K Number
K070947

Validate with FDA (Live)

Date Cleared
2008-02-14

(316 days)

Product Code
Regulation Number
876.1500
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Device Description

This special 510(k) is being submitted for an additional EndoWrist® instrument, the Cardiac Probe Grasper, to be utilized with probes and catheters used endoscopically during cardiac ablation procedures. This submission also includes the related minor labeling changes / clarifications on the da Vinci® Surgical Systems and EndoWrist Instruments. There are no changes in the technology, materials, manufacturing, performance, design, specifications or method of use for the da Vinci Surgical System or the CryoCath SurgiFrost Surgical CryoAblation System.

The da Vinci Surgical Systems consist of two integrated subsystems as follows: Intuitive Surgical® Endoscopic Instrument Control System: This sub-system is comprised of the Surgeon Console and Patient Side Cart. While seated at the Surgeon Console, the surgeon is able to view the surgical field and control critical aspects of the procedure, including movement of the endoscopic instruments and endoscope, within that field. Movements are controlled by the surgeon through use of Master Tool Manipulators (MTMs), two hand-operated mechanisms residing within the Surgeon Console. The EndoWrist® endoscopic instruments, stereo endoscope and camera attach to surgical arms located on the Patient Side Cart (PSC). The endoscopic instruments attach to arms known as Patient Side Manipulators (PSM). Another arm, known as the Endoscope Camera Manipulator (ECM), holds the stereo endoscope and camera. Commands from the Surgeon Console are relayed to the PSC, located adjacent to the patient, via various cables. Instrument and endoscope insertions, removals and exchanges are performed by another member of the surgical team positioned adjacent to the PSC. Intuitive Surgical® Stereo View Endoscopic System: The endoscopic vision system used with the da Vinci Surgical System, also known as the InSite® Vision System, consists of a stereo endoscope, endoscopic camera, and accessories including a light source and light guides. The InSite Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D) image of the surgical field. Additional monitors may be supplied by the user to allow other members of the surgical team to view the procedure.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments and the Intuitive Surgical® EndoWrist® Cardiac Probe Grasper, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria or detailed performance metrics. Instead, it relies on demonstrating substantial equivalence to predicate devices. The "performance" aspect is addressed by confirming that the new instrument maintains the basic functional characteristics of existing, cleared instruments within the da Vinci system.

Acceptance Criterion (Implicit)Reported Device Performance
Functional Equivalence: The Cardiac Probe Grasper must maintain the basic functional characteristics of existing EndoWrist instruments."Design analysis and comparison, as well as bench testing, have been conducted to confirm that basic functional characteristics of the subject devices are substantially equivalent to the predicate devices cited, and that design output meets the design input requirements." The document states the Cardiac Probe Grasper will be used to "position and hold probes used during ablative cardiac procedures, and will also facilitate precise and delicate blunt dissection of the pericardial reflections at the superior vena cava (SVC) and inferior vena cava (IVC)," which aligns with the general functions of EndoWrist instruments (grasping, blunt dissection). It also clarifies that "there are no changes in the instrument technology, materials, manufacturing or sterilization processes, requirements or compatibility with the da Vinci Surgical Systems." This implies the performance is consistent with the established system. Additionally, the da Vinci Surgical System itself receives a minor labeling change to include "delivery and placement of microwave and cryogenic ablation probes and accessories," and it's stated that "The use of the da Vinci Surgical System to deliver and place the SurgiFrost probe is equivalent to the currently cleared practice of using conventional minimally invasive surgical instruments." This suggests the system's performance in this new specific use case is also deemed equivalent.
Safety and Effectiveness: No new safety or effectiveness concerns should be introduced by the new instrument or labeling changes.The 510(k) submission aims to demonstrate substantial equivalence, which inherently means the device is as safe and effective as the predicate device(s). The fact that the FDA cleared the device implies they found no new safety or effectiveness concerns preventing market entry.

2. Sample Size Used for the Test Set and Data Provenance

The document only mentions "bench testing" and "design analysis and comparison." It does not provide specific details about a "test set" with a defined sample size for clinical or real-world performance evaluation.

  • Sample Size: Not specified.
  • Data Provenance: Not specified, but likely laboratory-based (bench testing). The regulatory submission is for a medical device in the US, so the "country of origin" of the data is implicitly internal to the manufacturer's testing or US-based contract labs. The study is presumed to be prospective as it's part of a pre-market submission, involving new tests on the Cardiac Probe Grasper.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. The document does not describe the establishment of a "ground truth" using experts in the context of the bench testing or design analysis mentioned. The evaluation is based on engineering principles and comparison to existing products.

