AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical® da Vinci® Surgical System is intended to assist in the accurate control of Intuitive Surgical® Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing and delivery and placement of microwave ablation probos and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Device Description

The da Vinci Surgical System, consists of three integrated sub-systems as follows: Endoscopic Instrument Control System: The Endoscopic Instrument Control System is comprised of two sub-systems - the Surgeon Console, Model IS2000-SSC, and Patient Side Cart. Model IS2000-PSC. While seated at the Surgeon Console, the surgeon controls critical aspects of the procedure, including movement of the endoscopic instruments and endoscope, within the operative field. Endoscopic instrument and camera movements are controlled by the surgeon through use of the Master Tool Manipulators (MTM), two hand operated mechanisms residing within the Surgeon Console. The endoscopic instruments are held in a fixed position (with respect to the patient) by either two (or optionally three) unique manipulators known as Patient Side Manipulators (PSM), which are located on the Patient Side Cart (PSC). The endoscope is also held in a fixed position (with respect to the patient) by another manipulator, similar to the PSM, known as the Endoscope Camera Manipulator (ECM) and also located on the PSC). The PSM and ECM are attached to surgical arms on the PSC known as Set-up Joint (SUJ) arms. Commands from the Surgeon Console are relayed to the PSC, which is located immediately adjacent to the patient, via a cable. Instrument and endoscope changes are performed by another provider positioned adjacent to the PSC. Stereo View Endoscopic Vision System: The endoscopic vision system used with the da Vinci Surgical System, also known as Intuitive Surgical Insite Vision System. Model VS1000, consists of a stereo endoscope, endoscopic camera, and various accessories, including a light source and light guides. The Insite Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D image) image of the surgical field.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Intuitive Surgical da Vinci Surgical System and Endowrist Instruments. However, the document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study with specific acceptance criteria and performance data for a new device.

Therefore, many of the requested items (acceptance criteria table, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, training set size and ground truth establishment) are not explicitly provided in this type of regulatory submission.

The document indicates that the submission is for design modifications to the existing da Vinci Surgical System, primarily affecting the Patient Side Manipulators (PSM), Endoscope Camera Manipulator (ECM), and the Patient Side Cart to enhance ease of use. Minor changes to the Surgeon Console and endoscopic instruments are also noted.

The core of the submission is to prove that these modifications do not alter the fundamental safety and effectiveness of the device compared to its already legally marketed predicate devices.

Here's an attempt to answer the questions based only on the provided text, highlighting what is implicitly or explicitly stated and what is missing:


1. A table of acceptance criteria and the reported device performance

This information is not explicitly provided in a quantified table format. The document states: "Design analysis and comparison, as well as bench testing and risk analysis activities, have been conducted to confirm that the characteristics of the modified device are substantially equivalent to the predicate devices cited."

This suggests that the "acceptance criteria" were likely related to demonstrating that the modified device's performance (functionality, safety, physical characteristics) was within acceptable parameters or identical to the predicate device, rather than achieving specific performance thresholds for a novel function. The "reported device performance" is essentially that it met the criteria for substantial equivalence.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided. The document mentions "bench testing" but does not specify the sample size of devices, components, or any test data used. The provenance (country of origin, retrospective/prospective) is also not stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided. In a substantial equivalence submission for a surgical system's mechanical modifications, "ground truth" as typically understood in AI/imaging studies (e.g., expert consensus on medical images) is not the primary focus. The "ground truth" here is more about engineering specifications, device functionality, and safety validated through design analysis and bench testing.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided. Adjudication methods are typically relevant in studies involving human interpretation or subjective assessments. Here, the assessment seems to be based on objective engineering comparisons and testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not conducted or mentioned. This type of study is relevant for AI-assisted diagnostic or interpretative devices, which this surgical system is not. The submission is for hardware modifications to a surgical control system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This concept is not applicable to the device described. The da Vinci Surgical System is a human-in-the-loop system by its very nature (a surgeon controls it). The modifications are to the physical components of the system, not a standalone algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" implicitly used for this type of submission would be the established performance and safety characteristics of the predicate device, against which the modified device's characteristics are compared. This is based on engineering specifications, functionality tests, and risk analysis. It does not involve medical "ground truth" like pathology or outcomes data in the way an AI diagnostic device would.

