K Number
K071899

Validate with FDA (Live)

Date Cleared
2007-10-22

(105 days)

Product Code
Regulation Number
882.1400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, CO, Pulse Oximetry, and EtCO2. The ENVOY may be used to monitor a wide range of patient conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

The Mennen Medical Envoy BIS module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The BIS Index, a processed parameter may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

  • The Bispectral Index is a complex technology, intended for use only as an adjunct to clinical judgment and training.
  • In addition, the clinical utility, risk/benefit, and application of this device have not undergone full evaluation in the pediatric population.
Device Description

The Envoy is a multiparameter physiological patient monitor, capable of monitoring:

  • ECG/Heart Rate .
  • Invasive blood pressure .
  • Non-invasive blood pressure ●
  • Respiration ●
  • Pulse oximetry
  • Two temperature channels ●
  • Cardiac output ●
  • EtCo2 .
  • Spirometry ●
  • EEG ●
  • BIS module (new subject of this application) .

The Envoy bedside patient monitor consists of a main processing unit, a mountable color monitor, and a module rack housing the various Mennen Medical plug-in vital signs modules. The modules monitor the patient's vital signs. Up to six internal modules can be plugged into a module rack. The Envoy can accommodate two module racks. The vital sign data derived from the modules by the Envoy are presented on the monitor as waveform and numeric displays. The Envoy vital signs modules acquire vital signs data from the patient, and display their waveforms and alarms indications on the Envoy display unit. Vital signs and waveform information are displayed simultaneously on the Envoy Display Unit. Up to 8 traces can be displayed at any one time.

The vital signs modules interface with readily available physiologic transducers through electrically isolated patient input connections. After amplification, the signals are digitized, analyzed and displayed. All processing and alarm determination for ECG, Respiration and Invasive Blood Pressure is made using proprietary algorithms and software based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions. The SpO2, Non-Invasive Blood Pressure, ExCO2 and Spirometry modules incorporate software and/or hardware technology developed by vendors whose products are marketed in the USA.

The BIS Module is used to monitor dual channel EEG waveform and the BIS index, used to estimate the level of conciseness of patient under anesthesia, or patients in the ICU that may be with limited conciseness.

The BIS index together with several quality parameters are displayed and stored by the Envoy monitor.

AI/ML Overview

The Mennen Medical Envoy BIS module is a modification to the existing Envoy Patient Monitor, adding Bispectral Index (BIS) monitoring capabilities. The device claims substantial equivalence to the predicate device, the Spacelab Medical Bispectral Index (BISx) Analysis Module 91482 (K060900).

The provided text outlines the functional description of the new BIS module and its comparison to the predicate device, focusing on display formats, data storage, and alarm functionalities. It also references "performance testing" and "SW Validation" but lacks specific details regarding acceptance criteria and the results of these studies.

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly present a table of acceptance criteria with corresponding reported device performance values. Instead, it relies on demonstrating substantial equivalence by comparing the functional and technical specifications of the Envoy BIS module with the predicate device. The primary claim for acceptance is that the Envoy BIS module's function to display BIS parameters and quality parameters, and its alarm capabilities, are equivalent to the predicate device.

The document highlights the following parameters and their ranges as displayed and stored by the Envoy monitor, which are derived from the BISx unit:

ParameterRangeDescription
BIS0 - 99Bispectral Index: The measure of consciousness of a patient, (0 = no brain activity), (100 = fully conscious).
EMG30 - 55 dBElectromyography: The absolute power of muscle activity and artifacts in the 70 - 110 Hz range. Value is in dB with respect to 0.0001 µV2.
SQI0 - 100 %Signal Quality Index: The percentage of good epochs and suppressed epochs in the last 120 epochs collected that could be used in the Bispectral Index calculation.
SR0 - 100 %Suppression Ratio: The percentage of epochs in the past 63 seconds in which the EEG signal is considered suppressed
BC0-30Burst Count: The number of EEG bursts in the last minute. An EEG burst is a momentary period of EEG activity among isoelectric or flat EEG. Blanked if SR is less than 5. Activated by connection of an Extend sensor.
SEF0.5 - 30 HzSpectral Edge Frequency: The frequency at which 95% of the total power lies below it and 5% lies above it.

