K Number
K072358

Validate with FDA (Live)

Date Cleared
2007-10-31

(70 days)

Product Code
Regulation Number
866.5660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioPlex™ 2200 Vasculitis kit is a multiplex flow immunoassay intended for the semi-quantitative detection of IgG autoantibodies to Myeloperoxidase (MPO), Proteinase 3 (PR3) and Glomerylar Basement Membrane (GBM) in human serum. In conjunction with clinical findings, the test system is used as an aid in the diagnosis of anti-neutrophil cytoplasmic antibodies (ANCA)associated vasculitides: Microscopic Polyangitis (MPA), Necrotising Glomerulonephritis, Churg-Strauss Syndrome, Wegener's Granulomatosis and the autoimmune renal disorder, Goodpasture's syndrome.

The BioPlex 2200 Vasculitis kit is intended for use with the Bio-Rad BioPlex 2200 System.

The BioPlex 2200 Vasculitis Calibrator Set is intended for the calibration of the BioPlex 2200 Vasculitis Reagent Pack.

The BioPlex 2200 Vasculitis Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex 2200 Vasculitis Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 Vasculitis Control Set has not been established with any other Vasculitis assays.

The BioPlex 2200 Vasculitis kit is a multiplex flow immunoassav intended for the semiquantitative detection of IgG autoantibodies to Myeloperoxidase (MPO), serine proteinase 3 (PR3) and Glomerular Basement Membrane (GBM) in human serum.

The BioPlex 2200 Vasculitis kit is intended for use with the Bio-Rad BioPlex 2200 System.

Uses:

The test system is used to detect the presence of antibodies in serum samples, as an aid in the diagnosis of certain autoimmune vasculitides such as Microscopic Polyangilitis (MPA), Necrotising Glomerulonephritis, Churg-Strauss Syndrome, Wegener's Granulomatosis and autoimmune renal disorders, such as Goodpasture's syndrome, in conjunction with clinical findings and other laboratory tests.

Device Description

The Vasculitis kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. Three (3) different populations of beads are coated with antigens associated with vasculitis disease (MPO, PR3 and GBM). The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel. The mixture is incubated at 37°C. After a wash cycle, antibody, conjugated to phycoerythin (PE), is added to the dyed beads and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity (RFI),

Three additional dyed beads, an Internal Standard Bead (ISB), a Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction vessel and the absence of significant non-specific binding in serum or plasma. Refer to the BioPlex 2200 System Operation Manual for more information.

The instrument is calibrated using a set of four (4) distinct calibrator vials, supplied separately by Bio-Rad Laboratories. A combination of four (4) vials representing four (4) different antibody concentrations are used for semi-quantitative calibration. The result for each of these antibodies is expressed as an antibody index (Al).

AI/ML Overview

The provided text describes the BioPlex 2200 Vasculitis kit, calibrators, and controls. The document focuses on the performance summary, including expected values, reproducibility studies, and comparative testing against predicate devices and IFA methods.

However, the document does not explicitly state "acceptance criteria" for performance metrics in a clear, tabulated format. It presents performance results and compares them to predicate devices, implying that these results met internal criteria for substantial equivalence.

Therefore, the response below will synthesize the implied acceptance criteria from the reported performance, as direct acceptance criteria are not explicitly stated.


Acceptance Criteria and Study to Prove Device Meets Criteria: BioPlex 2200 Vasculitis Kit

1. Table of Acceptance Criteria and Reported Device Performance

As explicit acceptance criteria are not stated, the table below infers the criteria based on the reported comparative performance in the 510(k) summary, specifically focusing on agreement with predicate EIA and IFA methods. The reported performance demonstrates substantial equivalence to predicate devices, which is the underlying requirement for 510(k) clearance.

Performance Metric (Implied Acceptance Criteria)Reported Device Performance (BioPlex 2200 Vasculitis Kit)
Normal Blood Donors (N=293) vs. EIA
Anti-MPO Positive AgreementN/A (0 positive per EIA); Negative Agreement: 100.0% (293/293)
Anti-PR3 Positive AgreementN/A (0 positive per EIA); Negative Agreement: 100.0% (293/293)
Anti-GBM Positive AgreementN/A (0 positive per EIA, 2 weak positive); Negative Agreement: 99.3% (289/291); Overall Agreement: 98.6% (289/293)
Unselected Patient Samples (N=300) vs. EIA
Anti-MPO Positive Agreement71.4% (5/7); Negative Agreement: 97.6% (284/291); Overall Agreement: 96.3% (289/300)
Anti-PR3 Positive Agreement100.0% (5/5); Negative Agreement: 99.0% (292/295); Overall Agreement: 99.0% (297/300)
Anti-GBM Positive AgreementNot Accurate (0/1); Negative Agreement: 99.7% (298/299); Overall Agreement: 99.3% (298/300)
Retrospective Positive Samples vs. EIA & IFA
Anti-MPO Positive Agreement (EIA)93.9% (92/98); Overall Agreement: 93.0% (93/100)
Anti-MPO Positive Agreement (IFA)93.3% (83/89); Overall Agreement: 84.0% (84/100)
Anti-PR3 Positive Agreement (EIA)100.0% (79/79); Overall Agreement: 83.0% (83/100)
Anti-PR3 Positive Agreement (IFA)94.9% (93/98); Overall Agreement: 93.0% (93/100)
Anti-GBM Positive Agreement (EIA)88.9% (16/18); Overall Agreement: 92.6% (25/27)
Reproducibility (Total %CV)Ranges from 5.6% to 11.5% for positive and near-cutoff samples (MPO, PR3, GBM), with higher %CVs for negative controls (e.g., Anti-PR3 Negative Control: 18.8%). (Implied acceptance: %CV values within acceptable limits for diagnostic assays, demonstrating precision across different sites, lots, and days).

