K Number
K071781

Validate with FDA (Live)

Device Name
PF4 IGG
Date Cleared
2007-12-19

(170 days)

Product Code
Regulation Number
864.7695
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PF4 IgG™ is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect antibodies reactive with Platelet Factor 4 (PF4) when it is complexed to polyanionic compounds such as polyvinyl sulfonate (PVS). These antibodies are found in some patients undergoing heparin therapy.

PF4 IgG™ is designed as a solid phase enzyme-linked immunosorbent assay (ELISA). This product is intended to be used as an in vitro diagnostic kit by hematology, coagulation, or other pathology laboratories to assist in screening patient samples for the presence of heparin-associated antibodies commonly found in patients with heparin-induced thrombocytopenia or thrombosis.

PF4 IgG™ is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum. The presence of heparin associated antibodies are commonly found in patients with Heparin Induced Thrombocytopenia (HIT).

Device Description

The PF4 IgGTM assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG™ ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG™ kit contains all of the reagents necessary to perform the assay.

In the PF4 IgG™ assay, a complex of PF4/PVS, which has been immobilized in the microwells serve as a target for the binding of antibodies associated with Type II HIT.

In the PF4 IgG™ assay, patient serum is first diluted (1:4), with the specimen diluent provided in the kit. The diluted sample is then added to microwells to which Platet Factor 4 (PF4) complexed to polyvinyl sulfonate (PVS) has been immobilized. The sample is then incubated for 30 minutes at 37°C. If an antibody which recognizes a site on PF4/PVS complex is present in the patient sample, binding will occur. Following this incubation, a wash step then removes any unbound antibodies. A goat anti-human IgG - alkaline phosphatase conjugate is then added to the wells. The conjugate is incubated for 30 minutes at 37°C. Following this incubation, a wash step then removes any unbound conjugate. The alkaline phosphate substrate, p-nitrophenyl phosphate (pNPP) is then added to the microwells. After a 30 minute incubation at room temperature (22 -- 25°C), the reaction is stopped by addition of the stopping solution (3 M sodium hydroxide). The optical density of the color that develops is measured in a spectrophotometer at 405 or 410 nm using a reference wavelength of 490 nm.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (PF4 IgG™ vs. SRA)Reported PF4 IgG™ Performance (vs. SRA)Acceptance Criteria (PF4 IgG™ vs. PF4 ENHANCED®)Reported PF4 IgG™ Performance (vs. PF4 ENHANCED®)
Sensitivity (Co-Positivity)Not explicitly stated (but compared)100% (95% CI: 84.5 - 100.0%)Not explicitly stated (but compared)74% (95% CI: 61.0 - 83.4%)
Specificity (Co-Negativity)Not explicitly stated (but compared)90% (95% CI: 85.1 - 93.3%)Not explicitly stated (but compared)100% (95% CI: 97.8 - 100.0%)
% AgreementNot explicitly stated (but compared)91%Not explicitly stated (but compared)93%
Assay Imprecision (O.D.)≤10% CV total imprecision≤10% CV total imprecisionNot applicableNot applicable
Reportable Results100% agreement100% agreementNot applicableNot applicable

Note: The document does not explicitly state pre-defined acceptance criteria values for sensitivity, specificity, and agreement. Instead, it presents the results of the comparison studies and concludes that the device "performs comparable to the predicate device." The conclusion for assay imprecision and reportable results states that the assay "showed acceptable assay imprecision" and "100% agreement."

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: 229 serum samples.
  • Data Provenance: The samples were obtained from the BloodCenter of Wisconsin (BCW) and were from patients receiving heparin treatment. The data is retrospective as the samples had been previously tested by BCW for PF4/heparin antibodies using the SRA. The country of origin is the USA.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number of experts or their qualifications for establishing the ground truth using the Serotonin Release Assay (SRA). It refers to the SRA as an "in house assay" at BCW.

4. Adjudication method for the test set

The document does not describe an adjudication method for conflicting results. The SRA results from BCW were used as the primary comparator (gold standard). For the PF4 IgG™ and PF4 ENHANCED® assays, samples were tested in duplicate, and the mean of the O.D. values was obtained. Results were considered positive if the mean O.D. value was ≥0.400.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This study describes an in vitro diagnostic (IVD) ELISA test, not an AI-assisted diagnostic tool that would involve human readers for interpretation of results.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This study describes a standalone assay's performance, which is an analogous concept to an algorithm-only performance for an IVD test. The PF4 IgG™ assay itself is the "algorithm only" in this context, providing a quantitative optical density value that is then interpreted as positive or negative based on a cutoff. No human-in-the-loop performance data is provided beyond implicitly trained lab technicians performing the assay.

