K Number
K073249

Validate with FDA (Live)

Date Cleared
2008-07-25

(249 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Solaris NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.

Device Description

The subject SOLARIS NT1 and NT1A Handheld Pulse Oximeters measure pulse rate and oxygen saturation. The signals are converted into digital data and processed; the pulse oximeter examines the data and displays the data. The subject pulse oximeters also provide operating control for the user. The pulse oximeters are intended for use in spot checking; the NT1A model is intended to both spot check and perform continuous monitoring. The pulse oximeters can be used in hospital clinical areas such as general wards to provide additional information to the medical and nursing staff about the physiological condition of the patient. The subject devices are intended to be used under supervision by clinical personnel. The intended location of use is clinics.

The subject pulse oximeters provide a rapid indication of a patient's level of oxygenation which reflects the effective ventilation. The NT1A Pulse Oximeter with alarm allows continuous and instantaneous monitoring of SpO2 and both the NT1 and NT1A Pulse Oximeters reduce the need for arterial puncture and bloodgas analysis.

The subject pulse oximeters are composed of two boards, a main control board and a physiological singal processing board. The main control board is composed of a LED display and its control and driver, data memory, audio circuit, UART communication interface, and DC/DC circuits. The physiological processing board is composed of a SpO2 sensor control and IR/RED LEDs driver, data collection, signal processing algorithm execution, UART communication interface and DC/DC circuits.

The oximeter is housed in a single package. The main components of the assembly are listed as following

  • Main Board .
  • SpO2 Module .
  • Key Pad .
  • Batterv ●
AI/ML Overview

The provided text describes a 510(k) summary for the SOLARIS NT1 and NT1A Handheld Pulse Oximeters. It states that laboratory testing was conducted to validate and verify that the device met all design specifications and was "substantially equivalent" to a predicate device, the Datex-Ohmeda TuffSat 3000 Pulse Oximeter.

However, the summary does not contain a specific table of acceptance criteria or reported device performance metrics in the format requested. It generally states that the device "met all design specifications" and demonstrated that it is "as safe, as effective, and performs as well as the legally marketed predicate device."

Here's a breakdown of the information that can be extracted or inferred from the provided text, along with what is not present:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Specific quantitative acceptance criteria (e.g., accuracy ranges for SpO2, pulse rate)Not explicitly stated in a quantitative manner. The document states the device "met all design specifications" and is "as safe, as effective, and performs as well as the legally marketed predicate device." This implies the performance is within acceptable limits for a pulse oximeter and comparable to the predicate.
Compliance with standards (IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, ISO14971)Device has been "tested to assure compliance" with these standards.

2. Sample Size for the Test Set and Data Provenance

The document mentions "clinical and laboratory testing" but does not specify the sample size for the test set or the data provenance (e.g., country of origin, retrospective/prospective).


3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the text. The document refers to "laboratory testing" and "clinical testing" but does not detail how ground truth was established for these tests, or the involvement/qualifications of experts.


4. Adjudication Method for the Test Set

The adjudication method (e.g., 2+1, 3+1, none) is not mentioned in the provided text.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not conducted or reported. The device is a "standalone" medical device (pulse oximeter) that provides direct readings, not an AI-assisted diagnostic tool that would typically involve human readers interpreting cases.


6. Standalone Performance

Yes, a standalone performance assessment was effectively done. The device's primary function is to measure and display SpO2 and pulse rate directly. The testing conducted aimed to show the device's inherent performance and its substantial equivalence to the predicate device without human-in-the-loop assistance influencing the device's measurement outputs. The "laboratory testing" and "clinical testing" would have assessed the device's accuracy and reliability as a standalone unit.


7. Type of Ground Truth Used

The type of ground truth used is not explicitly stated. For a pulse oximeter, ground truth typically involves comparing the device's readings against a gold standard method, such as arterial blood gas analysis for oxygen saturation. While "clinical testing" is mentioned, the specific method for establishing ground truth is not detailed.


8. Sample Size for the Training Set

This information is not applicable/relevant in the context of this device and testing described. Pulse oximeters operate based on well-established physiological principles and optical measurements. They do not typically involve "training sets" in the sense of machine learning algorithms. The device's core functionality is hardware-based signal processing.


9. How the Ground Truth for the Training Set Was Established

This information is not applicable/relevant as there is no mention or implication of a "training set" for an AI or machine learning model.

{0}------------------------------------------------

Exhibit #1 4073249

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 5l0(k) number is: _______________________________________________________________________________________________________________________________________________

Submitter's Identification: 1.

Alan P. Schwartz mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, N.Y. 11021

Date Summary Prepared: July 17, 2008

Sponsor Company Name/Address/Contact Person 2.

Solaris Medical Technology, Inc. 400 Oyster Point Blvd., Ste. 534 South San Francisco, CA 94080 Tel: (415) 221-4350 Fax: (415) 221-4360

Rachel Cheng Contact: Director, Regulatory Affairs Position:

Manufacturing Facility Name and Address 3.