2. Sample Size and Data Provenance for Test Set

  • Normal Blood Donors: N=293, tested with BioPlex 2200 and corresponding commercial EIA methods.
  • Unselected Patient Samples: N=300, previously tested with vasculitis tests, and tested with BioPlex 2200 and corresponding commercial EIA methods.
  • Retrospective Anti-MPO Positive Samples: N=100, tested with BioPlex 2200, corresponding commercial EIA, and ANCA IFA.
  • Retrospective Anti-PR3 Positive Samples: N=100, tested with BioPlex 2200, corresponding commercial EIA, and ANCA IFA.
  • Retrospective Anti-GBM Positive Samples: N=27, tested with BioPlex 2200 and corresponding commercial EIA.
  • Cross-reactivity study: 10 samples per cross-reactant (except 7 for anti-tTG), tested with BioPlex 2200 and corresponding commercial EIA.

Data Provenance: Not explicitly stated (e.g., country of origin, specific institutions). The samples are referred to as "normal blood donors," "unselected patient samples previously tested with vasculitis tests," and "retrospective samples positive for anti-MPO/PR3/GBM." The study was conducted at "two (2) US testing facilities and an internal site (Bio-Rad Laboratories)" for reproducibility, suggesting at least some US data. The studies are retrospective as they utilize "previously tested" and "retrospective" samples.

3. Number of Experts and Qualifications for Ground Truth

The document does not mention the use of experts to establish ground truth for the test set. Instead, it relies on comparative testing against "corresponding commercially available microplate EIA methods" and "ANCA IFA methods" which are presumably well-established and accepted diagnostic tests.

4. Adjudication Method for the Test Set

No adjudication method involving experts is mentioned. The ground truth for comparative effectiveness is established by the results of existing commercial assays (EIA and IFA). Discrepancies are reported (e.g., weak positive results, equivocal results) without specific details of an adjudication process beyond categorization.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was performed or described. This device is an in-vitro diagnostic kit for laboratory use, not typically subject to MRMC studies designed for imaging or complex diagnostic interpretations by multiple human readers. The study focuses on direct comparison of analyte detection performance between the new kit and predicate laboratory assays.

6. Standalone Performance

Yes, the standalone performance (algorithm only, without human-in-the-loop performance) is intrinsically what is presented. The device is a 'multiplex flow immunoassay' system designed to semi-quantitatively detect autoantibodies. Its performance metrics (e.g., agreement with predicate devices, reproducibility, cross-reactivity) are all measures of the device's standalone analytical capabilities.

7. Type of Ground Truth Used

The primary ground truth used is comparator assay results.

  • Results from "corresponding commercially available microplate EIA methods" for MPO, PR3, and GBM.
  • Results from "ANCA IFA method using ethanol-fixed slides" for anti-MPO (pANCA IFA) and anti-PR3 (cANCA IFA).

8. Sample Size for the Training Set

The document does not explicitly mention a separate "training set" or its sample size. The studies described are performance validation studies. In the context of IVD devices, a "training set" is not a standard concept as it might be for AI algorithms. The calibrator sets are used for instrument calibration, not for training a machine learning model.

9. How Ground Truth for the Training Set was Established

As no "training set" for an AI algorithm is mentioned, this question is not applicable. The device is an immunoassay kit, where calibration and control materials are used to ensure accurate measurement in routine use.

  • Calibrators: "supplied separately by Bio-Rad Laboratories. A combination of four (4) vials representing four (4) different antibody concentrations are used for semi-quantitative calibration." The ground truth for these calibrators would be established through careful characterization and quantification of the antibody concentrations by the manufacturer.
  • Controls: Used "to monitor the overall performance." The positive and negative controls would have their status (positive/negative for each antibody) established by the manufacturer through rigorous testing.

{0}------------------------------------------------

K072358

Image /page/0/Picture/1 description: The image shows the logo for Bio-Rad. The logo is white text on a black rounded rectangle. The text is in all capital letters and is bolded.

BIOPLEX 2200 VASCULITIS KIT, CALIBRATORS AND CONTROLS 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------510/k) NumberCANILIRenort DateIImmary !
A counts & com1/0700000000Atanarand and and and the driver and the children the distribution of the country

MANUFACTURER INFORMATION

Manufacturer
Manufacturer AddressBio-Rad Laboratories, Inc.Clinical Systems Division4000 Alfred Nobel DriveHercules, CA 94547
Telephone(510) 724-7000
Establishment Registration No.2915274
Owner / OperatorBio-Rad Laboratories, Inc.4000 Alfred Nobel DriveHercules, CA 94547
Owner / Operator No.9929003
Official Correspondent for the BioPlex 2200 Vasculitis
Official Correspondent AddressBio-Rad Laboratories6565 185th Ave NERedmond, WA 98052
Telephone425-881-8300
Establishment Registration No.3022521
Owner / OperatorBio-Rad Laboratories6565 185th Ave NERedmond, WA 98052
Official CorrespondentMr. Christopher Bentsen
Telephone(425) 498-1709
Fax(425) 498-1651

CLASSIFICATION INFORMATION

Classification NameTest System, Antineutrophil Cytoplasmic Antibodies (ANCA),Devices, Measure, Antibodies to Glomerular BasementMembrane (GBM)
Common Name:Multi-Analyte Detection System Vasculitis
Product Trade NameBioPlex 2200 Vasculitis kit on the BioPlex 2200 Multi-AnalyteDetection SystemBioPlex 2200 Vasculitis Calibrator SetBioPlex 2200 Vasculitis Control Set
Device ClassClass II
Classification PanelImmunology
Requlation Number866.5660-Multiple Autoantibodies Immunological Test System

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is black and white and features the company name in a bold, sans-serif font. The letters are slightly rounded and have a three-dimensional appearance. The logo is simple and recognizable.

LEGALLY MARKETED EQUIVALENT (SE) DEVICES

Comparative FDA ClearedPREDICATE DEVICE510(k) NumberDecisionDate
Phadia Varelisa™ MPO EIAK04104006/16/04
Phadia Varelisa™ PR3 EIAK04104307/02/04
INOVA QUANTA Lite™ GBM ELISAK98433602/08/99
INOVA NOVA Lite ANCA (Ethanol-fixed Slides)K96134007/10/96

DEVICE DESCRIPTION

The Vasculitis kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. Three (3) different populations of beads are coated with antigens associated with vasculitis disease (MPO, PR3 and GBM). The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead reagent into a reaction vessel. The mixture is incubated at 37°C. After a wash cycle, antibody, conjugated to phycoerythin (PE), is added to the dyed beads and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are re-suspended in wash buffer. The bead mixture then passes through the detector. The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity (RFI),

Three additional dyed beads, an Internal Standard Bead (ISB), a Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum or plasma to the reaction vessel and the absence of significant non-specific binding in serum or plasma. Refer to the BioPlex 2200 System Operation Manual for more information.

