K Number
K063662

Validate with FDA (Live)

Date Cleared
2008-07-21

(591 days)

Product Code
Regulation Number
862.1117
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test indicated for use as an in vitro diagnostic product used to measure N-terminal pro-brain natriuretic peptide (NT-proBNP) levels in EDTA whole blood. Measurement of NT-proBNP aids in the diagnosis and assessment of severity in individuals suspected of having heart failure and may aid in the risk stratification of patients with heart failure.

Device Description

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test indicated for use as an in vitro diagnostic product used with a RAMP reader to measure N-terminal pro-brain natriuretic peptide (NT-proBNP) levels in EDTA whole blood. Mixed EDTA whole blood is added to the sample well of the Test Cartridge which houses the immunochromatographic test strip. The red blood cells are retained in the sample pad, and the separated plasma migrates along the strip. Fluorescent-dyed latex particles coated with anti-NT-proBNP antibodies bind to NT-proBNP, if present in the sample. As the sample migrates along the strip. NT-proBNP bound particles are captured at the detection zone, and additional particles are captured at the internal standard zone.

The RAMP reader then measures the amount of fluorescence emitted by the complexes captured at the detection zone and at the internal standard zone. Using a ratio between the two fluorescence values, a quantitative reading is calculated.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the RAMP® NT-proBNP Assay, based on the provided 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as distinct pass/fail thresholds against which the device performance is measured. Instead, the document presents detailed performance characteristics (precision, linearity, limits, analytical specificity, interference) and then compares the clinical performance (method comparison, sensitivity/specificity, ROC analysis) against a predicate device (Roche Elecsys proBNP Assay) and using established cut-offs.

However, based on the conclusions and the comparative data, we can infer some desired performance characteristics.

Performance MetricAcceptance Criteria (Inferred/Implicit)Reported Device Performance
Precision (Control Material)Within Run CV: Low concentrations < 10%; Mid-High concentrations < 7% (Inferred from predicate's expected performance and current device's results)NT-proBNP ng/L (140): Within Run CV 9.4%, Total CV 10.3%NT-proBNP ng/L (449): Within Run CV 6.4%, Total CV 9.8%NT-proBNP ng/L (1675): Within Run CV 5.5%, Total CV 8.9%
Precision (Whole Blood Samples)CV: < 20% for very low concentrations, decreasing to < 5% for high concentrations (Inferred from industry standards for quantitative assays)Mean NT-proBNP ng/L (52): CV 20.7%Mean NT-proBNP ng/L (73): CV 16.6%Mean NT-proBNP ng/L (113): CV 12.1%Mean NT-proBNP ng/L (131): CV 10.3%Mean NT-proBNP ng/L (161): CV 6.6%Mean NT-proBNP ng/L (299): CV 7.4%Mean NT-proBNP ng/L (2306): CV 4.5%Mean NT-proBNP ng/L (4051): CV 4.4%Mean NT-proBNP ng/L (5889): CV 4.3%Mean NT-proBNP ng/L (8445): CV 5.4%Mean NT-proBNP ng/L (19504): CV 3.0%
Precision (End User)Acceptable variability across operators (Inferred from standard QC practices)Level 1 CV (%): 9.3%, 9.5%, 15.5%, 8.8%, 11.8%, N/A; combined 11.0%Level 2 CV (%): 3.4%, 6.7%, 4.4%, 0.4%, 12.4%, 2.5%; combined 5.0%
LinearityR-value near 1.00; Slope near 1.00; Offset near 0; Average recovery near 100% (Standard linearity requirements)High Concentration: R-value 1.00, Slope 1.06, offset -1.4 ng/L. Average recovery 108% (range 101-120%).Low Concentration: R-value 1.00, Slope 1.06, offset -2.0 ng/L. Average recovery 100% (range 85-110%).
Hook EffectNo high dose hook effect observed within clinically relevant range (Critical for patient safety)No high dose hook effect up to 350,000 ng/L NT-proBNP.
Limits of Detection (LoD)Clinically acceptable lower detection limit34 ng/L
Limits of Blank (LoB)Low enough to distinguish from true blank27 ng/L
Limits of Quantitation (LoQ)LoQ (20% CV) below clinical decision thresholds (Ensuring reliable measurement at relevant concentrations)57 ng/L (for 20% CV)
Reportable RangeCovers clinically relevant range27 - 22,000 ng/L (levels > 22,000 ng/L reported as > 22,000 ng/L).
Analytical Specificity (Cross-Rxn)Minimal or no cross-reactivity with specified physiological compounds (Ensuring accurate measurement of target analyte)No cross-reactivity observed with listed compounds up to maximum levels tested (e.g., ANP28, BNP32, CNP22, Angiotensins).
Interference (Therapeutic Drugs)Average difference < 10% from unspiked samples for common therapeutic compounds (Ensuring drug compatibility)Average difference < 10% from unspiked samples for all listed therapeutic compounds (e.g., Acetaminophen, Furosemide, Isosorbide Dinitrate).
Interference (Endogenous Substances)No interference at very high physiological concentrations (Ensuring robustness in various patient conditions)No interference observed with Hemoglobin (2 g/dL), Triglyceride (4 g/dL), Bilirubin (35 mg/dL), Cholesterol (500 mg/dL), Heparin (104 IU/mL).
Method Comparison (vs. Elecsys)Correlation coefficient (R) > 0.95; Slope near 1.0; Intercept near 0 (Demonstrating strong agreement with a legally marketed predicate device)Slope 0.97 (95% CI: 0.95 to 1.00)Intercept 19.39 ng/L (95% CI: 14.20 to 24.67)Correlation coefficient (R) 0.98 (95% CI: 0.97 to 0.98)
Clinical Sensitivity (Age-Stratified)High sensitivity for CHF patients (>85%) (Ensuring accurate identification of disease presence)Age < 75: Sensitivity 0.89 (95% CI: 0.84-0.93)Age > 75: Sensitivity 0.99 (95% CI: 0.92-1.0)
Clinical Specificity (No Comorbidity)High specificity for non-CHF patients without comorbidities (>70-80%) (Ensuring accurate identification of disease absence in healthy individuals)Age < 75: Specificity 0.85 (95% CI: 0.80-0.88)Age > 75: Specificity 0.72 (95% CI: 0.53-0.87)
Clinical Specificity (With Comorbidity)Clinically acceptable specificity, acknowledging the impact of comorbidities (Understanding device performance in complex patient populations)Age < 75: Specificity 0.43 (95% CI: 0.43-0.52)Age > 75: Specificity 0.48 (95% CI: 0.35-0.60) (Note: The wide CI for >75 suggests variability or smaller sample size, but the values are reported)
ROC Analysis AUCAUC > 0.85 and comparable to predicate device (Demonstrating good discriminatory power)AUC 0.87 (for both RAMP and Elecsys assays)
Correlation with NYHA ClassificationIncreasing NT-proBNP levels with increasing NYHA class (Supporting use in severity assessment)Mean NT-proBNP (ng/L) for NYHA Class I: 1686; II: 2831; III: 5737; IV: 8308 (shows clear trend of increasing mean with disease severity). This supports the claim that the system provides NT-proBNP results that correlate with severity of heart failure.

Study Details

2. Sample sizes for the test set and data provenance

  • Method Comparison Test Set (RAMP vs. Elecsys):
    • Total Patients Enrolled: 699
    • Samples analyzed within reportable range (34 ng/L to 22,000 ng/L): 580
    • HF Diagnosed: 274 (164 males, 110 females)
    • Non-HF Reference Group: 306 (124 males, 182 females)
    • Data Provenance: Not explicitly stated regarding country of origin, but implied to be from clinical sites where patients were enrolled. The use of "individual hospital criteria" suggests multiple sites. The study is prospective in nature, as samples were collected "for each of these subjects" and then analyzed by both devices.
  • Clinical Sensitivity and Specificity Test Set:
    • Total Subjects: 858
    • Diagnosed with HF: 299 (using local hospital criteria)
    • Non-HF with potentially confounding co-morbidity: 189
    • Reference Individuals (healthy, no co-morbidities): 370 (includes an additional 159 subjects from an additional clinical site without concomitant testing in the Elecsys system).
    • Data Provenance: Implied to be from clinical sites. The collection of "data collected from 858 subjects" suggests a mix of retrospective and prospective, however the descriptions of "presenting population" and "enrolled" subjects points more towards a prospective data collection. The additional 159 subjects were "added from an additional clinical site".

