Search Results
Found 2 results
510(k) Data Aggregation
(352 days)
The CARESCAPE ONE is a multi-parameter physiological patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility.
The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.
The CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.
The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.
Contraindications for using CARESCAPE ONE:
The CARESCAPE ONE is not intended for use within a controlled MR environment.
CARESCAPE ONE is a new patient monitor device based on GE Healthcare predicate devices, the Patient Data Module (K071073) and the CARESCAPE B450 (K132533).
CARESCAPE ONE, with CARESCAPE Software version 3 belongs to the CARESCAPE V3 patient monitor family. The concept of the CARESCAPE ONE is to provide a flexible bedside monitor that can also be used during intra-hospital transport. The flexibility of the CARESCAPE ONE allows the user to configure the monitor's vital sign acquisition for only the parameters they require. This is achieved using plug and play Active Cable Modules (ACM) that connect via medical grade USB ports on the CARESCAPE ONE monitor. Note that the USB ports are not compatible with commercial USB items on the market due to a custom connector design. Each ACM is dedicated to measuring a particular parameter, including ECG/Respiration, Invasive Blood Pressure, Temperature, SpO2, or CO2. The only exception is the Non-Invasive Blood Pressure (NIBP) measurement which does not require a separate ACM since the capability to measure NIBP is built-in to the CARESCAPE ONE monitor itself. The ACM's are CARESCAPE TEMP, CARESCAPE PRES, CARESCAPE ECG, CARESCAPE SpO2 (TruSignal), CARESCAPE SpO2 -Nellcor, CARESCAPE SpO2 - Masimo and CARESCAPE CO2 - LoFlo. CARESCAPE SpO2 - Nellcor, CARESCAPE SpO2 -Masimo, and CARESCAPE CO2 - LoFlo have been developed by their respective companies/manufacturers (OEM) for use with the CARESCAPE ONE. The technology from each OEM has received 510(k) clearance and is adapted to function with the CARESCAPE ONE. The OEM technologies are not new and are not a part of this submission, only their integration into the Parameters/Active Cable Modules for use with the CARESCAPE ONE is covered in this 510(k).
CARESCAPE ONE provides the users the acquired display values, waveforms, alarms and status messages in compact footprint monitor that runs on an internal battery as well as AC power when connected to the docking station.
Here's an analysis of the acceptance criteria and supporting studies for the CARESCAPE ONE device, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
The FDA 510(k) summary focuses on demonstrating substantial equivalence to predicate devices rather than providing a direct table of specific acceptance criteria with corresponding performance values for all features. However, it does highlight performance claims and comparisons for key components, specifically arrhythmia detection and SpO2.
Parameter/Feature | Acceptance Criteria (Implied by Predicate Equivalence & Standards) | Reported Device Performance (CARESCAPE ONE) |
---|---|---|
Arrhythmia Detection | Performance equivalent to EK-Pro V13 (predicate). | EK-Pro V14: Substantially equivalent EC57 results and performance values to EK-Pro V13. |
SpO2 Accuracy | Performance as per ISO 80601-2-61:2011 and FDA guidance. | TruSignal V3 SpO2: Accuracy validated in controlled desaturation study during motion and non-motion conditions, compared to CO-oximetry. |
Other Parameters | Equivalent performance to PDM (K071073) and B450 (K132533) | Same parameter acquisition technology as PDM, integrated differently. CO2 measurement technology previously cleared and used by GE Healthcare. |
Electrical Safety | Compliance with applicable consensus standards. | Bench testing demonstrated compliance. |
EMC | Compliance with applicable consensus standards. | Bench testing demonstrated compliance. |
Environmental (Mech.) | Compliance with applicable consensus standards. | Bench testing demonstrated compliance. |
Usability | Safe and effective for intended users/uses. | Extensive usability work, including critical task identification, formative, and summative testing. |
Software Level of Concern | Major | Software design, development, verification, validation, and traceability provided. |
Biocompatibility | Applicable for patient-contacting components. | Biocompatibility testing completed for patient cables with intermittent skin contact. |
Cybersecurity | Compliance with FDA guidance for medical device cybersecurity. | Security Risk Assessment, Threat model, Penetration testing, system integrity, access, audit, network, and remote service controls. |
2. Sample Size Used for the Test Set and Data Provenance
- Arrhythmia (EK-Pro V14): The document references ANSI/AAMI EC57:1998 (R)2012. This standard specifies databases (e.g., AHA, MIT-BIH) that are used for testing arrhythmia algorithms. The specific sample size from these databases used for EK-Pro V14 testing is not explicitly stated in this 510(k) summary, but it would be derived from the standard's requirements.
