(119 days)
The LDR Spine USA Easyspine System is a posterior, non-cervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis (with objective evidence of neurologic impairment), trauma (i. e., fracture or dislocation), spinal stenosis, deformities or curvatures (i. e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
Easyspine implants are single use devices for mono-and multi-segmental stabilization of the lumbar and thoracic vertebrae to promote fusion. The Easyspine System consists of sacral and pedicle screws, cross-connections, and rods of different rigidities. Specialized associated instrumentation is designed for implantation of these devices and for the distraction, compression or reduction of the lumbar and thoracic spine. The implants are made of surgical titanium alloy Ti6AI4V (ASTM F-136-92). Instruments used to implant Easyspine® are made of medical grade stainless steel. Both the Easyspine LP and Easyspine LP Multiaxial designs consist of various screws, rods, and connectors which are intended to provide temporary stabilization following surgery to fuse the thoraco-lumbar spine. The systems include side-loading screws which are offset as compared to the standard configuration. This offset allows separate rod tightening and polyaxiality locking. Additionally, the system facilitates conducting spondylolisthesis reduction by allowing bone screwing after rod tightening. The lateral offset loading feature incorporated into the designs has not altered the fundamental technology of the predicate Easyspine pedicle screw system. Operatively inserted implants serve to support the normal healing process. They are not intended to replace normal body structures, nor in cases of incomplete healing, to withstand the continually applied loading conditions.
The Easyspine® System, specifically the 6 mm and 7 mm diameter LP and LP Multiaxial pedicle screws, underwent non-clinical performance testing to demonstrate substantial equivalence to a predicate device (K043094 - LDR Spine USA Easyspine System).
Here's a breakdown of the acceptance criteria and the study details:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria | Reported Device Performance (Reference Standard: K043094) |
|---|---|
| Mechanical Performance | Tested in accordance with ASTM 1717-04. The new device showed performance comparable to the predicate device, indicating substantial equivalence. |
| Indications for Use | Same as predicate device. |
| Design | Posterior - pedicle screw/rod spine system, same as predicate device. |
| Sterility | Implants supplied sterile, instruments non-sterile, same as predicate device. |
| Implant Sterilization Methods | Gamma radiation, same as predicate device. |
| Implant Shelf Life | Five Years, same as predicate device. |
| Rod Diameter | 6 mm, same as predicate device. |
| Material | Titanium Alloy Ti6Al4V, same as predicate device. |
| Screw Sizes | 6 and 7 mm Diameters, same as predicate device. |
| Manufacturer | LDR Spine Medical, same as predicate device. |
| Product Code | MNI, MNH, same as predicate device. |
2. Sample Size Used for the Test Set and Data Provenance:
The study was a non-clinical performance study and involved mechanical testing in accordance with ASTM 1717-04. Therefore, it did not involve human patients or clinical data for a "test set" in the traditional sense of a clinical trial.
- Sample Size: Not explicitly stated in terms of a number of "samples" but implied to be sufficient for testing according to ASTM 1717-04. This would typically involve multiple test specimens of the screws and associated components to ensure statistical validity.
- Data Provenance: The data is from laboratory testing (mechanical performance testing) of the device. There is no country of origin for human data, as no human data was used.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
Not applicable. This was a non-clinical, mechanical performance study. "Ground truth" was established by the ASTM 1717-04 standard, which defines the testing methodology and performance parameters for spinal implant constructs. No human experts were required to establish ground truth for this type of testing.
4. Adjudication Method for the Test Set:
Not applicable. As a non-clinical mechanical performance study, there was no need for adjudication by multiple experts. The results were based on objective measurements against the specified ASTM standard.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:
Not applicable. This submission is for a medical device (pedicle screw system), not an AI-assisted diagnostic or imaging device. Therefore, no MRMC study or AI assistance was involved.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
Not applicable. This is a physical medical device, not an algorithm.
7. The Type of Ground Truth Used:
The ground truth used was mechanical performance data comparison against a legally marketed predicate device (K043094) and compliance with the ASTM 1717-04 standard. This standard serves as the established "ground truth" for evaluating the mechanical integrity and performance of spinal implant constructs.
8. The Sample Size for the Training Set:
Not applicable. As this is a non-clinical performance study for a physical device, there is no "training set" in the context of machine learning or AI models.
9. How the Ground Truth for the Training Set Was Established:
Not applicable. There was no training set involved in this non-clinical performance study.
{0}------------------------------------------------
K061017 (pg 1 x 3)
Image /page/0/Picture/1 description: The image shows the logo for LDR Spine. The letters "LDR" are stacked on top of the word "spine". The font is bold and sans-serif.
A Passion for Innovation
AUG 1 0 2006
510(k) Summary LDR Spine USA Easyspine® System
Lateral Polyaxial (LP) and LP Multiaxial Pedicle Screws 6 mm and 7 mm Diameters
Owner's Name & Address 1.