The "Similarities" and "Differences" sections (Page 9) act as an indirect comparison of performance criteria, stating that differences in display options between the two devices "do not change the efficiency of BIS and EEG displays of both monitors," implying an equivalence in performance.

2. Sample Size Used for the Test Set and Data Provenance

The document does not provide details on the sample size used for any specific test set related to the BIS module. There is no mention of human subject data, country of origin, or whether it was retrospective or prospective. The testing described primarily appears to be technical validation against the predicate device's performance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

No information is provided regarding the number or qualifications of experts used to establish ground truth for any test set. The validation primarily focuses on comparing technical specifications and functional equivalence.

4. Adjudication Method for the Test Set

No information on an adjudication method for a test set is provided.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No MRMC comparative effectiveness study is mentioned, nor is any effect size for human reader improvement with or without AI assistance. The device is a patient monitor, not an AI diagnostic tool involving human reader interpretation. The BIS module aids in monitoring the effects of anesthetic agents to help guide administration and potentially reduce awareness with recall, but this is a clinical utility, not a reader study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The BIS module itself processes EEG signals to generate the BIS index and related parameters. This processing can be considered "standalone" in the sense that the algorithm generates these outputs without direct human intervention in the calculation process. The device functions to "display the BIS parameter and the quality parameters" and to "provide alarm limits for the BIS index and provide visual and audible alarms." These outputs are generated by the device's algorithms based on the input from the BISx unit. However, the context of the question often implies a comparison to a clinical ground truth, which is not detailed here.

7. The Type of Ground Truth Used

The "ground truth" implicitly used in this submission is the established performance and output of the predicate device, the Spacelab Bispectral Index (BISx) Analysis Module 91482. The entire submission is built on demonstrating substantial equivalence to this legally marketed device. The "Substantial Equivalence Discussion" (Page 5) directly compares the Envoy BIS module's specifications and functionality to the predicate.

8. The Sample Size for the Training Set

No information is provided regarding a training set sample size. The device is not described as utilizing machine learning or AI that would typically involve a training set for model development. Its function appears to be based on established algorithms for processing EEG signals and deriving the BIS index, which are derived from the OEM (Aspect BISx).

9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned for the Envoy BIS module itself, there is no information on how its ground truth was established. The core processing capabilities for the BIS index are derived from the Aspect BISx unit (OEM agreement), suggesting that any fundamental ground truth for the BIS algorithm would have been established during the development and clearance of the original BISx device (K040183 and subsequently K060900 for the predicate).

{0}------------------------------------------------

CCT 2 2 2007

Name:Mennen Medical Ltd.
Registration Number:9611022
Operator Number:9069173
Address:4 Hayarden Street, Yavne, 81228, Israel
Postal Address:PO Box 102, .
Rehovot, 76100, Israel
Tel:+972-8-9323333
Fax:+972-8-9328510
Contact person:Ifat Oren, Regulatory Affairs

Traditional 510(k): Device Modification - Envoy Patient Monitor Terminology

Envoy = Subject of this 510(k). The Envoy Patient Monitor is a modified device, a system identical to of the Envoy Patient Monitor with the addition of BISx module.

SPACELAB MEDICAL BISPECTRAL INDEX (BISX) ANALYSIS - The predicate device. The SpaceLab monitor was cleared for marketing by the FDA (K060900)

Intended Use of the Envoy Patient Monitor

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, CO, Pulse Oximetry and EtCO2. The ENVOY may be used to monitor a wide range of patient conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

Indications for use - BIS Module

The Mennen Medical Envoy BIS module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The BIS Index, a processed parameter may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

  • The Bispectral Index is a complex technology, intended for use only as an adjunct to clinical judgment and training.

{1}------------------------------------------------

    • In addition, the clinical utility, risk/benefit, and application of this device have not undergone full evaluation in the pediatric population.
      *The Intended Use of the Envoy monitor as indicated above are same as the Indications For Use.

