K Number
K070144

Validate with FDA (Live)

Date Cleared
2007-02-01

(16 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems.

The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay.

Device Description

The Access Inhibin A Calibrators & Controls are designed for use with the Access Inhibin A Reagent for generation of the Inhibin A assay calibration curve on Beckman Coulter's Access immunoassay Systems. The Access Inhibin A Calibrator kit contains 7 X 2.5 mL bottles, one for each of seven calibrator levels. The Access Inhibin A Control kit contains 3 X 2.5 mL bottles, one for each of three control levels. The Access Immunoassay Systems utilize a competitive binding immunoenzymatic method for quantitative analyte measurement.

AI/ML Overview

The provided K070144 510(k) summary for the Access Inhibin A Calibrators & Controls indicates that the device met its acceptance criteria through performance characteristics testing, demonstrating substantial equivalence to the predicate device. However, the specific acceptance criteria and detailed performance results that would allow for a table comparing expected vs. reported performance are not explicitly defined in the provided document. The 510(k) summary generally states that the device "meet product claims and specifications" and "supports a finding of substantial equivalence."

1. Table of Acceptance Criteria and Reported Device Performance

As noted, the specific quantitative acceptance criteria (e.g., precision limits, linearity ranges, recovery percentages) and the detailed results demonstrating performance against these criteria are not provided in the submitted document. The 510(k) summary only offers a high-level statement that the device met its performance characteristics and claims.

Acceptance CriteriaReported Device Performance
Specific criteria for calibration, control range, precision, stability, etc., are not detailed in the provided document.The document states, "Based on the results of the performance characteristics testing, these calibrator materials meet product claims and specifications." and "Performance data from validation testing supports a finding of substantial equivalence to the Diagnostic Systems Laboratories (DSL) ACTIVE® Inhibin A ELISA Calibrators and Controls already in commercial distribution." No specific numerical performance metrics are given.

2. Sample Size Used for the Test Set and Data Provenance

The 510(k) summary does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). The document states that "Performance data from validation testing" was conducted, but it does not elaborate on the details of this testing.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Given that this is an in vitro diagnostic (IVD) device (calibrators and controls), the "ground truth" is typically established through analytical methods, reference materials, and comparison to a predicate device, rather than expert interpretation of images or clinical data. Therefore, the concept of "experts" establishing a ground truth for a test set in the way it might be for an AI diagnostic device is not applicable here. The ground truth for the calibrators and controls would be based on their assigned values and their performance in an assay, which is analytically determined.

4. Adjudication Method for the Test Set

As this is an IVD device, adjudication methods (like 2+1, 3+1) used for human interpretation of clinical cases are not applicable. Device performance is assessed through analytical testing and comparison to reference standards or predicate devices.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this type of IVD device (calibrators and controls). MRMC studies are typically performed for diagnostic imaging or AI devices where human readers interpret results, and the study aims to assess the impact of AI on reader performance. This device does not involve human interpretation of cases in that manner.

6. Standalone Performance Study

Yes, implicit in the submission is that a standalone (algorithm only without human-in-the-loop performance) was done. The "performance characteristics testing" mentioned in the summary refers to analytical studies evaluating the calibrators and controls themselves, independent of human interpretation or assistance. These studies would assess aspects like:

  • Calibration curve generation: How well the calibrators establish the assay's dose-response curve.
  • Control material accuracy and precision: How accurately and consistently the controls measure expected values.
  • Stability: The shelf-life and in-use stability of the calibrators and controls.
  • Matrix effects: Performance in different sample matrices.
  • Method comparison: Comparison of the Access Inhibin A assay (using these calibrators/controls) to the predicate device's assay.

The document states, "The Access Inhibin A Calibrators have been demonstrated to be equivalent to the predicate product. Based on the results of the performance characteristics testing, these calibrator materials meet product claims and specifications." This confirms standalone testing was performed.