The instrument is calibrated using a set of four (4) distinct calibrator vials, supplied separately by Bio-Rad Laboratories. A combination of four (4) vials representing four (4) different antibody concentrations are used for semi-quantitative calibration. The result for each of these antibodies is expressed as an antibody index (Al).

KIT COMPONENTS

BioPlex 2200 Vasculitis reagent pack (Catalog No. 665-1850). The reagent pack contains supplies sufficient for 100 tests.

VialDescription
Bead SetOne (1) 10 mL vial, containing dyed beads coated with Myeloperoxidase (MPO),Proteinase-3 (PR3) and Glomerular Basement Membrane (GBM); an InternalStandard bead (ISB), a Serum Verification bead (SVB), and a Reagent Blankbead (RBB) in a MOPS (3-[N-Morpholino] propanesulfonic acid) buffersupplemented with Glycerol and protein stabilizers (bovine). ProClin® 300(0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) are added aspreservatives.

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 2 of 21

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is displayed in white text on a black background. The text is bold and sans-serif. The logo is simple and recognizable.

ConjugateCONJOne (1) 5 mL vial, containing phycoerythrin conjugated murine monoclonal anti-human IgG and phycoerythrin conjugated murine monoclonal anti-human FXIIIin phosphate buffer supplemented with murine and bovine protein stabilizers.ProClin® 300 (0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) areadded as preservatives.
SampleDiluentOne (1) 10 mL vial, containing bovine and murine protein stabilizers intriethanolamine buffer. ProClin® 300 (0.3%), sodium benzoate (0.1%) andsodium azide (<0.1%) are added as preservatives.

ADDITIONAL REQUIRED ITEMS, AVAILABLE FROM BIO-RAD

Description
663-1800BioPlex 2200 Vasculitis Calibrator Set: Four (4) 500 µL vials, each containinghuman antibodies to MPO, PR3 and GBM, in a human serum matrix made fromdefibrinated plasma. All calibrators contain ProClin® 300 (0.3%), sodiumbenzoate (0.1%) and sodium azide (<0.1%) as preservatives.
663-1830BioPlex 2200 Vasculitis Control Set: Two (2) 1.5 mL Positive Control serum vials,each containing human antibodies to MPO, PR3 and GBM, in a human serummatrix made from defibrinated plasma; and two (2) 1.5 mL Negative Control serumvials, in a human serum matrix made from defibrinated plasma. All controlscontain ProClin® 300 (0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%)as preservatives.
660-0817BioPlex 2200 Sheath Fluid: Two (2) 4 L bottles containing Phosphate BufferedSaline (PBS). ProClin® 300 (0.03%) and sodium azide (<0.1%) are added aspreservatives.
660-0818BioPlex 2200 Wash Solution: One (1) 10 L bottle containing Phosphate BufferedSaline (PBS) and Tween 20. ProClin® 300 (0.03%) and sodium azide (<0.1%)are added as preservatives.
660-0000BioPlex 2200 Instrument and Software System

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Bio-Rad. The logo is white text on a black, rounded rectangle. The text is in a bold, sans-serif font.

INTENDED USE

BioPlex™ 2200 Vasculitis Kit

The BioPlex™ 2200 Vasculitis kit is a multiplex flow immunoassay intended for the semiquantitative detection of IgG autoantibodies to Myeloperoxidase (MPO), Proteinase 3 (PR3) and Glomerylar Basement Membrane (GBM) in human serum. In conjunction with clinical findings, the test system is used as an aid in the diagnosis of anti-neutrophil cytoplasmic antibodies (ANCA)associated vasculitides: Microscopic Polyangitis (MPA), Necrotising Glomerulonephritis, Churg-Strauss Syndrome, Wegener's Granulomatosis and the autoimmune renal disorder, Goodpasture's syndrome.

The BioPlex 2200 Vasculitis kit is intended for use with the Bio-Rad BioPlex 2200 System.

BioPlex 2200 Vasculitis Calibrator Set

The BioPlex 2200 Vasculitis Calibrator Set is intended for the calibration of the BioPlex 2200 Vasculitis Reagent Pack.

BioPlex 2200 Vasculitis Control Set

The BioPlex 2200 Vasculitis Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex 2200 Vasculitis Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 Vasculitis Control Set has not been established with any other Vasculitis assays.

INDICATIONS FOR USE

The BioPlex 2200 Vasculitis kit is a multiplex flow immunoassav intended for the semiquantitative detection of IgG autoantibodies to Myeloperoxidase (MPO), serine proteinase 3 (PR3) and Glomerular Basement Membrane (GBM) in human serum.

The BioPlex 2200 Vasculitis kit is intended for use with the Bio-Rad BioPlex 2200 System.

Uses:

The test system is used to detect the presence of antibodies in serum samples, as an aid in the diagnosis of certain autoimmune vasculitides such as Microscopic Polyangilitis (MPA), Necrotising Glomerulonephritis, Churg-Strauss Syndrome, Wegener's Granulomatosis and autoimmune renal disorders, such as Goodpasture's syndrome, in conjunction with clinical findings and other laboratory tests.

TECHNOLOGICAL CHARACTERISTICS

The following tables summarize similarities and differences between the BioPlex 2200 Vasculitis Kit and the predicate devices used in comparative studies with the BioPlex 2200 Vasculitis Kit.

A. BioPlex 2200 Vasculitis vs. Predicate Phadia Varelisa MPO ANCA ElA

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate Phadia Varelisa MPOANCA EIA
ReagentsWash Buffer, Sample DiluentWash Buffer, Serum Diluent

Table 1: Similarities between reagents and materials

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BIO-RAD. The logo is in white text on a black background. The text is bold and sans-serif. The logo is enclosed in a rounded rectangle.