7. The type of ground truth used

The primary ground truth for the comparison of methods study was the Serotonin Release Assay (SRA). The document states that the SRA is "considered to be the gold standard for testing for antibodies that can cause HIT."

8. The sample size for the training set

The document does not explicitly describe a "training set" in the context of machine learning or algorithm development. However, for determining the normal range and assay cutoff, 120 serum samples from normal healthy individuals were used. This set of samples was used to statistically establish the assay's cutoff, which plays a role similar to calibrating or "training" a threshold.

9. How the ground truth for the training set was established

For the "training set" (120 normal healthy individuals used for normal range and cutoff determination):

  • The ground truth was established by defining these individuals as "normal healthy" and their serum samples were tested with the PF4 IgG™ and PF4 ENHANCED® assays.
  • The normal range was determined non-parametrically using these samples, and the cutoff for the assays (≥0.400) was set based on the upper end of these calculated normal ranges.
  • Calculations for the normal range were performed by GTI's Manager of Clinical and Scientific Affairs (Melissa Pressman, Ph.D.), using the Med Calc software program. This implies statistical analysis of the optical density values from these known normal samples.

{0}------------------------------------------------

K071781

DEC 1 9 2007

Image /page/0/Picture/2 description: The image shows the logo for GTI Diagnostics. The logo consists of a stylized image of a person inside of a square. Below the image is the text "GTI. DIAGNOSTICS" in a bold, sans-serif font. The tagline "Good science starts with people." is below the company name.

800 . 233 . 1843 in usa 262 . 754 . 1000 TEL 262 . 754 . 9831 FAX gti@gtidiagnostics.com FMAII gtidiagnostics.com ﻟﻘﺎ ﻳﺪﻳﺮ ﺍﻟﻤﺪﻳﻨﺔ

20925 Crossroads Circle, Suite 200 Waukesha, Wi 53186 USA

510(k) Summary

  • I. Submitter:
Owner's Name:Genetic Testing Institute, Inc. (GTI)
Address:20925 Crossroads Circle, Waukesha, WI 53186
Phone:262.754.1000
Fax:262.754.9831
Name of Contact Person:Suzette C. Chance, Ph.D.
Date Summary Prepared:June 15, 2007

II. Name of Device:

Device Name:PF4 IgG™ Solid Phase ELISA
Proprietary Name:PF4 IgG™
Classification Name:Platelet Factor 4 Radioimmunassay
Product Code:LCO

III. Name of predicate device for claiming equivalence

GTI-PF4 ENHANCED® (K053559)

IV. Description of Device:

The PF4 IgGTM assay is an Enzyme Linked Immunosorbent Assay (ELISA). The PF4 IgG™ ELISA is intended to detect IgG antibodies in human serum that react with Platelet Factor 4 (PF4) when it is complexed to heparin or other polyanionic compounds. The PF4 IgG™ kit contains all of the reagents necessary to perform the assay.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains a logo for GTI Diagnostics. The logo features a stylized graphic above the text. Below the company name is the tagline "Good science starts with people."

Antibodies that react with PF4 when it is complexed to heparin are found in some patients undergoing heparin therapy. The presence of these antibodies has been shown to be associated with heparin induced thrombocytopenia Type II (Type II HIT). Heparin Induced Thrombocytopenia (HIT) is an adverse reaction that can occur in patients that are undergoing heparin therapy. HIT occurs in up to 3% of patients that are receiving heparin and can result from treatment with either low molecular weight heparin (LMWH) or unfractionated heparin (UFH). HIT is usually associated with thrombocytopenia and in some cases the more severe complications of arterial or venous thrombosis. The pathophysiology of HIT is linked to the formation of antibodies which are specific for epitopes formed when heparin binds to PF4. These antibodies can subsequently bind to platelets via the FcyRIIa receptor resulting in platelet activation.

It has been previously shown that antibodies which bind to PF4/heparin complexes also bind to PF4 when it is complexed with other polyanionic compounds such as polyvinyl sulfonate (PVS). In the PF4 IgG™ assay, a complex of PF4/PVS, which has been immobilized in the microwells serve as a target for the binding of antibodies associated with Type II HIT.