Device Description - Envoy Patient Monitor with BIS module

The Envoy is a multiparameter physiological patient monitor, capable of monitoring:

  • ECG/Heart Rate .
  • Invasive blood pressure .
  • Non-invasive blood pressure ●
  • Respiration ●
  • Pulse oximetry
  • Two temperature channels ●
  • Cardiac output ●
  • EtCo2 .
  • Spirometry ●
  • EEG ●
  • BIS module (new subject of this application) .

The Envoy bedside patient monitor consists of a main processing unit, a mountable color monitor, and a module rack housing the various Mennen Medical plug-in vital signs modules. The modules monitor the patient's vital signs. Up to six internal modules can be plugged into a module rack. The Envoy can accommodate two module racks. The vital sign data derived from the modules by the Envoy are presented on the monitor as waveform and numeric displays. The Envoy vital signs modules acquire vital signs data from the patient, and display their waveforms and alarms indications on the Envoy display unit. Vital signs and waveform information are displayed simultaneously on the Envoy Display Unit. Up to 8 traces can be displayed at any one time.

The vital signs modules interface with readily available physiologic transducers through electrically isolated patient input connections. After amplification, the signals are digitized, analyzed and displayed. All processing and alarm determination for ECG, Respiration and Invasive Blood Pressure is made using proprietary algorithms and software based on previously marketed Mennen Medical monitoring devices tested against well known and accepted data bases that present representative examples of waveform artifact to be encountered in real case conditions. The SpO2, Non-Invasive Blood Pressure, ExCO2 and Spirometry modules incorporate software and/or hardware technology developed by vendors whose products are marketed in the USA.

{2}------------------------------------------------

Food and Drug Administration Device Modification - Envoy Patient Monitor 510(k) for addition of BIS module

Information from each vital sign is presented in a separate portion of the display. Each vital sign is labeled for identification and numeric value. Displayed Vital sign information can include: Primary Vital Sign Name, waveform, Vital Sign Numeric Value, Alarm Status Message.

Operation of the Envoy is accomplished by interaction with front panel controls on the main processor unit. A quick-knob control allows direct interaction with displayed menus for direct parameter selection and setup. Where manual entry of alphanumeric information is required, a ` menu keyboard menu is displayed.

The Envoy is a reusable, software driven, patient monitor, intended for use as part of a physiological monitoring system in a hospital environment. As such it is not a life supporting, nor life sustaining device; nor is it implantable and therefore sterility is not a consideration.

The Envoy complies with (amongst others) the following voluntary standards:

  • IEC 60601-1: Medical Electrical Equipment General Requirements for Safety .
  • IEC 60601-1-2: General Requirements for Safety. Collateral standard: Electromagnetic . compatibility - Requirements and tests.
  • IEC 60601-2-27: Medical Electrical Equipment Part 2: Particular Requirements for the . Safety of Electrocardiographic Monitoring Equipment.
  • IEC 60601-2-30: Requirements for Automatic Cycling Indirect Blood Pressure . monitoring.
  • AAMI/ANSI SP-10/A I: Electronic or Automated Sphygmomanometers .
  • IEC 60601-2-34: Requirements for Invasive Blood Pressure monitoring. .
  • AAMVANSI ES1: Safe Current Limits for Electromedical Apparatus, and .
  • AAMI/ANSI EC13: Cardiac Monitors, Heart Rate Meters and Alarms. .

The Envoy is not a kit and does not contain any drug or biological products. The BIS module of the Envoy patient monitor is not sold as a stand alone device, but as part of a multiparameter physiological patient monitoring system (Envoy).

In chapter 1 of the Envoy Operating Manual, the following Prescription Notice appears: "Federal United States law restricts the sale and use of this instrument to qualified medical personnel only"

Functional description of the new Envoy BIS Module:

{3}------------------------------------------------

(Interface to Aspect BISx device cleared in K 040183)

The BIS Module is used to monitor dual channel EEG waveform and the BIS index, used to estimate the level of conciseness of patient under anesthesia, or patients in the ICU that may be with limited conciseness.

The BIS index together with several quality parameters are displayed and stored by the Envoy monitor.