CalibratorsCalibratorsCalibrators
ControlsNegative Control and Multi-AnalytePositive Control (MPO, PR3 and GBM)Negative Control, Positive Control

Table 2: Similarities between reagents with regard to function and use

Similarities betweenFunction and UseBioPlex 2200 VasculitisPredicate Phadia Varelisa MPOANCA EIA
Intended UseSemi-quantitative detection of IgGautoantibodies to MPO, PR3 and GBMin human serum.Semi-quantitative and qualitativedetermination of MPO Anti neutrophilcytoplasmic antibodies in human serumor plasma.
MatricesSerumSerum

Table 3: Differences between reagents and materials

Differences betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate Phadia Varelisa MPOANCA EIA
Solid PhaseBead reagent - dyed antigen coatedbeads96 well microplate - antigen coatedmicrowells
ReagentsConjugate: Anti-human IgG /PhycoerythrinConjugate: Anti-human IgG HRPconjugate/Substrate (TMB)
Sheath FluidSheath Fluid is used to suspend the beadreagent and introduce it into the detector.Not similar; not utilized in EIA 's

Table 4: Differences between reagents with regard to function and use

Differences betweenFunction and UseBioPlex 2200 VasculitisPredicate Phadia Varelisa MPOANCA EIA
Analyte DetectionMulti-Analyte Detection (human IgG autoantibodies to MPO, PR3 and GBM)Single Analyte Detection (Human IgG antibodies to MPO)
MatricesSerumPlasma

B. BioPlex 2200 Vasculitis vs. Predicate Phadia Varelisa PR3 ANCA EIA

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate Phadia Varelisa PR3ANCA EIA
ReagentsWash Buffer, Sample DiluentWash Buffer, Sample Diluent
CalibratorsCalibratorsCalibrators
ControlsNegative Control and Multi-AnalytePositive Control (MPO, PR3 and GBM)Negative Control, Positive Control

Table 5: Similarities between reagents and materials

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is in a bold, sans-serif font. The logo is surrounded by a black rounded rectangle.

Table 6: Similarities between reagents with regard to function and use

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate Phadia Varelisa PR3ANCA EIA
Intended UseSemi-quantitative detection of IgGautoantibodies to MPO, PR3 and GBMin human serum.Semi-quantitative and qualitativedetermination of PR3 Anti neutrophilcytoplasmic antibodies in human serumor plasma.
MatricesSerumSerum
Differences betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate Phadia Varelisa PR3ANCA EIA
Solid PhaseBead reagent - dyed antigen coatedbeads96 well microplate - antigen coatedmicrowells
ReagentsConjugate: Anti-human IgG /PhycoerythrinConjugate: Anti-Human IgG HRPConjugate /Substrate (TMB)
Sheath FluidSheath Fluid is used to suspend the beadreagent and introduce it into the detector.Not similar; not utilized in EIA's.

Table 7: Differences between reagents and materials

Table 8: Differences between reagents with regard to function and use
Differences betweenFunction and UseBioPlex 2200 VasculitisPredicate Phadia Varelisa PR3ANCA EIA
Analyte DetectionMulti-Analyte Detection (human IgGautoantibodies to MPO, PR3 and GBM)Single Analyte Detection (Human IgG antibodies to PR3)
MatricesSerumPlasma

C. BioPlex 2200 Vasculitis vs. Predicate INOVA QUANTA-Lite GBM ELISA

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA QUANTA-LiteGBM ELISA
ReagentsWash Buffer, Sample DiluentWash Buffer, Serum Diluent
ControlsNegative Control and Multi-AnalytePositive Control (MPO, PR3 and GBM)Negative Control, Positive Control

Table 9: Similarities between reagents and materials

Table 10: Similarities between reagents with regard to function and use

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA QUANTA-LiteGBM ELISA
Intended UseSemi-quantitative detection of IgGautoantibodies to MPO, PR3 and GBMin human serum.Semi-quantitative determination of antineutrophil cytoplasmic antibodies toGBM in human serum

Table 11: Differences between reagents and materials

Differences betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA QUANTA-LiteGBM ELISA
Solid PhaseBead reagent - dyed antigen coated beads96 well microplate - antigen coated microwells

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 6 of 21

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is displayed in white text on a black background. The text is bold and sans-serif, and the letters are spaced closely together. The logo is simple and recognizable.

ReagentsConjugate: Anti-human IgG /PhycoerythrinConjugate: Anti-Human IgG / Horse-radish Peroxidase, Substrate (TMB)
Sheath FluidSheath Fluid is used to suspend the beadreagent and introduce it into the detectorNot similar; not utilized in EIA's.
CalibratorsCalibratorsNone

Table 12: Differences between reagents with regard to function and use

Differences betweenFunction and UseBioPlex 2200 VasculitisPredicate INOVA QUANTA-LiteGBM ELISA
Analyte DetectionMulti-Analyte Detection (human IgGautoantibodies to MPO, PR3 and GBM)Single Analyte Detection (Human IgGantibodies to GBM)

D. BioPlex 2200 Vasculitis vs. Predicate INOVA Lite, ANCA, Ethanol Fixed Slides

Table 13: Similarities between reagents and materials

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA NOVA Lite,ANCA, Ethanol Fixed Slides
ControlsNegative Control and Multi-AnalytePositive Control (MPO, PR3 and GBM)IFA Negative Control, p-ANCAPositive Control, c-ANCA PositiveControl

Table 14: Similarities between reagents with regard to function and use

Similarities betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA NOVA Lite,ANCA, Ethanol Fixed Slides
Intended UseSemi-quantitative detection of IgGautoantibodies to MPO, PR3 and GBMin human serum.Semi-quantitative determination ofAnti neutrophil cytoplasmic antibodiesin human serum.
MatricesSerumSerum

Table 15: Differences between reagents and materials

Differences betweenComponents / MaterialsBioPlex 2200 VasculitisPredicate INOVA NOVA Lite,ANCA, Ethanol Fixed Slides
Solid PhaseBead reagent - dyed antigen coatedbeadsANCA, Ethanol fixed neutrophilsSubstrate slides
ReagentsConjugate: Anti-human IgG /PhycoerythrinConjugate: Anti-Human IgG / Horse-radish Peroxidase
Sheath FluidSheath Fluid is used to suspend the beadreagent and introduce it into the detector.Not similar; not utilized in EIA's.
CalibratorsCalibratorsNone

Table 16: Differences between reagents with regard to function and use

Differences betweenFunction and UseBioPlex 2200 VasculitisPredicate INOVA NOVA Lite,ANCA, Ethanol Fixed Slides
Analyte DetectionMulti-Analyte Detection (human IgGautoantibodies to MPO, PR3 and GBM)Single Analyte Detection (Human IgG antibodies)

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 7 of 21

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Bio-Rad. The logo is white text on a black background. The text is in a bold, sans-serif font and is slightly distorted.