8. The sample size for the training set

This information is not provided and is not applicable. This device is a surgical system with physical components and software, not an AI model that requires a "training set" of data in the machine learning sense.

9. How the ground truth for the training set was established

This information is not provided and is not applicable for the same reason as above.


Summary of what the document does convey regarding meeting acceptance criteria:

The document states its Conclusion: "Based upon the device's general specifications, intended use, and results of risk analysis and performance testing provided in this pre-market notification, the da Vinci Surgical System described herein has been shown to be substantially equivalent to current legally marketed predicate device."

This indicates that the implicit "acceptance criteria" were met through:

  • Design analysis and comparison to predicate devices.
  • Bench testing.
  • Risk analysis activities.

The core "performance" reported is that the "technology, materials, manufacturing methods, and performance are essentially the same for the predicate device as the subject device," with the primary differences being form factor and electromechanical components for ease of use, but not affecting core operation or functionality.

In essence, for this 510(k) submission for modifications to an existing surgical system, the "acceptance criteria" and "study" are focused on demonstrating that the changes do not negatively impact equivalence to the predicate device, rather than quantifying new performance metrics for a novel technology.

{0}------------------------------------------------

Section II

APR 2 9 2005

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

510(k) Number: K050369

DateApril 26, 2005
SubmitterIntuitive Surgical, Inc.950 Kifer RoadSunnyvale, CA 94086
ER Number2955842
ContactMike YramateguiDirector, Regulatory AffairsTelephone: (408) 523 - 2145Fax: (408) 523 - 1390e-mail: mike.yramategui@intusurg.com
SubjectDeviceName: Intuitive Surgical® da Vinci® Surgical System and Endowrist™InstrumentsClassification Name: System, Surgical, Computer Controlled Instrument(21 CFR 876.1500)Common Name: Endoscopic Instrument Control System, EndoscopicInstruments, and accessories
PredicateDevicesIntuitive Surgical da Vinci Surgical System and Endoscopic Instruments(legally marketed under K990144 / K002489 / K011002 / K013416 /K021036 / K022574 / K040237 / K043153 / K050005)
DeviceDescriptionThe da Vinci Surgical System, consists of three integrated sub-systems asfollows:
Endoscopic Instrument Control System: The Endoscopic InstrumentControl System is comprised of two sub-systems - the Surgeon Console,Model IS2000-SSC, and Patient Side Cart. Model IS2000-PSC. Whileseated at the Surgeon Console, the surgeon controls critical aspects of theprocedure, including movement of the endoscopic instruments andendoscope, within the operative field. Endoscopic instrument and cameramovements are controlled by the surgeon through

{1}------------------------------------------------

Device Description (continued)

use of the Master Tool Manipulators (MTM), two hand operated mechanisms residing within the Surgeon Console. The endoscopic instruments are held in a lixed position (with respect to the patient) by cither two (or optionally three) unique manipulators known as Patient Side Manipulators (PSM), which are located on the Patient Side Cart (PSC). The endoscope is also held in a fixed position (with respect to the patient) by another manipulator, similar to the PSM, known as the Endoscope Camera Manipulator (ECM) and also located on the PSC). The PSM and ECM are attached to surgical arms on the PSC known as Set-up Joint (SUJ) arms. Commands from the Surgeon Console arc relayed to the PSC, which is located immediately adjaccnt to the patient, via a cable. Instrument and endoscope changes are performed by another provider positioned adjacent to the PSC.

Stereo View Endoscopic Vision System: The endoscopic vision system used with the da Vinci Surgical System, also known as Intuitive Surgical Insite Vision System. Model VS1000, consists of a stereo endoscope, endoscopic camera, and various accessories, including a light source and light guides. The Insite Vision System provides two independent images that are relayed to the viewer located in the Surgeon Console, where they are fused to form a 3-D (or alternatively a 2-D image) image of the surgical field.

