automated analyzer for measuring various blood parameters. While it lists "Performance Data" categories, the document does not explicitly state acceptance criteria or present the results of a study demonstrating the device meets specific acceptance criteria.

Instead, the document primarily focuses on:

  • Device Description: How the ABL80 FLEX CO-OX system works, including its components and measurement principles.
  • Intended Use / Indications for Use: What the device measures and for what clinical purposes, and who should use it.
  • Substantial Equivalence: Comparing the ABL80 FLEX CO-OX to predicate devices (ABL80 FLEX and ABL700) and highlighting similarities in operational principles.
  • FDA 510(k) Clearance: The formal notification from the FDA approving the device for marketing.

The "PERFORMANCE DATA" section merely lists the types of studies that were performed (Method Comparison, Precision, Linearity, Calibration/Quality Control, Analytical Specificity, User Testing) and states that the ABL700 (a predicate device) was used for some of this data. It does not provide the results of these studies nor does it define what would be considered "acceptable" performance for each category.

Therefore, I cannot populate the table or answer the specific questions regarding acceptance criteria and study results based solely on the provided text. The document acts as an overview and regulatory clearance, not a detailed performance study report.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for SENDX MEDICAL, INC. The logo is in black and white and features a stylized font. Below the company name is the text "A RADIOMETER COMPANY" in a smaller font.

RADIOMETER

K080370

Section III - 510(k) Summary of Safety and Effectiveness

1. SUBMITTER INFORMATIONJUL 31 2008
A.Establishment Registration:2027541
B.Manufacturing Site:SenDx Medical, Inc.
C.Company Address:1945 Palomar Oaks WayCarlsbad, CA 92011
D.Date Prepared:February 5, 2008
2. CONTACT PERSON
A.Mark A. Dzendzel:Manager, Regulatory Affairs andQuality Systems
B.Phone:760-603-3412
C.Fax:760-900-6310
D.Email:mdzendzel@sendx.com
3. DEVICE IDENTIFICATION
A.Trade/Proprietary Name:ABL80 FLEX CO-OX
B.Classification:Class II (21CFR § 862.1120)
C.Product Code:75CHL
D.Subsequent Codes:75JGS 75CEM, 75JFP, 75CGZ,75CGA, 75GKR, 75DQA, 75GHS

4. DEVICE DESCRIPTION

The ABL80 FLEX CO-OX syste:n consists òf a modular analy.cer incorporating a user interface module with a large color touch screen interfacing to analyzer electronic and fluidis modules. The user interface module contains the analyzer CP1J and all of the required electronic interfaces for external communication and data stor:ge.

Sensors that measure pH, pCO2, pO2, cCa2', cCI, cK* cNa* and cGlu are contained in a cassette that also contains the sample inlet. This cassette attaches to the front of the analyzer.

The oximetry module measures ctHb, sO2, FO2Hb, F MetHb and FHHb. This module consists of a spectrometer, an ultrasonic hemolyzer and thermostatic components integrated into the analyzer.

10

{1}------------------------------------------------

The system also includes a reagent cartridge for the calibration and automatic quality control of the sensor and oximetry system. The calibration and quality control reagents are packaged in sealed foil pouches.

The analyzer and consumables incorporate "smart chip" technology for unique identification and lot specific calibration and quality control data.

5. INTENDED USE / INDICATIONS FOR USE

The ABL80 FLEX CO-OX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose and oximetry in whole blood. The ABL80 FLEX CO-OX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

SUBSTANTIAL EQUIVALENCE 6.

The ABL80 FLEX CO-OX is substantially equivalent in features and characteristics to the predicate ABL80 FLEX and the ABL700.

510(k) NumberDeviceManufacturer
K051804ABL80 FLEXSenDx Medical, Inc.
K980130ABL700 SeriesRadiometer Medical ApS

With each of the listed devices and the ABL80 FLEX CO-OX, the principles of operation are similar;

• AmperometrypO₂, Glucose
-----------------------------
  • pH, cNa*, cK*, cCa2+, cCl~, pCO2 . Potentiometry
  • . Spectrophotometry ctHb, sO2, FO2Hb, FCOHb, FMetHb and FHHb

7. PERFORMANCE DATA

The ABL700 will be used for the following data:

  • Method Comparison: A.
  • B. Precision
  • Linearity/Assay Reportable Range C.
  • D. Calibration/Quality Control
  • E. Analytical specificity - Interference
  • F. User Testing

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Re:

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SenDx Medical, Inc. c/o Mr. Mark A. Dzendzel Manager Regulatory Affairs & Quality Systems 1945 Palomar Oaks Way Carlsbad, CA 92011

JUL 3 1 2008

K080370 Trade Name: ABL80 Flex Co-Ox Regulation Number: 21 CFR 864.5620 Regulation Name: Automated hemoglobin system Regulatory Class: Class II Product Codes: GKR, GHS, GLY, JJY, CHL, JGS, CEM, JFP, CGZ, CGA Dated: June 16, 2008 Received: June 20, 2008

Dear Mr. Dzendzel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial squivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please nothe the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Jean M. Cooper, M.S., D.V.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K080370

Device Name: ABL80 FLEX CO-OX

Indications for Use:

The ABL80 FLEX CO-OX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose and oximetry in whole blood. The ABL80 FLEX CO-OX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.

These tests are only performed under a physician's order:

pH, pO2, and pCO2 : pH, pCO2 and pO2 measurements are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Potassium (cK'): potassium measurements are used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high blood potassium levels.

Sodium (cNa"): sodium measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, inappropriate antidiuretic secretion, or other diseases involving electrolyte imbalance.

Calcium (cCa2+): calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride (cCI): chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such a cystic fibrosis and diabetic acidosis.

Glucose (cGlu): glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

The following six parameters will be introduced with the ABL80 FLEX CO-OX:

Total Hemoglobin (ctHb): total hemoglobin measurements are used to measure the hemoglobin content of whole blood for the detection of anemia.

sOx oxygen saturation, more specifically the ratio between the concentration of oxyhemoglobin and oxyhemoglobin plus reduced hemoglobin.

Division Sign-Off

Office of In Vitro Diagnostic Device
Evaluation and Safety

510(k) 2090370

Page 1 of 2

{5}------------------------------------------------

FO₂Hb: oxyhemoglobin as a fraction of total hemoglobin.

FCOHb: carboxyhemoglobin measurements are used to determine the carboxyhemoglobin content of human blood as an aid in the diagnosis of carbon monoxide poisoning.

FMetHb: methemoglobin as a fraction of total hemoglobin.

FHHb: reduced hemoglobin as a fraction of total hemoglobin.

Prescription Use _ x (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Drvision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K080370

Page 2 of 2

§ 862.1120 Blood gases (P

CO2 , PO2 ) and blood pH test system.(a)
Identification. A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.(b)
Classification. Class II.