3. Number of experts used to establish the ground truth for the test set and their qualifications

The ground truth for Heart Failure (HF) diagnosis was primarily established "based on individual hospital criteria" and "using local hospital criteria." This implies that the diagnosis was made by treating physicians at the respective clinical sites. The document does not specify the number of experts or their specific qualifications (e.g., "radiologist with 10 years of experience").

4. Adjudication method for the test set

The document does not specify an explicit adjudication method (e.g., 2+1, 3+1, none) for the clinical diagnosis of heart failure. The ground truth ("local hospital criteria") suggests that hospital-based clinical judgment was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This device is a quantitative immunochromatographic test for measuring NT-proBNP levels, not an AI-assisted diagnostic imaging device or a decision support system for human readers. The comparison is between the RAMP device and a predicate laboratory device (Roche Elecsys proBNP Assay) and the device's performance metrics (sensitivity, specificity) in diagnosing HF.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, a standalone performance study was clearly done. The entire "Summary of Studies" section details the performance characteristics of the RAMP NT-proBNP Assay when used without human intervention in interpreting the raw data. The RAMP reader provides a quantitative result (ng/L) of NT-proBNP levels. The clinical efficacy (sensitivity, specificity) is calculated based on these quantitative results against a clinical diagnosis (ground truth), demonstrating the algorithm's standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The primary ground truth used for heart failure diagnosis was clinical diagnosis based on individual hospital criteria. This would typically involve a combination of patient history, physical examination, imaging (e.g., echocardiogram), and other laboratory tests as judged by the treating clinicians. The document does not mention pathology or long-term outcomes data being used as the primary ground truth for the diagnostic studies.

8. The sample size for the training set

The document does not explicitly describe a separate "training set" for the RAMP NT-proBNP Assay algorithm in the context of machine learning. For chemical assays like this, the "training" (or development and optimization) of the assay itself is done during product development using various known concentrations and interfering substances. The performance characteristics studies (precision, linearity, LoD, LoQ, analytical specificity, interference) could be considered part of the internal validation/optimization process which might indirectly involve a "training set" of samples, but these are presented as characterization studies rather than specific training data for a machine learning model. The clinical evaluation samples (699 for method comparison, 858 for sensitivity/specificity) serve as the test sets for demonstrating performance.

9. How the ground truth for the training set was established

As there is no explicitly defined "training set" in the machine learning sense, the method for establishing its ground truth is not applicable/not provided. The ground truth for the performance characterization studies (e.g., linearity, LoD) would be established by preparing samples with known, precise concentrations of NT-proBNP (spiked blood or control material where the true concentration is known).

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: _ K063662 .

1. Establishment:

Response Biomedical Corporation 1781 – 75" Avenue W.
Vancouver, British Columbia Canada, V6P 6P2

Tel: (604) 456-6010 Fax: (604) 456-6066

Contact: Ken Pilgrim Director - Quality / Regulatory

Prepared: July 2, 2008

2. Regulatory Information:

Trade Name:Response Biomedical Corporation RAMP® NT-proBNPAssay
Common Name:NT-proBNP immunological test system
Classification Name:NT-proBNP immunological test system
Regulation Number:862.1117, B-type Natriuretic Peptide Test System
Classification:Class II
Product Code:NBC
Panel:Clinical Chemistry (75)

3. Predicate Device:

Immunoassay:Elecsys® proBNP, (K022516, K032646, K051382) which iscurrently marketed by Roche Diagnostics GmbH.
Immunoassay:StatusFirst™ CHF NT-proBNP (K051596), which is currentlymarketed by Nanogen, Inc.

{1}------------------------------------------------

4. Description of the Device:

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test indicated for use as an in vitro diagnostic product used with a RAMP reader to measure N-terminal pro-brain natriuretic peptide (NT-proBNP) levels in EDTA whole blood. Mixed EDTA whole blood is added to the sample well of the Test Cartridge which houses the immunochromatographic test strip. The red blood cells are retained in the sample pad, and the separated plasma migrates along the strip. Fluorescent-dyed latex particles coated with anti-NT-proBNP antibodies bind to NT-proBNP, if present in the sample. As the sample migrates along the strip. NT-proBNP bound particles are captured at the detection zone, and additional particles are captured at the internal standard zone.

The RAMP reader then measures the amount of fluorescence emitted by the complexes captured at the detection zone and at the internal standard zone. Using a ratio between the two fluorescence values, a quantitative reading is calculated.

5. Comparison of Technological Characteristics:

The RAMP NT-proBNP Assay, Elecsys proBNP Assay and StatusFirst CHF NT-proBNP Test are all used for the quantitative measurement of NT-proBNP in human whole blood (RAMP) or plasma (Elecsys and StatusFirst). All three immunoassays utilize the binding of NT-proBNP to specific antibodies. The RAMP assay measures light production from a fluorescence reaction using a fluorometer, which is directly proportional to the amount of NT-proBNP present in a patient sample. The Elecsys assay is an electrochemiluminescence immunoassay utilizing the application of voltage to an electrode inducing an emission measured by a photomultiplier, which is directly proportional to the amount of NT-proBNP present in a patient sample. The StatusFirst test utilizes a biotin coupled antibody/streptavidin solid-phase chromatographic immunoassay in which the analyte concentration in the sample correlates with the intensity of the test band.

The RAMP and StatusFirst assays are both quantitative immunochromatographic tests, whereas the Elecsys proBNP Assay is a quantitative sandwich immunoassay.

The Elecsys proBNP Assay utilizes two sheep polyclonal antibodies that recognize epitopes located in the N-terminal part (1-76) of proBNP (1-108). The RAMP NT-proBNP Assay utilizes one sheep polyclonal and one mouse monoclonal antibody that recognize epitopes located in the N-terminal part (1-76) of proBNP (1-108). The StatusFirst test utilizes polyclonal and monoclonal antibodies to the N-terminal part (1-76) of proBNP (1-108).

The RAMP NT-proBNP Assay is for use in the central laboratory, stat-lab and point-ofcare facilities, while the Elecsys proBNP Assay is for use in the central and stat laboratories. The intended location for use of the StatusFirst is not indicated in the device labeling.

All three assays are indicated for use in the diagnosis and assessment of severity in individuals suspected of having congestive heart failure. All three assays may also aid in risk stratification.

A comparison between the RAMP, the Roche Elecsys and Nanogen StatusFirst assays is presented in the table below.