Intended Use

The Intuitive Surgical da Vinci Surgical System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, clectrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing and delivery and placement of microwave ablation probos and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

{2}------------------------------------------------

Comparison toPredicateDeviceThis 510(k) notification is submitted for design modifications to theEndoscopic Instrument Control System. These modifications affectprimarily the PSM, ECM, and the Patient Side Cart in order to enhancecase of use. There are also associated minor changes to the SurgeonConsole and endoscopic instruments. The user interface design isessentially identical to the predicate device except for minormodifications to accommodate the aforementioned changes.
Device operation, functionality, and methods of use for the subject deviceare identical to the predicate device. The technology, materials,manufacturing methods, and performance are essentially the same for thepredicate device as the subject device. The primary differences are 1)changes to the PSM and ECM form factor and associatedelectromechanical components; 2) changes to the Patient Side Cartmechanics and circuitry to accommodate the modified PSM and ECM;and 3) changes to the Surgeon Console and software to accommodate themodified Patient Side Cart. Minor mechanical modifications to theendoscopic instruments to interface with the modified PSM are alsodescribed.
TechnologicalCharacteristicsThe technological characteristics of the subject devices are essentially thesame as for the predicate devices.
PerformanceDataDesign analysis and comparison, as well as bench testing and riskanalysis activities, have been conducted to confirm that thecharacteristics of the modified device are substantially equivalent to thepredicate devices cited.
ConclusionBased upon the device's general specifications, intended use, and resultsof risk analysis and performance testing provided in this pre-marketnotification, the da Vinci Surgical System described herein has beenshown to be substantially equivalent to current legally marketed predicatedevice.

1

:

:

·


{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a stylized human figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 9 2005

Mr. Michael H. Yramategui Director, Regulatory Affairs Intuitive Surgical, Inc. 950 Kifer Road Sunnyvale, California 94086

Re: K050369

K050507
Trade/Device Name: Intuitive Surgical da Vinci Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NA Y Dated: February 10, 2005 Received: February 14, 2005

Dear Mr. Yramategui:

We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(x) premainter is substantially equivalent (for the indications referenced above and nave determined the arrest are devices marketed in interstate for use stated in the enclosure) to tegally manced profithe Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 20, 1970, the enactition with the provisions of the Federal Food, Drug, devices that have been recrassified in accordance while approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a provisions of the Act . The and Cosmetic Act (Act) that do not require apperal controls provisions of the Act. The Act. The
You may, therefore, market the device, subject to the general consucted pos You may, inerefore, market the device, belyer to meet general controls provisions of the Fet merade requirements of the management misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mis chairs are regulations affecting your device can
may be subject to such additional controls. Existing major regulation FDA may may be subject to such additional controlis. Entroling and on 898. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to Register be found in the Code of Peacharies.
publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualice of a substitive only other requirements of the Act
that FDA has made a determination that your device acteral agencies. You must that FDA has made a determination that your arrest by other Federal agencies. You must and or any Federal statutes and regulations and united to: registration and listing (21)
comply with all the Act's requirements, including, but not one provine moutrements as set comply with all the Act s requirements, mendanturing practice requirements as set
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFR Part 807); labeling (21 CFR 1 at 601); god manat 820); and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR 1000-1050 forth in the quality systems (QS) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000

{4}------------------------------------------------

Page 2 – Mr. Michael H. Yramategui

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your article equivalence of your device to a legally
premarket notification. The FDA finding of sybstantial end this permits yo premarket notification. The FDA multing of substantial equive and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
rs of the may of the more of 0115. All any desse not the requiries entitled If you desire specific advice for your ac not on on success note the regulation entitled, the regulation entitled, contact the Office of Complaned at (210) 276 - 1 - 1 - (21)FR Part 807.97). You may obtain "Misbranding by reference to premarket nothleadon (en oet from the Division of Small
other general information on your responsibilities under the Act from number (800) 638-20 other general information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-findex html Manufacturers, international and Consulter Abolokation and Consultances of the new onlines.html.

Sincerely yours,

Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K050369

Intuitive Surgical® da Vinci® Surgical System and Endowrist 60 Device Name: Instrumicats

Indications For Use:

The Intuitive Surgical" da Vinci" Surgical System is intended to assist in the accurate control of Intuitive Surgical® Endoscopic Instruments including rigid endoscopes, blunt and sharp cadoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, clectrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and shapp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave ablation probos and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general noncardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system can also be employed, with adjunctive mediastinotomy to norform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating toom environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDCO)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

tive

Ko 50369

Page 1 of 1

Intuitive Surgical" da Vinci Surgical System

SIO(k) Section III, Page 1 of 1

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.