- Data Provenance: The databases mentioned in EC57 are typically retrospective, internationally recognized ECG databases. The exact country of origin for the data used for this specific EK-Pro V14 validation is not specified, but the standard's databases often compile data from various global sources.
- SpO2 (TruSignal V3): The document states a "controlled desaturation study." The sample size is not explicitly stated in this summary.
- Data Provenance: A controlled desaturation study is a prospective clinical study designed to test pulse oximeter accuracy. The country of origin for this study is not specified.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Arrhythmia (EK-Pro V14): When using standard databases like those referenced by ANSI/AAMI EC57, the ground truth (arrhythmia annotations) is typically established by multiple, highly experienced cardiologists or cardiac electrophysiologists. The exact number and qualifications for the original annotation of the standard databases are not specified in this document, as the standard itself defines these.
- SpO2 (TruSignal V3): For controlled desaturation studies, the "ground truth" for SpO2 is provided by a CO-oximeter measurement, which itself is a highly accurate laboratory method for oxygen saturation. This is an objective measurement rather than expert consensus. Therefore, "experts" in the traditional sense for establishing ground truth are not directly applicable here.
4. Adjudication Method for the Test Set
- Arrhythmia (EK-Pro V14): For standard databases, ground truth annotation often involves multiple experts reviewing and adjudicating discrepancies, sometimes with a "majority rules" or senior expert decision process. The specific adjudication method for the creation of the databases referenced by EC57 is not detailed in this summary, but would adhere to the methodologies defined by the creators of those databases.
- SpO2 (TruSignal V3): As the ground truth comes from CO-oximetry, an objective measurement, adjudication by experts is not applicable.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was NOT done. The document explicitly states: "Clinical studies of the CARESCAPE ONE device performance were not required to establish substantial equivalence." This indicates that human-in-the-loop performance with or without AI assistance was not assessed in this submission for primary equivalence.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done
- Yes, standalone performance was assessed for key algorithms:
- EK-Pro V14 Arrhythmia Detection algorithm: Its performance was evaluated against the ANSI/AAMI EC57 standard, comparing it to the predicate's EK-Pro V13 algorithm. This is a standalone algorithm performance evaluation.
- TruSignal V3 SpO2 algorithm: Its accuracy was validated in a controlled desaturation study against CO-oximetry. This is also a standalone algorithm performance evaluation.
7. The Type of Ground Truth Used
- Arrhythmia (EK-Pro V14): The ground truth would be expert consensus annotations of ECG waveforms from standard arrhythmia databases (e.g., those specified in ANSI/AAMI EC57).
- SpO2 (TruSignal V3): The ground truth was objective laboratory measurement using CO-oximetry in a controlled study. This is considered a gold standard for blood oxygen saturation.
8. The Sample Size for the Training Set
- The document does not explicitly state the sample size used for training the EK-Pro V14 arrhythmia detection algorithm or the TruSignal V3 SpO2 algorithm. In 510(k) submissions, training set details (especially for existing, updated algorithms) are often not required to be as extensively disclosed as test set performance, particularly when demonstrating equivalence to a predicate.