{2}------------------------------------------------

Substantial Equivalence Comparison Table

Features/ TechnicalInformationResponse BiomedicalCorporationRAMP NT-proBNP AssayRoche DiagnosticsElecsys proBNP AssayNanogen, Inc.StatusFirst NT-proBNP Test
Indication for UseThe RAMP NT-proBNP Assay isa quantitativeimmunochromatographic testindicated for use as an in vitrodiagnostic product used tomeasure N-terminal pro-brainnatriuretic peptide (NT-proBNP)levels in EDTA whole blood.Measurement of NT-proBNP aidsin the diagnosis and assessment ofseverity in individuals suspectedof having congestive heart failureand may aid in the riskstratification of patients with heartfailure.Elecsys proBNP is used as an aid inthe diagnosis of individuals suspectedof having congestive heart failure.The test is further indicated for therisk stratification of patients withacute coronary syndrome andcongestive heart failure. The test mayalso serve as an aid in the assessmentof increased risk of cardiovascularevents and mortality in patients at riskfor heart failure who have stablecoronary disease.The device is intended for use with theDXpress™ Reader to providequantitative results as an aid in thediagnosis of CHF.
Quantitative/QualitativeQuantitativeQuantitativeQuantitative
Test PrincipleImmunochromatographicfluorescence immunoassayElectrochemiluminescentimmunoassayBiotin coupled antibody/streptavidinsolid-phase chromatographicimmunoassay
TraceabilityStandardizationReference standard - purifiedsynthetic NT-proBNP (1-76) inhuman serum matrixReference standard - purifiedsynthetic NT-proBNP (1-76) inhuman serum matrixUnknown
Antibodies UsedOne sheep polyclonal and onemouse monoclonal antibody,recognizing epitopes located inthe N-terminal part (1-76) ofproBNP (1-108).Two sheep polyclonal antibodies,recognizing epitopes located in the N-terminal part (1-76) of proBNP (1-108).Polyclonal and monoclonal antibodies tothe N-terminal part (1-76) of proBNP (1-108) of unknown animal origin.
Site of UseCentral laboratory, stat-lab andpoint-of-care facilitiesCentral laboratory and stat-labUnknown
Specimen TypeWhole blood (EDTA)Serum and Plasma (lithium andsodium heparin, EDTA)Plasma (EDTA)
Reported Range27 - 22,000 ng/L5 - 35,000 ng/L20 - 5,000 ng/L
Test Time15 minutes after Test Componentscome to room temperature18 minutes15 minutes after Test Device comes toroom temperature
InstrumentRAMP readersElecsys 1010, Elecsys 2010 andMODULAR analytics E 170 family ofanalyzersDXpress Reader
Assay Reagent StorageRefrigerated (2 to 8°C); must bebrought to room temperature priorto running assay.Refrigerated (2 to 8°C) for 12 weeks.Refrigerated (2 to 8°C); must be broughtto room temperature prior to runningassay.
Stability at RoomTemperature14 days unopened; but not beyondthe expiration date.On E170 & 2010 for 8 weeks, on1010 for 4 weeks (at 20-25°C up to 20hours opened in total).14 days unopened; but not beyond theexpiration date.
Standard CurveLot specific; provided on LotCard.Generate with each reagent lot.Lot specific; provided on Data Chip.
Features/ TechnicalInformationResponse BiomedicalCorporationRAMP NT-proBNP AssayRoche DiagnosticsElecsys proBNP AssayNanogen, Inc.StatusFirst NT-proBNP Test
Quality ControlsProvided in every Test Cartridge.Built in performance controls forroutine QC requirements.Indicates sufficient sample wasapplied and the cartridge wasinserted and read properly by theinstrument. Antibody quality,system function and assay timingare checked on cach assay run.An unacceptable result from thecontrol displays an error messageon the instrument.Recommended at least once each dayof useProvided in every Test Device. Thecontrol line is an internal positiveprocedural control. A distinct reddish-purple control line should appear at thecontrol position if the test is performedproperly, an adequate sample volume isused, the sample and reagent are wickingon the membrane, and the reagents at thecontrol line are reacting with theconjugate-color indicator. In addition, aclear background may be considered anegative procedural control. TheDxpress Reader will report "Control:Valid" and the test results for NT-proBNP (pBNP) when Internal ControlQC is satisfied.
Liquid ControlAudit MicroFD BNP ControlElecsys PreciControl proBNPElecsys PreciControl CardiacCommercially available NT-proBNPControls
Result Interpretation125 pg/mL for patients 75 yearsand younger and 450 pg/mL forpatients older than 75 years.125 pg/mL for patients younger than75 years and 450 pg/mL for patients75 years and older.Decision thresholds: Patients under 75years of age: 125 pg/mL, Patients 75years of age and older: 450 pg/mL.NT-proBNP results less than or equal tothe decision threshold values areconsidered normal values representativeof patients without CHF.Results greater than the above stateddecision threshold values are consideredabnormal and suggestive of patients withCHF

{3}------------------------------------------------

6. Summary of Studies:

PERFORMANCE CHARACTERISTICS

Precision

The intra-assay and inter-assay precision of the RAMP NT-proBNP Assay were determined by one operator assaying duplicates of three concentrations of control material (140, 449 and 1675 ng/L NT-proBNP) twice each day over a 10-day period. The mean, standard deviation and %CV were calculated for each reported concentration of NT-proBNP. The results of this precision analysis are shown below.

PrecisionNT-proBNP Standards
NT-proBNP ng/L1404491675
Within Run CV9.4%6.4%5.5%
Total CV10.3%9.8%8.9%

EDTA anticoagulated whole blood samples spanning the reportable range of the RAMP NTproBNP Assay were analyzed. Ten replicate measurements were carried out by a single operator in one day. The results of twelve samples from this precision analysis are shown below.

PrecisionEDTA Whole Blood Samples
Mean NT-proBNP ng/L5273113131161299230640515889844519504
CV (%)20.716.612.110.36.67.44.54.44.35.43.0

{4}------------------------------------------------

Precision in the hands of the end user was evaluated at each clinical site (N=4). Each operator performed three (3) replicates of a plasma based control product. The results are presented for each operator and across operators below:

Operator11. 11. 11. 11. 11. 11.combinea
Level 1 CV (%)9.3%15.5%------------------------------------------------------------------------------------------------------------------------------------------------------------------------------8.8%11.8%1
Level 2 CV (%)4%10/04%0 10/12.4%2.5%Comments of the couple of a cases(ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮ

Linearitv

A high NT-proBNP antigen concentration was prepared in normal donor EDTA blood and determined to contain 21,921 ng/L NT-proBNP by assaying the sample in duplicate. The sample was serially diluted six times. Regression analysis using the method of Passing-Bablok of actual versus expected NT-proBNP concentration resulted with an R-value of 1.00, a slope of 1.06 and an offset of -1.4 ng/L. The recovery of NT-proBNP antigen at the six dilutions ranged from 101 to 120% with an average of 108%.

A low NT-proBNP antigen concentration was prepared in normal donor EDTA blood and determined to contain 264 ng/L NT-proBNP by assaying the sample in duplicate. The sample was serially diluted four times. Regression analysis using the method of Passing-Bablok of actual versus expected NT-proBNP concentration resulted with an R-value of 1.00, a slope of 1.06 and an offset of -2.0 ng/L. The recovery of NT-proBNP antigen at the four dilutions ranged from 85 to 110% with an average of 100%

Hook Effect

There is no high dose hook effect in the RAMP NT-proBNP Assay up to the highest level tested (350,000 ng/L NT-proBNP).

Limits of Detection and Quantitation

Following CLSI EP-17A, limit of detection (LoD) was determined to be 34 ng/L, the limit of blank (LoB) was calculated as the 95" percentile from forty replicates of a blank sample run using the RAMP NT-proBNP Assay and was determined to be 27 ng/L ..

NT-proBNP levels in excess of 22,000 ng/L are reported as greater than (>) 22,000 ng/L.

The limit of quantitation (LoQ) is defined as the NT-proBNP level at which the test method displays a particular coefficient of variation (%CV). As shown below, the 20% LoQ for the RAMP NT-proBNP Assay was determined from whole blood analyses to be 57 ng/L.

Image /page/4/Figure/11 description: The image is a scatter plot titled "RAMP NT-proBNP Low Range Precision Profile EDTA Whole Blood". The x-axis is labeled "[NT-proBNP] (ng/L)" and ranges from 0 to 300. The y-axis is labeled "Precision (%CV)" and ranges from 0 to 40. The data points are scattered, showing a general trend of decreasing precision as NT-proBNP concentration increases.

{5}------------------------------------------------

Analytical Specificity

Human anti-mouse antibodies (HAMA) and Rheumatoid Factor (RhF) appear to have minimal cross-reactivity with the RAMP NT-proBNP Assay. Possible cross-reactivity of other substances was evaluated by spiking different concentrations of the potential cross-reactants into EDTA blood which had NT-proBNP added. No cross reactivity was observed with the RAMP NTproBNP assay up to the maximum levels tested for the compounds listed in the following table.