4. Adjudication Method for the Test Set

Not applicable, as no expert-based ground truth establishment or adjudication process is described for a test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. This type of study is not mentioned in the document. The device is a surgical instrument and system, not an imaging AI or diagnostic algorithm that would typically undergo an MRMC study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a surgical instrument and system that inherently requires human-in-the-loop operation. It is not an algorithm that operates autonomously.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to diagnostic AI algorithms (e.g., pathology, outcomes data) is not directly applicable here. For this surgical instrument, the "ground truth" for its performance is derived from its ability to meet design input requirements and demonstrate functional equivalence and compatibility with the existing da Vinci Surgical System, which is already legally marketed and established as safe and effective for its intended uses. This is verified through:

  • Design analysis and comparison: Engineering review against predicate devices.
  • Bench testing: Laboratory tests to confirm functional characteristics.

8. The Sample Size for the Training Set

Not applicable. This device is a mechanical surgical instrument and system, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Section II

:4

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

DateFebruary 8, 2008
SubmitterIntuitive Surgical, Inc.950 Kifer RoadSunnyvale, CA 94086
ER Number2955842
ContactMichael YramateguiSr. Director, Regulatory & Quality AffairsTelephone: (408) 523 - 2145Fax: (408) 523 - 1390E-mail: mike.yramategui@intusurg.com
SubjectDeviceTrade Name(s):1. Intuitive Surgical ® da Vinci ® Surgical System and Endoscopic Instruments2. Intuitive Surgical ® EndoWrist ® Cardiac Probe GrasperClassification Name:System, Surgical, Computer Controlled Instrument (21 CFR 876.1500)Common Name:Endoscopic Instrument Control System, Endoscopic Instruments and AccessoriesDevice Class:Intuitive Surgical ® da Vinci ® Surgical System and Endoscopic Instruments: Class II, NAY, System, Surgical, Computer Controlled Instrument

510(k) Number: K070947

{1}------------------------------------------------

11

PredicateDevicesIntuitive Surgical da Vinci® Surgical System and EndoscopicInstruments(legally marketed underK990144, K002489, K011002, K013416, K021036,K022574, K040237, K43153, K043288, K050404, K050369, K050802 andK063220)
Intuitive Surgical EndoWrist® Long Tip Forceps and EndoWristPericardial Dissector(legally marketed under K990144, K013416, K021036, and K050369)
CryoCath Inc. SurgiFrost® Surgical CryoAblation System(legally marketed under K062140, K053436 and K040690)
DeviceDescriptionThis special 510(k) is being submitted for an additionalEndoWrist® instrument, the Cardiac Probe Grasper, to be utilized

with probes and catheters used endoscopically during cardiac ablation procedures. This submission also includes the related minor labeling changes / clarifications on the da Vinci® Surgical Systems and EndoWrist Instruments. There are no changes in the technology, materials, manufacturing, performance, design, specifications or method of use for the da Vinci Surgical System or the CryoCath SurgiFrost Surgical CryoAblation System.

EndoWrist® Cardiac Probe Grasper

The Cardiac Probe Grasper will be used with the da Vinci Surgical Systems to position and hold probes used during ablative cardiac procedures, and will also facilitate precise and delicate blunt dissection of the pericardial reflections at the superior vena cava (SVC) and inferior vena cava (IVC).

This Intuitive Surgical instrument will be compatible with the previously approved da Vinci Endoscopic Instrument Control Systems Models IS1200 and IS2000.

To address use of the Cardiac Probe Grasper with cardiac ablation probes, an Instructions for Use addendum sheet will be included with the general IFU for instruments that are compatible with the specified probes.

{2}------------------------------------------------

12

Device Description (continued)

da Vinci® Surgical Systems

The da Vinci Surgical Systems consist of two integrated subsystems as follows: Intuitive Surgical® Endoscopic Instrument Control System:

This sub-system is comprised of the Surgeon Console and Patient Side Cart. While seated at the Surgeon Console, the surgeon is able to view the surgical field and control critical aspects of the procedure, including movement of the endoscopic instruments and endoscope, within that field. Movements are controlled by the surgeon through use of Master Tool Manipulators (MTMs), two hand-operated mechanisms residing within the Surgeon Console.