Solid PhaseBead reagent - dyed antigen coated beadsANCA, Ethanol fixed neutrophilsSubstrate slides
------------------------------------------------------------------------------------------------------------

PERFORMANCE SUMMARY

A. Expected Values

Expected values for the BioPlex 2200 Vasculitis kit are presented in the following tables for serum samples from normal blood donors (N=293) and unselected patient samples previously tested with vasculitis tests (N=300). A total of 300 serum samples from the normal blood donor population were tested. Seven (7) samples from the normal blood donor population were excluded due to "Serum Verification Bead (SVB) signal too low" analysis error messages during BioPlex 2200 Vasculitis kit testing. For all analytes, results of <1.0 Al are negative and results of 1.0 Al or greater are reported as positive.

Table A. BioPlex 2200 Vasculitis Kit -Normal Blood Donors (N=293)

BioPlex ResultPositive# (%)Negative# (%)
Anti-MPO0/293 (0.0%)293/293 (100.0%)
Anti-PR30/293 (0.0%)293/293 (100.0%)
Anti-GBM2/293 (0.7%)291/293 (99.3%)

Table B. BioPlex 2200 Vasculitis Kit -Unselected Patient Samples Previously Tested With Vasculitis Tests (N=300)

BioPlex ResultPositive# (%)Negative# (%)
Anti-MPO14/300 (4.7%)286/300 (95.3%)
Anti-PR38/300 (2.7%)292/300 (97.3%)
Anti-GBM1/300 (0.3%)299/300 (99.7%)

B. Reproducibility Studies

A reproducibility panel, consisting of ten (10) serum panel members, was prepared by Bio-Rad Laboratories. The positive panel members were prepared by combining patient samples positive for

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 8 of 21

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is bold and sans-serif.

antibodies to MPO, PR3 and GBM. Two (2) of the ten (10) had high levels of the antibodies to MPO, PR3 and GBM; two (2) members had low levels of the antibodies to MPO, PR3 and GBM; and two (2) members had antibody levels near the cutoff. There was also one (1) high negative and one (1) low negative panel member. In addition, a BioPlex 2200 Vasculitis positive control (antibody positive for MPO, PR3 and GBM) and a negative control (antibody negative for all 3 analytes) were included and tested as panel members.

Reproducibility testing was performed at two (2) US testing facilities and an internal site (Bio-Rad Laboratories) on a total of two (2) lots of the BioPlex 2200 Vasculitis kit. The ten (10) panel members were provided to the each of the testing sites. Two (2) of the three (3) testing facilities evaluated reproducibility using one (1) kit lot of the BioPlex 2200 Vasculitis kit and the third site evaluated the second lot of the BioPlex 2200 Vasculitis kit. Each of the ten (10) panel members was tested in duplicate (x2) on two runs per day for three days at each testing site (2 times x 2 runs x 3 days x 3 sites = 36 replicates per panel member and controls). The data were then analyzed for intra assay and inter-assay reproducibility according to the Clinical and Laboratory Standards Institute quidance (formerly NCCLS) EP5-A2, revised November 2004 and ISO/TR 22971:2005. The mean Antibody Index (AI), standard deviation (SD), and percent coefficient of variation (%CV) for each panel member were calculated. Results can be found in the below table:

VasculitisKit PanelMembersBioPlex 2200 Vasculitis Kit
SampleNMeanAIWithin-RunBetween-RunBetween-DayBetween-Site*Total
Vasculitis Anti-MPOHighPositive1364.00.2777.0%0.1563.9%0.0220.6%0.3268.2%0.45611.5%
HighPositive2365.50.2955.4%0.2494.5%0.0000.0%0.2544.6%0.4628.4%
LowPositive1361.50.0654.4%0.0412.8%0.0433.0%0.0171.2%0.0896.2%
LowPositive2361.90.1698.6%0.0000.0%0.0000.0%0.0251.3%0.1708.7%
Near361.00.0585.6%0.0000.0%0.0323.1%0.0484.6%0.0827.9%
Table. Reproducibility; BioPlex 2200 Vasculitis

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 9 of 21

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for BIO-RAD. The text is in white and the background is black. The logo is in a rounded rectangle.

Cutoff 1
NearCutoff 2361.30.0533.9%0.0332.5%0.0413.1%0.0000.0%0.0755.6%
PositiveControl362.90.1986.8%0.0000.0%0.0000.0%0.1023.5%0.2237.7%
NegativeControl360.20.0000.0%0.0000.0%0.0000.0%0.0000.0%0.0000.0%
HighPositive1364.20.3017.2%0.0000.0%0.1012.4%0.0761.8%0.3267.8%
HighPositive2364.50.1713.8%0.2024.5%0.0000.0%0.2676.0%0.3768.4%
LowPositive1361.40.0674.9%0.0241.7%0.0755.5%0.0332.4%0.1087.9%
Vasculitis Anti-PR3LowPositive2361.50.1288.3%0.0322.1%0.0000.0%0.0855.5%0.15710.2%
NearCutoff 1361.20.0625.1%0.0000.0%0.0574.7%0.0272.2%0.0887.3%
NearCutoff 2361.10.0534.7%0.0242.1%0.0453.9%0.0000.0%0.0736.4%
PositiveControl362.30.1345.8%0.0000.0%0.0954.1%0.0321.4%0.1677.3%
NegativeControl360.20.0198.9%0.0000.0%0.02712.6%0.02410.9%0.04118.8%

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Bio-Rad. The logo is white text on a black rounded rectangle. The text is in all caps and is a sans-serif font.

