K Number
K063035

Validate with FDA (Live)

Date Cleared
2007-10-01

(367 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Transoma Medical Sleuth Implantable ECG Monitoring System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for:

  • patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or .
  • patients who experience transient symptoms that may suggest a cardiac arrhythmia.
Device Description

The Transoma Medical Sleuth Implantable ECG Monitoring System is designed to measure and store the patient s electrocardiogram (ECG) during symptomatic and asymptomatic events. The Sleuth System consists of the Model 2010 Implantable Monitoring Device (IMD), the Model 4000 Activator, and the Model 5000 Base Station (Bluetooth Modem). The IMD is an anatomically-shaped Litanium structure which is implanted subcutancously, typically in the left pectoral region. The Activator is a hand-held device carried by the patient.

The IMD continuously measures the R-R interval and automatically stores ECG segments when this interval falls outside of pre-defined limits. Using the Activator, the patient can also request storage of ECG segments when symptoms are experienced. The IMD stores a limited quantity of events, which are · ansferred to the Activator via telemetry throughout the day. On a daily basis, the Activator transfers events through the Base Station to a service center for review and analysis.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study details:

Note: The provided text is a 510(k) Summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting formal acceptance criteria and a detailed study report for meeting those criteria. Therefore, explicit "acceptance criteria" and "reported device performance" in a quantitative table format are not directly stated or fully detailed as they would be in a comprehensive study report. However, I will infer what can be gleaned about performance and success based on the summary.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Inferred from Study Goals)Reported Device Performance (Summary of Findings)
Diagnostic viability of ECG signalsECG waveform data was analyzable and categorizable by the cardiac monitoring center. "Diagnostic viability" was achieved, with 13 patients discovered to have arrhythmias, and Sleuth records aiding diagnosis.
Performance of the system (overall function)Successfully captured events initiated manually.
Successfully detected and recorded asymptomatic events automatically.
Successfully stored and transferred trending data.
Safety ProfileOccurrence of two adverse events in >4,700 patient days was "within expectation" and not serious or definitively linked to the Sleuth system.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 28 patients.
  • Data Provenance: Prospective clinical study conducted at 4 major medical centers in Panama.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document states, "The cardiac monitoring center has been able to analyze and categorize the ECG waveform data." and "Sleuth records were important in helping the physicians arrive at a firm diagnosis."

  • Number of Experts: Not explicitly stated. It refers to a "cardiac monitoring center" and "physicians." It's likely involved multiple clinical staff, but a specific count is not given.
  • Qualifications of Experts: Not explicitly stated, but inferred to be medical professionals at a "cardiac monitoring center" and "physicians" capable of diagnosing arrhythmias from ECG data.

4. Adjudication Method for the Test Set

Not explicitly stated. The document mentions a "cardiac monitoring center" analyzing and categorizing data and "physicians" using the records to arrive at a diagnosis. This suggests a clinical interpretation process, but no specific adjudication method (e.g., 2+1, 3+1) is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not described. The study focused on the performance of the Sleuth device itself, not on comparing human readers with and without AI assistance from the device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is an "Implantable ECG Monitoring System" that automatically detects and records events. While it automatically detects asymptomatic events, the "cardiac monitoring center" and "physicians" are involved in the "analysis and categorization" of data and "diagnosis." The clinical study focused on the system's performance in a real-world clinical setting with human involvement in the final diagnostic step. Therefore, it implicitly includes human-in-the-loop for diagnosis, even though the device performs automatic detection. A purely "standalone" algorithm-only performance evaluation, separate from human interpretation, is not explicitly detailed.

7. The Type of Ground Truth Used

The ground truth appears to be based on expert consensus/clinical diagnosis derived from the analyzed ECG waveforms. The statement "thirteen of the patients were discovered during monitoring to have arrhythmias, and the Sleuth records were important in helping the physicians arrive at a firm diagnosis" indicates that the physicians' final diagnoses, informed by the Sleuth data, served as the clinical ground truth for arrhythmia detection.

8. The Sample Size for the Training Set

The document does not provide information about a separate training set. The clinical study described with 28 patients appears to be a validation or performance study, not a training study for the algorithm. It's likely the device's algorithms were developed and potentially trained on other, unspecified data prior to this validation study.

9. How the Ground Truth for the Training Set Was Established

Since no information on a training set is provided, the method for establishing its ground truth is also not described in this document.

