K Number
K042318

Validate with FDA (Live)

Date Cleared
2004-12-09

(105 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Audit™ MicroCV" General Chemistry Linearity Set consists of five levels of human based serum. Each level contains the following analytes: Acid Phosphatase, Albumin, Alkaline Phosphatase, ALT, Amylase, AST, Bilirubin (Total and Direct), BUN, Calcium, Chloride, Cholesterol, CO2 Creatine Kinase, Creatinine, Gamma-GT, Glucose, HDL Cholesterol, fron, LDH, LDL Cholesterol, Lactate, Lipase, Magnesium, Phosphorus, Potassium, Sodium, Total Protein, Triglycerides and Uric Acid. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents and instruments.

This product may be used for proficiency testing in interlaboratory surveys and to perform CLIA directed calibration verification for these same analytes with similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

In addition, levels B - E of this product may be used as unassayed quality control material for these analytes or as an assayed quality control material for the analyzer systems specified in the package insert. It is not intended to be used as an assayed quality control material for any other analyzer systems.

Device Description

The Device is a QC material that will be used to determine the linearity of clinical chemistry assays as defined by CLIA-88 in Federal Register 42 CFR Part 493, Department of Health and Human Services, January 24, 2003.

Audit™ MicroCV™ General Chemistry Linearity Set is assayed quality control material consisting of human based serum. It is intended to simulate human patient serum samples for the purpose of monitoring the precision and to detect systematic analytical deviations of laboratory testing procedures for Acid Phosphatase, Albumin, Alkaline Phosphatase, ALT, Amylase, AST, Bilirubin (Total and Direct), BUN, Calcium, Chloride, Cholesterol, CO2 Creatine Kinase, Creatinine, Gamma-GT, Glucose, HDL Cholesterol, Iron, LDH, LDL Cholesterol, Magnesium, Comments, Potassium, Sodium, Total Protein, Lipase, La Crete, Triglycerides and Uric Acid. These five levels demonstrate a linear relationship to each other for their respective analytes, reagents and instruments.

This product may also be used as unassayed quality control material for these same analytes and may be used for proficiency testing in interlaboratory surveys. In addition, this product may also be used to perform CLIA directed calibration for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

The five levels QC material is to be assayed as an unknown for the determination of linearity.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Audit™ MicroCV™ General Chemistry Linearity Set, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Internal Standard)Reported Device Performance (Target Values)
Linearity: Coefficient of correlation (R²) ≥ 0.975 (for five linearity levels)Achieved for all analytes. (Implied by the conclusion of "linearity confirmed.")
Storage Stability (Unopened Vial): 90-110% recovery of concentration/activity compared to reference vial (0 day) at 2-8°C.Predicted 12 years at 2-8°C.
Open Vial Stability: 90-110% recovery of concentration/activity compared to reference vial (0 day).Predicted 24 hours at 2-8°C.

Note: The document specifies manufacturing targets for each analyte at five different levels (A-E) as an internal standard for production and value assignment. However, the overarching performance acceptance criteria for the device itself (as a linearity set) are linearity and stability. The individual analyte target values and ranges (e.g., Bicarbonate Level A: 10 ± 1 mmol/L) represent the intended composition of the product, which then must demonstrate linearity across these levels.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: For stability studies:
    • Stress Stability: Selected vials (Level A and Level E) from a lot were removed at different time intervals. The exact number of vials or lots is not specified beyond "selected vials."
    • Open Bottle Study: "TEST" vials (not a specific number, implying multiple vials from a lot) were reconstituted and repeatedly accessed.
  • Data Provenance: Retrospective, as the stress stability study was a "backward performed study." The "real time stability data is be collected," indicating the initial clearance relied on accelerated studies. The origin of the base matrix is a mixture of bovine serum and normal processed human serum.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Not applicable. This device is an in vitro diagnostic quality control material, not an AI or imaging device requiring expert interpretation of results. The "ground truth" for its performance is established by analytical methods and statistical calculations. The "value assignment protocol" describes how target values are derived, which involves instrumental measurements rather than expert consensus on a diagnostic outcome.