CompoundConcentrationCompoundConcentrationCompoundConcentration
ANP283.1 µg/mLpreproANP 104-1231 ng/mLEndothelin20 ng/L
BNP323.5 µg/mLAldosterone0.6 ng/mLArg-Vasopressin1 ng/mL
CNP222.2 µg/mLAngiotensin I0.6 ng/mLRenin50 ng/mL
preproANP26-553.5 µg/mLAngiotensin II0.6 ng/mLAndrenomedullin1 ng/mL
preproANP56-921 ng/mLAngiotensin III1 ng/mLUrodilatin3.5 µg/mL

Interference

Potentially interfering substances were evaluated by spiking different concentrations of potential interferents into EDTA whole blood with NT-proBNP added to provide levels of 125 and 450 ng/L. Different blood samples were used for each potential interferent with an average difference of less than 10% from the unspiked samples observed in each case. The therapeutic compounds tested (at concentrations up to and including those indicated) are listed in the following table:

CompoundConcentrationCompoundConcentration
Acetaminophen20 mg/dLFurosemide6 mg/dL
Acetylsalicylic acid100 mg/dLHydralazine20 µg/mL
Allopurinol4 mg/dLHydrochlorothiazide20 µg/mL
Amiodarone20 µg/mLIndomethacin40 µg/mL
Amiodipine Besylate4 µg/mLInsulin120 µU/mL
Ampicillin sodium salt100 mg/dLIsosorbide Dinitrate15 mg/dL
Ascorbic acid30 mg/dLLisinopril4 mg/dL
Atenolol1 mg/dLMethyldopa2.5 mg/dL
Caffeine10 mg/dLMetoprolol Tartrate2 mg/dL
Captopril15 mg/dLNicotine2 mg/dL
Carvedilol5 mg/dLNifedipine6 mg/dL
Chloramphenicol25 mg/dLNitroglycerin19.2 mg/dL
Clopidogrel Hydrogensulfate7.5 mg/dLOxytetracycline100 µg/mL
Cyclosporin A0.5 mg/dLProbenecid600 µg/mL
Diclofenac60 µg/mLPropranolol0.2 mg/dL
Digitoxin0.03 mg/dLQuinidine20 µg/mL
Digoxin0.05 mg/dLSimvastatin4 mg/dL
Diltiazem120 µg/mLTheophylline100 mg/dL
Phenytoin10 mg/dLTrimethoprim60 µg/mL
Dipyridamole30 µg/mLVerapamil16 mg/dL
Enalapril Maleate4 mg/dLWarfarin20 µg/mL
Erythromycin20 mg/dL

Hemoglobin, triglyceride, bilirubin, cholesterol, and heparin at levels of very high physiological concentrations were also investigated for possible interference. No interference was observed when tested at the concentrations up to and including those shown in the following table

CompoundConcentrationCompoundConcentration
Hemoglobin2 g/dLCholesterol500 mg/dL
Triglyceride4 g/dLHeparin104 IU/mL
Bilirubin35 mg/dL

{6}------------------------------------------------

CLINICAL EVALUATIONS OF ANALYTICAL PERFORMANCE

Method Comparison

Six hundred and ninety-nine (699) patients were enrolled in the method comparison study. The presenting population included 46% (323) subjects with hypertension, 30% (208) with shortness of breath, 22% (152) with diabetes, 14% (99) with pulmonary disorders, 12% (84) with coronary disease, 8% (56) with atrial fibrillation, 4% (31) with renal failure, 19% (133) were healthy, and the remainder had diagnoses not believed to be cardiac related (hepatitis, HIV, cancer, etc.).

EDTA and heparin whole blood samples were obtained for each of these subjects. An aliquot of EDTA whole blood was used for the RAMP NT-proBNP Assay and heparinized plasma was prepared for the Roche Elecsys proBNP Assay. From these analyses it was determined that 580 samples contained between 34 ng/L (RAMP LoD) and 22,000 ng/L of NT-proBNP. Of these, 274 were diagnosed with heart failure (HF) based on individual hospital criteria (164 males and 110 females) and 306 were non-HF reference group patients (124 males and 182 females). Regression analysis data of RAMP NT-proBNP versus Elecsys proBNP using the method of Passing-Bablok is presented in the table below.

Comparative MethodSlopeIntercept (ng/L)Correlation coefficient (R)
Roche Elecsys0.9719.390.98
95% CI0.95 to 1.0014.20 to 24.670.97 to 0.98

Clinical Sensitivity and Specificity

Clinical sensitivity and specificity were calculated using data collected from 858 subjects. Of these, 299 were diagnosed with HF using local hospital criteria, 189 individuals without HF but with potentially confounding co-morbidity (diabetes, renal insufficiency, hypertension or chronic obstructive pulmonary disease) and 370 reference individuals. This reference group includes an additional 159 subjects added from an additional clinical site without concomitant testing in the Elecsys system. Of these, 55% (87) were male and 8% (12) were more than 75 years old. None of these patients had reported co-morbidities. These subjects were healthy individuals with no clinical indications for natriuretic peptide testing.

The use of the cut-offs of 125 ng/L for ≤75 years of age and 450 ng/L for >75 years of age was evaluated for the RAMP data (N = 858; 299 with CHF) stratifying by the presence or absence of co-morbidities (diabetes, renal insufficiency, hypertension or chronic obstructive pulmonary disease).

CHF PatientsRAMPRAMP
Age (years)< 75> 75
N21782
Sensitivity0.890.99
95% CI(0.84-0.93)(0.92-1.0)
Non-CHFno comorbidityRAMPRAMP
Age (years)< 75> 75
N34030
Specificity0.850.72
95% CI(0.80-0.88)(0.53-0.87)

Age Stratified Sensitivity and Specificity: 125/450 ng/L by Age

{7}------------------------------------------------

Non-CHF with comorbidityRAMP
Age (years)< 75> 75
N12465
Specificity0.430.48
95% CI(0.43-0.52)(0.35-0.60)

The Receiver Operator Characteristics (ROC)

The ROC analyses for both the RAMP NT-proBNP and Roche Elecsys proBNP assays for the parallel clinical study population are shown below. The additional 159 non-CHF RAMP patients are excluded from this analysis to allow direct comparison to the Elecsys system. The area under the curve (AUC) for both the RAMP NT-proBNP Assay and Elecsys proBNP assay is 0.87.

Image /page/7/Figure/3 description: The image is a plot of sensitivity versus 1-specificity. There are three curves plotted on the graph. The first curve represents 'No discrimination', the second curve represents 'RAMP NT-proBNP result (ng/L)', and the third curve represents 'Elecsys NT-proBNP result (ng/L)'. The RAMP and Elecsys curves are very similar, and both are much higher than the 'No discrimination' curve.

HF Population by NYHA Classification

The 299 subjects diagnosed with heart failure were evaluated using the RAMP NT-proBNP Assay. The descriptive statistics for NT-proBNP concentrations (ng/L) are presented according to NYHA Functional Classification in the table below.

All
NYHA ClassIIIIIIIV
n58918466
Mean1686283157378308
SD3161435659397090
Median832147936086628
95th percentile5560810420177>22000
Male
n32565540
Mean1737287057998855
SD3924464161827612
Median724131836235772
95th percentile47221074221068>22000

{8}------------------------------------------------

Female
n26352926
Mean1624277156187466
SD1918392155516251
Median907162235986937
95th percentile543873061672721839

Non-HF and HF Group Descriptive Statistics

The overall incidence of disease in the presenting population (n=858) included 38% (323) subjects with hypertension, 24% (208) who presented with shortness of breath, 18% (152) with diabetes, 12% (99) with pulmonary disorders, 10% (84) with coronary disease, 7% (56) with atrial fibrillation, 4% (31) with renal failure, 34% (292) were healthy, and the remainder had diagnoses not believed to be cardiac related (hepatitis, HIV, cancer, etc.).

