K Number
K072526

Validate with FDA (Live)

Date Cleared
2007-10-05

(28 days)

Product Code
Regulation Number
870.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Micrus Courier microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into peripheral, coronary, and neuro vasculature.

Device Description

The Micrus® Courier Enzo™ Microcatheter 0.0190" in a diagnostic intravascular catheter with an in-vivo shapeable tip which deflects under operator control at an angle of ±90° from the an m-110 shapedere up was @ Courier Enzo™ Microcatheter 0.0190"is designed with a feature which allows the clinician to adjust the catheter tip shape in-vivo by turning a knob. No shaping while anovired. This design feature allows the clinician to adjust the catheter tip shape to toor to require variations in patient anatomy without having to remove the catheter from the patient's body in order to re-shape it.

AI/ML Overview

The provided text describes a 510(k) submission for the Micrus® Courier Enzo™ Microcatheter 0.0190". This submission aims to demonstrate substantial equivalence to a predicate device, the Micrus® Courier Enzo™ Microcatheter 0.0170".

Here's an analysis of the acceptance criteria and the study as described in the document, based on the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state specific quantifiable acceptance criteria (e.g., tensile strength greater than X, or burst pressure greater than Y). Instead, it relies on demonstrating substantial equivalence to a predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence in intended use, shape, materials, design, and function compared to predicate device (Micrus® Courier Enzo™ Microcatheter 0.0170").The Micrus® Courier Enzo™ Microcatheter 0.0190" is stated to be "substantially equivalent to the predicate device with regard to intended use, shape of distal tip angle, materials, design, and function." The intended use is clearly defined as assisting in the delivery of diagnostic and therapeutic agents into peripheral, coronary, and neuro vasculature. The device description highlights its in-vivo shapeable tip, a feature also present in the predicate, allowing adjustment of the catheter tip shape under operator control. The accessories listed are also consistent with those for the predicate device.
Verification that the device does not raise new questions regarding safety and effectiveness."Verification testing conducted on the Micrus® Courier Enzo™ Microcatheter 0.0190" demonstrates the device is substantially equivalent to the predicate device and does not raise new questions regarding safety and effectiveness with respect to diagnostic intravascular catheters when used in accordance with its Instructions for Use." While specific test results are not detailed, the statement implies that the testing confirmed safety and effectiveness within the defined parameters.

2. Sample Size Used for the Test Set and Data Provenance

The document states that "in-vitro tests" were conducted.

  • Sample Size for Test Set: Not explicitly stated.
  • Data Provenance: The document implies the data is from in-vitro testing performed by the manufacturer, Micrus Endovascular Corporation, based in San Jose, California, USA. It is therefore prospective as it was conducted for the purpose of this submission. No human or patient data is mentioned, only laboratory testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a microcatheter, and the evaluation is based on engineering and performance criteria through in-vitro testing, not on interpretation of complex medical images or diagnoses requiring expert consensus for ground truth.

4. Adjudication Method for the Test Set

Not applicable. As described above, the evaluation relies on in-vitro testing against pre-defined performance characteristics, not on subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or mentioned. This type of study is relevant for diagnostic imaging AI systems where human reader performance is being evaluated with and without AI assistance. This document describes a medical device (microcatheter), not an AI diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a medical device, not an algorithm or AI system. Its performance inherently involves human handling and operation. The "in-vivo shapeable tip which deflects under operator control" explicitly indicates human-in-the-loop operation.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is based on engineering specifications and functional performance criteria established for medical catheters, primarily against the predicate device. It's not derived from expert consensus, pathology, or outcomes data in the traditional sense for diagnostic or prognostic devices. The substantial equivalence argument relies on demonstrating that its physical and functional characteristics meet or are comparable to those of the predicate device, which is already legally marketed.

8. The Sample Size for the Training Set

Not applicable. This is a medical device (microcatheter), not a machine learning model. Therefore, there is no "training set" in the context of AI or algorithm development.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As mentioned, there is no training set for this type of medical device.

{0}------------------------------------------------

K072526

OCT 5 2007

510(k) Summary

Micrus Endovascular Corporation Micrus® Courier Enzo™ Microcatheter 0.0190''

This 510(k) summary for the Micrus® Courier Enzo™ Microcatheter 0.0190"is submitted in accordance with the requirements of 21 C.F.R. § 807.92.