7. Type of Ground Truth Used

The ground truth for an IVD calibrator and control kit is based on analytical characterization and alignment to reference standards. Specifically, the document states: "Both are aligned to WHO international standard 91/624." This indicates that the values assigned to the calibrators and controls are traceable to an internationally recognized standard. The performance testing would then verify that the device accurately reflects these ground truth values.

8. Sample Size for the Training Set

The concept of a "training set" in the context of machine learning (AI) is not applicable to this traditional IVD device. Calibrators and controls are manufactured to specific specifications and then analytically validated. There is no algorithm "trained" in the sense of AI.

9. How the Ground Truth for the Training Set Was Established

Since there is no "training set" in the AI sense, this question is not applicable. The values for the calibrators and controls are established during their manufacturing process, quality control, and by alignment to recognized reference standards, not through an AI training process.

{0}------------------------------------------------

K070144

Access Inhibin A Calibrators & Controls 510(k) Summary

1.0 Submitted By:

Tyler Foutch Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 368-1653 FAX: (952) 368-7610

2.0 Date Submitted

January 15, 2007

3.0 Device Name(s):

  • 3.1 Proprietary Names
    Access® Inhibin A Calibrators Access® Inhibin A Controls

3.2 Classification Names

Calibrator (21 CFR § 862.1150) Control (21 CFR § 862.1660)

4.0 Legally Marketed Device

The Access Inhibin A Calibrators & Controls claim substantial equivalence to the Diagnostic Systems Laboratories (DSL) ACTIVE® Inhibin A ELISA Calibrators and Controls currently in commercial distribution, FDA 510(k) Number K002128.

5.0 Device Description

The Access Inhibin A Calibrators & Controls are designed for use with the Access Inhibin A Reagent for generation of the Inhibin A assay calibration curve on Beckman Coulter's Access immunoassay Systems. The Access Inhibin A Calibrator kit contains 7 X 2.5 mL bottles, one for each of seven calibrator levels. The Access Inhibin A Control kit contains 3 X 2.5 mL bottles, one for each of three control levels. The Access Immunoassay Systems utilize a competitive binding immunoenzymatic method for quantitative analyte measurement.

FEB - 】2007

{1}------------------------------------------------

6.0 Intended Use

The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems.

The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay.

7.0 Comparison to the Predicate

The Access Inhibin A Calibrators and the predicate calibrator utilize different test methodologies, automated competitive binding immunoenzymatic method for the Access and manual ELISA method for the DSL. The Access Inhibin A calibrators and controls are provided in Bovine Serum Albumen, while the predicate calibrator and control are provided in Fetal Bovine Serum. Both are aligned to WHO international standard 91/624. Both devices come in a liquid solution format. The Access kit contains three controls while the predicate contains two controls. The measured range differs between the two devices, the Access assay range is 0-1500 pg/mL, while the predicate range is 0-1000 pg/mL

8.0 Summary of Performance Data

The Access Inhibin A Calibrators have been demonstrated to be equivalent to the predicate product. Based on the results of the performance characteristics testing, these calibrator materials meet product claims and specifications.

Performance data from validation testing supports a finding of substantial equivalence to the Diagnostic Systems Laboratories (DSL) ACTIVE® Inhibin A ELISA Calibrators and Controls already in commercial distribution.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure, with three curved lines representing its wings or body. The logo is black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Tyler Foutch Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318

FEB - 1 2007

Re: K070144

Trade/Device Name: Access Inhibin A Calibrators and QC Regulation Number: 21 CFR § 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIT, JJX Dated: January 15, 2007 Received: January 18, 2007

Dear Mr. Foutch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please othe the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll the from the (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean m. Cooper, M.S., D.V.M.

Sean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K070144

Device Name:_Access Inhibin A Calibrators and QC

Indications For Use:

The Access Inhibin A calibrators are intended to calibrate the Access Inhibin A assay for the quantitative determination of dimeric Inhibin A levels in human serum and plasma using the Access Immunoassay Systems.

The Access Inhibin A QC is intended for monitoring system performance of the Access Inhibin A assay.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C Benson
Vision Sign-Off

Office of In Vitro Diagnostic Device Braluation and Safety

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

K070144

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.