4. Adjudication Method for the Test Set

  • Not applicable. See point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. This is not an AI or imaging device; it is a laboratory quality control material. MRMC studies are used to evaluate the diagnostic performance of devices, often in comparison to human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

  • No. This is not an algorithmic device. Its performance is intrinsic to its chemical composition and how it performs across analytical instruments.

7. The Type of Ground Truth Used

  • Analytical Measurements and Statistical Calculation:
    • Linearity Ground Truth: Established by plotting target values against designated levels and calculating the coefficient of correlation (R²). An R² ≥ 0.975 indicates linearity.
    • Stability Ground Truth: Established by comparing the concentration/activity of analytes in stressed/opened vials against reference vials (zero day) using linear regression and modified Arrhenius equations to predict stability. The acceptable range is 90-110% recovery.
    • Value Assignment Ground Truth: For each analyte, 30 values were collected (10 individual assays from a single vial, assayed in the morning and afternoon over five days from three instruments/labs if available). The average of these 30 data points served as the target value.

8. The Sample Size for the Training Set

  • Not applicable. This product is a physical quality control material, not a machine learning model. It does not have a "training set" in the computational sense. The manufacturing targets and value assignments are based on internal analytical data from the production of the material itself.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. See point 8.

{0}------------------------------------------------

DEC - 9 2004

K042318

Revised Summary

In Vitro Diagnostic Devices: Aalto Scientific, Ltd. 510(k) Submissions Summary Form

Name: Aalto Scientific, Ltd. Address: 1959 Kellogg Ave, Carlsbad, CA 92008 Telephone: 760 - 431-7922 Fax Number: 760 - 431-6942 Contact Person: Alan Vekich Ext: 120 Establishment Registration Number: 2022832

Identification of the Device

Device Name:Audit™ MicroCV™ General Chemistry Linearity Set
Proprietary/Trade name:Audit™ MicroCV™ Linearity Set
Common Name:General Chemistry Linearity
Classification Name:Assay QC Material
Device Classification:Class I
Regulation Number:21 CFR § 862.1660
Panel:75
Product Code:JJY

Identification of the Predicate Device

Predicate Device Name: Manufacturer: 510(k) Number or Clearance Information: Predicate Device Labeling:

Validate Chem 10 Calibration Verification Test Set Maine Standards Company K023410

Description of the Device and reason for the submission

The Device is a QC material that will be used to determine the linearity of clinical chemistry assays as defined by CLIA-88 in Federal Register 42 CFR Part 493, Department of Health and Human Services, January 24, 2003.

{1}------------------------------------------------

Intended Use(s) and Indication(s) for Use of the subject device

Audit™ MicroCV™ General Chemistry Linearity Set is assayed quality control material consisting of human based serum. It is intended to simulate human patient serum samples for the purpose of monitoring the precision and to detect systematic analytical deviations of laboratory testing procedures for Acid Phosphatase, Albumin, Alkaline Phosphatase, ALT, Amylase, AST, Bilirubin (Total and Direct), BUN, Calcium, Chloride, Cholesterol, CO2 Creatine Kinase, Creatinine, Gamma-GT, Glucose, HDL Cholesterol, Iron, LDH, LDL Cholesterol, Magnesium, Comments, Potassium, Sodium, Total Protein, Lipase, La Crete, Triglycerides and Uric Acid. These five levels demonstrate a linear relationship to each other for their respective analytes, reagents and instruments.

This product may also be used as unassayed quality control material for these same analytes and may be used for proficiency testing in interlaboratory surveys. In addition, this product may also be used to perform CLIA directed calibration for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

Name:Validate® Chem 10 CalibrationVerification Test SetAudit™ MicroCV™ GeneralChemistry Linearity Set
510(k) Number:K023410K042318
Intended Use:Validate Chem 10 CalibrationVerification Test Set solutionsare intended for in vitrodiagnostic use in verifyingcalibration, validating reportableranges, and determining linearityin automated, semi-automatedand manual chemistry systems.Audit™ MicroCV™ GeneralChemistry Linearity Set isassayed quality control materialconsisting of human basedserum. It is intended to simulatehuman patient serum samples forthe purpose of monitoring theprecision and to detect systematicanalytical deviations oflaboratory testing procedures.This product may also be used asunassayed quality controlmaterial for these same analytesand may be used for proficiencytesting in interlaboratory surveys.In addition, this product may alsobe used to perform CLIAdirected calibration verificationfor these same analytes withsimilar reagents on similarinstrumentation in accordancewith current CLIA-88 guidelinesand regulations.
Stability Claims:Stable until the expiration dateprinted on the bottle and storagecontainer when stored at -10 to 20° C and handled according toinstructions.Stored at 2-8° C and will remainstable in the unopened vial fortwelve months from the date ofmanufacture. After opening, thecontents should be reconstitutedimmediately and used according