ParameterRangeDescription
BIS0 - 99Bispectral Index: The measure of consciousness of apatient, (0 = no brain activity), (100 = fully conscious).
EMG30 - 55 dBElectromyography: The absolute power of muscleactivity and artifacts in the 70 - 110 Hz range. Value is indB with respect to 0.0001 µV2.
SQI0 - 100 %Signal Quality Index: The percentage of good epochsand suppressed epochs in the last 120 epochs collectedthat could be used in the Bispectral Index calculation.
SR0 - 100 %Suppression Ratio: The percentage of epochs in thepast 63 seconds In which the EEG signal is consideredsuppressed
BC0-30Burst Count: The number of EEG bursts in the lastminute. An EEG burst is a momentary period of EEGactivity among isoelectric or flat EEG. Blanked if SR is lessthan 5. Activated by connection of an Extend sensor.
SEF0.5 - 30 HzSpectral Edge Frequency: The frequency at which 95%of the total power lies below it and 5% lies above it.

The parameters displayed and stored by the Envoy monitor are the following:

To get these parameters we use Aspect sensor and BISx unit attached to the Envoy BIS module with an interface cable produced by Aspect.

The BISx Is a product of Aspect and is sold to Mennen Medical under OEM agreement.

The BIS sensors, that consist of a set of 4 electrodes, attached to the patients forehead, are Aspect products and will be sold to end users by Aspect directly and not by Mennen Medical.

The function of the Envoy BIS module and Envoy display and storage capabilities is to display the BIS parameter and the quality parameters. To provide alarm limits for the BIS index and provide visual and audible alarms in case of the BIS index being out of the preset range, or in cases that the quality parameters are in such range that they limit the reliability of the BIS Index.

The decision on artifacts and limited rellability are provided to the Envoy BIS module by the BISx. Those can not be set by Mennen Medical Envoy monitor. The clinical BIS alarm limits are set by the user, on the Envoy monitor.

{4}------------------------------------------------

The Description of the BIS measurement system is described below:

  • The BIS sensor (Made and sold by Aspect) is attached to the patients forehead. (See attached
    Aspect document: "The BIS"" Family of Sensors") 1.
  • BISx (Figure 1 OEM by Aspect) is receiving, amplifying and digitizing the EEG brain signals. It 2. analyses the EEG and creates the BIS index and the quality parameters and sends them to the BIS module by RS232 protocol
  • Envoy BIS Module receives the BISx output and transfers it to the Envoy bedside computer for 3. display and storage.
    1. BIS Display and storage: BIS is displayed as a number. EMG and SQI are displayed as vertical bars, and SR and BC as numbers in the BIS display area.
  • Clinical and technical alarm messages are displayed in the BIS display area. 5.
  • Continuous graphic trend of the BIS inclex and quality parameters are also available on the screen 6. under the label CRG.
    1. The real time EEG signal is continuously displayed on the monitor screen.
  • In edition, the Envoy provides long term storage and display of the BIS index and its related 8. parameters as numeric Charts and graphic Trends.

Image /page/4/Picture/10 description: The image shows a black and white photo of an electrical device with a cord. The device has a circular shape with a lighter-colored center and darker edges. The cord is coiled on either side of the device. The cord appears to be thick and insulated.

Figure 1: BISx™ device

Substantial Equivalence Discussion

Comparison of the Envoy BIS Module with the predicated device Spacelab Bispectral Index (BISx) analysis module 91482

Page 5 of 14

{5}------------------------------------------------

Comparison of the indication for use

Spacelab Indication for Use

The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482

is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482

may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

Mennen Indication for use

The Mennen Medical Envoy BIS module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The Mennen Medical BIS module may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

Conclusion

The indication for use of Mennen Medical BIS module and of the predicated device, has the same wording. The only difference between the two is the module and company name.

