K Number
K063091

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-01-16

(463 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AIGISRx is intended to securely hold a pacemaker pulse generator or defibrillator in order to create a stable environment when implanted in the body. AIGISRX contains the antimicrobial agents, rifampin and minocycline, which have been shown to reduce infection in an in-vivo model of bacterial challenge following surgical implantation of the generator or defibrillator. This device is only intended to be used in conjunction with pacemakers and implantable defibrillators.

Device Description

AIGISRX is a dual component (resorbable and non-resorbable), sterile prosthesis designed to hold a pacemaker pulse generator or defibrillator to create a stable environment when implanted in the body. AIGISRX is constructed of knitted filaments of polypropylene that are coated with a bioresorbable polyarylate polymer. AIGISRX bioresorbable polymer coating contains the antimicrobial agents, rifampin and minocycline in concentrations of 86 ug/cm2.

AI/ML Overview

Here's an analysis of the provided text regarding the AIGISRX™ Anti-bacterial Envelope, focusing on acceptance criteria and the supporting study information:

It's important to note that the provided text is a 510(k) summary and not a comprehensive study report. Therefore, some information, particularly detailed methodological aspects typically found in full study publications, is either summarized or not explicitly stated.


Acceptance Criteria and Device Performance for AIGISRX™ Anti-bacterial Envelope

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary does not explicitly list formal "acceptance criteria" in the manner of specific pre-defined thresholds for performance metrics (e.g., "sensitivity must be > 90%"). Instead, the document outlines the device's intended functions and then presents performance data demonstrating that these functions are met, thereby supporting its substantial equivalence to predicate devices.

Here's an interpretation of the implicit "acceptance criteria" based on the device's intended use and the reported performance:

Acceptance Criterion (Implicit)Reported Device Performance
Mechanical Stability: Securely hold a pacemaker pulse generator or defibrillator to create a stable environment when implanted in the body."AIGISRX is a dual component (resorbable and non-resorbable), sterile prosthesis designed to hold a pacemaker pulse generator or defibrillator to create a stable environment when implanted in the body." The permanent polypropylene mesh is "incorporated into the tissue," implying long-term mechanical stability after the resorbable coating is gone.
Antimicrobial Agent Delivery: Release antimicrobial agents (rifampin and minocycline) for a sufficient duration to reduce infection risk."AIGISRX releases the antimicrobial agents, rifampin and minocycline for a minimum of 7 days to reduce the risk of infection of the implanted pulse generator following surgery."
Antimicrobial Efficacy (In Vitro): Demonstrate antimicrobial activity against relevant bacterial strains."In in vitro studies, AIGISRX demonstrated antimicrobial activity against Methicillin Resistant Staphylococcus aureus (MRSA), Staphylococcus aureus, Staphylococcus epidermidis, Acinetobacter baumanii, Enterobacter aerogenes and Proteus mirabilis." (Note: A caveat is mentioned: "the in vitro and in vivo activity of the AIGISR antimicrobials is variable against non-epidermidis strains of coagulase-negative staphylococci.")
Antimicrobial Efficacy (In Vivo): Reduce infections in an in-vivo model of bacterial challenge following surgical implantation of the generator or defibrillator."AIGISRX™ also demonstrated in vivo effectiveness in reducing infections in a series of studies in which a pulse generator canister placed into a AIGISRX™ pouch and generator canister alone (Control) were implanted into appropriate models of infectivity (dogs or rabbits)." "The bacteria tested included Staph aureus, Staph epidermidis, Acinetobacter baumanii, and E coli which represent a majority of the infections reported in pacemaker-related endocarditis." (Note: The same caveat regarding non-epidermidis strains applies here as well regarding variability.)
Bioresorption: The resorbable component functions as a carrier for antimicrobial agents and resorbs within a reasonable timeframe, leaving the permanent mesh."The purpose of the resorbable coating is to act as a carrier for the antimicrobial agents. Once placed, the polymer resorbs in approximately 140 days leaving a light-weight permanent mesh incorporated into the tissue."
Biocompatibility/Safety: As implied by its 510(k) clearance and classification as "Polymeric Surgical Mesh," the device must be safe for implantation and compatible with the body, similar to predicate devices. (No specific safety data is detailed in this summary).The device's clearance and classification indicate that it meets the general safety requirements for its type, and no adverse events or safety concerns are highlighted in the summary as a basis for rejection.

