K Number
K070223

Validate with FDA (Live)

Date Cleared
2007-03-13

(48 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Audit™ MicroCV™ Lipids Linearity Set (Low) consists of five levels. Each level contains the following analytes: Apolipoprotein A1, Apolipoprotein B, Cholesterol, HDL Cholesterol. LDL Cholesterol and Triglyceride and may be used for proficiency testing in interlaboratory surveys and to perform CLIA directed calibration verification for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

In addition, Level A - E of this product may be used as unassayed quality control material for these analytes or as an assayed quality control material for the analyzer systems specified in the package insert. It is not intended to be used as an assayed quality control material for any other analyzer systems.

Device Description

The Audit™ MicroCV™ Lipids Linearity Set (Low) is a human based, lyophilized, five level set of QC material, with each level containing 6 analytes. It is used to confirm the proper calibration, linear operating range, and reportable range of Lipids methods for the analytes listed. Level A is near the lower limit level and Level E has concentrations near the upper limit of instruments. Levels B - D are related by linear dilution of Level A and Level E.

AI/ML Overview

The Audit™ MicroCV™ Lipids Linearity Set (Low) is a quality control material used to verify and validate the Analytical Measurement Range for non-waived Lipids testing methods.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Reconstituted Stability24 hours when stored tightly capped at 2 - 8° C.
Shelf LifeTwo years, when stored unopened at 2 - 8° C.
Substantial EquivalenceDeemed substantially equivalent to predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

The provided document does not explicitly state the sample size used for the stability studies. It only mentions that "All supporting data is retained on file at Aalto Scientific, Ltd." The data provenance is internal to Aalto Scientific, Ltd. and the studies are described as "real time studies" which implies a prospective nature for shelf life, while reconstituted stability would also be prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided. The study focuses on product stability and equivalence to a predicate device, rather than diagnostic accuracy requiring expert consensus on a test set.

4. Adjudication Method for the Test Set:

Not applicable. The study design does not involve adjudication of results by multiple experts as it is focused on the chemical stability and performance characteristics of a quality control material.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

Not applicable. This type of study is relevant for diagnostic devices that involve human interpretation of results. The Audit™ MicroCV™ Lipids Linearity Set (Low) is a quality control material.

6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance):

This concept is not directly applicable. The device itself is a quality control material, not an algorithm. Its performance is measured by its stability and ability to perform as a linearity set, which is an inherent characteristic of the material. The studies performed are to establish these inherent characteristics.

7. Type of Ground Truth Used:

The ground truth for the stability studies is implied to be the direct measurement of analyte concentrations over time under specified storage conditions. For the substantial equivalence claim, the ground truth is the established performance and characteristics of the predicate device (Audit™ MicroCV™ General Chemistry Linearity Set K042318).

8. Sample Size for the Training Set:

Not applicable. The product is a quality control material, not a machine learning algorithm that requires a training set. The "study" mentioned refers to stability testing, not algorithm training.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this type of product. The ground truth for the stability studies would be established through standard analytical laboratory procedures, measuring the chemical composition of the control material over time.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten sequence of characters and numbers. The sequence starts with the letters 'KO', followed by the numbers '70223'. The handwriting is somewhat rough, with thick strokes, and the characters are closely spaced together.

Image /page/0/Picture/1 description: The image shows a logo for Aalto Scientific, Ltd. The logo features a triangular shape composed of many small dots, resembling a pyramid or mountain. Below the triangular shape, the text "Aalto Scientific, Ltd." is written in a simple, sans-serif font. The overall design is clean and minimalist.

510(k) Summary

A. Submitter

MAR 1 3 2007

Aalto Scientific, Ltd. 1959 Kellogg Ave. Carlsbad, CA 92008 (760) 431-7922 Telephone: (760) 431-6824 Fax:

B. Contact Person

Alan Vekich ي س Regulatory Affairs Specialist (760) 431-7922 ext. 12 1760) Telephone: avekich@aaltoscientific.com E-mail:

C. Date of Summary Preparation

December 1, 2006

D. Device Identification

Product Trade Name:Audit™ MicroCV™ Lipids Linearity Set (Low)
Common Name:Lipids Linearity
Classification Name:Assay QC Material
Device Classification:Class I
Regulation Number:21 CFR 862.1660
Panel:75
Product Code:JJY

E. Device to Which Substantial Equivalence is Claimed

Audit™ MicroCVTM General Chemistry Linearity Set Aalto Scientific, Ltd., Carlsbad, CA K042318

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for Aalto Scientific, Ltd. The logo features a stylized triangular shape at the top, composed of many small dots or squares, giving it a textured appearance. Below the triangular shape, the text "Aalto Scientific, Ltd." is printed in a simple, sans-serif font.

. .

F. Description of the Device

The Audit™ MicroCV™ Lipids Linearity Set (Low) is a human based, Iyophilized, five level set of QC material, with each level containing 6 analytes. It is used to confirm the proper calibration, linear operating range, and reportable range of Lipids methods for the analytes listed. Level A is near the lower limit level and Level E has concentrations near the upper limit of instruments. Levels B - D are related by linear dilution of Level A and Level E.