Substantial Equivalence

{2}------------------------------------------------

to the instrument manufacturer'sinstructions. After reconstitutionmaterial is stable 24 hours.
Constituents:Alkaline phosphates (ALP)Alanine aminotransferase (ALT)Amylase (AMY)Aspartate aminotransferase(AST)Creatine kinase (CK)x-glutamyl transferase (GGT)Lactate dehydrogenase (LD)Lipase (LIP)Bilirubin, totalAlkaline phosphates (ALP)Alanine aminotransferase (ALT)Amylase (AMY)Aspartate aminotransferase(AST)Creatine kinase (CK)x-glutamyl transferase (GGT)Lactate dehydrogenase (LD)Lipase (LIP)Bilirubin, totalAcid Phosphatase, TotalAlbuminBilirubin, ConjugatedCalciumLactateChlorideCholesterolCarbon dioxideCreatinineGlucoseHDL CholesterolIron, TotalLDL CholesterolMagnesiumPhosphorusPotassiumSodiumTotal ProteinTriglyceridesUrea (BUN)Uric Acid
Levels Available:FiveFive
Contents:10 x 5mls5 x 5 mls

Technological Characteristics of the Device

The five levels QC material is to be assayed as an unknown for the determination of linearity.

Performance Characteristics and Data'

Specific Performance Characteristics and Data

Method Comparison - a method comparison for each sample or matrix claimed for analysis, e.g., whole blood, capillary whole blood, serum, plasma, urine, etc.

{3}------------------------------------------------

Matrix:

    1. The base matrix is mixture of two animal sera: Bovine serum and Normal Processed Human Serum that has had the lipids removed.
    1. NaCl, LiCl, Lactic Acid, BUN (Urea), CaCl2, Creatinine, Dextose, MgCl2, NaHPO4, Uric Acid, KCI, iron, Na Acetate, Conjugate Bilirubin, Bilirubin, are ACS or Reagent Grade Commercially available chemicals. Used as an analyte adjustor.
    1. Pyridoxal-5-Phosphate is an ACS or Reagent Grade Commercially available chemicals and used either as weighed crystal or dissolved in DI water and measured out as a liquid. Used as a stabilizer.
    1. LD, AST/SGOT, ALT/SGPT, CK, GGT, Amylase, Lipase, Alkaline Phosphatase, and Acid Phosphatase are plant or animal derived material from approved vendors. The Vendors supply a COA and Aalto Scientific, Ltd.'s OA department confirms the claims.
    1. Sucrose and Sorbitol are ACS or Reagent Grade Commercially available chemicals and used as a Freeze/Dry bulking agent and preservative.
  • The QC materials and methods used in the manufacturing of the General Chemistry Linearity Set. Level A though E where Class A type general chemistry equipment (such as graduated cylinders) and calibrated balances. Testing materials where the reagents, calibrators and controls that were obtained from the respected manufacturers.
Precision/Reproducibility:N/A
Linearity/assay reportable range:N/A

{4}------------------------------------------------

Manufacturing:

The Linearity set is made according to NCCLS protocol #EP6-A for a five level linearity set. The dilution schemes can be found in Appendix A for a five level sample with equally spaced concentrations. Just like a Proficiency Sample Set where the actual concentrations are not important for the purposes of assessing linearity as long as dilutions are proportional. The linear range will be defined by the highest and lowest measured concentrations where the response is linear. The manufacturing targets for each analyte is listed in Table 1 and instrument and reagent used to value assignment analyte is listed in Table 2.