The circulating NT-proBNP concentration was determined in 858 individuals with and without HF. The HF patients included those with prior established heart failure that were not acutely destabilized at the time of enrollment (and thus similar to those who might be tested in the outpatient setting). Patients (N=17) for whom the measured NT-proBNP level was greater than 22,000 ng/L have been excluded. The descriptive statistics for the Non-HF and HF groups are presented in the following table:

No co-morbidityWith co-morbidity
Age (years)>75<75ALLAge (years)>75<75ALL
n30340370n65124558
Mean449.7132.8158.5Mean1013.0870.5349.4
SD810.9671.2687.8SD1524.63445.1996.0
Median88.024.528.0Median512.0185.9566.0
95th percentile2447.4216.4451.095th percentile3986.02463.21543.9
% <125 ng/L84% <125 ng/L44
% < 450 ng/L74% < 450 ng/L48

A. Non-HF Patients -RAMP Results (ng/L)

B. HF Patients - RAMP Results (ng/L)

Age (years)>75<75ALL
N80203283
Mean4970.13133.13652.4
SD5185.83755.04280.2
Median3300.51735.02040.0
95th percentile19005.011373.312800.0
% >125 ng/L89
% >450 ng/L100

7. Conclusion:

The RAMP NT-proBNP device demonstrates performance equivalent to that stated in the Nanogen StatusFirst CHF labeling and strong correlation to the predicate Roche Diagnostics Elecsys® proBNP method.

These data show that:

{9}------------------------------------------------

  • The RAMP NT-proBNP Assay has excellent sensitivity for less acute populations as . might be seen in an outpatient setting when the 125/ 450 age dependent cut-offs presented in cleared NT-proBNP assay labeling is applied.
  • The RAMP NT-proBNP Assay shows equivalent clinical sensitivity and specificity to that . stated in the predicate Nanogen StatusFirst CHF NT-proBNP labeling using these cutoffs.
  • The RAMP NT-proBNP Assay has excellent correlation to the predicate Roche . Diagnostics Elecsys proBNP.
  • The RAMP system provides NT-proBNP results that correlate with severity of heart . failure, with significantly higher values of NT-proBNP as patients from NYHA Class I through IV are examined.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for "RAMP" with a registered trademark symbol. Below the logo, the text "NT-proBNP" is displayed in a bold font. The text is black and the background is white.

Failure to follow RAMP NT-proBNP Assay procedures may result in invalid and/or erroneous results. Read the entire Package Insert prior to use.

NAME RAMP® NT-proBNP Assay

INTENDED USE

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test indicated for use as an in vitro diagnostic product used to measure N-terminal pro-brain natriuretic peptide (NT-proBNP) levels in EDTA whole blood. Measurement of NT-proBNP aids in the diagnosis and assessment of severity in individuals suspected of having heart failure and may aid in the risk stratification of patients with heart failure.

Image /page/10/Picture/5 description: The image contains the letters 'IVD' in a simple, sans-serif font. The letters are black and are enclosed in a thin, black rectangular border. The background is white, providing a clear contrast to the text and border.

For in vitro Diagnostic Use Only

Image /page/10/Picture/7 description: The image shows the logo and contact information for Response Biomedical Corporation. The address is 1781-75th Avenue W., Vancouver, British Columbia, Canada V6P 6P2. The telephone number is 001-604-456-6010, and the fax number is 001-604-456-6066. The email address is techsupport@responsebio.com, and the website is www.responsebio.com.

Printed in Canada Response Biomedical Corporation

CE

Page 1 of 11

{11}------------------------------------------------

BACKGROUND AND INTRODUCTION

Heart failure (HF) is a chronic, progressive disease in which the heart muscle weakens and its function becomes impaired, thus impeding the heart's ability to pump enough blood to support the body's metabolic demands. When cardiac muscle is stretched, such as with elevated ventricular filling pressure, the inactive prohormone B-type natriuretic peptide (proBNP) is released and rapidly cleaved into physiologically active BNP and the N-terminal fragment NT-proBNP. Natriuretic peptides can be used for the diagnosis of clinical problems associated with left ventricular dysfunction. The advent of testing for BNPs has improved the ability of physicians to make a qualified diagnosis of heart failure and to monitor the success of treatment.3 Being able to test the levels of NT-proBNP in patient blood samples is very useful as these levels are indicative of the degree of HF, and when combined with clinical judgment they provide superior diagnostic performance than clinical judgment alone. NTcinneal judghen they provide superior ulagitostic perroritance from and IFF. " "Time and IF"." "The " has also been shown to aid in the assessment of increased risk of cardiovascular events and mortality in patients at risk for HF who have stable coronary artery disease. 910,1

Point-of-care (POC) or "near-patient" testing allows for diagnostic assays to be performed at the site of patient care delivery such as the emergency room (ER), chest pain evaluation center, or intensive care unit (ICU). Compared with centralized laboratory testing provides for rapid clinical decision making by reducing the time spent ordering tests and transporting samples, as well as retrieving data.

PRINCIPLES OF THE TEST

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test for the determination of NT-proBNP in EDTA whole blood. EDTA whole blood is mixed with buffer and antibody-coated, labeled particles, and applied into the sample well of the Test Cartridge. The red blood cells are retained in the sample pad, and the separated plasma migrates along the strip. Fluorescent-dyed particles coated with anti-NT-proBNP antibodies bind to NT-proBNP, if present in the sample. As the sample migrates along the strip, NT-proBNP bound particles are captured at the detection zone, and excessive fluorescent-dyed particles are captured at the internal standard zone.

The RAMP reader then measures the amount of fluorescence emitted by the complexes bound at the detection zone and at the internal standard zone. Using a ratio between the two fluorescence values, a quantitative reading is calculated. For further information on the use of the RAMP reader refer to the Operator's Manual.

EQUIPMENT, REAGENTS, MATERIALS

Materials ProvidedAmountEquipment and Material required, but not providedEquipment recommended but not required
Pouches containing one each RAMP NT-proBNP Test Cartridge and Assay Tip25RAMP InstrumentPrinter and accessories
RAMP NT-proBNP Sample Buffer Vials25Barcode scanner
Transfer Device1PC with RS-232 connector
Lot Card1Commercial Quality Control Material
Package Insert1

Reagents

  • The RAMP NT-proBNP Assay kit contains all the reagents necessary for the quantification of NTproBNP in EDTA whole blood using the RAMP reader.
  • The RAMP NT-proBNP Sample Buffer contains phosphate buffer, animal protein, surfactant, and ◆ ProClin® 300 / ProCline 950 as preservatives.

ProClin is a registered trademark of Rohm and Haas Company.

{12}------------------------------------------------

WARNINGS AND PRECAUTIONS

  • The device contains material of animal origin and should be handled as a potential biohazard. ◆
  • ProClin is a potential skin sensitizer. Avoid spilling or splashing reagents containing ProClin on skin or clothing. In case of contact, thoroughly flush with water.
  • Use appropriate precautions in the collection, handling, storage and disposal of blood specimens, . including prepared blood specimens, and used kit contents.
    • (Refer to institutional quidelines for biological waste management.)
  • Discard and do not use any visibly damaged cartridges, or the contents of any Cartridge/Assay Tip � pouch with a damaged seal.
  • Do not use kit contents after the "Use By" date. .
  • Do not mix components from different kits. .
  • The RAMP NT-proBNP Test Cartridge, Assay Tip, and Sample Buffer Vial should be discarded . after a single-use. Do not reuse.
  • Dispose of unpouched, unused Test Cartridges and Assay Tips within 60 minutes. .
  • Sample/Buffer mixture must be applied to cartridge sample well only. .
  • Blood samples that show gross hemolysis may interfere with the test and cause erroneous results. . If this occurs, another blood sample should be obtained and tested.
  • The following symbols are used on the kit packaging: �
SymbolDefinition
LOTBatch Code
REFReference Number
Use By
Do Not Reuse
Consult Instructions For Use
Caution, Consult Accompanying Documents
Storage Temperature

NOTE: Do not insert a Test Cartridge that is externally wet with blood or other liquid into the RAMP reader, as this may cause contamination or damage to the reader.

