(136 days)
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner. These gloves are not intended to be used as a chemical barrier.
Shanghai China Star Corp Powder free blue Nitrile Examination Glove is a class I device having product code 80LZA. It is a disposable device powdered with absorbable dusting powder that meets all requirements of ASTM D 631900a-05 and is tested with chemotherapy drugs.
Here's a breakdown of the acceptance criteria and study information for the Shanghai China Star Corp Powder Free Blue Nitrile Patient Examination Glove, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Test / Characteristic | Acceptance Criteria (from predicate/standard) | Reported Device Performance (New Device) |
|---|---|---|
| Physical Specifications | ||
| Width (size medium) | 92mm (Predicate K040841) | 92mm |
| Overall length | 240mm (Predicate K040841) | 240mm |
| Palm thickness | 0.17mm (Predicate K040841) | 0.17mm |
| Finger thickness | 0.18mm (Predicate K040841) | 0.18mm |
| Tensile strength pre-aging min | 21mpa (Predicate K040841) | 21mpa |
| Tensile strength after-aging min | 16mpa (Predicate K040841) | 16mpa |
| Ultimate elongation pre-aging min | 500 (Predicate K040841) | 500 |
| Ultimate elongation after-aging min | 500 (Predicate K040841) | 500 |
| Performance Testing | ||
| Dermal Sensitization Test | Passes (Predicate K040841, K022765) | Passes |
| Primary Skin irritation | Passes (Predicate K040841, K022765) | Passes |
| Permeation testing | Not explicitly stated as "Passes" criteria, but ASTM D 6978-05 standard implies expected performance | Passes |
| Iodine Test | Not explicitly stated as "Passes" criteria | Passes |
| Tensile strength (ASTM D6319-00a) | Meets requirements of ASTM D6319-00a | Gloves meet the requirements of ASTM D6319-00a. |
| Barrier strength (ASTM D6319-00a) | Meets requirements of ASTM D6319-00a | Gloves meet the requirements of ASTM D6319-00a. |
| Residual powder test | Passes (Predicate K040841, K022765) | Passes |
| Biocompatibility standards | Yes (Predicate K040841, K022765) | Yes |
| Duration of biocompatibility | Limited (Predicate K040841, K022765) | Limited |
| Pinholes | AQL 2.5 (ASTM D 6319-00a) | AQL 2.5 |
| General Inspection Level | S-2 (ASTM D 6319-00a) | S-2 |
| General AQL for Inspection | AQL 4.0 (ASTM D 6319-00a) | AQL 4.0 |
Study Details:
The provided document describes a 510(k) premarket notification which aims to demonstrate that a new device is substantially equivalent to legally marketed predicate devices. This is a regulatory pathway, not a traditional clinical study designed to "prove" device performance in the same way a randomized controlled trial would. Instead, it relies on demonstrating that the new device meets established standards and performs similarly to predicate devices.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document repeatedly references "Inspection level S-2, AQL 4.0" and "pinholes at AQL 2.5" as per ASTM D 6319-00a. These are sampling plans for quality control, not a fixed sample size for a "test set" in the context of clinical studies. ASTM D 6319-00a specifies statistical sampling procedures for examination gloves. The exact sample size would depend on the lot size being tested according to the standard.
- Data Provenance: The document does not explicitly state the country of origin for the testing data, although the applicant and manufacturer are based in Shanghai, China. The testing appears to be retrospective in the sense that it's performed on manufactured gloves to demonstrate compliance with standards, rather than a prospective study observing usage in a real-world setting.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts
- This information is not applicable (N/A) in this context. The "ground truth" for this device's performance is established by conformance to recognized international standards (ASTM D 6319-00a, ASTM D 6978-05) and comparison to existing predicate devices. There isn't a need for expert consensus on interpreting complex medical images or clinical outcomes here. The tests are objective measurements against defined criteria.
4. Adjudication Method for the Test Set
- N/A. As explained above, this is not a study requiring adjudication of expert opinions or clinical outcomes. The tests are objective and deterministic.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No. An MRMC study is typically used for diagnostic imaging devices to assess how human readers' performance with and without AI assistance changes. This document describes examination gloves, which do not involve diagnostic interpretation by human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- N/A. This device is a physical product (gloves), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.
7. The Type of Ground Truth Used
- The "ground truth" used for this device is primarily established industry standards (ASTM D 6319-00a and ASTM D 6978-05) and the performance specifications of legally marketed predicate devices (K040841, K022765). These standards define the acceptable physical properties and performance characteristics (e.g., tensile strength, barrier integrity, biocompatibility).
8. The Sample Size for the Training Set
- N/A. This is not a machine learning or AI device that requires a "training set." The device is a manufactured product tested against defined specifications.
9. How the Ground Truth for the Training Set was Established
- N/A. No training set is involved.
