K Number
K071164

Validate with FDA (Live)

Device Name
LIFE+CEL BATTERY
Date Cleared
2007-12-07

(225 days)

Product Code
Regulation Number
870.1025
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To power the functions of various devices for which batteries or battery packs are configured.

Since rechargeable batteries and battery packs are "device specific", and are designed to operate and fit into the equipment for which they were manufactured, only qualified personnel should evaluate, test, charge, or install these devices.

This battery is shipped only to customers who request a replacement battery for a particular device or to replace a competitor's replacement battery. Biomedical equipment service professionals, EMT's, etc. therefore know that the intended use is as a replacement battery.

Only replacement battery 2L226 in this submission (Patient Monitor) is provided Over-the-Counter (OTC).

All other replacement batteries, specifically; 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 in this submission (Defib/AED's) are provided by prescription only.

Device Description

Amco rechargeable and non-rechargeable battery packs are utilized as a primary direct current (d-c) power source or as a standby or backup d -c power source for portable as well as stationary medical equipment.

AI/ML Overview

The provided document is a 510(k) summary for "life+cel" or "lifecel" Battery Packs, which are replacement battery packs for various medical devices. The document describes the testing protocols and performance criteria used to demonstrate substantial equivalence to predicate devices.

Based on the information provided, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Incoming InspectionCorrect specification, no visible damage, and randomly voltage tested cells. Lot numbers recorded. Cases inspected for form, fit, function, and cosmetics."All cells are inspected for correct specification, visible damage, and randomly voltage tested prior to acceptance. The lot numbers are recorded for tracking purposes should any fail during final assembly and inspection activities. Cases are also inspected for form, fit, function, and cosmetics." (Implies all cells met criteria for acceptance, lot numbers were recorded, and cases passed inspection).
Cadex TestingBattery cells and core packs tested using a Cadex Electronics Battery Analyzer Model C7400ER in "Auto Mode" through three (3) full charge/discharge cycles. Target capacity (percentage of battery capacity compared to nominal) must meet or exceed a required threshold of 90%. Custom test parameters (C-Rates, delta V, volts-per-cell) can be used."Voltage and capacity of battery cells and core packs are tested using a Cadex Electronics Battery Analyzer Model C7400ER (Extended Range) in the "Auto Mode" prior to installation into cases or shrink wrap. This exercises the batteries in order to identify performance characteristics by running them through three (3) full charge/discharge cycles... Target capacity is a pass/fail mark and our batteries must meet or exceed a required threshold of 90%, or higher, prior to final Quality Control inspection. Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner." (Indicates all tested units met the 90% or higher target capacity threshold. Failed samples and entire lots were rejected as per protocol).
Voltage Testing (Completed Packs)100% testing for correct voltage/polarity prior to shipment."All Battery Packs are tested 100% for correct voltage / polarity prior to shipment. Those devices that fail are rejected and quarantined." (Implies all shipped battery packs passed this 100% inspection).
Defibrillator TestingIndependent (Beta Tests) and random tests on finished packs to ensure they meet the expected number of shocks as specified by the OEMs, using NETECH Model Delta 2000 Defibrillator Analyzers."Independent testing (Beta Tests), as well as random tests on finished packs, are performed using NETECH Model Delta 2000 Defibrillator Analyzers to insure that they meet the expected number of shocks as specified by the OEM's." (Implies the battery packs met the OEM specifications for the number of shocks).
Life CycleThe replacement battery must provide as many or more charge and discharge cycles as the original (OEM) battery."The replacement battery must provide as many or more charge and discharge cycles as the original." (This is stated as a requirement that must be met; thus, it implies the device meets this criterion, although it is noted as an "ongoing process" and not part of standard QC final inspection).
TemperatureThe replacement battery must function correctly over the same temperature range as the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively)."The replacement battery must function correctly over the same temperature range as compared to the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively)." (Stated as a requirement that must be met, implying functional correctness was demonstrated at these temperatures).
Mechanical & Electrical Component IntegrityNormal testing involves drop tests from a predetermined height (usually 2-3 feet) onto a hard, uniform surface. Battery packs inspected for case cracks, cell separation, and electrical/electronic component damage. Root cause analysis if damage occurs. If no visible damage, tested for form, fit, and function."Normal testing would involve drop tests from a predetermined height, usually 2-3 feet, onto a hard, uniform surface. Battery packs are inspected for case cracks, cell separation, and electrical/electronic component damaqe. Root cause analysis is performed should any damage occur. If there is no visible damage, the battery is tested for form, fit, and function." (Implies the battery packs underwent and passed these tests, demonstrating integrity as per the protocol. The absence of reported failures indicates acceptance).
Active Safety DevicesInspected and tested before use and after installation."Active Safety devices are inspected and tested before use and after installation." (Implied to have passed these inspections and tests).