2. Sample size used for the test set and the data provenance

  • Test Set Description: The "test set" for the in vivo efficacy was described as "a series of studies in which a pulse generator canister placed into a AIGISRX™ pouch and generator canister alone (Control) were implanted into appropriate models of infectivity (dogs or rabbits)."
  • Sample Size: The exact number of animals (dogs or rabbits) used in these in vivo studies is not specified in the provided text.
  • Data Provenance: The in vivo data was obtained from animal models (dogs or rabbits). The country of origin is not specified. The studies appear to be prospective experimental studies designed to test the device's efficacy. The in vitro studies were laboratory-based experiments.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable and not provided in the context of this device and studies. The studies described are in vitro antimicrobial tests and in vivo animal infection models. "Ground truth" in these cases would be established by direct observation of microbial growth (in vitro) or clinical signs of infection, bacterial cultures, and potentially histological analysis (in vivo animal models), rather than expert reader interpretations of medical images.


4. Adjudication method for the test set

This information is not applicable and not provided. Given the nature of microbiology and animal model studies, adjudication methods (like 2+1 or 3+1 consensus) are typically used for human retrospective image analysis or clinical endpoint determination from human trials, not for direct laboratory or animal observations.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The AIGISRX™ Anti-bacterial Envelope is a physical medical device (an implantable envelope with antimicrobial agents), not an AI-powered diagnostic or assistive technology for human readers. Therefore, an MRMC study and analysis of AI assistance is irrelevant to this product.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. As stated above, this is a physical medical device, not an algorithm.


7. The type of ground truth used

  • In Vitro Studies: The ground truth for antimicrobial activity was established by direct observation of microbial growth or inhibition in laboratory settings using various bacterial strains.
  • In Vivo Studies (Animal Models): The ground truth for infection presence/absence was established by direct evidence of infection in the animal models. This likely involved:
    • Bacterial inoculation: Directly infecting the implanted control and device groups.
    • Clinical observation: Monitoring animals for signs of infection.
    • Microbial culture: Culturing implant sites or tissues from the animals to validate the presence and identify the bacteria causing infection.
    • The statement "observed for a minimum of 7 days to validate the presence of infection in the animals" suggests a direct endpoint.

8. The sample size for the training set

This information is not applicable in the context of this device. A "training set" typically refers to data used to train a machine learning algorithm. Since this is a physical medical device, there is no algorithm to train. The preclinical in vitro and in vivo studies are direct performance evaluations, not training phases for an algorithm.


9. How the ground truth for the training set was established

This information is not applicable, as there is no "training set" in the sense of data used for machine learning.

{0}------------------------------------------------

TYRX PHARMA INC.

K063091

January 14, 2008

SECTION 5. 510(K) SUMMARY

JAN 16 2008

510(K) SUMMARY (AS REQUIRED BY 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)3A of the Safe Medical Devices Act of 1990, TyRx Pharma, Inc. is providing a summary of the safety and effectiveness information available for TyRx AIGIS,xx" Anti-bacterial Envelope, as well as the substantial equivalence decision making process used for TyRx AIGISRX™ Anti-bacterial Envelope.

Sponsor/Applicant Name and Address: TyRx Pharma, Inc. 1 Deer Park Drive, Suite G Monmouth Junction, NJ 08852

Sponsor Contact Information: Mark Citron Vice President, Clinical and Regulatory Affairs Phone: 732-246-8676 Fax: 732-246-8677 E-mail: mark@tyrxpharma.com

Date of Preparaton of 510(k) Summary: January 14, 2008

New Device Trade/Proprietary Name: AIGISRX" Anti-bacterial Envelope

Device Common/Classification Name: Bag, Polymeric Mesh, Pacemaker

Predicate Devices Name and 510(k) Numbers:

  • . Bard Parsonnet Pacemaker Pouch (Listing Number 2212754)
  • Pivit AB (TyRx Antimicrobial Mesh) (K053656) .
  • OrthoGuard AB Antimicrobial Sleeve (K012198) .