GENERAL INFORMATION

Manufacturer:Micrus Endovascular Corporation821 Fox LaneSan Jose, California 95131Phone: (408) 433-1400Est. Registration No. 2954740
Contact Person:R. Michael CromptonVice President, Regulatory / Clinical Affairs& Quality
Date Prepared:September 6, 2007
DEVICE DESCRIPTION
Classification:Class II
Trade Name:Micrus® Courier Enzo™ Microcatheter 0.0190"
Generic/Common Name:Diagnostic intravascular catheter (21 CFR § 870.1200)

PREDICATE DEVICE

Micrus® Courier Enzo™ Microcatheter 0.0170" (reference: K070456)

.

ﻴﺔ

INTENDED USE

The Micrus Courier microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into peripheral, gorionary, and neuro vasculature.

{1}------------------------------------------------

DEVICE DESCRIPTION

The Micrus® Courier Enzo™ Microcatheter 0.0190" in a diagnostic intravascular catheter with an in-vivo shapeable tip which deflects under operator control at an angle of ±90° from the an m-110 shapedere up was @ Courier Enzo™ Microcatheter 0.0190"is designed with a feature which allows the clinician to adjust the catheter tip shape in-vivo by turning a knob. No shaping while anovired. This design feature allows the clinician to adjust the catheter tip shape to toor to require variations in patient anatomy without having to remove the catheter from the patient's body in order to re-shape it.

The accessories for the Micrus Courier Enzo Microcatheter 0.0190", which are not supplied as Phe development of the sales to those for the Micrus Courier Pre-Shaped Microcatheters and include:

  • Guiding catheter (generally, 5-7F) ●
  • Guidewire compatible with the microcatheter ●
  • Rotating Hemostatic Valves (RHV); two (2) required .
  • 3-way stopcock ●
  • 1-way valve .
  • Femoral sheath .
  • Continuous Saline Flush Set-ups with Pressure Bags, one as a flush for the guiding . catheter and the other as a flush for the microcatheter

SUBSTANTIAL EQUIVALENCE

The Micrus® Courier Enzo™ Microcatheter 0.0190"is substantially equivalent to the predicate device identified previously. The Micrus® Courier Enzo™ Microcatheter 0.0190"is substantially equivalent to the predicate device with regard to intended use, shape of distal tip angle, materials, design, and function.

Verification testing conducted on the Micrus® Courier Enzo™ Microcatheter 0.0190" demonstrates the device is substantially equivalent to the predicate device and does not raise new questions regarding safety and effectiveness with respect to diagnostic intravascular catheters when used in accordance with its Instructions for Use.

CONCLUSION

As described in this 510(k) Summary, Micrus Endovascular Corporation considers the Micrus® Courier Enzo™ Microcatheter 0.0190" substantially equivalent to the Micrus® Courier Enzo™ Microcatheter 0.0170" based on a comparison of intended uses and the results of in-virro tests.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and three wave-like shapes below it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Micrus Endovascular Corporation .OCT 5 * 2007 c/o Mr. R. Michael Crompton Vice President, Regulatory/Clinical Affairs & Quality 821 Fox Lane San Jose, CA 95131

Re: K072526

Trade/Device Name: Micrus® Courier Enzo™ Microcatheter 0.0190" Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: DQO Dated: September 6, 2007 Received: September 7, 2007

Dear Mr. Crompton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Crompton

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Singerely Nours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K072526

Statement of Indications for Use

510(k) Number (if known): K

Device Name: Micrus® Courier Enzo™ Microcatheter - 0.0190"

Indications for Use: The Micrus Courier microcatheters are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils, into peripheral, coronary, and neuro vasculature.

Prescription Use × (Per 21 C.F.R. 801 Subpart D) AND/OR

Over-the-Counter Use (21 C.F.R. 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign Off)
Division of Cardiovascular Device
510(k) Nubber
K072526

Page 1 of 1

§ 870.1200 Diagnostic intravascular catheter.

(a)
Identification. An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.(b)
Classification. Class II (performance standards).