2. Sample Size Used for the Test Set and Data Provenance:

The document describes several testing protocols:

  • Incoming Inspection: "All cells are inspected..." and "Cases are also inspected...". This implies 100% inspection of incoming cells and cases.
  • Cadex Testing: "Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner." This suggests lot-based testing where if a sample fails, the entire lot is tested. The specific sample size for a given lot is not provided, but it's not a fixed test set in the traditional sense, but rather an ongoing QC process.
  • Voltage Testing (Completed Packs): "All Battery Packs are tested 100% for correct voltage / polarity prior to shipment." This indicates a 100% test set of all manufactured units.
  • Defibrillator Testing: "Independent testing (Beta Tests), as well as random tests on finished packs..." This suggests both an unspecified "Beta Test" sample and a "random" sample from finished production.
  • Life Cycle, Temperature, Mechanical & Electrical Component Integrity: The wording "Normal testing would involve..." and "The replacement battery must..." suggests design validation and verification activities, likely involving a representative sample, but the specific sizes are not mentioned.

Data Provenance: The testing appears to be conducted by Amco International Manufacturing & Design, Inc./Phoenix Technology, likely at their facility in Elizabeth, Colorado, USA. The data is prospective in the sense that it's part of their manufacturing and quality control process for the battery packs being submitted.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the document. The "ground truth" for these battery packs is based on objective measurements against specified performance criteria (e.g., voltage, capacity, number of shocks, temperature function, mechanical integrity) and comparison to OEM specifications. It does not involve expert clinical interpretation or consensus.

4. Adjudication Method for the Test Set:

This information is not applicable as the testing described relates to objective engineering and performance verification against predefined specifications, not subjective interpretation requiring adjudication among experts. The "adjudication" is essentially a pass/fail based on quantitative measurements.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The device is a battery pack, not an AI-assisted diagnostic or therapeutic tool for human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The device is a battery pack, not an algorithm. The performance described is its inherent function as a power source.

7. The Type of Ground Truth Used:

The ground truth used for evaluating the battery packs is based on objective engineering and performance specifications, primarily:

  • Quantitative measurements: Voltage, capacity (e.g., 90% threshold for Cadex testing), number of shocks for defibrillator batteries.
  • Functional performance: Correct function across specified temperature ranges.
  • Structural integrity: Resistance to mechanical damage (e.g., drop tests).
  • Comparison to OEM specifications: The replacement devices "must meet or exceed these benchmark results consistently" from OEM battery packs.

8. The Sample Size for the Training Set:

This information is not applicable as the device is a battery pack and does not involve AI or machine learning models that require a training set. The descriptions are for manufacturing and quality control testing.

9. How the Ground Truth for the Training Set was Established:

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left and consists of a stylized graphic element. The FDA logo is on the right and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 16, 2021

Amco International Manufacturing & Design, Inc. Alex Henderson Technical Consultant 377 Zane Court Hillside Village Elizabeth, Colorado 80107

Re: K071164

Trade/Device Name: Life+cel Battery Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX

Dear Alex Henderson:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated December 7, 2007. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code MHX.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih, Office of Cardiovascular Devices, 301-796-5813, Jennifer.Shih@fda.hhs.gov.

Sincerely,

Jennifer W. Shih -S

Jennifer Shih Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OEC 7 2007

Amco International Manufacturing & Design, Inc. c/o Mr. Alex Henderson Phoenix Technology 377 Zane Court, Hillside Village Elizabeth, CO 80107

Re: K071164

Trade/Device Name: life+cel™ (or lifecel™) Battery Packs with Model Numbers: 2L226. 5L375. 6L577. 7L991. 9L977. 9L990 and 9L998 Regulation Number: Unclassified Regulation Name: Unclassified Regulatory Class: Unclassified Product Code: MOY, MOX Dated: December 4, 2007 Received: December 5, 2007

Dear Mr. Henderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Alex Henderson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Blzimima fh

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

Image /page/3/Picture/1 description: The image shows the logo for Phoenix Technology. The logo features a phoenix bird with its wings spread, surrounded by a circular halo. The text "PHOENIX TECHNOLOGY" is written in a serif font to the right of the bird.

INDICATIONS FOR USE STATEMENT

510(K) Number: K071164

Device Name: Box, Battery

Indications for Use:

To power the functions of various devices for which batteries or battery packs are configured.

Since rechargeable batteries and battery packs are "device specific", and are designed to operate and fit into the equipment for which they were manufactured, only qualified personnel should evaluate, test, charge, or install these devices.