LIMITATIONS OF PROCEDURE

  • Specimens from patients who have received preparations of mouse monoclonal antibodies for diagnosis or therapy may contain human anti-mouse antibodies. In addition, other heterophilic antibodies may be present in patient samples. Such specimens may show either falsely elevated or depressed values when tested with immunoassays. The RAMP NT-proBNP Assay has been formulated to reduce the effects of heterophilic antibodies: however results from patients known to have such antibodies should be interpreted with caution.
  • Blood specimens using anti-coaqulants other than EDTA (e.g. heparin) and serum have not been . evaluated, and should not be used.
  • . A test result that is inconsistent with the clinical signs and symptoms should be interpreted with caution.
  • Functions such as technical or procedural errors, or the addition of substances not evaluated for . may interfere with the RAMP NT-proBNP Assay.

The results obtained from the use of this product should be used only as an adjunct to other diagnostic procedures and information available to the physician.

Page 3 of 11

{13}------------------------------------------------

STORAGE INSTRUCTIONS

The Test Cartridges and Sample Buffer Vials should be stored at 2 - 8℃. Prior to sample preparation, components should be removed from the refrigerator and allowed to equilibrate to room temperature for at least 15 minutes. Once removed from refrigeration the components are stable for up to 14 days when stored at room temperature (15 - 25℃), but not beyond the expiration date printed on the pouch. If the test is not used on the same day it is removed from refrigeration, use a permanent marker to write the date of removal from the refrigerator and the discard date on the foil pouch and sample buffer vial and/or the kit box.

NOTE: Do Not Freeze.

SPECIMEN COLLECTION AND PREPARATION

Testing should be completed within 2 hours of phlebotomy. However, if this is not possible, the EDTA whole blood can be stored for up to 2 days at 2 - 8℃.

NOTES: The sample should be well-mixed before use.

  • Use only EDTA whole blood. Do not use heparinized whole blood, or plasma, or serum.
    Do not use samples that have been frozen

CALIBRATION PROCEDURES

Each RAMP NT-proBNP kit includes a Lot Card that is individually packaged in an anti-static pouch. The Lot Card provides all calibration information specific to the kit Test Cartridge lot, including lot number, expiration date, and standard curve information. For further details on loading lot-specific information, see the reader Operator's Manual. No additional calibration, beyond insertion of the Lot Card, is necessary. This operation is required only once per Assay kit lot.

SET-UP

If not previously done, remove the Lot Card for this kit lot from its pouch and insert it into the Lot Card slot on the RAMP instrument. Once the Lot Card information has been uploaded, return the Lot Card to its pouch.

WARNING: Avoid touching the contacts at the end of the Lot Card.

  • 2 For additional information on the operation of the RAMP reader, please refer to the Operator's Manual.

SAMPLE ANALYSIS PROCEDURE

NOTE: Prior to sample preparation, remove all necessary components from refrigeration and allow to equilibrate to room temperature for at least 15 minutes

You will need the Transfer Device and for each sample being tested, one Sample Buffer Vial and a pouch containing one Test Cartridge and one Assay Tip.

  • Place the Sample Buffer Vial upright on a clean, dry, level surface or in a sample vial holder. 1.
  • Open a pouch containing one NT-proBNP Test Cartridge and one NT-proBNP Assay Tip. Place 2. the Test Cartridge on a clean, dry, level surface.
  • Firmly attach the Assay Tip to the supplied Transfer Device. 3.
  • Ensure that the EDTA whole blood sample is well mixed (by inversion). Remove the cap from the র্য Sample Buffer Vial.
  • ક. Fully depress the Transfer Device plunger and insert the Assay Tip into the blood sample. Do not prime or rinse the Assay Tip in the blood sample.
  • Gently release the plunger to fill the Assay Tip and immediately transfer the Assay Tip into the 6. Sample Buffer Vial close to, but not touching the bottom of the Vial. Do not press against the bottom of the Vial as this may block the Tip.

Page 4 of 11

{14}------------------------------------------------

  • Mix the sample slowly by pressing and releasing the plunger 10 times (2 seconds per cycle), 7. taking care each time to eject entire sample into the Vial and to draw only liquid and no air into the Assay Tip. This will provent foaming.
  • Position the filled Assay Tip directly over the sample well of the Test Cantridge and fully depress 8. the plunger to dispense the mixed blood sample well. (Disregard any remaining droplet in the Assay Tip.) Dispose of used Assay Tip and Sample Buffer Vial according to local biohazard procedures.
  • Immediately (within 30 seconds) insert the Test Cartridge into the reader, and then press until ರು firm resistance is felt. Delay will lead to an error message.
  • Do not insert a Test Cartridge that is externally wet with blood or other liquid into the 10. RAMP reader, as this may cause contamination or damage to the reader.
  • Test result is complete approximately 15 minutes from Test Cartridge insertion. 11.
  • Record the result. 12.
  • I For additional information on printing and/or uploading results, please refer to the Operator's Manual.
  • Remove the used Test Cartridge when prompted to do so by the instrument display. Dispose of 13. the used Test Cartridge according to local biohazard procedures.

QUALITY CONTROL

System Quality Control (QC)

  • 21 Refer to the reader Operator's Manual for full details on quality control measures and troubleshooting.
  • The instrument has error checking and self-diagnostic functions that assure procedural control. . These include algorithms and measurements used to confirm acceptable operator technique, sample handling, and assay performance.
  • If a problem is detected, a message is displayed. Contact Response Biomedical Technical . Support (604-219-6119) for assistance.
  • Valid results are displayed only after all performance requirements have been met. e

Procedural Controls

The RAMP NT-proBNP Assay has built in (procedural) controls. Each Test Cartridge has an internal standard zone that is scanned as part of the test protocol to ensure proper sample flow. Control limits for each lot of Test Cartridges are established during the manufacturing process and are incorporated in the specific Lot Card. If a control result does not meet specifications, the sample result is not reported and a message is displayed.

External Quality Control

  • It is recommended that two levels of quality control material be run in the RAMP NT-proBNP Assay upon receipt of a new lot number or shipment of reagent and in conformance with federal, state and local requirements for quality control testing.
  • · To run a QC sample, follow the instructions under the SAMPLE ANALYSIS PROCEDURE section. Treat the control as a whole blood sample.

Test Run Messages

When the RAMP reader is unable to continue a specific task it will emit an audio alarm and display a message.

  • A Refer to the Operator's Manual Troubleshooting Guide section for a full description of all Messages. If repeated tests give unexpected or inconsistent results, contact our Technical Support for assistance.
    Page 5 of 11

{15}------------------------------------------------

PERFORMANCE CHARACTERISTICS

Precision

The intra-assay and inter-assay precision of the RAMP NT-proBNP Assay were determined by one operator assaving duplicates of three concentrations of control material (140, 449 and 1675 ng/L NTproBNP) twice each day over a 10-day period. The mean, standard deviation and %CV were calculated for each reported concentration of NT-proBNP. The results of this precision analysis are shown below.

PrecisionNT-proBNP Standards
NT-proBNP ng/L1404491675
Within Run CV9.4%6.4%5.5%
Total CV10.3%9.8%8.9%

EDTA anticoagulated whole blood samples spanning the reportable range of the RAMP NT-proBNP Assay were analyzed. Ten replicate measurements were carried out by a single operator in one day. The results of twelve samples from this precision analysis are shown below.

PrecisionEDTA Whole Blood Samples
Mean NT-proBNP ng/L5273113131161299230640515889844519504
CV (%)20.716.612.110.36.67.44.54.44.35.43.0

Precision in the hands of the end user was evaluated at each clinical site (N=4). Each operator performed three (3) replicates of a plasma based control product. The results are presented for each operator and across operators below:

Operator123456combined
Level 1 CV (%)9.3%9.5%15.5%8.8%11.8%11.0
Level 2 CV (%)3.4%6.7%4.4%0.4%12.4%2.5%5.0

Limits of Detection and Quantitation

Following CLSI EP-17A, limit of detection (LoD) was determined to be 34 ng/L, the limit of blank (LoB) was calculated as the 95" percentile from forty replicates of a blank sample run using the RAMP NTproBNP Assay and was determined to be 27 ng/L ..