In the PF4 IgG™ assay, patient serum is first diluted (1:4), with the specimen diluent provided in the kit. The diluted sample is then added to microwells to which Platet Factor 4 (PF4) complexed to polyvinyl sulfonate (PVS) has been immobilized. The sample is then incubated for 30 minutes at 37°C. If an antibody which recognizes a site on PF4/PVS complex is present in the patient sample, binding will occur. Following this incubation, a wash step then removes any unbound antibodies. A goat anti-human IgG - alkaline phosphatase conjugate is then added to the wells. The conjugate is incubated for 30 minutes at 37°C. Following this incubation, a wash step then removes any unbound conjugate. The alkaline phosphate substrate, p-nitrophenyl phosphate (pNPP) is then added to the microwells. After a 30 minute incubation at room temperature (22 -- 25°C), the reaction is stopped by addition of the stopping solution (3 M sodium hydroxide). The optical density of the color that develops is measured in a spectrophotometer at 405 or 410 nm using a reference wavelength of 490 nm.

V. Intended Use

PF4 IgG™ is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect antibodies reactive with Platelet Factor 4 (PF4) when it is complexed to polyanionic compounds such as polyvinyl sulfonate (PVS). These antibodies are found in some patients undergoing heparin therapy.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for GTI Diagnostics. The logo features a stylized human figure within a square shape. Below the logo, the text "GTI DIAGNOSTICS" is printed in a bold, sans-serif font, followed by the tagline "Good science starts with people."

VI. Support of substantial equivalence based on comparison of features, characteristics and components to the predicate device:

A comparison of the features and characteristics of the two devices can be summarized as follows:

PF4 ENHANCED®PF4 IgGTM
Intended UsePF4 ENHANCED® is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect antibodies reactive with Platelet Factor 4 (PF4) when it is complexed to polyanionic compounds such as polyvinyl sulfonate (PVS). These antibodies are found in some patients undergoing heparin therapy.PF4 IgGTM is a qualitative solid phase enzyme linked immunosorbent assay (ELISA) designed to detect IgG antibodies reactive with Platelet Factor 4 (PF4) when it is complexed to polyanionic compounds such as polyvinyl sulfonate (PVS). These antibodies are found in some patients undergoing heparin therapy.
Indications for UsePF4 ENHANCED® is designed as a solid phase enzyme-linked immunosorbent assay (ELISA). This product is intended to be used as an in vitro diagnostic kit by hematology, coagulation, or other pathology laboratories to assist in screening patient samples for the presence of heparin-associated antibodies commonly found in patients with heparin-induced thrombocytopenia or thrombosis.PF4 IgGTM is designed as a solid phase enzyme-linked immunosorbent assay (ELISA). This product is intended to be used as an in vitro diagnostic kit by hematology, coagulation, or other pathology laboratories to assist in screening patient samples for the presence of heparin-associated antibodies commonly found in patients with heparin-induced thrombocytopenia or thrombosis.
TechnologyELISA with a colorimetric measurement systemELISA with a colorimetric measurement system
Reportable ResultsQualitative assay; results arereported as positive ornegativeQualitative assay; results arereported as positive ornegative
Packaging Configuration13 and 45 Test kits13 and 45 Test kits
Reagents:
Microwell stripsImmobilized PF4/PVSComplexImmobilized PF4/PVSComplex
Concentrated Wash Solution10X Tris Buffer, NaCl,Tween 20, 1% NaN310X Tris Buffer, NaCl,Tween 20, 1% NaN3
Specimen DiluentPhosphate Buffered Saline,0.05% NaN3Phosphate Buffered Saline,0.05% NaN3
Substrate BufferDiethanolamine andmagnesium chloride, 0.02%NaN3Diethanolamine andmagnesium chloride, 0.02%NaN3
SubstratePNPP (crystalline powder)PNPP (crystalline powder)
Stopping Solution3 M NaOH3 M NaOH
Positive Serum ControlHuman serum containingBovine Albumin and 0.1%NaN3Human serum containingBovine Albumin and 0.1%NaN3
Negative Serum ControlHuman serum containing0.1% NaN3Human serum containing0.1% NaN3
ConjugateGoat anti-human Ig G+A+Mconjugated to alkalinephosphatase enzymeGoat anti-human Ig Gconjugated to alkalinephosphatase enzyme

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for GTI Diagnostics. The logo consists of a black abstract figure above the text "GTI. DIAGNOSTICS". Below the text is the phrase "Good science starts with people."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for GTL Diagnostics. The logo consists of a stylized graphic above the text "GTL DIAGNOSTICS". Below the company name is the tagline "Good science starts with people:".