AnalyteLevel ALevel BLevel CLevel DLevel E
AlbuminAs measuredAs measuredAs measuredAs measuredAs measured
Bicarbonate (mmol/L)10 ± 117.5 ± 1.7525 ± 2.532.5 ± 3.2540 ± 4
Bili, Total (mg/dL)As measured5 ± 210 ± 215 ± 220 ± 2
Bili, Conj. (mg/dL)As measured2.5 ± 0.255 ± 0.57.5 ± 0.7510 ± 1
Calcium (mg/dL)1.5 ± 0.25.2 ± 0.58.8 ± 0.912.4 ± 116 ± 1.6
Chloride (mmol/L)50 ± 582 ± 8.5113 ± 11.3144 ± 14.5175 ± 17.5
Cholesterol (mg/dL)2 ± 1125 ± 15250 ± 25375 ± 38500 ± 50
HDL-Chol (mg/dL)1 ± 137.5 ± 575 ± 8112 ± 15150 ± 15
Creatinine (mg/dL)0.2 ± 0.16.3 ± 0.712.5 ± 1.2518.8 ± 1.925 ± 2.5
Glucose (mg/dL)7 ± 1200 ± 20394 ± 40587 ± 59780 ± 78
Iron (µg/dL)As measured144 ± 15288 ± 30432 ± 45576 ± 58
Lactic Acid (mmol/L)0.5 ± 0.13.25 ± 0.46 ± 0.69 ± 0.912 ± 1.2
Magnesium (mg/dL)2 ± 0.22.8 ± 0.33.5 ± 0.44.25 ± 0.45 ± 0.5
OsmolalityAs measuredAs measuredAs measuredAs measuredAs measured
Phosphorus (mg/dL)1 ± 0.52.8 ± 0.34.5 ± 4.56.3 ± 0.78 ± 0.8
Potassium (mEq/dL)1 ± 0.54 ± 0.87 ± 0.910 ± 113 ± 1.3
Protein, Total (g/dL)2 ± 0.34.3 ± 0.56.5 ± 0.79 ± 0.911 ± 1.1
Sodium (mmol/L)80 ± 8103 ± 11125 ± 13148 ± 15170 ± 17
Triglycerides (mg/dL)1 ± 0.5220 ± 22439 ± 44658 ± 66876 ± 88
Uric Acid (mg/dL)0.5 ± 0.25.4 ± 0.710.3 ± 1.115.2 ± 1.620 ± 2
Urea Nitrogen (mg/dL)0.5 ± 0.232 ± 3.562.5 ± 6.594 ± 9.5125 ± 8
A. Phosphatase (IU/L)4 ± 227 ± 450 ± 673 ± 7.895 ± 9.5
ALT/GPT (IU/L)3 ± 1240 ± 24477 ± 48714 ± 72950 ± 95
ALP (IU/L)3 ± 2303 ± 31602 ± 61901 ± 911200 ± 120
AMY (IU/L)2 ± 1502 ± 501001 ± 1001501 ± 1502000 ± 200
CK (IU/L)3 ± 1503 ± 511002 ± 1011501 ± 1502000 ± 200
AST/GOT (IU/L)3 ± 1240 ± 24477 ± 48714 ± 72950 ± 95
x GT (IU/L)2 ± 1339 ± 34676 ± 681013 ± 1011350 ± 135
LDH (IU/L)2 ± 1239 ± 24476 ± 48713 ± 72950 ± 95
LIPASE (IU/L)2 ± 177 ± 8151 ± 15226 ± 23300 ± 30
LDL Chol (mg/dL)1 ± 0.5101 ± 11201 ± 21301 ± 31400 ± 40

{5}------------------------------------------------

*The manufacturing targets are based on the performance of the listed instrument and reagent that correspond to the listed analyte.

Value Assignment Protocol: To obtain the target values for each analyte listed in table 2, Aalto Scientific, Ltd. will perform Value Assignment according to Aalto's SOP Q-0051, titled "Value Assignment Protocol for Aalto Label Control". Protocol is to collect 30 values for each analyte listed on the instrument listed by the reagent listed. The SOP calls for three instruments or labs (if available). Ten individual assays from single evaluated vial (bottle). Require that one bottle be assayed in the morning and another in afternoon over five days time span. The thirty data points will be compiled and the average used as the target value.