This battery is shipped only to customers who request a replacement battery for a particular device or to replace a competitor's replacement battery. Biomedical equipment service professionals, EMT's, etc. therefore know that the intended use is as a replacement battery .

Only replacement battery 2L226 in this submission (Patient Monitor) is provided Over-the-Counter (OTC).

All other replacement batteries, specifically; 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 in this submission (Defib/AED's) are provided by presc ription only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blummo
(Division-Sign-Off)

Division of Cardiovascular Devices
510(k) Number K071164

377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson(@phoenixsolutions.org

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Phoenix Technology. The logo features a stylized phoenix bird with its wings spread, surrounded by a halo. To the right of the bird, the words "PHOENIX TECHNOLOGY" are written in a serif font, with "PHOENIX" in bold and "TECHNOLOGY" in a lighter, less prominent font.

OEC - 7 2007

510(k) Summary – K071164 -1 of 5-

Replacement Battery Packs 2L226, 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998

Submitter:Amco International Manufacturing & Design, Inc.Attn: Mr. Adam Milewski69-81 108th Street, Suite 6GForest Hills, New York 11375
Contact Person:Alex HendersonPhoenix Technology377 Zane Court,Hillside VillageElizabeth, CO 80107Tel / Fax: 303-646-3715Email: alex_henderson@phoenixsolutions.org
Date Prepared:April 16, 2006
Device Name:Trade/Proprietary Name: life+celTM or lifecelTM Battery PackCommon/Generic Name: Box Battery OR Pack BatteryClassification Name: Box Battery, Rechargeable
Classification:Cardiovascular PanelClass
21CFR 870.1025Arrhythmia Detector and AlarmII
21CFR 870.1110Blood Pressure ComputerII
21CFR 870.1130Systems, Measurement, Blood Pressure, Non-InvasiveII
21CFR 870.2300Monitor, Cardiac (Including Cardio tachometer & Rate Alarm)II
21CFR 870.2340ElectrocardiographII
No Regulatory ReferenceRechargeable Batteries for Class III DevicesIII

377 Zane Court • Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 ● Email: alex_henderson@phoenixsolutio ns.org

{5}------------------------------------------------

510(k) Summary -2 of 5 -

Legally Marketed Predicate Devices:

This submission compares the specifications and functionality of various rechargeable battery packs with those of similar devices that were included as part of the following original predicate devices and submissions:

    1. The AMCO 2L226 life+cel™ is equivalent to that manufactured by:
    • I Alpha Source / Access Battery, cleared under K032051
    • Unipower Corp., cleared under K940842 and K935111
    • . R & D Battery, cleared under K951281
    • I OEM, GE Marquette - Patient Monitor Dash 3000/4000 - Reference: K051367, K020290 (Battery included with device)
    1. The AMCO 5L375 life+cel™ is equivalent to that manufactured by:
    • प्र Alpha Source / Access Battery, cleared under K920964
    • 트 Unipower Corp., cleared under K940842 and K935111
    • I R & D Battery, cleared under K951281
    • 프 OEM, Physio Control / Medtronic's - Monitor/Defibrillator (FastPak) LP5 - Reference: K890079, K931053 (Battery included with device)
    1. The AMCO 6L577 life+cel™ is equivalent to that manufactured by:
    • l Alpha Source / Access Battery, cleared under K0920964
    • . Unipower Corp., cleared under K940842 and K935111
    • 트 R & D Battery, cleared under K951281
    • OEM, Physio Control - Monitor/Defibrillator (Non-FastPak) LP5 / LP10 - Reference: K890079, K931053 (Battery included with device)
    1. The AMCO 7L991 life+cel™ is equivalent to that manufactured by:
    • Alpha Source / Access Battery, cleared under K920964
    • . Unipower Corp., cleared under K940842 and K935111
    • L OEM, Physio Control / Medtronic's - Monitor/Defibrillator LP12 (NiCad) - Reference: K973486, K010918 (Battery included with device)

377 Zanc Court • Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 ● Email: alex henderson@phoenixsolutions.org

{6}------------------------------------------------

510(k) Summary -3 of 5 -

Legally Marketed Predicate Devices (continued):

    1. The AMCO 9L977 life+cel™ is equivalent to that man ufactured by:
    • Alpha Source / Access Battery, cleared under K920964 .
    • r Unipower Corp., cleared under K940842 and K935111
    • R & D Battery, cleared under K945869
    • OEM, Health Dyne / Vickers Respironics Smart Monitor 900 Series 트
      • Reference: K874695 (Battery included with device)
    1. The AMCO 9L990 life+cel™ is equivalent to that manufactured by:
    • · Alpha Source / Access Battery, cleared under K032051
    • Unipower Corp., cleared under K940842 and K935111 .
    • OEM, CAS Medical NIBP Monitor 740 .
      • Reference: K033048 (Battery included with device)
    1. The AMCO 9L998 life+cel™ is equivalent to that manufactured by:
    • Alpha Source / Access Battery, cleared under K920964 .
    • . Unipower Corp., cleared under K940842 and K935111
    • OEM, Physio Control / Medtronic's Monitor/Defibrillator LP12 (SLA) - Reference: K973486. K010918 (Battery included with device)

Description:

Amco rechargeable and non-rechargeable battery packs are utilized as a primary direct current (d-c) power source or as a standby or backup d -c power source for portable as well as stationary medical equipment.