Spacelab Bispectral Index(BISx) AnalysisEnvoy BIS Module
Part NumberModule 91482551-147-000
510KK060900
Features
Display format:Single zoneBIS Numeric and BIS TrendBIS numeric + one EEG +EMG vertical bar + EMGvertical bar + SR, BCnumeric
Spacelab Bispectral Index(BISx) AnalysisEnvoy BIS Module
Single zoneWaveform + trendBIS numeric + Two EEG +EMG vertical bar + EMGvertical bar + SR, BCnumeric + Trend of BISand EMG
Dual zoneN/ABIS numeric + Two EEG +EMG vertical bar + EMGvertical bar + SR, BCnumeric
Three zoneBIS + EMG trends +one EEGBIS numeric + Two EEG +EMG vertical bar + EMGvertical bar + SR, BCnumeric + Trend of BISand EMG
Four zoneN/ABIS numeric + Two EEG +EMG vertical bar + EMGvertical bar + SR, BC,numeric + Trend of BISand EMG + Trend of anytwo other parameters
Waveform + TrendN/AEEG waveform and BIS +EMG graphic trend
Wave scale+/- 2, 5, 10, 20, 50, 100, 200,400 μV5, 10, 25, 50, 100 µV/cm
EEG sweep speed12.5, 25, 50 mm/sec6.25, 12.5, 25 mm/sec
BIS Task windowBIS, EEG waveform, EMG, SQI,SR, SEF, MF, Alarm limitsBIS, EEG waveform,EMG, SQI, SR, BC,Alarm limits,Alarm messages
Data Storage24 hours of:BIS,Spectral Edge Frequency (SEF),Median Power Frequency (MF)Electromyography strength(EMG)Signal Quality Index (SQI)Suppression Ratio (SR)EEG waveform - 45 days3 month of:BIS,Spectral Edge Frequency(SEF),Electromyography strength(EMG)Signal Quality Index (SQI)
Spacelab Bispectral Index(BISx) AnalysisEnvoy BIS Module
Burst Count (BC)Suppression Ratio (SR)
Burst Count (BC)
ElectrodeImpedance testingManual or Auto by BISxAuto on connection ofsensor and Manual anytime
PatientcompatibilityAdult and pediatricSame
Input SpecificationSet By BISxSet By BISx
Number of channelsTwoSame
Sweep speed15, 30, 50 mm/second6.25, 12.5, 25 mm/second
EEG Input signal+/- 1 µV to +/-400 µVSame
EEG Bandwidth0.25 to 100 HzSame
EMG bandwidth70 to 110 HzSame
DC offset+/- 300 mVSame
Input Impedance>50 MΩSame
Input Capacitance<100 pFSame
Common Moderejection> 110dBSame
Input Noise<0.3 µV RMS (2.0µV peak topeak to peak)Same
Smoothing rate10, 15, or 30 secondsSame
ElectricalSpecification
Patient leakagecurrant< 100 μV
Isolation4000VAC
Operating Voltage+5 VDC , +/- 12 VDCSame
Power consumption4.5 Watt maximum1.5 Watt
AlarmsAudible for High and Low BISAudible and visual (redbackground)
Caution alarmAudio and visual (yellowbackground) and alarm
Spacelab Bispectral Index(BISx) AnalysisEnvoy BIS Module
message
ClassificationMDD Class IIbEN 60601-1Externally powered, rated forcontinuous operationSame
Aspect BISx podType BF, defibrillator proof,Body floating applied partSame
EnvironmentalRequirement
StorageTemperature: -25° to 60° CHumidity : 95%(non-condensing)Altitude: 0 to 12,192 mTemperature: -15° to 60° CHumidity : 10 to 95%(non-condensing)Altitude: -350 to 5,000m
OperatingTemperature: 0° to 40° CHumidity : 95% (non-condensing)Altitude: 0 to 4,572 mTemperature: +5° to 40° CHumidity : 10 to 95% (non-condensing)Altitude: -350 to 3,050 m

Specification Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Envoy BIS module and Spacelab BIS analysis module: Similarities and Differences:

Similarities:

The following technological and other characteristic/features apply to both Envoy BIS module and Spacelab 91482 Bispectral Index (BISx) module:

  • . Both use Aspect Medical Systems BISx pod as their input device
  • Both provide the BISx pod its power (5 Volt)
  • Both receive from the BISx by serial data communication the EEG waveform and the ● BIS parameter together with its quality and related parameters.
  • . Both have the capability to display EEG waveform and BIS numeric parameters
  • Both enable correlation between BIS and other vital signs displayed and stored by the � monitor to which the BIS module is inserted.

Differences:

The difference between the BIS parameters display of the Envoy monitor and the Spacelab monitors are mainly due to the format of the display of the different monitors.