9. How the Ground Truth for the Training Set was Established
- The document does not explicitly state how the ground truth for the training sets of the EK-Pro V14 or TruSignal V3 SpO2 algorithms was established. For complex algorithms like these, training data is typically meticulously annotated by a combination of clinical experts (for arrhythmia) or using highly accurate reference methods (for SpO2) and then verified. Again, in a 510(k) for an updated algorithm or component, the full developmental history and training data specifics may not be detailed if the focus is on demonstrating equivalent performance post-modifications.
Ask a specific question about this device
(225 days)
To power the functions of various devices for which batteries or battery packs are configured.
Since rechargeable batteries and battery packs are "device specific", and are designed to operate and fit into the equipment for which they were manufactured, only qualified personnel should evaluate, test, charge, or install these devices.
This battery is shipped only to customers who request a replacement battery for a particular device or to replace a competitor's replacement battery. Biomedical equipment service professionals, EMT's, etc. therefore know that the intended use is as a replacement battery.
Only replacement battery 2L226 in this submission (Patient Monitor) is provided Over-the-Counter (OTC).
All other replacement batteries, specifically; 5L375, 6L577, 7L991, 9L977, 9L990, and 9L998 in this submission (Defib/AED's) are provided by prescription only.
Amco rechargeable and non-rechargeable battery packs are utilized as a primary direct current (d-c) power source or as a standby or backup d -c power source for portable as well as stationary medical equipment.
The provided document is a 510(k) summary for "life+cel" or "lifecel" Battery Packs, which are replacement battery packs for various medical devices. The document describes the testing protocols and performance criteria used to demonstrate substantial equivalence to predicate devices.
Based on the information provided, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Incoming Inspection | Correct specification, no visible damage, and randomly voltage tested cells. Lot numbers recorded. Cases inspected for form, fit, function, and cosmetics. | "All cells are inspected for correct specification, visible damage, and randomly voltage tested prior to acceptance. The lot numbers are recorded for tracking purposes should any fail during final assembly and inspection activities. Cases are also inspected for form, fit, function, and cosmetics." (Implies all cells met criteria for acceptance, lot numbers were recorded, and cases passed inspection). |
Cadex Testing | Battery cells and core packs tested using a Cadex Electronics Battery Analyzer Model C7400ER in "Auto Mode" through three (3) full charge/discharge cycles. Target capacity (percentage of battery capacity compared to nominal) must meet or exceed a required threshold of 90%. Custom test parameters (C-Rates, delta V, volts-per-cell) can be used. | "Voltage and capacity of battery cells and core packs are tested using a Cadex Electronics Battery Analyzer Model C7400ER (Extended Range) in the "Auto Mode" prior to installation into cases or shrink wrap. This exercises the batteries in order to identify performance characteristics by running them through three (3) full charge/discharge cycles... Target capacity is a pass/fail mark and our batteries must meet or exceed a required threshold of 90%, or higher, prior to final Quality Control inspection. Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner." (Indicates all tested units met the 90% or higher target capacity threshold. Failed samples and entire lots were rejected as per protocol). |
Voltage Testing (Completed Packs) | 100% testing for correct voltage/polarity prior to shipment. | "All Battery Packs are tested 100% for correct voltage / polarity prior to shipment. Those devices that fail are rejected and quarantined." (Implies all shipped battery packs passed this 100% inspection). |
Defibrillator Testing | Independent (Beta Tests) and random tests on finished packs to ensure they meet the expected number of shocks as specified by the OEMs, using NETECH Model Delta 2000 Defibrillator Analyzers. | "Independent testing (Beta Tests), as well as random tests on finished packs, are performed using NETECH Model Delta 2000 Defibrillator Analyzers to insure that they meet the expected number of shocks as specified by the OEM's." (Implies the battery packs met the OEM specifications for the number of shocks). |
Life Cycle | The replacement battery must provide as many or more charge and discharge cycles as the original (OEM) battery. | "The replacement battery must provide as many or more charge and discharge cycles as the original." (This is stated as a requirement that must be met; thus, it implies the device meets this criterion, although it is noted as an "ongoing process" and not part of standard QC final inspection). |