Vasculitis Anti-GBMHighPositive1364.30.2495.8%0.0000.0%0.0000.0%0.2044.7%0.3227.5%
HighPositive2364.80.1623.4%0.1944.1%0.0000.0%0.2805.9%0.3777.9%
LowPositive1361.40.0815.6%0.0000.0%0.0644.5%0.0876.0%0.1359.4%
LowPositive2361.70.0935.5%0.0523.1%0.0000.0%0.0472.8%0.1166.9%
NearCutoff 1361.10.0373.4%0.0454.1%0.0000.0%0.0797.2%0.0999.0%
NearCutoff 2361.20.0605.0%0.0413.4%0.0171.4%0.0816.7%0.1109.2%
PositiveControl362.80.1394.9%0.0000.0%0.0000.0%0.0632.3%0.1535.4%
NegativeControl360.20.0000.0%0.0000.0%0.0000.0%0.0000.0%0.0000.0%
  • Between-site variance includes between lot variance

C. Comparative Testing

Three hundred (300) normal blood donors and three hundred (300) unselected patient samples previously tested with vasculitis tests were tested with the BioPlex 2200 Vasculitis kit. Seven (7) of the three hundred (300) normal blood donor samples were excluded due to "Serum Verification Bead (SVB) signal too low" analysis error during BioPlex 2200 Vasculitis kit testing. All samples were also tested by the corresponding commercially available microplate EIA methods. The results can be observed in Tables A - F.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for BIO-RAD. The logo is white text on a black rounded rectangle. The text is in all caps and bolded.

EIA ResultPositiveNegativeTotalPositive (+) % Agreement95% Confidence IntervalNegative (-) % Agreement95% Confidence IntervalOverall % Agreement95% Confidence Interval
Positive000N/AN/A100.0%(293/293)98.7%,100%100.0%(293/293)98.7%,100%
Equivocal000
Negative0293293
Total0293293

Table A. BioPlex 2200 vs. Anti-MPO EIA - Normal Blood Donors (N=293)

N/A = Not Applicable

Table B. BioPlex 2200 vs. Anti-PR3 EIA - Normal Blood Donors (N=293)

BioPlex 2200 Vasculitis Anti-PR3 Result
PositiveNegativeTotalPositive (+)% Agreement95% ConfidenceIntervalNegative (-)% Agreement95% ConfidenceIntervalOverall% Agreement95% ConfidenceInterval
EIA ResultPositive000N/AN/A
Equivocal000100.0%(293/293)98.7%,100%100.0%(293/293)98.7%,100%
Negative0293293
Total0293293

N/A = Not Applicable

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for BIO-RAD. The logo is in white text on a black background. The text is bold and sans-serif.

PositiveNegativeTotalPositive (+) % Agreement95% Confidence IntervalNegative (-) % Agreement95% Confidence IntervalOverall % Agreement95% Confidence Interval
EIA ResultPositive02*2N/AN/A99.3% (289/291)97.5%,99.8%98.6% (289/293)96.5%,99.5%
Negative2289291
Total2291293

Table C. BioPlex 2200 vs. Anti-GBM EIA - Normal Blood Donors (N=293)

  • Two (2) positive anti-GBM EIA results were weak positive. N/A = Not Applicable

Table D. BioPlex 2200 vs. Anti-MPO EIA - Unselected Patient Samples Previously Tested With Vasculitis Tests (N=300)

BioPlex 2200 Vasculitis Anti-MPO Result
EIA ResultPositiveNegativeTotalPositive (+) % Agreement95% Confidence IntervalNegative (-) % Agreement95% Confidence IntervalOverall % Agreement95% Confidence Interval
Positive52771.4%(5/7)35.9%,91.8%97.6%(284/291)95.1%,98.8%96.3%(289/300)93.6%,97.9%
Equivocal*202
Negative7284291
Total14286300
  • Two (2) anti-MPO EIA equivocal results are included in the Overall Agreement.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black rounded rectangle. The text is bold and sans-serif.

BioPlex 2200 Vasculitis Anti-PR3 Result
PositiveNegativeTotalPositive (+)% Agreement95%ConfidenceIntervalNegative (-)% Agreement95%ConfidenceIntervalOverall% Agreement95%ConfidenceInterval
EIA ResultPositive505100.0%(5/5)56.5%,100%99.0%(292/295)97%,99.7%99.0%(297/300)97.1%,99.7%
Equivocal000
Negative3292295
Total8292300

Table E. BioPlex 2200 vs. Anti-PR3 EIA – Unselected Patient Samples Previously Tested With Vasculitis Tests (N=300)

Table F. BioPlex 2200 vs. Anti-GBM EIA - Unselected Patient Samples Previously Tested With Vasculitis Tests (N=300)

BioPlex 2200 Vasculitis Anti-GBM Result
PositiveNegativeTotalPositive (+) % Agreement95% Confidence IntervalNegative (-) % Agreement95% Confidence IntervalOverall % Agreement95% Confidence Interval
EIA ResultPositive01*1Not Accurate (0/1)Not Accurate99.7% (298/299)98.1%, 99.9%99.3% (298/300)97.6%, 99.8%
Negative1298299
Total1299300
  • One (1) anti-GBM EIA results was a weak positive.

The BioPlex 2200 Vasculitis kit was further evaluated by testing 227 retrospective samples positive for anti-MPO (N=100), anti-PR3 (N=100), and anti-GBM (N=27). All samples were also tested by the corresponding commercially available microplate EIA methods. In addition, the anti-MPO and anti-PR3 positive samples were tested by an ANCA IFA method using ethanol-fixed slides. The

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 14 of 21

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is bold and sans-serif. The logo is enclosed in a rounded rectangle.

results can be observed in Tables G - K.

Table G. BioPlex 2200 vs. Anti-MPO EIA – Retrospective Anti-MPO Positive Samples (N=100)

PositiveNegativeTotalPositive (+)% Agreement95% ConfidenceIntervalNegative (-)% Agreement95% ConfidenceIntervalOverall% Agreement95% ConfidenceInterval
EIA ResultPositive9269893.9%(92/98)87.3%,97.2%N/AN/A93.0%(93/100)86.2%,96.6%
Equivocal*101
Negative011
Total937100
  • One (1) anti-MPO EIA equivocal result is included in the Overall Agreement.