{9}------------------------------------------------

  • The Spacelab BIS module is compatible with a family of Ultraview monitors, while the . Mennen Medical BIS module is for the Envoy monitor only.
  • There are differences in the display ontions of the Spacelab Ultraview BIS display that . has two display formats: Single Zone and Three-zone while the Envoy monitor has at least 6 options of display for BIS value and parameters. The differences, however, do not change the efficiency of BIS and EEG displays of both monitors.
  • Data storage of the Spacelab Ultraview monitor is for shorter duration of the Envoy. .

Conclusion of comparison

We consider the Envoy BIS module to be substantially equivalent to the Spacelab 91482 Bispectral Index module.

Any differences between to two monitors for the display of BIS, EEG and related parameters do not raise any new issues of safety and effectiveness.

The Intended use of the Mennen medical Envoy BIS module and the predicated device is the same.

Verification, Validation and Testing

The Envoy BIS module has been subject to extensive performance testing to ensure that: 1. The acquisition and display of the patient data and waveforms by the Envoy with BIS - 550-OPT-400 are equivalent to the predicate device Spacelab with BIS module.

At the system level, SW Validation of the performance of the Envoy with 550-OPT-400 as compared to the Spacelab with BIS module, was carried out in accordance with the test plan described in the Mennen Medical Validation Test Procedure for the Envoy BIS module.

The SW Test Description for the Envoy with BIS - 550-OPT-400 was derived from the SW Test Description for the Envoy, with the necessary addition of the BIS measurements

Final testing for the Envoy BIS module included performance tests designed to ensure that the device meets all functional requirements and performance specifications, in accordance with the requirements of the Final Test Procedure for the Envoy BIS module. Electrical Safety testing and BMC testing were performed by an independent testing laboratory (Standard Institute of Israel SII) to ensure that the device complies to applicable industry and safety standards.

Proposed Labeling

The system will be called Envoy with BIS module p/n 550-OPT-400

Page 2-2 of the introduction to the Envoy User Guide contains the following Prescription Notice: "Federal United States law restricts the sale and use of this instrument to qualified medical personnel only."

The following symbols appear on page 2-4 of the Envoy User's Guides under the section entitled "Label Symbols".

{10}------------------------------------------------

SymbolDescriptionLocation of Symbol
Image: [Alternating Current Symbol]Alternating CurrentRear of the Processing unit, IsolationTransformer, and Module Rack
Image: [Equipotential Symbol]EquipotentialOn the rear of the Processing unit
Image: [Attention Symbol]Attention, consult accompanying documents (Service to be performedby qualified technician, consult servicemanual before removing cover)On Isolation Transformer and Processing unit.
Image: [Off Symbol]Off (power disconnection from mainpower supply)On right of Processing unit
Image: [On Symbol]On (power connection to the mainpower supply)On the right of the Processing unit.
Image: [Type BF Symbol]Type BF applied partdefibrillator-proofOn NIBP and Sp02 modules.
Image: [Type CF Symbol]Type CF applied part - direct cardiacapplicationdefibrillator-proofOn ECG, and Dual BP and CO/2TMP modules
Image: [Fuse Symbol]FuseOn rear of Processing unit and Isolation Transformer

. .

Image /page/10/Picture/2 description: The image shows a three-dimensional rectangular prism. The letters "BIS" are written on the front face of the prism. The front face of the prism is covered in horizontal lines.

Page 11 of 14

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows a product label for Mennen Medical Ltd. The model number is BIG, and the part number is 551-147-000. The serial number is 12348. The label also includes the CE marking, the revision number REV-A, and the manufacturing date JUNE 2007.

Voluntary Standards

Appropriate voluntary standards for this device, to which conformance have been demonstrated:

* IEC 60601-1 (2005)

General Requirement for Safety for Medical Electrical Systems - part 1, Amendment 1 - 1991-11 Amendment 2 - 1995-03

* IEC 60601-1-1 (2000)

Medical electrical equipment - Part 1-1: General requirements for safety - Collateral standard: Safety requirements for medical clectrical systems

* IEC 60601-1-2 (2007):

Medical electrical equipment. General requirements for Safety. Electromagnetic Compatibility Requirements and Tests.