NT-oroBNP levels in excess of 22.000 ng/L are reported as greater than (>) 22.000 ng/L.

The limit of quantitation (LoQ) is defined as the NT-proBNP level at which the test method displays a particular coefficient of variation (%CV). As shown below, the 20% LoQ for the RAMP NT-proBNP Assay was determined from whole blood analyses to be 57 ng/L.

Image /page/15/Figure/12 description: The image shows a scatter plot titled "RAMP NT-proBNP Low Range Precision Profile EDTA Whole Blood". The x-axis is labeled "[NT-proBNP] (ng/L)" and ranges from 0 to 300. The y-axis is labeled "Precision (%CV)" and ranges from 0 to 40. The data points are scattered across the plot, showing the relationship between NT-proBNP concentration and precision.

Page 6 of 11

{16}------------------------------------------------

Hook Effect

There is no high dose hook effect in the RAMP NT-proBNP Assay up to the highest level tested (350,000 ng/L NT-proBNP).

Linearity

A high NT-proBNP antigen concentration was prepared in normal donor EDTA blood and determined to contain 21,921 ng/L NT-proBNP by assaying the sample in duplicate. The sample was serially diluted six times. Regression analysis using the method of Passing-Bablok of actual versus expected NT-proBNP concentration resulted with an R-value of 1.06 and an offset of -1.4 ng/L. The recovery of NT-proBNP antigen at the six dilutions ranged from 101 to 120% with an average of 108%.

A low NT-proBNP antigen concentration was prepared in normal donor EDTA blood and determined to contain 264 ng/L NT-proBNP by assaying the sample in duplicate. The sample was serially diluted four Regression analysis using the method of Passing-Bablok of actual versus expected NTtimes. proBNP concentration resulted with an R-value of 1.00, a slope of 1.06 and an offset of -2.0 ng/L. The recovery of NT-proBNP antigen at the four dilutions ranged from 85 to 110% with an average of 100%.

Analytical Specificity

Human anti-mouse antibodies (HAMA) and Rheumatoid Factor (RhF) appear to have minimal crossreactivity with the RAMP NT-proBNP Assay. Possible cross-reactivity of other substances was evaluated by spiking different concentrations of the potential cross-reactants into EDTA blood which had NT-proBNP added. No cross reactivity was observed with the RAMP NT-proBNP Assay up to the maximum levels tested for the compounds listed in the following table.

CompoundConcentrationCompoundConcentrationCompoundConcentration
ANP283.1 µg/mLpreproANP 104-1231 ng/mLEndothelin20 ng/L
BNP323.5 µg/mLAldosterone0.6 ng/mLArg-Vasopressin1 ng/mL
CNP222.2 µg/mLAngiotensin I0.6 ng/mLRenin50 ng/mL
preproANP26-553.5 µg/mLAngiotensin II0.6 ng/mLAndrenomedullin1 ng/mL
preproANP56-921 ng/mLAngiotensin III1 ng/mLUrodilatin3.5 µg/mL

Interference

Potentially interfering substances were evaluated by spiking different concentrations of potential interferents into EDTA whole blood with NT-proBNP added to provide levels of 125 and 450 ng/L. Different blood samples were used for each potential interferent with an average difference of less than 10% from the unspiked samples observed in each case. The therapeutic compounds tested (at concentrations up to and including those indicated) are listed in the following table:

CompoundConcentrationCompoundConcentration
Acetaminophen20 mg/dLFurosemide6 mg/dL
Acetylsalicylic acid100 mg/dLHydralazine20 µg/mL
Allopurinol4 mg/dLHydrochlorothiazide20 µg/mL
Amiodarone20 µg/mLIndomethacin40 µg/mL
Amiodipine Besylate4 µg/mLInsulin120 µU/mL
Ampicillin sodium salt100 mg/dLIsosorbide Dinitrate15 mg/dL
Ascorbic acid30 mg/dLLisinopril4 mg/dL
Atenolol1 mg/dLMethyldopa2.5 mg/dL
Caffeine10 mg/dLMetoprolol Tartrate2 mg/dL
Captopril15 mg/dLNicotine2 mg/dL
Carvedilol5 mg/dLNifedipine6 mg/dL
Chloramphenicol25 mg/dLNitroglycerin19.2 mg/dL
Clopidogrel Hydrogensulfate7.5 mg/dLOxytetracycline100 µg/mL
Cyclosporin A0.5 mg/dLProbenecid600 µg/mL
Diclofenac60 µg/mLPropranolol0.2 mg/dL

Page 7 of 11

{17}------------------------------------------------

CompoundConcentrationCompoundConcentration
Digitoxin0.03 mg/dLQuinidine20 µg/mL
Digoxin0.05 mg/dLSimvastatin4 mg/dL
Diltiazem120 µg/mLTheophylline100 mg/dL
Phenytoin10 mg/dLTrimethoprim60 µg/mL
Dipyridamole30 µg/mLVerapamil16 mg/dL
Enalapril Maleate4 mg/dLWarfarin20 µg/mL
Erythromycin20 mg/dL

Hemoglobin, triglyceride, bilirubin, cholesterol, and heparin at levels of very high physiological concentrations were investigated for possible interference. No interference was observed when tested at the concentrations up to and including those shown in the following table:

CompoundConcentrationCompoundConcentration
Hemoglobin2 g/dLCholesterol500 mg/dL
Triglyceride4 g/dLHeparin104 IU/mL
Bilirubin35 mg/dL

CLINICAL EVALUATIONS OF ANALYTICAL PERFORMANCE

Method Comparison

Six hundred and ninety-nine (699) patients were enrolled in the clinical evaluation. The presenting population included 46% (323) with hypertension, 30% (208) who presented with shortness of breath, 22% (152) with diabetes, 14% (99) with pulmonary disorders, 12% (84) with coronary disease, 8% (56) with atrial fibrillation, 4% (31) with renal failure, 19% (133) were healthy, and the remainder had diagnoses not believed to be cardiac related (hepatitis, HIV, cancer, etc.). EDTA and hepain whole blood samples were obtained for each of these subjects.

EDTA and heparin whole blood samples were obtained for each of these subjects. An aliquot of EDTA whole blood was used for the RAMP NT-proBNP Assay and heparinized plasma was prepared for the Roche Elecsys proBNP Assay. From these analyses it was determined that 580 samples contained between 34 ng/L (RAMP LoD) and 22,000 ng/L NT-proBNP. Of these, 274 were diagnosed with heart failure (HF) based on individual hospital criteria (164 males and 110 females) and 306 were non-HF reference group patients (124 males and 182 females). Regression analysis data of RAMP NT-proBNP versus Elecsys proBNP using the method of Passing-Bablok is presented in the table below.

Comparative MethodSlopeIntercept (ng/L)Correlation coefficient (R)
Roche Elecsys0.9719.390.98
95% CI0.95 to 1.0014.20 to 24.670.97 to 0.98

Clinical Sensitivity and Specificity

Clinical sensitivity and specificity were calculated using data collected from 858 subjects. Of these, 299 were diagnosed with HF using local hospital criteria, 189 individuals without HF but with potentially confounding co-morbidity (diabetes, renal insufficiency, hypertension or chronic obstructive pulmonary disease) and 370 reference individuals. This reference group includes an additional 159 subjects added from an additional clinical site without concomitant testing in the Elecsys system. Of these, 55% (87) were male and 8% (12) were more than 75 years old. None of these patients had reported co-morbidities. These subjects were healthy individuals with no clinical indications for natriuretic peptide testing.

The use of the cut-offs of 125 ng/L for ≤75 years of age and 450 ng/L for >75 years of age was evaluated across all patients stratifying by the presence of co-morbidities (diabetes, renal

{18}------------------------------------------------

insufficiency, hypertension or chronic obstructive pulmonary disease). Sensitivity and specificity are shown in the following table.