The similarities between these two products can be summarized as follows:

  • Both PF4 IgG™ and PF4 ENHANCED® have similar intended uses and the same . indications for use
  • PF4 IgG™ and PF4 ENHANCED® use the same technology (ELISA) and assay . steps
  • . PF4 IgG™ and PF4 ENHANCED® have identical reagents with the exception of the conjugate used for antibody detection

The difference between the two products can be summarized as follows:

The PF4 ENHANCED® kit utilizes a conjugate that detects bound human IgG, IgA, and/or IgM antibodies (goat anti-human IgG+A+M conjugated to alkaline phosphatase). The PF4 IgG™ kit utilizes a conjugate that detects bound human IgG (goat anti-human IgG conjugated to alkaline phosphatase).

VI. Support of substantial equivalence with performance data:

The details of each of the following studies are covered in Section 7: Performance Data of this 510(k). Only a brief summary of these studies is provided in this section.

Assay Precision

Description of Study

Three samples of varying antibody concentrations were prepared by diluting a patient sample (sera) containing a high level of an anti-PF4/heparin antibody into a pool of normal serum containing no PF4/heparin antibody. This sample was diluted to obtain 3 separate samples that had low positive reactivity (approximately 0.5 O.D. or 0.1 O.D. units above the cutoff), medium positive reactivity, and high positive reactivity. In addition to these samples, the positive and negative controls provided in the kit were used for this study. The positive control consists of a human serum containing a PF4/heparin IgG antibody, whereas the negative control consists of a sera sample containing no PF4/heparin antibody.

Each sample or control was tested in duplicate in the PF4 IgGTM assay in 10 separate assavs.

Results and Analysis

To obtain imprecision of the O.D. values, the data were analyzed by ANOVA according to CLSI Document EP-5A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. In addition, the reportable result (positive or negative) was analyzed for agreement within and between runs according to CLSI Document

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains a logo and text. The logo is a stylized black and white image that resembles a person with outstretched arms inside a shield. Below the logo, the text "GTI. DIAGNOSTICS" is printed in a simple, sans-serif font. Underneath this, the phrase "Good science starts with people." is written in a smaller font size.

EP 12-A Vol. 22, No 14, User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline.

The calculations of imprecision for the O.D. values showed that the assay demonstrated ≤10% cv total imprecision for samples of all levels of PF4/PVS antibody reactivity. In addition, the correct reportable result was obtained for each result of each assay. There was 100% agreement between all reportable results (within-run and between-run) for each sample tested.

Conclusions

The PF4 IgG™ assay showed acceptable assay imprecision of the O.D. values as well as the reportable results.

Normal Range and Assay Cutoff

Description of Study:

One hundred and twenty serum samples were obtained from normal healthy individuals and were tested (in duplicate) in the PF4 IgG™ and the PF4 ENHANCED® assays. The mean of the duplicates was obtained for each sample tested and the results were analyzed for normality of the distribution of the O.D. values.

Results and Analysis:

The results showed that neither set of data were normally distributed. A nonparametric analysis was used to determine the normal range for each assay. The normal ranges were calculated to be 0.142 - 0.352 for PF4 IgG™ and 0.332 - 0.407 for PF4 ENHANCED®. The calculations were based on a non-parametric 95% reference interval with a 90% confidence. The cutoff for the assays was then taken to be the upper end of the normal range (0.352 for PF4 IgG™ and 0.407 for PF4 ENHANCED®). Calculations for the normal range were performed by GTI's Manager of Clinical and Scientific Affairs (Melissa Pressman, Ph.D.), using the Med Calc software program.

Conclusions:

While statistically different, the upper end of the normal ranges for the PF4 ENHANCED® and the PF4 IgG™ assays were not significantly different. Therefore, the cutoff of ≥0.400 was confirmed for the PF4 ENHANCED® assay and established for the PF4 IgGTM assay.

Comparison of Methods Study

Accuracy was demonstrated by a study in which PF4 IgG™ was compared to both the PF4 ENHANCED® assay and the Serotonin Release Assay (SRA). The following provides a

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains a logo for CTI. DIAGNOSTICS. The logo consists of a stylized human figure above the text "CTI. DIAGNOSTICS". Below the company name is the tagline "Good science starts with people."

description of the study, the results, and conclusions. This study was conducted as an internal study at GTI.

