K Number
K071652

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-10-25

(129 days)

Product Code
Regulation Number
864.5200
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HemoCue WBC system is indicated for use for quantitative determination of white blood cell (WBC) count in capillary or venous whole blood. The HemoCue WBC system is for In Vitro Diagnostic use only. The HemoCue WBC Analyzer is only to be used with HemoCue WBC Microcuvettes. The HemoCue WBC system is indicated for use in clinical laboratories and for point-of-care settings.

Device Description

The system consists of the HemoCue WBC Analyzer together with specially designed microcuvettes, the HemoCue WBC Microcuvettes. The microcuvette serves both as a sample container and a reaction chamber. A blood sample of approximately 10 uL is drawn into the cavity by capillary action. A hemolysing agent lyses the red cells in the microcuvette and a staining agent colors the white blood cells. An image is taken of the stained white blood cells and the number of cells is counted by image analysis. The result is presented within 3 minutes on the analyzer's display. The system reports results in the measuring range 0.3 - 30.0 x107/L. The system is factory calibrated and needs no further calibration.

The HemoCue WBC Analyzer is a portable device. The main parts are the cuvette holder (in which the microcuvette is placed), the cuvette moving arm (brings the microcuvette into correct measuring position), a magnifying optic unit, a camera, image processing software, a display and a power adapter.

The HemoCue WBC Microcuvette is made of polystyrene plastic and contains saponin that hemolyzes the red blood cells, methylene blue that stains the white blood cells and nonactive reagents. A blood sample of approximately 10 µL is drawn into the cavity by capillary action. The microcuvette serves as a sample container and a reaction chamber. No dilution of the sample is required.

AI/ML Overview

The provided text for HemoCue WBC System (K071652) is a 510(k) summary and approval letter, which serves to establish substantial equivalence to predicate devices rather than fully detailing acceptance criteria and studies demonstrating performance. The document states that "Studies were conducted in-house, in clinical laboratory settings and point of care settings to demonstrate the performance with intended specifications of the HemoCue WBC system and to validate that the intended user can easily operate the system and obtain results as expected." However, it does not provide specific details on the acceptance criteria, reported performance, or the methodologies of these studies.

Therefore, much of the requested information cannot be definitively extracted from the provided text.

Here's a breakdown of what can and cannot be answered based on the provided text:

1. A table of acceptance criteria and the reported device performance

This information is not explicitly provided in the given text. While it states that studies were conducted to "demonstrate the performance with intended specifications," it does not list these specifications or the corresponding results in a table format or any other detailed manner.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not explicitly provided. The text mentions "in-house, in clinical laboratory settings and point of care settings" for studies but does not specify sample sizes or data provenance (country, retrospective/prospective).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. The text describes the device's technical characteristics and its indications for use, but it does not detail how ground truth was established for its validation studies.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable as the HemoCue WBC System is an automated cell counter, not an AI-assisted diagnostic tool for human readers. The comparison is against predicate devices (Sysmex XS-1000i and manual light microscopy), not human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the HemoCue WBC System is described as a "semi-automated device" that performs its analysis via "image analysis" and presents a result. Therefore, the performance described would inherently be standalone performance of the device's algorithm without a human in the loop for the actual WBC count determination. The process involves filling a microcuvette, placing it in the analyzer, and receiving a result.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not explicitly provided. The comparison is made against a "Sysmex XS-system" and "manual light microscopic WBC method." It's highly probable that these predicate methods served as the ground truth or reference methods, but the text doesn't explicitly state how discrepancies were resolved or what constituted the ultimate "ground truth."

8. The sample size for the training set

This information is not provided.

9. How the ground truth for the training set was established

This information is not provided. The text focuses on the device's operation and regulatory equivalence, not the specifics of its development and training data.

{0}------------------------------------------------

007 2 5 2007

K071652

5.0 510(k) Summary

Submitter:HemoCue ABBox 1204Ängelholm, Sweden SE-262 23+46 431 45 82 00 (Telephone)+46 431 45 82 25 (FAX)
Contact:Mr. Allan White (Official Correspondent)HemoCue, Inc.40 Empire DriveLake Forest, CA 92630-2244(949) 859-2630 x110 (Telephone)(949) 598-8630 (FAX)allan@hemocue.com
Date of Preparation:June 14, 2007
Common name of device:Semi-Automated Cell Counter
Proprietary Name:HemoCue WBC System
Class:Class II
Panel:Hematology (81)
Regulation number:Automated and Semi-Automated Hematology Devices,21 CFR § 864.5200 Automated Cell Counter or Semi-Cell Counter
Product Code:GKL
Equivalent to:HemoCue AB claims substantial equivalence to the currentlegally marketed device: Sysmex XS-1000i, Automatedhaematology analyzer (K060656) and Manual lightmicroscopic WBC method (class I exempt)

5.1 Device Description

The system consists of the HemoCue WBC Analyzer together with specially designed microcuvettes, the HemoCue WBC Microcuvettes. The microcuvette serves both as a sample container and a reaction chamber. A blood sample of approximately 10 uL is drawn into the cavity by capillary action. A hemolysing agent lyses the red cells in the microcuvette and a staining agent colors the white blood cells. An image is taken of the stained white blood cells and the number of cells is counted by image analysis. The result is presented within 3 minutes on the analyzer's display. The system reports results in the measuring range 0.3 - 30.0 x107/L. The system is factory calibrated and needs no further calibration.