Newtech, Inc. R1-B1, Hi-Tech Industrial Park Nanshan District, Shenzhen Guanadong 518057 P.R. China

Name of the Device: 4.

SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories

{1}------------------------------------------------

5. Common or Usual Name:

Oximeter. Pulse 74 DQA. 21 CFR Part 870.2700

6. Predicate Device Information:

K# 001688, Datex-Ohmeda TuffSat 3000 Pulse Oximeter, Datex-Ohmeda Inc.

7. Device Description:

Purpose and Function of Device:

The subject SOLARIS NT1 and NT1A Handheld Pulse Oximeters measure pulse rate and oxygen saturation. The signals are converted into digital data and processed; the pulse oximeter examines the data and displays the data. The subject pulse oximeters also provide operating control for the user. The pulse oximeters are intended for use in spot checking; the NT1A model is intended to both spot check and perform continuous monitoring. The pulse oximeters can be used in hospital clinical areas such as general wards to provide additional information to the medical and nursing staff about the physiological condition of the patient. The subject devices are intended to be used under supervision by clinical personnel. The intended location of use is clinics.

The subject pulse oximeters provide a rapid indication of a patient's level of oxygenation which reflects the effective ventilation. The NT1A Pulse Oximeter with alarm allows continuous and instantaneous monitoring of SpO2 and both the NT1 and NT1A Pulse Oximeters reduce the need for arterial puncture and bloodgas analysis.

The subject pulse oximeters are composed of two boards, a main control board and a physiological singal processing board. The main control board is composed of a LED display and its control and driver, data memory, audio circuit, UART communication interface, and DC/DC circuits. The physiological processing board is composed of a SpO2 sensor control and IR/RED LEDs driver, data collection, signal processing algorithm execution, UART communication interface and DC/DC circuits.

The oximeter is housed in a single package. The main components of the assembly are listed as following

  • Main Board .
  • SpO2 Module .

{2}------------------------------------------------

  • Key Pad .
  • Batterv ●

8. Intended Use:

The SOLARIS NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive, spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.

Comparison to Predicate Devices: 9.

The basic intended use of the subject devices and Datex-Obmeda TuffSat 3000 Pulse Oximeter are the same:

  • I. Both the NT1 Pulse Oximeter and Datex-Ohmeda TuffSat 3000 Pulse Oximeter are non-invasive, spot-checking monitors. The difference between the NT1 and Datex-Ohmeda devices and the NT1A device is that in addition to being noninvasive spot-checking monitors, the NT1A device performs continuous monitoring.
    1. The patient parameters for both the subject devices and Datex-Ohmeda TuffSat 3000 Pulse Oximeter are the same: oxygen saturation (SpO2), pulse rate (PR) and pulse strength.
    1. The target populations range for both the subject devices and the Datex-Ohmeda Tuffsat 3000 Pulse Oximeter is the same: for pediatrics and adults.

Both the NT1 and NT1A Pulse Oximeters are substantially equivalent to Datex-Ohmeda TuffSat 3000 Pulse Oximeter. The pulse oximeters provide a means for interfacing with a patient and collecting parameter specific physiological signals. Then the signals are converted into digital data and processed, and the SpO2 and pulse rate values are calculated and displayed on LED screen.

10. Testing

Laboratory testing was conducted to validate and verify that the SOLARIS NT1

{3}------------------------------------------------

and NT1A Handheld Pulse Oximeters with Sensor Accessories met all design specifications and were substantially equivalent to the predicate device, Datex-Ohmeda TuffSat 3000 Pulse Oximeter. SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories have also been tested to assure compliance to the requirements of various published standards, including IEC60601-1, IEC60601-2, IEC60601-1-4, EN865, EN475, and ISO14971.

11. Conclusions:

The conclusions drawn from clinical and laboratory testing of the SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories demonstrates that the device is as safe, as effective, and performs as well as the legally marketed predicate device, the Datex-Ohmeda TufSet 3000 Pulse Oximeter, K001688.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 25 2008

Solaries Medical Technology, Incorporated C/O Mr. Alan P. Schwartz mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021

Re: K073249

Trade/Device Name: SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Sensor Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: July 7, 2008 Received: July 9, 2008

Dear Mr. Schwartz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Schwartz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

TN Samuels, Rnd, MD for //

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Exhibit B

Indications for Use

1 Page

510(k) Number (if known):

SOLARIS NT1 and NT1A Handheld Pulse Oximeters with Device Name: Sensor Accessories

Indications For Use:

Solaris NT1 and NT1A Handheld Pulse Oximeters with sensor accessories are non-invasive spot-check, oxygen saturation and pulse rate monitors. They operate on battery power using SOLARIS reusable Sp02 sensors for pediatric and adult patients.

Mr. Vml

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K073249 510(k) Number: _______________________________________________________________________________________________________________________________________________________________

Prescription Use × (Per 21 CFR 801 Subpart D)

OR

Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).