Description of Study:

The samples used in the method comparison study consisted of a set of 229 different sera obtained from the BloodCenter of Wisconsin (BCW). Each sample was provided in a single aliquot and was stored frozen at -80°C until the time it was tested. These samples were obtained from patients receiving heparin treatment and were originally tested by the BCW for the presence of PF4/heparin antibodies by the SRA. The SRA is an "in house assay" that detects the presence of antibodies in serum that are capable of activating platelets. The SRA is considered to be the gold standard for testing for antibodies that can cause HIT.

Samples were tested in duplicate in the PF4 IgG™ and PF4 ENHANCED® assays at GTI. The mean of the O.D. values for each sample was obtained. Results were considered to be positive in the PF4 IgG™ and PF4 ENHANCED® assays if the mean of the O.D. value was ≥0.400. Results from the SRA were based on the data provided by the BCW.

Results and Analysis:

Analysis of the data was performed using 2x2 tables. The PF4 IgG™ assay results were compared to the PF4 ENHANCED® and the SRA. In addition, the PF4 ENHANCED® assay results were compared to those of the SRA. The calculations for co-positivity, conegativity, and % agreement for each comparison are shown in the table below.

PF4 IgGTMVersus SRAPF4ENHANCED®Versus SRAPF4 IgGTMversusPF4ENHANCED®
Sensitivity (Co-Positivity)100%100%74%
95% Confidence Interval84.5 - 100.0%84.5 - 100.0%61.0 - 83.4%
Specificity (Co-negativity)90%83%100%
95% Confidence Interval85.1 - 93.3%77.0 - 87.2%97.8 - 100.0%
% Agreement91%84%93%

Conclusions:

The PF4 IgG™ assay showed excellent specificity (co-negativity), and agreement with the predicate device (PF4 ENHANCED®). Although the sensitivity (co-positivity) of the PF4 IgGTM assay compared to the predicate device was only 74%, some of these discordant results could be explained by the fact that some of these samples may contain only IgM and/or IgA antibodies. However, more importantly, the PF4 IgG™ assay showed an improved specificity (co-negativity) over the PF4 ENHANCED® assay when compared to

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for GTI Diagnostics. The logo consists of a stylized human figure inside of a shield shape. Below the logo is the text "GTI. DIAGNOSTICS" and the tagline "Good science starts with people."

the Serotonin Release Assay (SRA). The SRA is considered to be the gold standard for detection of antibodies that result in HIT.

Stability

No additional stability studies were conducted on the PF4 IgG™ kit since this kit consists of reagents that are used in other kits manufactured by GTI for which stability has already been demonstrated. All of the components of PF4 IgGTM kit with the exception of the anti-human IgG conjugate are identical to those of the PF4 ENHANCED® for which stability of each component has already been established. The anti-human IgG conjugate used in the PF4 IgG™ kit is used in other GTI kits for which stability has also been established. These kits have previously cleared by FDA (GTI QuikScreen®; BK950005 and GTI BScreen®; BK 990043). The expiration of the PF4 IgG™ kit is determined by the expiration dating of the component with the least expiration date.

VIII. Conclusion:

Based on comparison with the predicate device, (PF4 ENHANCED®), these data demonstrate that PF4 IgG™ performs comparable to the predicate device and the PF4 IgGTM kit does not present new issues of safety and effectiveness.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like bird with three curved lines forming its body and wings. The bird is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

DEC 1 9 2007

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Genetic Testing Institute C/O Suzette C. Chance 20925 Crossroads Circle Suite 200 Waukesha, Wisconsin 53186-4054

Re: K071781

Trade/Device Name: PF4 IGG Regulation Number: 21 CFR 864.7695 Regulation Name: Platelet Factor 4 Radioimmunassay Regulatory Class: Class II Product Code: LCO Dated: June 15, 2007 Received: July 2, 2007

Dear Ms. Chance:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{9}------------------------------------------------

Page 2 - Genetic Testing Institute

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Robert Booker

Robert L. Becker, Jr., M.D., Ph.I Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use

510(k) Number (if known): K071781

Device Name: PF4 IgG™

Indications For Use: PF4 IgG™ is a qualitative screening assay for the detection of heparin associated IgG antibodies in human serum. The presence of heparin associated antibodies are commonly found in patients with Heparin Induced Thrombocytopenia (HIT).

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of 1

510(k) K071781

§ 864.7695 Platelet factor 4 radioimmunoassay.

(a)
Identification. A platelet factor 4 radioimmunoassay is a device used to measure the level of platelet factor 4, a protein released during platelet activation by radioimmunoassay. This device measures platelet activiation, which may indicate a coagulation disorder, such as myocardial infarction or coronary artery disease.(b)
Classification. Class II (performance standards).