Table 2: The Instrument and Reagent used for Value Assigning Analytes in Table 1.
Analyte*InstrumentReagents Vendor
pHOrionBeckman Coulter, Inc
Acid PhosphataseHitachi 911Roche Diagnostic Corp
AlbuminHitachi 911Roche Diagnostic Corp
ALPHitachi 911Thermo Trace, Ltd.
ALT/GPTHitachi 911Thermo Trace, Ltd.
AMYHitachi 911Thermo Trace, Ltd.
AST/GOTHitachi 911Thermo Trace, Ltd.
BicarbonateCobas bioThermo Trace, Ltd.
Bilirubin, ConjugateCobas bioHemagen Diagnostics, Inc. (Raichem)
Bilirubin, TotalHitachi 911Pointe Scientific, Inc.
CalciumHitachi 911Pointe Scientific, Inc.
ChlorideHitachi 911Roche Diagnostic Corp
CholesterolHitachi 911Thermo Trace, Ltd.
CKHitachi 911Pointe Scientific, Inc.
CreatinineHitachi 911Roche Diagnostic Corp
Gamma GTHitachi 911Roche Diagnostic Corp
GlucoseHitachi 911Roche Diagnostic Corp
HDL cholesterolHitachi 911Roche Diagnostic Corp
IronHitachi 911Roche Diagnostic Corp
Lactic AcidHitachi 911Pointe Scientific, Inc.
LDHHitachi 911Thermo Trace, Ltd.
LDL cholesterolHitachi 911Roche Diagnostic Corp
LipaseHitachi 911Roche Diagnostic Corp
MagnesiumCobas bioPointe Scientific, Inc.
OsmolalityWescorWescor
PhosphorusHitachi 911Roche Diagnostic Corp
PotassiumHitachi 911Roche Diagnostic Corp
Protein, TotalHitachi 911Pointe Scientific, Inc.
SodiumHitachi 911Roche Diagnostic Corp

{6}------------------------------------------------

TriglyceridesHitachi 911Thermo Trace, Ltd.
Urea NitrogenHitachi 911Roche Diagnostic Corp
Uric AcidHitachi 911Roche Diagnostic Corp
*Not all analytes tested are claimed. Aalto Scientific, Ltd. presented a complete list ofQA/QC testing. Aalto Scientific, Ltd. also performs Microbial growth, Moisture andTurbidity plus an abbreviated stressed shelf life and open bottle check on each lot.

Check Linearity: The target values will be plotted against X-axis value called level 1, 2, 3, 4, and 5 and then determine that the coefficient of correlation (R4) greater or equal to 0.975 and if so, declare that linear relationship of the five levels has been confirmed.

Traceability (controls, calibrators, or methods):

The reagents, calibrators and controls used were those purchased from the manufacturer. The instrumentations used were the Roche Diagnostic Corp Hitachi 911, Roche Diagnostic Corp Cobas Bio, Wescor 5500 Osmometer and Orion 720A pH meter. An independent control was included in each assay batch and is monitored under Aalto Scientific, Ltd. QA SOP protocols.

The stability characteristics of Audit™ MicroCV™ General Chemistry Stability: Linearity Set is determined using stress stability studies (stressed at 37°C) to estimate product storage stability at 2 to 8° C by linear regression and modified Arrhenus equations. The following assumption was made: if 90 to 110% of the concentration and/or activity of the listed constituent is retained when compared to a reference vial (zero day) of the same lot stored at 2 to 8° C and tested in the same a run that constituent is considered to be stable for the corresponding real time period at 2 to 8° C.

There are five levels in the General Chemistry Linearity Set. Level "A" though "E". The Level "A" concentration is set at the lowest concentration measurable by the assay and Level "E" is set at the highest concentration measurable. All other levels are merely dilutions of these two levels. After each of the five levels of Audit™ MicroCV™ General Chemistry Linearity Set was filled, freeze/dried and labeled, vials were placed at 2 to 8° C for storage. The stress stability study was a backward performed study were selected vials (Level A and Level E) where removed from storage at different time intervals and placed in a 37° C incubator. On the day of testing, all the vials were removed from the incubator and move vials removed from storage (zero day) and all vials were tested in the same assay with QC control material according to manufacturer instructions.

