K Number
K073178

Validate with FDA (Live)

Date Cleared
2007-12-27

(44 days)

Product Code
Regulation Number
864.8625
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

It is an established laboratory procedure to use a stable control to monitor the performance of diagnostic tests. HCT Extended Hematology Control is an assayed control designed to monitor values obtained from automated, semi-automated and manual methods. For in vitro Diagnostic Use Only

Device Description

HCT Extended Hematology Control is an in vitro diagnostic reagent composed of human erythrocytes suspended in a plasma-like fluid with preservatives. It is an assayed whole blood control designed to monitor values obtained from automated, semi-automated and manual methods. It is sampled in the same manner as a patient specimen.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the HCT Extended Hematology Control:

Please note: The provided text is a 510(k) summary for a diagnostic test control (Hematology Quality Control Mixture), not an AI/ML powered device. Therefore, many of the typical acceptance criteria and study design elements found in AI/ML performance studies (e.g., sample size for test set, number of experts, MRMC studies, standalone performance, training set details) are not applicable or explicitly mentioned in detail in this type of submission. The focus for a control device is typically on demonstrating stability, precision, and equivalence to a predicate device.


Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance
Stability & PerformanceRemaining within the assay range over the stated life of the product."The HCT Extended Hematology Control passed the acceptance criteria of remaining within the assay range over the stated life of the product." "substantial equivalence in performance, precision and stability to the predicate device."

Study Details

  1. Sample size used for the test set and the data provenance:

    • Sample Size: The document states that "Laboratory testing of 3 validation lots" was performed. This refers to manufacturing lots of the control material, not a dataset of patient samples. The number of individual test runs or replicates within these lots is not specified.
    • Data Provenance: Not explicitly stated, however, as a laboratory control, the testing would have been conducted internally by R&D Systems, Inc. The document does not refer to patient data from a specific country or whether it was retrospective or prospective in the context of typical AI/ML studies.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable / Not stated. For a hematology control, "ground truth" is typically established by the manufacturer through precise analytical methods and quality control procedures, not by expert review of individual cases. The "assay range" is determined by the expected values on various instruments.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • None. This concept is not relevant for a quality control material where performance is assessed analytically against predefined ranges, not through subjective interpretation requiring adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI/ML device and therefore no MRMC study was performed. The device is a diagnostic reagent (control), not an AI-assisted interpretation tool.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI device. Its "performance" refers to its analytical characteristics (stability, precision) when tested in a laboratory setting using standard laboratory instruments.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Manufacturers' established assay ranges and analytical precision. The "ground truth" for a control device is its expected value range and stability, which is determined by the manufacturer based on analytical testing, calibration to reference methods (if applicable), and established quality control principles.
  7. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.
  8. How the ground truth for the training set was established:

    • Not applicable. This is not an AI/ML device that requires a training set.

Summary Context:

This 510(k) pertains to a hematology quality control product. Its purpose is to ensure laboratory instruments are performing correctly by providing a known sample with expected assay ranges. The "study" described is a validation of the control's performance, precision, and stability against the manufacturer's established criteria and in comparison to a previously cleared predicate device. The core requirement for clearance is demonstrating "substantial equivalence" to the predicate, meaning it performs similarly and has the same intended use and technological characteristics. The provided text confirms that the device passed its internal acceptance criteria, primarily revolving around maintaining its assay range over its specified lifespan.

{0}------------------------------------------------

K073178

510(k) Summary

Submitter Information

R&D Systems, Inc. 614 McKinley Place N.E. Minneapolis, MN 55413

Contact: Nancy Ring Phone: 612-656-4533 Fax: 612-379-6580

DEC 2 7 2007

Date Prepared: 11/12/07

Device Information

Proprietary Name: HCT Extended Control Common Name: Hematology Controls Classification 21 CFR 864.8625 Classification Name: Hematology Quality Control Mixture Product Code: JPK Device Class: -Hematology (81) Panel:

Predicate Device

R&D Systems CBC-7 Hematology Control (K843962)

Description of Device

HCT Extended Hematology Control is an in vitro diagnostic reagent composed of human erythrocytes suspended in a plasma-like fluid with preservatives. It is an assayed whole blood control designed to monitor values obtained from automated, semi-automated and manual methods. It is sampled in the same manner as a patient specimen.

Intended Use:

HCT Extended is a control designed to monitor values obtained from automated, semiautomated and manual methods. Please refer to the for specific instrument models.

Technological Comparison to Predicate

The new device has the same technological characteristics and intended use as the legally marketed predicate device. Both products are used to monitor values from automated, semi-automated and manual methods. Both are assayed controls that can be used to monitor hematocrit.

{1}------------------------------------------------

Summary of Performance Data

Laboratory testing of 3 validation lots has shown the HCT Extended Hematology Control to have substantial equivalence in performance, precision and stability to the predicate device. The HCT Extended Hematology Control passed the acceptance criteria of remaining within the assay range over the stated life of the product.

Substantial Equivalence Conclusion

The data demonstrates that the HCT Extended Hematology Control is substantially equivalent to the legally marketed predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of a human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the graphic. The text is in all caps and appears to be in a sans-serif font.

Public Health Service

DEC 2 7 2007

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

R & D Systems, Inc. C/O Nancy C. Ring 614 McKinley Place, N.E. Minneapolis, MN 55413

Re: K073178

Trade/Device Name: HCT Extended Hematology Control Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology Quality Control Mixture Regulatory Class: Class II Product Code: JPK Dated: November 12, 2007 Received: November 13, 2007

Dear Ms. Ring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 -

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Peter Becker

Robert L. Becker, Jr., MD, PhD Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 3 --

.

:

cc: HFZ-401 DMC

HFZ-404 510(k) Staff HFZ- 440 Division D.O.

{5}------------------------------------------------

R & D Systems, Inc. HCT Extended Hematology Control November 12, 2007

Indications for Use

510(k) Number (if known): __ KO 73/ 9 8

Device Name: HCT Extended Hematology Control

Indications for Use:

It is an established laboratory procedure to use a stable control to monitor the performance of diagnostic tests. HCT Extended Hematology Control is an assayed control designed to monitor values obtained from automated, semi-automated and manual methods.

For in vitro Diagnostic Use Only

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Josephine Bautista
Division Sign-Off

e of In Vitro Diagnostic Device Evaluation and Safety

510(k) K073178

Page 1 of 1 ..

§ 864.8625 Hematology quality control mixture.

(a)
Identification. A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).(b)
Classification. Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.