N/A = Not Applicable

Table H. BioPlex 2200 vs. Anti-PR3 EIA - Retrospective Anti-PR3 Positive Samples
(N=100)
BioPlex 2200 Vasculitis Anti-PR3 Result
PositiveNegativeTotalPositive (+)% Agreement95% ConfidenceIntervalNegative (-)% Agreement95% ConfidenceIntervalOverall% Agreement95% ConfidenceInterval
EIA ResultPositive79079100.0%(79/79)95.4%,100%N/AN/A83.0%(83/100)74.4%,89.1%
Equivocal*9110
Negative7411
Total955100
  • Ten (10) anti-PR3 ElA equivocal results are included in the Overall Agreement.

  • N/A = Not Applicable

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 15 of 21

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is in a bold, sans-serif font. The logo is surrounded by a rounded rectangle.

BioPlex 2200 Vasculitis Anti-GBM Result
PositiveNegativeTotalPositive (+)% Agreement95% ConfidenceIntervalNegative (-)% Agreement95% ConfidenceIntervalOverall% Agreement95% ConfidenceInterval
EIA ResultPositive16*2**1888.9%(16/18)67.2%,96.9%N/AN/A92.6%(25/27)76.6%,97.9%
Negative099
Total161127

Table I. BioPlex 2200 vs. Anti-GBM ElA - Retrospective Anti-GBM Positive Samples (N=27)

*Two (2) of the sixteen (16) anti-GBM ElA positive results were weak positive.

** One (1) of the two (2) anti-GBM EIA positive results were weak positive.

The remaining fifteen (15) of the eighteen (18) anti-GBM EIA results were moderate to strong positive.

N/A = Not Applicable

Table J. BioPlex 2200 vs. pANCA IFA - Retrospective Anti-MPO Positive Samples
(N=100)
BioPlex 2200 Vasculitis Anti-MPO Result
PositiveNegativeTotalPositive (+) % Agreement95% Confidence IntervalNegative (-) % Agreement95% Confidence IntervalOverall % Agreement95% Confidence Interval
pANCA IFA ResultPositive8368993.3%(83/89)86.1%,96.9%N/AN/A84.0%(84/100)75.6%,89.9%
Negative10111
Total937100

N/A = Not Applicable

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is in a bold, sans-serif font. The logo is surrounded by a rounded rectangle.

BioPlex 2200 Vasculitis Anti-PR3 Result
PositiveNegativeTotalPositive (+)% Agreement95%ConfidenceIntervalNegative (-)% Agreement95%ConfidenceIntervalOverall% Agreement95%ConfidenceInterval
cANCA IFA ResultPositive9359894.9%(93/98)88.6%,97.8%N/AN/A93.0%(93/100)86.2%,96.6%
Negative202
Total955100

Table K. BioPlex 2200 vs. cANCA IFA – Retrospective Anti-PR3 Positive Samples (N=100)

N/A = Not Applicable

Tables L and M compare EIA results from the retrospective positive samples for anti-MPO EIA (N=100) and anti-PR3 EIA (N=100) with ANCA IFA results.

Table L. pANCA IFA vs. Anti-MPO EIA - Retrospective Anti-MPO Positive Samples
(N=100)
Anti-MPO EIA Result
PositiveEquivocal*NegativeTotalPositive (+)% Agreement95%ConfidenceIntervalNegative (-)% Agreement95%ConfidenceIntervalOverall% Agreement95%ConfidenceInterval
pANCA IFA ResultPositive87118997.8%(87/89)92.2%-99.4%N/AN/A87.0%(87/100)79%,92.2%
Negative110011
Total9811100
  • One (1) anti-MPO EIA equivocal result is included in the Overall Agreement. N/A = Not Applicable

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black, rounded rectangle. The text is bold and slightly slanted.

Anti-PR3 EIA Result
PositiveEquivocal*NegativeTotalPositive (+)% Agreement95%ConfidenceIntervalNegative (-)% Agreement95%ConfidenceIntervalOverall% Agreement95%ConfidenceInterval
cANCA IFA ResultPositive7810109879.6%(78/98)70.6%,86.4%N/AN/A79.0%(79/100)70.0%-85.8%
Negative1012
Total791011100

Table M. cANCA IFA vs. Anti-PR3 EIA - Retrospective Anti-PR3 Positive Samples (N=100)

  • Ten (10) anti-PR3 EIA equivocal results are included in the Overall Agreement. N/A = Not Applicable

A combined positive agreement of 94.1% (176/187) was observed between BioPlex 2200 Vasculitis anti-MPO/anti-PR3 and ANCA IFA results (from Tables L and M), compared to a combined positive agreement of 88.2% (165/187) between anti-MPO/anti-PR3 EIA and ANCA IFA results (from Tables N and O). Also, a combined overall agreement of 88.5% (177/200) was observed between BioPlex 2200 Vasculitis anti-MPO/anti-PR3 and ANCA IFA results compared to a combined overall agreement of 83.0% (166/200) observed between anti-MPO/anti-PR3 and ANCA IFA results.

Table N summarizes the test results (positive percent agreement, negative percent agreement and overall percent agreement) for each of the three antibodies (anti-MPO, anti-PR3 and anti-GBM) from normal blood donors (N=293), unselected patient samples previously tested with vasculitis tests (N=300), and retrospective positive samples (N=100 for anti-MPO, N=100 for anti-PR3 and N=27 for anti-GBM). These sample populations were tested by the BioPlex 2200 Vasculitis kit and anti-MPO, anti-PR3 and anti-GBM ElAs. In addition, anti-MPO and anti-PR3 known positive samples were tested by an ANCA IFA method.