For the Open Bottle Study, "TEST" vials were reconstituted with the diluents and not re-frozen. The reconstituted vials were stored at 2 to 8 ° C. The product is removed each working day from the refrigerator and open to the

{7}------------------------------------------------

atmosphere from 10 to 60 minutes. On specified days, testing performed for all the analytes listed according to instrument and reagents specifications for their respective analyte.

The "Calculated Months to Failure" values were derived using linear regression analysis to estimate failure at less than 90% (-10%) or greater than 110% (+10%) of original recovery of each constituent for each level. Based on these failure estimates from the stress stability data, the predicted storage stability of the product of Unopened Vial at 2 to 8 ° C equals 12 years, and Open Vial at 2 to 8 ° C equals 24 hours.

Name/place of manufacturer, packer, or distributor

Manufacturer is Aalto Scientific, Ltd. 1959 Kellogg Ave Carlsbad, CA

92008

Distributor is Audit MicroControls, Inc. 3540 W. Sahara Ave., #086 Las Vegas, NV 89102-5833

Date of last labeling revision Date: N/A

Conclusion:

The submitted material in this premarket notification for Audit™ MicroCV™ General Chemistry Linearity Set supports a substantially equivalence decision based on the performance/stability and stress stability data. Real time stability data is be collected.

Attachments

Summary of Safety and Effectiveness in the format specified under 21 CFR 807.92 or 510(k) Statement in the content and format specified under 21 CFR 807.93 Truthful and Accurate Statement

Indications for Use Form

1 A discussion of the performance data that was generated in support of the device including protocols, raw data points, graphical representation, and analyses or conclusions for the following parameters as applicable.

4Provide a copy of the labeling in draft. Labeling should be conformance to the format and order of 21 CFR 809.10(b) (see below). An explanation should be provided for each element that is either absent or not applicable. 21 CFR § 809.10 Labeling for In Vitro Diagnostic Devices. 21 CFR § 809.10(b) - labeling accompanying each product (package insert) shall bear ... in the format and order specified below except where not applicable:

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Audit MicroControls. The logo consists of a pattern of black dots on the left and the words "AUDIT" and "MicroControls" on the right. The word "AUDIT" is in a larger font than "MicroControls" and is underlined.

Audit™ MicroCV™ General Chemistry Linearity Set

For In Vitro Diagnostic Use Only.

@Audit" MicroControls", Inc., Las Vegas, NV 89102, (866) 252-8348

INTENDED USE

Audit" MicroCV" General Chemistry Linearity Set consists of five levels of human based serum. Each level contains the following analytes: Acid Phosphatase, Albumin, Alkaline Phosphatase, ALT, Amylase, AST, Bilirubin (Total and Direct), BUN, Calcium, Chloride, Cholesterol, CO2 Creatine Kinase, Creatinine, Gamma-GT, Glucose, HDL Cholesterol, Iron, Lactate, LDH, LDL Cholesterol, Lipase, Magnesium, Phosphorus, Potassium, Sodium, Total Protein, Triglycerides and Uric Acid. These five levels demonstrate a linear relationship to each other for their respective analytes, reagents and instruments'

This product may also be used as unassayed quality control material for these analytes or as an assayed quality control material for the analyzer systems specified in this package insert. It is not intended to be used as an assayed quality control material for any other analyzer systems. When used for quality control purposes, it is recommended that each laboratory establish its own means and acceptable ranges and use the values provided only as guides. In addition, it may be used for proficiency testing in interlaboratory surveys and to perform CLIA directed calibration verification2 for these same analytes with similar reagents on similar instrumentation in accordance with current CLIA-88 guidelines and regulations

SUMMARY AND PRINCIPLE

As defined in the Clinical Laboratory Improvement Amendments of 1988 (CLIA) by the Centers for Medicare and Medicaid Services (CMS) and the Centers for Disease Control (CDC), each laboratory must revalidate each test method's analytical measurement range (AMR) at least every six months as well as following changes in lots of analytically critical reagents or major system components". Good laboratory practices require that stable reference materials be used to verify the accuracy and precision of testing methods and techniques. Audit" MicroCV" General Chemistry Linearity Set may be used as one would use human serum to verify and validate the AMR.