CHF Patients
Age (years)< 75> 75
N21782
Sensitivity0.890.99
95% CI(0.84-0.93)(0.92-1.0)
Non-CHF no comorbidity*
Age (years)< 75>75
N34030
Specificity0.850.72
95% CI(0.80-0.88)(0.53-0.87)
Non-CHF with comorbidity
Age (years)<75>75
N12465
Specificity0.430.48
95% CI(0.43-0.52)(0.35-0.60)

Age Stratified Sensitivity and Specificity: 125/450 ng/L for Ages ≤ 75 / >75

udes additional 159 healthy subjects

Receiver Operator Characteristic (ROC) Analyses

The ROC analyses for both the RAMP NT-proBNP and Roche Elecsys proBNP assays for the clinical study population are shown below. The additional 159 patients without Elecsys results were excluded from this comparison. The area under the curve (AUC) for both the RAMP NT-proBNP Assay and Elecsys proBNP assays is 0.87.

Image /page/18/Figure/6 description: The image is a graph that plots sensitivity versus 1-specificity. There are two curves on the graph, one for RAMP NT-proBNP result (ng/L) and one for Elecsys NT-proBNP result (ng/L). The x-axis ranges from 0.0 to 1.0, and the y-axis ranges from 0.0 to 1.0. There is also a line indicating 'No discrimination'.

CHF Population by NYHA Classification

The 299 subjects diagnosed with heart failure were evaluated using the RAMP NT-proBNP Assay. The descriptive statistics for NT-proBNP concentrations are presented according to NYHA Functional Classification in the table below.

Page 9 of 11

{19}------------------------------------------------

All
NYHA ClassIIIIIIIV
n58918466
Mean1686283157378308
SD3161435659397090
Median832147936086628
95th percentile5560810420177>22000
Male
n32565540
Mean1737287057998855
SD3924464161827612
Median724131836235772
95th percentile47221074221068>22000
Female
n26352926
Mean1624277156187466
SD1918392155516251
Median907162235986937
95th percentile543873061672721839

Non-HF and HF Group Descriptive Statistics

The overall incidence of disease in the presenting population (n=858) included 38% (323) subjects with hypertension, 24% (208) who presented with shortness of breath, 18% (152) with diabetes, 12% (99) with pulmonary disorders, 10% (84) with coronary disease, 7% (56) with atrial fibrillation, 4% (31) with renal failure, 34% (292) were healthy, and the remainder had diagnoses not believed to be cardiac related (hepatitis, HIV, cancer, etc.).

The circulating NT-proBNP concentration was determined in 858 individuals with and without HF. The HF patients included those with prior established heart failure that were not acutely destabilized at the time of enrollment (and thus similar to those who might be tested in the outpatient setting). Patients (N=17) for whom the measured NT-proBNP level was greater than 22,000 ng/L have been excluded. The descriptive statistics for the Non-HF with and without co-morbidities and the HF groups are presented in the following tables:

No co-morbidityWith co-morbidity
Age (years)>75<75ALLAge (years)>75<75ALL
n30340370n65124558
Mean449.7132.8158.5Mean1013.0870.5349.4
SD810.9671.2687.8SD1524.63445.1996.0
Median88.024.528.0Median512.0185.9566.0
95th percentile2447.4216.4451.095th percentile3986.02463.21543.9
% <125 ng/L84% <125 ng/L44
% < 450 ng/L74% < 450 ng/L48

A. Non-HF Patients -RAMP Results (ng/L)

Page 10 of 11

{20}------------------------------------------------

B. HF Patients - RAMP Results (ng/L)

Age (years)>75<75ALL
N80203283
Mean4970.13133.13652.4
SD5185.83755.04280.2
Median3300.51735.02040.0
95th percentile19005.011373.312800.0
% >125 ng/L89
% >450 ng/L100

BIBLIOGRAPHY

  • Hall C. Essential biochemistry and physiology of (NT-pro)BNP. Eur J Heart Fail. 2004;6:257-260.
  • 2 Fichards AM, Nicholls GM, Yandle TG, et al. Plasma N-terminal pro-brain natriuretic peptide and adrenomedulin: New hormonal predictors of left ventricular function and prognosis atter myccardial infarction. Circ. 1998;97:1921-1929.
    1. Rademaker MT, and Richards MA. Cardiac natriuretic peptides for cardiac health. Clin Sci. 2005;108:23-36.
  • ه Januzzi JL, Camargo CA, Anwaruddin S, et al. The N-terminal Pro-BNP investigation of dyspnea in the emergency department (PRIDE) study. Am J Cardiol. 2005;95:948-954.
  • 5, Baggish AL, Cameron R, Anwaruddin S, et al. A clinical and biochemical oftical pathway for the evaluation of palients with suspected acute congestive heart failure. The proBNP investigation of dyspnea in the emergency department (PRIDE) algorithm. Crit Path Cardiol. 2004;3:171-176.
    1. James SK, Lindahl B, Siegbahn A, et al. N-terminal pro-brain natriuretic peptide and other risk markers for the separate prediction of mortality and subsequent myocardial infarction in patients with unstable coronary artery disease: a global utilization of strategies to open occluded arteries (GUSTO)-IV substudy. Circ. 2003;108:275-81.
    1. Jernberg T, Stridsberg M, Venge P, et al. N-terminal pro brain natriurelic peptide on admission for early risk stratification of patients with chest pain and no ST-segment elevation. J Am Coll Cardiol. 2002;40:437-45.
    1. Fisher C, Berry C, Blue L, et al. N-terminal pro B type natriuretic peptide, but not the new putative cardiac hormone relaxin, predicts prognosis in patients with chronic heart failure. Heart. 2003;89:879-81.
  • ರ್ಥಿ Schnabel R, Rupprecht HJ, Lackner KJ, et al. Analysis of N-terminal-pro-brain natriurelic peptide and C-reactive protein for risk stratification in stable and unstable coronary artery disease: Results from the AtheroGene study. Eur Heart J. 2005;26:241-249.
    1. Kragelund C, Groenning B, Kober L, et al. N-terninal pro-B-type natriuretic peptide and long-term mortality in stable coronary heart disease. N Eng J Med. 2005;352:666-675.
  • 11 Ndrepepa G, Braun S, Niemoller K, et al. Prognostic value of N-terminal pro-brain natriuretic peptide in patients with chronic stable angina. Circ. 2005; 112:2102-2107.

Formatted: German (Germany)

Formatted: Italian (Italy)

Formatted: English (U.S.)

Formatted: Dutch (Netherlands)

Formatted: Dutch (Netherlands)

Page 11 of 11

{21}------------------------------------------------

Image /page/21/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract eagle design, with three stylized lines forming the body and wings. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Response Biomedical Corporation c/o Dr. Ken Pilgrim, Director - Quality/Regulatory 1781 - 75th Avenue W. Vancover, British Columbia Canada V6P 6P2

JUL 21 2008

Re: K063662

Trade/Device Name: Response Biomedical Corporation RAMP® NT-proBNP Assay Regulation Number: 21 CFR 862.1117 Regulation Name: B-type Natriuretic peptide test system Regulatory Class: Class II Product Code: NBC Dated: April 22, 2008 Received: April 23, 2007

Dear Dr. Pilgrim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{22}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Jean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{23}------------------------------------------------

Statement of Indications for Use

Indications for Use

510(k) Number (if known): K063662

Device Name: RAMP® NT-proBNP Assay

Indications for Use:

The RAMP NT-proBNP Assay is a quantitative immunochromatographic test indicated for use as an in vitro diagnostic product used to measure N-terminal pro-brain natriuretic peptide (NTproBNP) levels in EDTA whole blood. Measurement of NT-proBNP aids in the diagnosis and assessment of severity in individuals suspected of having heart failure and may aid in the risk stratification of patients with heart failure.

Prescription Use × (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Sean Cooper
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K063662

§ 862.1117 B-type natriuretic peptide test system.

(a)
Identification. The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.(b)
Classification. Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”