{0}------------------------------------------------
SHANGHAI CHINASTAR COR
No#283 Che Xin Rd, Che Dun Zhen, Song Jaing County Shanghai, China C/O: 11F No.201, Nanking East Rd., Sec 3, Taipei, Taiwan, R.O.C Tel:886-2-2713-6677 Fax:886-2-2546-2480
510(k) Summary
Prepared: March 28, 2007
| Applicant | Shanghai China Star CorpNo. 283, Che Xing Road, Che Dun Zhen,Song JiangCounty: Shanghai, China |
|---|---|
| Manufacturer | Shanghai China Star CorpNo. 283, Che Xing Road, Che Dun Zhen,Song JiangCounty: Shanghai, China |
| Submitter | Romil RambhiaOfficial Correspondent forShanghai China Star Corp |
| Address | mdi Consultants, Inc.55 Northern Blvd., Suite 200Great Neck, New York 11021Tel: 516-482-9001Fax: 516-482-0186romil@mdiconsultants.com |
| Trade/proprietary Name | Shanghai China Star CorpPowder free Blue Nitrile PatientExamination Glove tested withchemotherapy drugs. |
| Common Names | POWDER-FREE Patient ExaminationGlove |
| Classification name | Patient Examination Glove |
| Classification number | 21 CFR 880.6250 |
Device Description:
Shanghai China Star Corp Powder free blue Nitrile Examination Glove is a class I device having product code 80LZA. It is a disposable device powdered with absorbable dusting powder that meets all requirements of ASTM D 631900a-05 and is tested with chemotherapy drugs.
Intended Use:
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or to prevent finger examiner. These gloves are not intended to be used as a chemical barrier
{1}------------------------------------------------
Substantial Equivalence Discussion:
| Characteristicand parameters | Shanghai China StarCorp(New Device) | MEDLINEINDUSTRIES, INC(K) 040841 | ALLEGIANCEHEALTHCARECORP.(K) 022765 |
|---|---|---|---|
| Product Code | LZA | LZA | LZA |
| Intended Use | A powder free patientexamination glove(Tested for Use withChemotherapy Drugs)is a disposable deviceintended for medicalpurposes that is wornon the examiners handor finger to preventcontamination betweenpatient and examiner.These gloves are notintended to be used as achemical barrier | Medline Powder-FreeBlue NitrileExamination Gloves(Tested for Use withChemotherapyDrugs) is adisposable deviceintended for medicalpurposes that is wornon the examiner'shand or finger topreventcontaminationbetween patient andexaminer. | A patient examinationglove is a disposabledevice intended formedical purposes thatis worn on theexaminer's hand orfinger to preventcontaminationbetween patient andexaminer. In addition,these gloves are wornto protect the weareragainst exposure tochemotherapy drugs.The gloves with testedfor use withchemotherapy drugslabeling claims. |
| Width (sizemedium) | 92mm | 92mm | N/A |
| Overall length | 240mm | 240mm | N/A |
| Palm thickness | 0.17mm | 0.17mm | N/A |
| Finger thickness | 0.18mm | 0.18mm | N/A |
| Tensile strengthpre aging min | 21mpa | 21mpa | N/A |
| Tensile strengthafter aging min | 16mpa | 16mpa | N/A |
| Ultimateelongation preaging min | 500 | 500 | N/A |
| Ultimateelongation afteraging min | 500 | 500 | N/A |
| MeetsBiocompatibilitystandards | Yes | Yes | Yes |
| Duration of bio-compatibility | Limited | Limited | Limited |
| Skin irritation test | Passes | Passes | Passes |
| Dermalsensitization | Passes | Passes | Passes |
| Residual powdertest | Passes | Passes | Passes |
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is substantially equivalent to the predicate devices.
{2}------------------------------------------------
N/A is we don't know the specifications.
Summary of Testing:
| Test | Results |
|---|---|
| 1. Dermal Sensitization Test | Passes |
| 2. Primary Skin irritation | Passes |
| 3. Permeation testing per ASTM D 6978-05 | Passes |
| 4. Iodine Test | Passes |
| 5. Tensile strength | Gloves meets the requirements ofASTM D63 19-00a. |
| 6. Barrier strength | Gloves meets the requirements ofASTM D63 19-00a. |
The standards used by Shanghai Chinstar Corp to determine substantial equivalence are based on ASTM D 631900a-2005. All testing meets requirements for physical specifications and dimensions conducted on gloves, Inspection level S-2, AQL 4.0, pinholes at AQL 2.5
There are no special labeling claims and we do not claim our gloves to be hypoallergenic.
Conclusion:
Powder free Blue Nitrile Patient Examination Glove tested with chemotherapy drugs performance was equivalent to any other conventional method evaluated. Our evaluation concluded that our device raises no new issues of Safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a circular design with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem depicting a stylized eagle with three overlapping profiles, symbolizing health, services, and people. The eagle is a common symbol of the United States, representing freedom and strength.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shanghai China Star Corporation C/O Mr. Romil Rambhia Official Correspondent mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11001
AUG 3 0 2007
Re: K071072
Trade/Device Name: Shanghai China Star Corp Powder Free Blue Nitrile Patient Examination Glove Tested with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: August 21, 2007 Received: August 22, 2007
Dear Mr. Rambhia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -Mr. Rambhia
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): Not Assigned as of now
1/07/072
Applicant: Shanghai China Star Corp
Device Name: Shanghai China Star Corp Powder free Blue Nitrile Patient Examination Glove tested with chemotherapy drugs.
Indications for Use:
A powder free patient examination glove (Tested for Use with Chemotherapy Drugs) is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between patient and examiner. These gloves are not intended to be used as a chemical barrier.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Shula A Murphy, 6
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K021072
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.