{0}------------------------------------------------

510(k) Summary

K063035
page 1/3

General Provisions

Submitter's Name and Address

Transoma Medical, Inc. 4358 West Round Lake Road Arden Hills, Minnesota 55112 Phone: 651-481-7410 Fax: 651-481-7415

Contact Person

Angela Raun Phone: 651-481-7460 Fax: 651-481-7415 Fmail: araun(@TransomaMedical.com

OCT 1 * 2007

Date of Summary

September 18, 2007

Proprietary Name of DeviceSleuth Implantable ECG Monitoring System
Common/Usual NameImplantable ECG Monitoring System
Classification NameCardiac Implantable Event RecorderProduct Code - MXCRegulation Number 21 CFR Part 870.2800Device Class II
Legally Marketed Device(s) toWhich Substantial Equivalenceis ClaimedMedtronic, Inc., Reveal® Plus InsertableLoop Recorder (ILR) System (K003667,K994331 and K972242)
Brentwood Medical Technology Corporation- IOmark™ Digital Holter, (K031466)

Device Description

The Transoma Medical Sleuth Implantable ECG Monitoring System is designed to measure and store the patient s electrocardiogram (ECG) during symptomatic and asymptomatic events. The Sleuth System consists of the Model 2010 Implantable Monitoring Device (IMD), the Model 4000 Activator, and the Model 5000 Base Station (Bluetooth Modem). The IMD is an anatomically-shaped Litanium structure which is implanted subcutancously, typically in the left pectoral region. The Activator is a hand-held device carried by the patient.

The IMD continuously measures the R-R interval and automatically stores ECG segments when this interval falls outside of pre-defined limits. Using the Activator, the patient can also request storage of ECG segments when symptoms are experienced. The IMD stores a limited quantity of events, which are · ansferred to the Activator via telemetry throughout the day. On a daily basis, the Activator transfers events through the Base Station to a service center for review and analysis.

{1}------------------------------------------------

Intended Use

The Transoma Medical Sleuth 2010 System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for:

KOG: 3035

page 2/3

  • patients with clinical syndromes or situations at increased risk of cardiac arrhythmias or .
  • patients who experience transient symptoms that may suggest a cardiac arrhythmia. 133035 .

Summary of Technological Characteristics

The Transoma Medical Sleuth 2010 System incorporates substantially equivalent technology. comparable features, labeling, and intended use, and is similar to the predicate devices currently available on the market.

Non-clinical Test Summary

The substantial equivalence of the Transoma Medical Sleuth 2010 System has been demonstrated via bench and animal testing. Testing included:

In vitro/ bench testing of the IMD, the Activator, the Base Station and the system In vivo canine testing Biocompatibility testing Software testing Electromagnetic compatibility testing Electrical safety testing Sterilization validation Shelf-life testing

Clinical Study Summary

A clinical study was conducted at 4 major medical centers in Panama. The primary objectives were to evaluate the diagnostic viability of the ECG signals and the performance of the system. Twenty-eight (28) patients were enrolled in the study. The patients had unexplained syncope/ pre-syncope or were at risk of arrhythmias. Patients were monitored continuously (24/7) by the Sleuth device for arrhythmic events. Clinic follow-ups were 7-10 days. 1, 3, and 6 months after implantation.

The cardiac monitoring center has been able to analyze and categorize the ECG waveform data. The Sleuth system has successfully captured events mitiated manually, as well as asymptomatic events which are detected and recorded automatically by the system. As designed, the Sleuth system has also successfully stored and transferred trending data.

The Sleuth ECG system has performed well following implantation in the initial 28 patients. The system has diagnostic viability, since thirtcen of the patients were discovered during monitoring to have arrhythmus, and the Sleuth records were important in helping the physicians arrive at a firm diagnosis. The occurrence of two adverse events within > 4,700 collective patient days of use is within expectation, and these events were neither serious, nor definitively shown to derive from the Sleuth system.

Performance Standards

Performance standards have not been established by the Food and Drug Administration for these devices under Section 514.

Conclusion (Statement of Equivalence)

Extensive bench and animal testing have demonstrated the Sleuth Model 2010 System functions in accordance with product specifications. Additionally, testing demonstrated the Sleuth System functions equivalently to comparative systems, i.e., the Brentwood IQmark Digital Holter (K031466) and the Medtronic Reveal Plus Insertable Loop Recorder (K003667, K994331 and K972242). These data

{2}------------------------------------------------

support a determination of substantial equivalence and subsequently market clearance of the Transoma Medical Sleuth System, comprised of the Model 2010 IMD, Model 4000 Activator, and Model 5000 Base Station.

Kax 3035
page 3/3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wing segments, creating a sense of movement or flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Transoma Medical, Inc. c/o Ms. Angie Raun Vice President Regulatory Affairs and Quality Assurance 4358 West Round Lake Road Arden Hills, MN 55112

2007

Re: K063035

Trade/Device Name: Sleuth Implantable ECG Monitoring System Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: MXC Dated: June 29, 2007 Received: July 2, 2007

Dear Ms. Raun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bfimmimorfor

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

Section 4: Indications for Use Statement

510(k) Number_ K 063035

Device Name: Transoma Medical Sleuth Implantable ECG Monitoring System

Indications for Use:

The Transoma Medical Sleuth Implantable ECG Monitoring System is an implantable, patient- and automatically-activated monitoring system that records subcutaneous ECG and is indicated for:

  • Patients with clinical syndromes or situations at increased risk of cardiac . arrhythmias
  • Patients who experience transient symptoms that may suggest a cardiac . arrhythmia

Prescription Use X (21 CFR 801 Subpart D)

OR

Over The Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

.. .--------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blymlamor

ardiovascular Devic

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).