{1}------------------------------------------------

The test principle consists of three simple steps:

    1. Fill the microcuvette with a drop of blood
    1. Place the microcuvette in the analyzer
    1. Receive a result within 3 minutes.

5.2 Indications for Use

The HemoCue WBC system is indicated for use for quantitative determination of white blood cell (WBC) count in capillary or venous whole blood. The HemoCue WBC system is for In Vitro Diagnostic use only. The HemoCue WBC Analyzer is only to be used with HemoCue WBC Microcuvettes. The HemoCue WBC system is indicated for use in clinical laboratories and for point-of-care settings.

5.3 Summary of Technological Characteristics The HemoCue WBC Analyzer

The HemoCue WBC Analyzer (figure 1) is a portable device. The main parts are the cuvette holder (in which the microcuvette is placed), the cuvette moving arm (brings the microcuvette into correct measuring position), a magnifying optic unit, a camera, image processing software, a display and a power adapter.

Image /page/1/Picture/9 description: The image shows a black electronic device, possibly a printer or a scanner, from a high-angle perspective. The device has a rectangular shape with a slightly curved front edge. The top surface of the device appears to have some labels or markings, but they are not clearly visible due to the image quality. The device is placed on a white surface, which provides a contrasting background.

Figure 1. The HemoCue WBC Analyzer

The HemoCue WBC Microcuvette

The HemoCue WBC Microcuvette (Figure 2) is made of polystyrene plastic and contains saponin that hemolyzes the red blood cells, methylene blue that stains the white blood cells and nonactive reagents. A blood sample of approximately 10 µL is drawn into the cavity by capillary action. The microcuvette serves as a sample container and a reaction chamber. No dilution of the sample is required.

Image /page/1/Picture/13 description: The image is a black and white photograph. The top portion of the image is lighter in color, and the bottom portion is darker. The image is slightly blurry, and there are some small white spots scattered throughout the image.

Figure 2. The HemoCue WBC Microcuvette

{2}------------------------------------------------

5.4 Similarities and Differences with Predicate Devices

The HemoCue WBC system is a semi-automated device intended for use for the quantification of White Blood Cells in clinical laboratories and point of care settings.

The Sysmex XS 1000i (predicate device) is an automated hematology analyzer for use in clinical laboratories.

Light microscopes (predicate device) are used as to enumerate leukocytes in a haemocytometer mounted on the stage of the microscope and read manually

Similarities and Differences: All three devices quantify leucocvtes (WBCs), but the Sysmex XS device also quantifies red blood cells, platelets, hemoglobin, hematocrit and provides a differential cell count.

Assessment of Performance 5.5

Studies were conducted in-house, in clinical laboratory settings and point of care settings to demonstrate the performance with intended specifications of the HemoCue WBC system and to validate that the intended user can easily operate the system and obtain results as expected.

5.6 Conclusion

Based on the information and performance data presented in this premarket notification, the HemoCue WBC system meets the manufacturer's intended use specifications and is substantially equivalent to the Sysmex XS-system and the manual light microscopic method of counting WBC.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles an abstract eagle or a caduceus, rendered in black.

Food and Druq Administration 2098 Gaither Road Rockville MD 20850

OCT 2 5 2007

Hemocue AB Hemocue Inc. c/o Allan White 40 Empire Drive Lake Forest, California 92630

Re: K071652

Trade/Device Name: Hemocue WBC System Regulation Number: 21 CFR 864.5200 Regulation Name: Automated cell counter Regulatory Class: Class II Product Code: GKL Dated: June 14, 2007 Received: June 18, 2007

Dear Mr. White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter

{4}------------------------------------------------

Page 2 - Allan White

will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Robert Becker

Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number: K071652

Device Name: HemoCue® WBC system

Indications For Use:

The HemoCue WBC system is indicated for use for quantitative determination of white blood cell (WBC) count in capillary or venous whole blood. The HemoCue WBC system is for In Vitro Diagnostic use only. The HemoCue WBC Analyzer is only to be used with HemoCue WBC Microcuvettes. The HemoCue WBC system is indicated for use in clinical laboratories and for point-of-care settings.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off
Josephine Bautista

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K071652

§ 864.5200 Automated cell counter.

(a)
Identification. An automated cell counter is a fully-automated or semi-automated device used to count red blood cells, white blood cells, or blood platelets using a sample of the patient's peripheral blood (blood circulating in one of the body's extremities, such as the arm). These devices may also measure hemoglobin or hematocrit and may also calculate or measure one or more of the red cell indices (the erythrocyte mean corpuscular volume, the mean corpuscular hemoglobin, or the mean corpuscular hemoglobin concentration). These devices may use either an electronic particle counting method or an optical counting method.(b)
Classification. Class II (performance standards).