The EndoWrist® endoscopic instruments, stereo endoscope and camera attach to surgical arms located on the Patient Side Cart (PSC). The endoscopic instruments attach to arms known as Patient Side Manipulators (PSM). Another arm, known as the Endoscope Camera Manipulator (ECM), holds the stereo endoscope and camera. Commands from the Surgeon Console are relayed to the PSC, located adjacent to the patient, via various cables. Instrument and endoscope insertions, removals and exchanges are performed by another member of the surgical team positioned adjacent to the PSC.

Intuitive Surgical® Stereo View Endoscopic System: The endoscopic vision system used with the da Vinci Surgical System, also known as the InSite® Vision System, consists of a stereo endoscope, endoscopic camera, and accessories including a light source and light guides. The InSite Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D) image of the surgical field. Additional monitors may be supplied by the user to allow other members of the surgical team to view the procedure.

Intended Use

EndoWrist® Cardiac Probe Grasper

EndoWrist Instruments, including scissors, scalpels, forceps, needle drivers and electrocautery are intended for endoscopic manipulation of tissue, including: grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery and suturing.

{3}------------------------------------------------

Intended Use (continued) The Cardiac Probe Grasper is an addition to the EndoWrist Instrument family, and there are no changes to be made in the intended use of the device.

da Vinci® Surgical Systems

The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures. and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

This minor labeling change is a clarification within the currently cleared indications for use, and does not alter the intended use of the device.

EndoWrist® Cardiac Probe Grasper Comparison to Predicate The design of the predicate devices' distal end effectors, or Device grips, was modified in order to better grasp and manipulate cardiac ablation probes. There will be no change in the instrument technology. materials. manufacturing or sterilization processes, requirements or compatibility with the da Vinci Surgical Systems.

{4}------------------------------------------------

Comparison toPredicateDevice(continued)Labeling modification for the Cardiac Probe Grasper willconsist of new box labeling and a supplement to the InstrumentInstructions For Use describing the use of the instrument withspecified cardiac ablation probes.
da Vinci® Surgical Systems, CryoCath CryoAblation System
There are no changes in the design, technology, materials,manufacturing, performance, specifications, or method of usefor either the Intuitive da Vinci Surgical Systems or theSurgiFrost® Surgical CryoAblation System.
The only modification consists of an update to the Indicationstatement which provides a clarification of tasks that can beperformed with the da Vinci Surgical Systems to add deliveryand placement of cryogenic ablation probes. There will be nomodifications to CryoCath SurgiFrost labeling.
The use of the da Vinci Surgical System to deliver and placethe SurgiFrost probe is equivalent to the currently clearedpractice of using conventional minimally invasive surgicalinstruments. This change does not impact the operationalcharacteristics or change the intended use of either device.
TechnologicalCharacteristicsThe technological characteristics of the subject devices are thesame as the predicate devices.
PerformanceDataDesign analysis and comparison, as well as bench testing, havebeen conducted to confirm that basic functional characteristics ofthe subject devices are substantially equivalent to the predicatedevices cited, and that design output meets the design inputrequirements.
ConclusionBased upon the technical information, intended use andperformance information provided in this pre-market notification,the EndoWrist® Cardiac Probe Grasper instrument and the daVinci Surgical Systems described herein have been shown to besubstantially equivalent to current legally marketed predicatedevices.

14

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA", which is arranged in a circular fashion around the top and left side of the caduceus.

FEB 1 4 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Intuitive Surgical, Inc. c/o Mr. Michael Yramategui Senior Director, Regulatory Affairs 950 Kifer Rd. Sunnyvale, CA 94086

Re: K070947

Trade/Device Name: Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments; Intuitive Surgical® EndoWrist® Cardiac Probe Grasper Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II (two) Product Code: NAY, GEI, GEH Dated: December 10, 2007 Received: December 11, 2007

Dear Mr. Yramategui:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Michael Yramategui

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at ((240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Blommor for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section III

Indications for Use

510(k) Number (if known): K070947

Intuitive Surgical® da Vinci® Surgical System and Endoscopic Instruments Device Name: Intuitive Surgical® EndoWrist® Cardiac Probe Grasper

Indications For Use:

The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Prescription Use × (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

E. Mcd

(Division Sign-Off)
Division of Cardiovascular Devices

Koloqui

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.