Temperature | The replacement battery must function correctly over the same temperature range as the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively). | "The replacement battery must function correctly over the same temperature range as compared to the original. Testing is done at 0, 25, and 40°C (32, 77, and 104°F respectively)." (Stated as a requirement that must be met, implying functional correctness was demonstrated at these temperatures). |
Mechanical & Electrical Component Integrity | Normal testing involves drop tests from a predetermined height (usually 2-3 feet) onto a hard, uniform surface. Battery packs inspected for case cracks, cell separation, and electrical/electronic component damage. Root cause analysis if damage occurs. If no visible damage, tested for form, fit, and function. | "Normal testing would involve drop tests from a predetermined height, usually 2-3 feet, onto a hard, uniform surface. Battery packs are inspected for case cracks, cell separation, and electrical/electronic component damaqe. Root cause analysis is performed should any damage occur. If there is no visible damage, the battery is tested for form, fit, and function." (Implies the battery packs underwent and passed these tests, demonstrating integrity as per the protocol. The absence of reported failures indicates acceptance). |
Active Safety Devices | Inspected and tested before use and after installation. | "Active Safety devices are inspected and tested before use and after installation." (Implied to have passed these inspections and tests). |
2. Sample Size Used for the Test Set and Data Provenance:
The document describes several testing protocols:
- Incoming Inspection: "All cells are inspected..." and "Cases are also inspected...". This implies 100% inspection of incoming cells and cases.
- Cadex Testing: "Any samples that do not meet the criteria are rejected, and subsequently, the entire lot is tested in this manner." This suggests lot-based testing where if a sample fails, the entire lot is tested. The specific sample size for a given lot is not provided, but it's not a fixed test set in the traditional sense, but rather an ongoing QC process.
- Voltage Testing (Completed Packs): "All Battery Packs are tested 100% for correct voltage / polarity prior to shipment." This indicates a 100% test set of all manufactured units.
- Defibrillator Testing: "Independent testing (Beta Tests), as well as random tests on finished packs..." This suggests both an unspecified "Beta Test" sample and a "random" sample from finished production.
- Life Cycle, Temperature, Mechanical & Electrical Component Integrity: The wording "Normal testing would involve..." and "The replacement battery must..." suggests design validation and verification activities, likely involving a representative sample, but the specific sizes are not mentioned.
Data Provenance: The testing appears to be conducted by Amco International Manufacturing & Design, Inc./Phoenix Technology, likely at their facility in Elizabeth, Colorado, USA. The data is prospective in the sense that it's part of their manufacturing and quality control process for the battery packs being submitted.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
This information is not provided in the document. The "ground truth" for these battery packs is based on objective measurements against specified performance criteria (e.g., voltage, capacity, number of shocks, temperature function, mechanical integrity) and comparison to OEM specifications. It does not involve expert clinical interpretation or consensus.
4. Adjudication Method for the Test Set:
This information is not applicable as the testing described relates to objective engineering and performance verification against predefined specifications, not subjective interpretation requiring adjudication among experts. The "adjudication" is essentially a pass/fail based on quantitative measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable. The device is a battery pack, not an AI-assisted diagnostic or therapeutic tool for human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable. The device is a battery pack, not an algorithm. The performance described is its inherent function as a power source.
7. The Type of Ground Truth Used:
The ground truth used for evaluating the battery packs is based on objective engineering and performance specifications, primarily:
- Quantitative measurements: Voltage, capacity (e.g., 90% threshold for Cadex testing), number of shocks for defibrillator batteries.
- Functional performance: Correct function across specified temperature ranges.
- Structural integrity: Resistance to mechanical damage (e.g., drop tests).
- Comparison to OEM specifications: The replacement devices "must meet or exceed these benchmark results consistently" from OEM battery packs.
8. The Sample Size for the Training Set:
This information is not applicable as the device is a battery pack and does not involve AI or machine learning models that require a training set. The descriptions are for manufacturing and quality control testing.
9. How the Ground Truth for the Training Set was Established:
This information is not applicable for the reasons stated in point 8.
Ask a specific question about this device
Page 1 of 1