WARNINGS AND PRECAUTIONS

Because this product is of human origin, it should be handled as though capable of transmitting infectious diseases. Each serum, plasma or whole blood donor unit used in the preparation of this material was tested by United States Food and Drug Administration (FDA) approved methods and found to be negative for antibodies to HIV and HCV and nonreactive for HBSAg. Because no test method can offer complete assurance that HIV, hepatitis B virus, and hepatitis C virus or other infectious agents are absent, this material should be handled as though capable of transmitting infectious diseases. This product may also contain other human source material for which there is no approved test. The FDA recommends such samples be handled at the Centers for Disease Control's Biosafety Level 2.

This product contains less than 0.1% sodium azide that may react with lead and copper plumbing to form potentially explosive metal azides. On disposal, flush with a large volume of water to prevent azide build-up.

Audit" MicroCV" General Chemistry Linearity Set is intended solely for in vitro diagnostic use for the purpose described on the labeling. Audit" MicroControls, Inc. shall not be liable for any unclaimed damages arising from any other usage.

Cat. No. K701M-5

Contents 5 x 5 mL plus diluent

Lot No. XXXXX, XXXXX, XXXXX, XXXXX, XXXXX

Expires XX/XX/XX

STORAGE AND STABILITY

Audit" MicroCV" General Chemistry Linearity Set is stored at 2-8°C and will remain stable in the unopened vial for twelve months from the date of manufacture. After opening, the contents should be reconstituted immediately and used according to the instrument manufacturer's instructions.

It is recommended that Audit "MicroCV" General Chemistry Linearity Set be used within twenty-four (24) hours after reconstitution and stored tightly capped at 2-8°C. Leaving the vial uncapped, or prolonging its time at room temperature, will void this reconstituted stability claim. Make sure the contents of the vial are well mixed before use.

PROCEDURE

Follow the manufacturer's instructions provided for quality control and for verifying and validating the AMR. Verify that the lot number on each vial matches the package insert. To avoid evaporation, do not leave the vial uncapped. Q.C. requirements should be performed in conformance with local, state and/or federal regulations or accreditation requirements. CLIA directed calibration verification linearity material should be run4:

  • every six (6) months. 1.
  • when a complete change of reagents for a procedure is introduced
  • 3 when there is major preventive maintenance or replacement of critical parts that may influence test performance
    1. when control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits.
    1. when the laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification.

Materials provided

  • · Audit™ MicroCV™ General Chemistry Linearity Set, 5 x 5 mL
  • · Diluent, 5 x 5 ml.

INSTRUCTIONS FOR USE

    1. Remove a vial from the package, reconstitute with 5ml of the diluent provided, and gently swirl occasionally for 10 minutes. Do not shake. Do not mix mechanically
    1. Refer to instrument or assay instruction manual for quality control on verifying and validating the AMR
    1. After sampling, replace stopper and return to original package at 2-8°C to obtain the maximum 24-hour reconstituted stability

CALCULATIONS OF RESULTS

Each set of Audit™ MicroCV" General Chemistry Linearity Set is prepared in a manner such that an equal distance exists between each consecutive level. This dilution scheme is consistent with the NCCLS recommendation2 for preparing linearity sets.

{9}------------------------------------------------

LIMITATIONS OF THE PROCEDURE

Make sure that each vial is brought to room temperature before testing. If the contents of any of the vials become frozen, discard all vials and request a replacement set, as the results will not be valid. If the diluent becomes cloudy, do not use as bacterial contamination may be suspected.

2Federal Register 42 CFR Part 493, Department of Haath and Human Services, January 24, 2003; p.3691.