☆ IEC 60601-2-27 (2005): Medical electrical equipment, Part 2, Requirements for safety of electrocardiograph monitoring equipment.

{12}------------------------------------------------

    • IEC 60601-2-30 (1999): Medical electrical equipment, Part 2 - requirements for safety of automatic cycling indirect blood pressure monitoring equipment
    • IEC 60601-2-34 (2005): Medical electrical equipment, Part 2 - Particular requirements for the safety of direct blood pressure monitoring equipment
  • や IEC 60601-2-49 (2006): Particular Requirements for the safety of multifunction patient monitoring equipment

Indications for use - BIS Module

The Mennen Medical Envoy BIS module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The Mennen Medical BIS module may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

  • The Bispectral Index is a complex technology, intended for use only as an adjunct to clinical judgment and training.
  • In addition, the clinical utility, risk/benefit, and application of this device have not undergone full evaluation in the pediatric population.

*The Intended Use of the Envoy monitor as indicated above are same as the Indications For Use.

Confidentiality

Mennen Medical Ltd. considers its intent to market the Envoy with 550-OPT-400 BIS module option, to be confidential commercial information. The company has not disclosed its intent to market this device to anyone except its employees, others with a financial interest in the company, its advertising and law firms, and its consultants. Mennen Medical, therefore, requests the FDA not disclose the existence of this application until such time as final action on the submission is taken.

In addition, some of the material in this application may be trade secret or confidential commercial or financial information within the meaning of 21 CFR § 20.61 and therefore not disclosable under the Freedom of Information Act, even after the existence of the application becomes public. We ask that FDA consult with the Company as provided in 21 CFR § 20.45 before making any part of this submission publicly available.

{13}------------------------------------------------

Attachment:

Envoy Patient Monitor, User Manual – part 13 d FTP for Bis option in Envoy Patient Monitor - Part 16 j 2 Changes to Envoy Patient Monitor - BIS Risk Analysis - part 16 d

Page 14 of 14

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mennen Medical Ltd. c/o Mr. Ifat Oren 4 Ha-Yarden Street, Yavne P.O. Box 102 Rehovot 76100 Israel

  • 9 2012 APR

Re: K071899

Trade/Device Name: BIS Module for Envoy Patient Monitor Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, OMC, OLT, MHX, ORT Dated (Date on orig SE ltr): September 17, 2007 Received (Date on orig SE ltr): September 21, 2007

Dear Mr. Oren:

This letter corrects our substantially equivalent letter of October 22, 2007.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{15}------------------------------------------------

Page 2 - Mr. Ifat Oren

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{16}------------------------------------------------

Indications for Use

510(k) Number (if known): K071899

Device Name: BIS Module for Envoy Patient monitor

Indications For Use:

We submit that the intended use and the indications for use of the Envoy have been affected by the changes only by the addition of the new BIS parameter. Other aspects of the Envoy monitor where not changed.

Envoy Indications for Use:

The ENVOY Monitor is a physiological patient monitor intended to be used for monitoring vital signs of critically ill adult and pediatric patients in the hospital environment, such as: ECG/Heart Rate, Invasive Blood Pressure, Respiration, Temperature, Noninvasive Blood Pressure, CO, Pulse Oximetry, and EtCO2. The ENVOY may be used to monitor a wide range of patient conditions in many different clinical specialties within the hospital. The device is intended for use by qualified health care providers, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.

Indications for use - BIS Module

The Mennen Medical Envoy BIS module

is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.

The BIS Index, a processed parameter may be used as an aid in monitoring the effects of ccrtain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

    • The Bispectral Index is a complex technology, intended for use only as an adjunct to clinical judgment and training.
    • In addition, the clinical utility, risk/benefit, and application of this device have not undergone full evaluation in the pediatric population.

*The Intended Use of the Envoy monitor as indicated above is same as the Indications For Use.

Prescription UseAND/OROver-The-Counter Use
YES (Part 21 CFR 801 Subpart D)NO (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off) Office of Device Evaluation (ODE)

Division of General, Restorative, and Neurological Devices

Page 1 of 1

510(k) NumberK071895
------------------------

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).