K Number
K063648

Validate with FDA (Live)

Date Cleared
2007-03-07

(90 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgi-Wrap MAST Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues.

The Surgi-Wrap MAST Tendon Sheet is also indicated for reinforcement of soft tissues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Surgi-Wrap MAST Tendon Sheet is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The Surgi-Wrap MAST Tendon Sheet reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients own soft tissue.

The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.

Device Description

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is a resorbable implant in sheet form manufactured from poly (L-lactide-co-D,L-lactide) 70:30 (PLA). The Surgi-Wrap MAST Tendon Sheet can be cut with scissors to the desired shape and size. The Surgi-Wrap MAST Tendon Sheet Sheet is fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The Surgi-Wrap MAST Tendon Sheet can be used either alone or in conjunction with soft tissue fixation devices such as resorbable sutures, which can also serve to fixate the Surgi-Wrap MAST Tendon Sheet and prevent dislocation. The Surgi-Wrap MAST Tendon Sheet may be used in conjunction with various MAST Biosurgery Class I manual instruments (forceps, scissors, clamps, etc.).

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in various shapes such as rectangles, ovals, and circles and will be provided in other shapes and sizes as needed for particular surgical procedures. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in sheets of 25mm x 25mm to 500mm x 500mm and will be provided in other shapes and sizes as needed for particular surgical procedures. The thickness of the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet ranges from 0.02 mm to 1.0 mm according to the region to be treated. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in solid sheets. The borders of the sheets may be aligned with holes to attach suture material.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called the Surgi-Wrap MAST Tendon Sheet. This type of document is for premarket notification to demonstrate substantial equivalence to a legally marketed predicate device, rather than a clinical study establishing novel acceptance criteria and performance against those.

Therefore, the document does not contain a traditional study that defines specific acceptance criteria with corresponding performance metrics that would be typically found in a clinical trial or a performance study for a new, non-substantially equivalent device.

Instead, the submission demonstrates "substantial equivalence" to existing, legally marketed predicate devices. This means the device is considered as safe and effective as a predicate device and does not raise different questions of safety and effectiveness. The "acceptance criteria" here are effectively met by demonstrating similarity to the predicate devices through material composition, design characteristics, intended use, and various testing (in vitro and in vivo).

However, I can extract the information that is present and frame it as closely as possible to your request.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) submission based on substantial equivalence, explicit quantitative acceptance criteria with corresponding performance metrics are not defined in the same way as a de novo clinical study. Instead, the "acceptance criteria" can be inferred as similarity or equivalence to the predicate devices in terms of properties and function. The "reported device performance" is essentially the evidence presented to support this equivalence.

Acceptance Criteria (Implied)Reported Device Performance
Material Properties: Suitable for surgical implantation, resorbable.Fabricated from polylactic acid (PLA), specifically poly (L-lactide-co-D,L-lactide) 70:30. Demonstrated bioresorbable.
Design Characteristics: Flexible, malleable, cuttable, configurable.Provided in sheet form, can be cut with scissors to desired shape/size. Malleable when heated to ~55°C for 3D conformation. Available in various shapes (rectangles, ovals, circles) and sizes (25mm x 25mm to 500mm x 500mm), thicknesses (0.02mm to 1.0mm).
Mechanical Properties (Post-Heating): Maintain appropriate viscosity and sufficient strength after heating.In Vitro Testing: Prolonged heating in saline at 60°C for 120 minutes showed viscosity within an appropriate range. Aging testing demonstrated sufficient strength for indicated uses. Mechanical testing determined substantial equivalence to predicate devices' mechanical strengths.
Safety and Efficacy (Biological): Safe and efficacious for tendon injury management and soft tissue reinforcement.In Vivo Testing: Animal studies conducted to demonstrate safety and efficacy of the material for the indications for use. These studies specifically concluded the materials are "safe and efficacious."
Intended Use: Consistent with predicate devices.Indicated for management and protection of tendon injuries (no substantial tissue loss) and reinforcement of soft tissues during tendon repair surgery (rotator cuff, patellar, Achilles, biceps, quadriceps, etc.). Shares identical indications for use language with predicate devices. Indicated for open and laparoscopic/endoscopic procedures (thickness limited for latter).
Substantial Equivalence: To legally marketed predicate devices.Found substantially equivalent to: MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473), Integra LifeSciences Tendon Wrap (K053655), TEI Biosciences TissueMend Soft Tissue Repair Matrix (K060989), and S. Jackson Supra FOIL Smooth Nylon foil Sheets (K973379).

Due to the nature of a 510(k) submission for substantial equivalence based on in vitro and animal studies, most of the detailed information requested regarding sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, and ground truth for human-centric clinical trials is not applicable or not provided in this document.