LDR Spine USA 4030 West Braker Lane, Suite 360 Austin, TX 78759
Phone: (512) 344-3333 (512) 344-3350 Fax:
2. Contact Person
James W. Burrows Director of Clinical Marketing LDR Spine USA 4030 West Braker Lane, Suite 360 Austin, TX 78759
Phone: (512) 344-3307 Fax: (512) 344-3350 Email: jamesburrows@ldrspine.com
Date 510(k) Summary Prepared: April 12, 2006 3.
- Easyspine® System 4. Trade Name: Pedicle Screw Spinal System Common Name: Pedicle Screw Spinal System (21 CFR §888.3070, Classification Name: Product Codes MNI and MNH)
Legally Marketed Equivalent Predicate Device 5. K043094 - LDR Spine USA Easyspine System consisting of rods, connectors, and Standard and a pedicle screws.
{1}------------------------------------------------
K06i017 (pg 2 of 3)
6. Device Description
Easyspine implants are single use devices for mono-and multi-segmental stabilization of the lumbar and thoracic vertebrae to promote fusion. The Easyspine System consists of sacral and pedicle screws, cross-connections, and rods of different rigidities. Specialized associated instrumentation is designed for implantation of these devices and for the distraction, compression or reduction of the lumbar and thoracic spine.
The implants are made of surgical titanium alloy Ti6AI4V (ASTM F-136-92). Instruments used to implant Easyspine® are made of medical grade stainless steel.
Both the Easyspine LP and Easyspine LP Multiaxial designs consist of various screws, rods, and connectors which are intended to provide temporary stabilization following surgery to fuse the thoraco-lumbar spine. The systems include side-loading screws which are offset as compared to the standard configuration. This offset allows separate rod tightening and polyaxiality locking. Additionally, the system facilitates conducting spondylolisthesis reduction by allowing bone screwing after rod tightening.
The lateral offset loading feature incorporated into the designs has not altered the fundamental technology of the predicate Easyspine pedicle screw system.
Operatively inserted implants serve to support the normal healing process. They are not intended to replace normal body structures, nor in cases of incomplete healing, to withstand the continually applied loading conditions.
7. Intended Use of the Device
The LDR Spine USA Easyspine System is a posterior, non-cervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis (with objective evidence of neurologic impairment), trauma (i. e., fracture or dislocation), spinal stenosis, deformities or curvatures (i. e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
{2}------------------------------------------------
Kcl017 (pg 3 of 3)
Summary of Technological Characteristics 8.
| Feature | Easyspine System w/ Ø6.0 mm, Ø7 mm LP & LP MultiaxialScrew | EasyspineSystem(K043094) | SubstantiallyEquivalent? |
|---|---|---|---|
| Indications forUse: | The LDR Easyspine® System is a posterior, non-cervical, pedicle systemintended to provide immobilization and stabilization of spinal segments inskeletally mature patients as an adjunct to fusion in the treatment of thefollowing acute and chronic instabilities or deformities of the thoracic,lumbar, and sacral spine: spondylolisthesis (grades 3 and 4) of the L5-S1vertebra, degenerative spondylolisthesis (with objective evidence ofneurologic impairment), trauma (i. e., fracture or dislocation), spinalstenosis, deformities or curvatures (i. e., scoliosis, kyphosis, and/orlordosis), tumor, pseudoarthrosis, and failed previous fusion. | Same | YES |
| Design: | Posterior - pedicle screw/rod spine system | Same | YES |
| Sterile: | Implants supplied sterileInstruments supplied non-sterile | Same | YES |
| ImplantSterilizationMethods: | Gamma radiation | Same | YES |
| Implant ShelfLife: | Five Years | Same | YES |
| Rod Diameter: | 6 mm | Same | YES |
| Material: | Titanium Alloy Ti6Al4V | Same | YES |
| Screw Sizes: | 6 and 7 mm Diameters | Same | YES |
| Manufacturer: | LDR Spine Medical | Same | YES |
| Product Code: | MNI, MNH | Same | YES |
Non-Clinical Performance Data 9.
Testing was conducted in accordance with ASTM 1717-04 Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model. Based upon the test data presented, risk analysis, and comparison of features, the LDR Spine 6 mm and 7 mm diameter Easyspine LP and LP Multiaxial screws are substantially equivalent to the LDR Spine 6mm and 7 mm diameter Easyspine Standard and Easyspine a pedicle screws that were cleared in premarket notification K043094.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, symbolizing the department's mission to protect and promote the health and well-being of Americans. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple and recognizable, representing the HHS's role in public health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 0 2006
LDR Spine USA % Mr. James W. Burrows Director of Clinical Marketing 4030 West Braker Lane, Suite 360 Austin, Texas 78759
Re: K061017 Trade Name: Easyspine® System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNI, MNH Dated: July 19, 2006 Received: July 20, 2006
Dear Mr. Burrows:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. James W. Burrows
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbery Buchum
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
KO61017 510(k) Number:
Easyspine® System Device Name(s):
Indications for Use:
The LDR Spine USA Easyspine System is a posterior, noncervical, pedicle system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: spondylolisthesis (grades 3 and 4 at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment), trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.
Prescription Use _ X
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Barbara Hutchins for MKM
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K061012
(Per 21 CFR 801.109)
(Optional format 1-2-96)
N/A