Device Description

AIGISRX is a dual component (resorbable and non-resorbable), sterile prosthesis designed to hold a pacemaker pulse generator or defibrillator to create a stable environment when implanted in the body. AIGISRX is constructed of knitted filaments of polypropylene that are coated with a bioresorbable polyarylate polymer. AIGISRX bioresorbable polymer coating contains the antimicrobial agents, rifampin and minocycline in concentrations of 86 ug/cm2.

Device Intended Use:

{1}------------------------------------------------

3

ind 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . create a stable environment when implanted in the body. AIGISRX contains the antimicrobial agents, rifampin and minocycline, which have been shown to reduce infection in an in-vivo model of bacterial challenge following surgical implantation of the generator or defibrillator. This device is only intended to be used in conjunction with pacemakers and implantable defibrillators.

Performance Data:

AIGISpx is constructed of knitted filaments of polypropylene that are coated with a bioresorbable polyarylate polymer. The purpose of the resorbable coating is to act as a carrier for the antimicrobial agents. Once placed, the polymer resorbs in approximately 140 days leaving a light-weight permanent mesh incorporated into the tissue.

AIGISRX releases the antimicrobial agents, rifampin and minocycline for a minimum of 7 days to reduce the risk of infection of the implanted pulse generator following surgery. In in vitro studies, AIGISRX demonstrated antimicrobial activity against Methicillin Resistant Staphylococcus aureus (MRSA), Staphylococcus aureus, Staphylococcus epidermidis, Acinetobacter baumanii. Enterobacter aerogenes and Proteus mirabilis.

AIGISRX™ also demonstrated in vivo effectiveness in reducing infections in a series of studies in which a pulse generator canister placed into a AIGISRX™ pouch and generator canister alone (Control) were implanted into appropriate models of infectivity (dogs or rabbits). Both AIGISRY™ and the controls were inoculated and observed for a minimum of 7 days to validate the presence of infection in the animals. The bacteria tested included Staph aureus, Staph epidermidis, Acinetobacter baumanii, and E coli which represent a majority of the infections reported in pacemaker-related endocarditis. It should be noted that the in vitro and in vivo activity of the AIGISR antimicrobials is variable against non-epidermidis strains of coagulase-negative staphylococci

Conclusions:

Based on the 510(k) summaries (21 CFR 807) and the information provided herein, we conclude that the new device is substantially equivalent to the Predicate Devices under the Federal Food Drug and Cosmetic Act.

{2}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 6 2008

TyRx Pharma, Inc. c/o Mr. Mark Citron Vice President 1 Deer Park Drive Monmouth Junction, NJ 08852

Re: K063091 AIGISRX Anti-Bacterial Envelope Regulation Number: 21 CFR 878.3300 Regulation Name: Polymeric Surgical Mesh Regulatory Class: II Product Code: FTL Dated: January 3. 2008 Received: January 4, 2008

Dear Mr. Citron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Mr. Mark Citron

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Blymmmon for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

1. INDICATIONS FOR USE STATEMENT

Page 1 of 1

510(k) Number (if known):K063091
------------------------------------

Device Name: AIGISRXTM Anti-Bacterial Envelope

INDICATIONS/CONTRAINDICATIONS:

INDICATIONS FOR USE

AIGISRx is intended to securely hold a pacemaker pulse generator or defibrillator in order to create a stable environment when implanted in the body. AIGISRX contains the antimicrobial agents, rifampin and minocycline, which have been shown to reduce infection in an in-vivo model of bacterial challenge following surgical implantation of the generator or defibrillator. This device is only intended to be used in conjunction with pacemakers and implantable defibrillators.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Conpungente
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)

Division of Cardiovascular Devices
510(k) Number K063091

OR

Prescription Use (Per 21 CFR 801.109)

Over-The-Counter Use

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.