Here's an assessment based on the available information:

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size (Test Set): Not specified for the in vitro or in vivo (animal) studies. The document only states that "testing was performed" and "animal studies were conducted."
  • Data Provenance:
    • In Vitro Testing: Performed on the device material itself, likely in a laboratory setting.
    • In Vivo Testing: Animal studies. The country of origin for these studies is not specified, but the submission is to the US FDA.
    • Retrospective or Prospective: Not explicitly stated, but in vitro and in vivo animal studies are generally prospective by design.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Not applicable / Not provided. The studies described (in vitro and animal) do not involve human expert interpretation of clinical data in the way implied by this question. The "ground truth" for these studies would be defined by the experimental setup and measurements (e.g., direct measurement of viscosity, mechanical strength tests, histological analysis in animals).

4. Adjudication Method for the Test Set

  • Not applicable / Not provided. Adjudication methods like "2+1" or "3+1" are relevant for human reader studies interpreting complex data (e.g., medical images). The testing described here does not involve such a process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. An MRMC study is not mentioned or implied. This type of study would involve human readers evaluating clinical cases with and without AI assistance to measure improvement in diagnostic performance. The Surgi-Wrap MAST Tendon Sheet is a physical implantable device, not an AI diagnostic tool.

6. If a Standalone (Algorithm Only) Performance Study was done

  • Not applicable / Not provided. This question generally pertains to AI algorithms. The "standalone" performance here would refer to the device's inherent physical and biological properties as demonstrated in the in vitro and in vivo (animal) tests, without human interaction during the test itself (though human surgeons implant the device).

7. The Type of Ground Truth Used

  • In Vitro Testing: The "ground truth" for in vitro tests would be the direct measurements of properties like inherent viscosity and mechanical strength using standardized laboratory methods and equipment.
  • In Vivo Testing (Animal Studies): The "ground truth" from these studies would be based on observations of safety (e.g., biocompatibility, adverse reactions) and efficacy (e.g., tissue integration, healing response, absence of substantial adverse events) in the animal models, likely confirmed by macroscopic and microscopic (histopathological) examination.

8. The Sample Size for the Training Set

  • Not applicable / Not provided. The concept of a "training set" is relevant for machine learning algorithms. The provided document describes testing of a physical medical device.

9. How the Ground Truth for the Training Set was Established

  • Not applicable / Not provided. As explained above, there is no "training set" in the context of this device's submission.

{0}------------------------------------------------

510(k) Summarv Surgi-Wrap MAST Tendon Sheet

ADMINISTRATIVE INFORMATION

Manufacturer Name:

Official Contact:

MAST Biosurgery, Inc.

MAR 0 7 2007

Page 1 of 3

6749 Top Gun Street, Suite C San Diego, CA 92121

Kenneth K. Kleinhenz Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994

DEVICE NAME

Classification Name:

Trade/Proprietary Name:

Surgical Mesh, Polymeric

Surgi-Wrap MAST Tendon Sheet

ESTABLISHMENT REGISTRATION NUMBER 3004661493

DEVICE CLASSIFICATION AND PRODUCT CODE

As shown in 21CFR 878.3300, Surgical Mesh are polymeric screens intended to be implanted to reinforce soft tissues. These devices are classified as Class II. Surgical Mesh have been assigned Product Code FTL and FTM.

INTENDED USE

The Surgi-Wrap MAST Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues.

The Surgi-Wrap MAST Tendon Sheet is also indicated for reinforcement of soft tissues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Surgi-Wrap MAST Tendon Sheet is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The Surgi-Wrap MAST Tendon Sheet reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients own soft tissue.

The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.

{1}------------------------------------------------

Surgi-Wrap MAST Tendon Sheet 510(k) Summary

Page 2 of 3

DEVICE DESCRIPTION

Design Characteristics

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is a resorbable implant in sheet form manufactured from poly (L-lactide-co-D,L-lactide) 70:30 (PLA). The Surgi-Wrap MAST Tendon Sheet can be cut with scissors to the desired shape and size. The Surgi-Wrap MAST Tendon Sheet Sheet is fully malleable when heated to approximately 55℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The Surgi-Wrap MAST Tendon Sheet can be used either alone or in conjunction with soft tissue fixation devices such as resorbable sutures, which can also serve to fixate the Surgi-Wrap MAST Tendon Sheet and prevent dislocation. The Surgi-Wrap MAST Tendon Sheet may be used in conjunction with various MAST Biosurgery Class I manual instruments (forceps, scissors, clamps, etc.).