3Faderal Register 42 CFR Part 483, Department of Health and Human Services, January 24, 2003; 5483.2

"Federal Register 42 CFR Part 483, Department of Health and Human Services, January 24, 2003, 5483, 120 (1) (1) (1)

on guidelines provided by The National Committee for Clinical Laboratory Standards (NCCCS) in

ivaluation of the Linearity of Quantifative Heasurement Procedures: A Statistica approved guideline
Guideline", April 20

EXPECTED VALUES

Each lot of product is manufactured such that a linear relationship exists among levels. The analyte concentrations in this insert were derived from multiple replicate analyses on the instruments indicated. Actual results obtained may vary depending on instrumentation, methodology and assay temperature. Results may also be dependent on the accuracy of the instrument/reagent system calibration. The degree of acceptable non-linearity is an individual judgment based on methodology, clinical significance and medical decision levels of the test analyte.

The material and information presented here in no manner constitutes an overruling of any federal, state or other regulatory body's regulations and/or guidelines

ORDERING INFORMATION

ProductNumberProductDescriptionProductPackaging
K701M-5Audit MicroCV General Chemistry Linearity Set5 x 5 mL

Analytes include:

InstrumentUnitsmeanrangemeanrangemeanrangemeanrangemeanrange
AcidPhosphatase,TotalHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
AlbuminHitachi 911g/dLxx-xxx-xxx-xxx-xxx-x
AlkalinePhosphataseHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
ALTHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
AmylaseHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
ASTHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
Bilirubin, DirectHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Bilirubin,TotalHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
CalciumHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
CO2Hitachi 911mEq/Lxx-xxx-xxx-xxx-xxx-x
ChlorideHitachi 911mEq/Lxx-xxx-xxx-xxx-xxx-x
CholesterolHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Creatine KinaseHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
CreatinineHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Gamma-GTHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
GlucoseHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
HDL CholesterolHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Iron, TotalHitachi 911ug/Lxx-xxx-xxx-xxx-xxx-x
LactateHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
LDHHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
LDL CholesterolHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
LipaseHitachi 911IU/Lxx-xxx-xxx-xxx-xxx-x
MagnesiumHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
PhosphorusHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
PotassiumHitachi 911mEq/Lxx-xxx-xxx-xxx-xxx-x
SodiumHitachi 911mEq/Lxx-xxx-xxx-xxx-xxx-x
Total ProteinHitachi 911g/dLxx-xxx-xxx-xxx-xxx-x
TriglyceridesHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Urea (BUN)Hitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x
Uric AcidHitachi 911mg/dLxx-xxx-xxx-xxx-xxx-x

*AUDIT MicroControls, Inc. does not recommend using Level A for quality control purposes as the levels may be too low to measure

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Alan J. Vekich Scientist Aalto Scientific, Ltd. 1959 Kellogg Avenue Carlsbad, CA 92008

DEC - 9 2004

Re: K042318

Trade/Device Name: Audit™ MicroCVIM General Chemistry Linearity Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: November 22, 2004 Received: November 22, 2004

Dear Mr. Vekich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{11}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sincerely yours,

Cornelia B. Parks

Cornelia B. Rooks, MA Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

Indications for Use

510(k) Number (if known): K042318

Audit™ MicroCV™ General Chemistry Linearity Set Device Name:

Indications For Use:

The Audit™ MicroCV" General Chemistry Linearity Set consists of five levels of human based serum. Each level contains the following analytes: Acid Phosphatase, Albumin, Alkaline Phosphatase, ALT, Amylase, AST, Bilirubin (Total and Direct), BUN, Calcium, Chloride, Cholesterol, CO2 Creatine Kinase, Creatinine, Gamma-GT, Glucose, HDL Cholesterol, fron, LDH, LDL Cholesterol, Lactate, Lipase, Magnesium, Phosphorus, Potassium, Sodium, Total Protein, Triglycerides and Uric Acid. The five levels demonstrate a linear relationship to each other for their respective analytes, reagents and instruments.

This product may be used for proficiency testing in interlaboratory surveys and to perform CLIA directed calibration verification for these same analytes with similar instrumentation in accordance with current CLIA-88 guidelines and regulations.

In addition, levels B - E of this product may be used as unassayed quality control material for these analytes or as an assayed quality control material for the analyzer systems specified in the package insert. It is not intended to be used as an assayed quality control material for any other analyzer systems.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Bemeen
Division Sign-Off

of in Vince Digan

510441 K04 23/18

Page 1 of

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.