{18}------------------------------------------------

Table N. Summary of Positive, Negative and Overall Percent Agreement for Normal Blood Donors (N=293), Unselected Patient Samples Previously Tested With Vasculitis Tests (N=300), and Retrospective Positive Samples (N=100 for anti-MPO, N=100 for anti-PR3 and N=27 for anti-GBM)

Anti-MPO EIAResultpANCA IFAResultAnti-PR3 EIAResultcANCA IFAResultAnti-GBM EIAResult
% PosAgreement% NegAgreement% OverallAgreement% PosAgreement% NegAgreement% OverallAgreement% PosAgreement% NegAgreement% OverallAgreement% PosAgreement% NegAgreement% OverallAgreement% PosAgreement% NegAgreement% OverallAgreement
BioPlex 2200 Vasculitis ResultNormal BloodDonors*N/A293/293100.0%293/293100.0%NTNTNTN/A293/293100.0%293/293100.0%NTNTNTN/A289/29199.3%289/29198.6%
UnselectedPatientSamplesPreviously5/771.4%284/29197.6%289/30096.3%NTNTNT5/5100.0%292/29599.0%297/30099.0%NTNTNTNotAccurate(0/1)298/29999.7%298/30099.3%
RetrospectivePositiveSamples92/9893.9%N/A93/10093.0%83/8993.3%N/A84/10084.0%79/79100.0%N/A83/10083.0%93/9894.9%N/A93/10093.0%16/1888.9%N/A25/2792.6%

Normal blood donors and unselected patient samples previously tested with vascultis tests were not tested by ANCA IFA. N/A = Not Applicable NT = Not Tested

E. Cross-Reactivity

A cross-reactivity study was performed to determine if samples from various disease states and other potentially interfering factors interfere with test results when tested with the BioPlex 2200 Vasculitis kit. A panel of ten (10) specimens positive for each cross reactant were evaluated for possible cross reactivity with the BioPlex 2200 Vascultis kit for each of the three (3) autoantibodies. Samples were also tested on a corresponding commercially available microplate ElAs. Most of the samples evaluated were high positive for each disease state. The results demonstrated that the various disease state samples evaluated do not cross react with the three (3) autoantibodies in the BioPlex 2200 Vasculitis kit. Results can be found in the below table:

Table. Cross-Reactivity
---------------------------
CrossReactivesNMethodAnti-MPOAnti-PR3Anti-GBM
ANA10BioPlex 2200000
ANA10EIA000
ANA10Discrepants000
Anti-SaccharomycesCerevisiae(ASCA)10BioPlex 2200010
EIA010
Discrepants000
Anti-Cardiolipin10BioPlex 2200000
EIA100
Discrepants100
Anti-dsDNA10BioPlex 2200000
EIA000
Discrepants000
Anti-Histone10BioPlex 2200220
EIA100
Discrepants120
RheumatoidFactor (RF)10BioPlex 2200000
EIA000
Discrepants000
Anti-ThyroidPeroxidase(TPO)10BioPlex 2200000
EIA000
Discrepants000
Anti-tissueTransglutaminase (tTG)7*BioPlex 2200000
EIA000
Discrepants000
Anti-SmoothMuscle (ASMA)10BioPlex 2200000
EIA000
Discrepants000
HCV10BioPlex 2200000
EIA000
Discrepants000
HIV10BioPlex 2200000
EIA000
Discrepants000

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07) Page 19 of 21

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for Bio-Rad. The logo is white text on a black background. The text is in a bold, sans-serif font. The logo is surrounded by a black oval shape.

BioPlex 2200 Vasculitis Kit,Calibrator Set and Control Set 510(k) Summary (10-24-07)
Page 20 of 21

{20}------------------------------------------------

Image /page/20/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is in a bold, sans-serif font. The logo is enclosed in a rounded rectangle.

  • Due to limited availability of samples, only seven tTG specimens were evaluated.

{21}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/21/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Bio-Rad Laboratories c/o Ms. Priya Bondre Regulatory Affairs Representative 6565 185th Avenue NE Redmond, WA 98052

Re: K072358

Trade/Device Name: BioPlex™ 2200 Vasculitis kit BioPlex™ 2200 Vasculitis Calibrator Set BioPlex™ 2200 Vasculitis Control Set Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test System Regulatory Class: Class II Product Code: MOB, MVJ, JIX, JJY Dated: August 20, 2007 Received: August 22, 2007

Dear Ms Bondre:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

OCT 3 1 2007

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket

{22}------------------------------------------------

Page 2 -

notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Fobert Becker

Robert L. Becker, Jr., M.D., Ph. D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{23}------------------------------------------------

Indications for Use

510(k) Number: K072358

Device Name: BioPlex™ 2200 Vasculitis kit on the BioPlex™ 2200 Multi-Analyte Detection System BioPlex™ 2200 Vasculitis Calibrator Set BioPlex™ 2200 Vasculitis Control Set

Indications For Use:

BioPlex™ 2200 Vasculitis kit

The BioPlex™ 2200 Vasculitis kit is a multiplex flow immunoassay intended for the semi-quantitative detection of IgG autoantibodies to Mveloperoxidase (MPO), serine Proteinase 3 (PR3) and Glomerular Basement Membrane (GBM) in human serum.

The BioPlex 2200 Vasculitis kit is intended for use with the Bio-Rad BioPlex 2200 System.

Uses:

The test system is used to detect the presence of antibodies in serum samples, as an aid in the diagnosis of certain autoimmune vasculitides such as Microscopic Polyangiitis (MPA), Necrotising Glomerulonephritis, Churg-Strauss Syndrome, Wegener's Granulomatosis and autoimmune renal disorders, such as Goodpasture's syndrome, in conjunction with clinical findings and other laboratory tests.

BioPlex™ 2200 Vasculitis Calibrator Set

The BioPlex 2200 Vasculitis Calibrator Set is intended for the BioPlex 2200 Vasculitis Reagent Pack.

BioPlex™ 2200 Vasculitis Control Set

The BioPlex 2200 Vasculitis Control Set is intended for use as an assayed quality control to monitor the overall performance of the BioPlex 2200 Instrument and BioPlex 2200 Vasculitis Reagent Pack in the clinical laboratory. The performance of the BioPlex 2200 Vasculitis Control Set has not been established with any other Vasculitis assays.

Prescription Use _ V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD)

Maria M Chen


Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic
Device Evaluation and Safety

Page 1 of _1

510(k) KO12358

§ 866.5660 Multiple autoantibodies immunological test system.

(a)
Identification. A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).(b)
Classification. Class II (performance standards).