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in various shapes such as rectangles, ovals, and circles and will be provided in other shapes and sizes as needed for particular surgical procedures. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in sheets of 25mm x 25mm to 500mm x 500mm and will be provided in other shapes and sizes as needed for particular surgical procedures. The thickness of the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet ranges from 0.02 mm to 1.0 mm according to the region to be treated. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is provided in solid sheets. The borders of the sheets may be aligned with holes to attach suture material.

Material Composition

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is fabricated from polylactic acid (PLA).

In Vitro Testing

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures. Therefore, testing was performed to determine the effect of prolonged heating in saline at 60°C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. The relatively brief exposure anticipated during the surgical preparation of MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is not expected to have a significant effect on its mechanical properties.

Aging testing was performed on MAST Biosurgery Surgi-Wrap MAST Tendon Sheet. Testing demonstrated that the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is strong enough for the indications for use.

Mechanical testing was performed on the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet which determined the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.

{2}------------------------------------------------

Surgi-Wrap MAST Tendon Sheet 510(k) Summary

Page 3 of 3

In Vivo Testing

Animal studies were conducted to demonstrate safety and efficacy of the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet material. The animal studies demonstrated that the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet materials are safe and efficacious for the indications for use.

EQUIVALENCE TO MARKETED PRODUCT

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473). Integra LifeSciences Tendon Wrap (K053655), TEI Biosciences TissueMend Soft Tissue Repair Matrix, (K060989) and S. Jackson Supra FOLL Smooth Nylon foil Sheets (K973379); Class II medical devices that were cleared for marketing in the United States under K061473, K053655, K060989 and K973379 respectively.

Indications For Use

The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet shares indications for use principles with the predicate devices as all devices are indicated for use in the same surgical procedures involving tendons. Furthermore, the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet shares identical indications for use language with each of the predicate devices.

Design and Materials

The physical designs of the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet and the MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473), Integra LifeSciences Tendon Wrap (K053655), TEI Biosciences Tissue Repair Matrix, (K060989) and S. Jackson Supra FOIL Smooth Nylon foil Sheets (K973379) predicate devices are substantially equivalent as they are all flexible and malleable sheets provided in various sizes that they can be cut to shape intra-operatively and subsequently placed on, under, and around tendon anatomy. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet and the MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473), Integra LifeSciences Tendon Wrap (K053655), and TEI Biosciences TissueMend Soft Tissue Repair Matrix, (K060989) predicates all share the common design feature of being bioresorbable. The MAST Biosurgery Surgi-Wrap MAST Tendon Sheet is substantially equivalent to the S. Jackson Supra FOLL Smooth Nylon foil Sheets (K973379) and MAST Biosurgery Surgi-Wrap MAST Bioresorbable Sheet (K061473) predicates as they are all thin, semi-rigid sheets provided in sizes ranging from 25mm and 500mm x 500mm with various thicknesses ranging from 0.02mm to 1.0mm. The mechanical characteristics of the MAST Biosurgery Surgi-Wrap MAST Tendon Sheet are substantially equivalent to the predicate devices with respect to mechanical characteristics.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAST Biosurgery, Inc. % Mr. Kenneth K. Kleinhenz Regulatory Affairs 6749 Top Gun Street, Suite C San Diego, California 92121

MAR 0 7 2007

Re: K063648

Trade/Device Name: Surgi-Wrap MAST Tendon Sheet Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: December 4, 2006 Received: December 7, 2006

Dear Mr. Kleinhenz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -- Mr. Kenneth K. Kleinhenz

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark McMullen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K0653648

Indications for Use

510(k) Number :

Device Name: Surgi-Wrap MAST Tendon Sheet

Indications for Use:

The Surgi-Wrap MAST Tendon Sheet is indicated for the management and protection of tendon injuries in which there has been no substantial loss of tendon tissue. The resorbable Protective Film minimizes tissue attachment to the device in case of direct contact with the tissues.

The Surgi-Wrap MAST Tendon Sheet is also indicated for reinforcement of soft tissues repaired by suture or suture anchors, during tendon repair surgery, including reinforcement of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Surgi-Wrap MAST Tendon Sheet is not intended to replace normal body structure or provide the full mechanical strength to support tendon repair of the rotator cuff, patellar, Achilles, biceps, quadriceps, or other tendons. Sutures used to repair the tear and sutures or bone anchors used to attach the tissue to the bone provide biomechanical strength for the tendon repair. The Surgi-Wrap MAST Tendon Sheet reinforces soft tissue and provides a remodelable scaffold that is replaced by the patients own soft tissue.

The device is indicated for open and laparoscopic / endoscopic procedures. Laparoscopic / endoscopic procedures are limited to sizes from 0.02mm - 0.2mm in thickness.

Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Exaluation (ODE)

(Division Sign-off)
Division of General, Restorative,
and Neurological Devices

K013648

Page 1 of 1

70(k) Number K065678

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.