G. Statement of Intended Use

The Lipids Linearity Set is intended to simulate human patient serum samples for the purpose of verifying and validating the Analytical Measurement Range for non-waived Lipids testing methods as identified in the package insert.

I. Summary of Performance Data

Stability studies have been performed to determine the reconstituted stability and shelf life for the Audit™ MicroCVTM Lipids Linearity Set (Low). All supporting data is retained on file at Aalto Scientific, Ltd. Product claims are as follows:

Reconstituted Stability: Once a vial has been reconstituted, all analytes will be stable for 24 hours when stored tightly capped at 2 - 8° C.

Shelf Life: Two years, when stored unopened at 2 - 8° C. Note: Real time studies are ongoing to support the shelf life of this product.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Aalto Scientific, Ltd. The logo consists of the text "Aalto Scientific, Ltd." in a simple font, with the "A" in "Aalto" slightly larger than the other letters. Above the text is a triangular shape made up of many small circles or dots, creating a textured, pyramid-like design.

H. Technical Characteristics Compared to Predicate Device

CharacteristicsAudit™ MicroCV™ Lipids Linearity Set(Low)(New Device)Audit™ MicroCV™ General ChemistryLinearity Set(K042318)
Intended UseAudit™ MicroCV™ Lipids Linearity Set(Low) is assayed quality control materialconsisting of Delipidized Human Serum.It is intended to simulate human patientserum samples for the purpose ofmonitoring the precision and to detectsystematic analytical deviations oflaboratory testing procedures. Thisproduct may also be used as unassayedquality control material for these sameanalytes and may be used for proficiencytesting in inter-laboratory surveys. Inaddition, this product may also be usedto perform CLIA directed calibrationverification for these same analytes withsimilar reagents on similarinstrumentation in accordance withcurrent CLIA-88 guidelines andregulations.Audit™ MicroCV™ General ChemistryLinearity Set is assayed quality controlmaterial consisting of human basedserum. It is intended to simulate humanpatient serum samples for the purpose ofmonitoring the precision and to detectsystematic analytical deviations oflaboratory testing procedures. Thisproduct may also be used as unassayedquality control material for these sameanalytes and may be used for proficiencytesting in inter-laboratory surveys. Inaddition, this product may also be used toperform CLIA directed calibrationverification for these same analytes withsimilar reagents on similarinstrumentation in accordance withcurrent CLIA-88 guidelines andregulations.
Number ofAnalytes per vial630
Number of levelsper set55
Contents5 x 2 mls5 x 5 mls
MatrixDelipidized Human SerumHuman Based Serum
Type of AnalytesLipidsGeneral Chemistry
FormLyophilizedLyophilized
StabilizersSucroseNone
PreservativesSorbitolSorbitol
Storage2 to 8° CUntil expiration date2 to 8° CUntil expiration date
ReconstitutedStability24 hours at 2 to 8° C24 hours at 2 to 8° C

·

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Aalto Scientific, Ltd. The logo features a triangular shape at the top, composed of many small dots. Below the triangular shape, the text "Aalto Scientific, Ltd." is written in a simple, sans-serif font. The text is left-aligned and positioned directly under the triangular shape.

J. Conclusions

Based upon the purpose of the device, the descriptions and labeling of the predicate device, the safety and efficacy, and the stability data generated, the product is substantially equivalent to the predicate device.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The background is white.

Public Health Service

Food and Druq Administration 2098 Gaither Road Rockville MD 20850

Aalto Scientific, Ltd. c/o Mr. Alan Vekich Regulatory Affairs Specialist 1959 Kellogg Avenue Carlsbad, CA 92008

MAR 1 3 2007

Re: K070223 Trade/Device Name: Audit™ Microcv™ Lipids Linearity Set (Low) Regulation Number: 21 CFR§862.1660 Regulation Name: Quality control material (assayed and unassayed). Regulatory Class: Class I Product Code: JJY Dated: January 17, 2007 Received: January 24, 2007

Dear Mr. Vekich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean m. Cooper, M.S., D.v.M.

Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): unknown

Device Name: Audit™ MicroCV™ Lipids Linearity Set (Low)

Indications For Use:

The Audit™ MicroCV™ Lipids Linearity Set (Low) consists of five levels. Each level contains the following analytes: Apolipoprotein A1, Apolipoprotein B, Cholesterol, HDL Cholesterol. LDL Cholesterol and Triglyceride and may be used for proficiency testing in interlaboratory surveys and to perform CLIA directed calibration verification for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

In addition, Level A - E of this product may be used as unassayed quality control material for these analytes or as an assayed quality control material for the analyzer systems specified in the package insert. It is not intended to be used as an assayed quality control material for any other analyzer systems.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Benson
Sign-Off

ffice of In Vitro Diagnostic Device Valuation and Safety

K070223

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.