Statement of Intended Use:

To power the functions of the specific devices for which the batteries or battery packs are intended for back-up power or portable use.

Comparison of Technological Characteristics

The design components and functionality of the various battery packs listed are similar to those of their predicate devices. All these devices provide a means of supplying electrical power through chemical reaction. The energy provided depends upon the voltage and capacity rating of a particular pack and the amount of current used by the device into which they are installed. The performance and life span of these batteries depends on operating conditions of temperature, current drain, and the charge/discharge method (if applicable). These parameters are taken into account in designing such batteries. The goal is to develop battery packs that maintain capacity for as high and as long as possible. Typical cell chemistries are Sealed -Lead Acid (SLA), Nickel-Cadmium (NiCad), Nickel-Metal Hydride (NiMH), and Lithium based.

377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson@phoenixsolutio ns.org

{7}------------------------------------------------

510(k) Summary -4 of 5 -

BATTERY PACK TESTING - GENERAL PROTOCOL

INCOMING INSPECTION

All cells are inspected for correct specification, visible damage, and randomly voltage tested prior to acceptance. The lot numbers are recorded for tracking purposes should any fail during final assembly and inspection activities. Cases are also inspected for form, fit, function, and cosmetics.

CADEX TESTING

Voltage and capacity of battery cells and core packs are tested using a Cadex Electronics Battery Analyzer Model C7400ER (Extended Range) in the "Auto Mode" prior to installation into cases or shrink wrap. This exercises the batteries in order to identify performance characteristics by running them through three (3) full charge/discharge cycles. Tests typically take 12 hours for each battery pack.

All battery chemistries can be tested using custom test parameters, depending on Quality Control and customer requirements. This allows for various C -Rates, delta V (ΔV), and volts-per-cell to be entered into the test protocol through the Cadex Battery Shop Software utility (Reference Cadex Test Report Examples - Exhibit A).

Target capacity is the percentage of the battery capacity compared to nominal capacity and serves as a threshold, This threshold, or target capacity, can be set to any desired range (90 - 95% is typical).

Target capacity is a pass/fail mark and our batteries must meet or exceed a required threshold of 90%, or higher, prior to final Quality Control inspection. Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner.

Battery packs are not shipped fully charged. There are specific DOT, FAA, and EPA requlations and quidelines that address these concerns.

VOLTAGE TESTING - Completed Packs

All Battery Packs are tested 100% for correct voltage / polarity prior to shipment. Those devices that fail are rejected and quarantined.

DEFIBRILLATOR TESTING

Independent testing (Beta Tests), as well as random tests on finished packs, are performed using NETECH Model Delta 2000 Defibrillator Analyzers to insure that they meet the expected number of shocks as specified by the OEM's.

377 Zane Court · Elizabeth, Colorado USA 80107 Telephone: 303-646-3715 · Email: alex henderson@phoenixsolutions.org

{8}------------------------------------------------

510(k) Summary -5 of 5 -

BATTERY PACK TESTING (continued)

SAFETY and PERFORMANCE

Safety and performance testing of battery packs are performed to ensure that these devices meet all functional requirements and performance specifications.

In comparison analysis. OEM Battery Packs set the benchmark. Replacement devices must meet or exceed these benchmark results consistently.

Concerns that are addressed during bench test comparison analysis are:

  • = Life cycle
    The replacement battery must provide as many or more charge and discharge cycles as the original. This is an ongoing process and is not part of the standard QC final inspection protocol. Shelf life, on the other hand, is based on the original cell manufacturer's specification sheets and Certificates of Conformance.

I Temperature

The replacement battery must function correctly over the same temperature range as compared to the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively).

. Mechanical & Electrical Component Integrity

Normal testing would involve drop tests from a predetermined height, usually 2-3 feet, onto a hard, uniform surface. Battery packs are inspected for case cracks, cell separation, and electrical/electronic component damaqe. Root cause analysis is performed should any damage occur.

  • If there is no visible damage, the battery is tested for form, fit, and function.

  • Active Safety devices are inspected and tested before use and after installation.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.