K Number
K972242

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-01-16

(214 days)

Product Code
Regulation Number
870.2800
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic Model 9525 Reveal Insertable Loop Recorder is an implantable, patientactivated monitoring system that records subcutaneous ECG and is designed for diagnostic evaluation of patients who experience transient symptoms that may suggest a cardiac arrhythmia.

Device Description

The Reveal ILR system is designed to record and store subcutaneous electrocardiogram (ECG) during symptomatic events. The system consists of the Model 9525 implanted recorder and the Model 6190 Activator. A Medtronic Model 9790 programmer equipped with a Medtronic Model 9766A radio frequency telemetry head and Model 9802E software is required for programming and retrieving data from the recorder.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Medtronic Reveal™ Insertable Loop Recorder (K972242), based on the provided text:

Acceptance Criteria and Device Performance

The provided document doesn't explicitly state quantitative acceptance criteria in a dedicated section with specific numerical targets. Instead, it describes a series of tests performed to ensure the device meets its design and performance requirements, and then concludes with a statement of substantial equivalence to predicate devices. The "performance" is implicitly demonstrated by the successful completion of these tests.

Therefore, the table below will list the types of testing performed and the reported outcome.

Acceptance Criterion (Implicit)Reported Device Performance
In Vitro/Bench Testing:
Model 9525 implanted recorder testing (hybrid qualification, electromagnetic compatibility, mechanical, battery)Passed all in vitro requirements.
Model 6190 Activator testingPassed all in vitro requirements.
Package testingPassed all in vitro requirements. (Both package configurations were fully validated.)
In Vivo Canine Testing:
Comparison to surface ECG monitorData demonstrates the Reveal ILR system is safe and provides ECG information substantially equivalent to that provided by the surface monitor.
Comparison to competitive devices (Instromedix King of Hearts Express® 3X™, Instromedix HeartCard 3X™)Data demonstrates the Reveal ILR system is safe and provides ECG information substantially equivalent to that provided by the cardiac event recorders.
Biocompatibility:Not required as blood-contacting materials are the same as the predicate Medtronic Minix implantable pulse generator.
Sterilization Validation:A process appropriate for sterilizing the Model 9525 implanted recorder (using 100% Ethylene Oxide (EtO)) was validated.
Overall Performance (Substantial Equivalence):The testing provides reasonable assurance that the Reveal ILR system will perform as intended. Based on similarities in design, materials, in vitro test data, and canine in vivo electrical performance, Medtronic considers the Reveal ILR system to be substantially equivalent to the Instromedix King of Hearts Express 3X and the Medtronic Minix implantable pulse generator.

Study Information Details

The provided 510(k) summary focuses on demonstrating substantial equivalence through in vitro and in vivo animal testing, rather than a clinical trial with human subjects from a "test set" in the typical sense for algorithm performance.

  1. Sample size used for the test set and the data provenance:

    • In Vitro/Bench Testing: The sample sizes for these tests are not specified but would typically involve multiple units of each component (recorder, activator, packaging) tested under various conditions to ensure robustness and adherence to specifications. The data provenance is internal laboratory testing by Medtronic.
    • In Vivo Canine Testing: "Two comparative studies of the Reveal ILR system were performed with canines." The exact number of canines is not specified. The data provenance is animal (canine) studies, presumably conducted by Medtronic or a contracted research organization.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For the canine studies, the "ground truth" for ECG information was likely established by the "surface ECG monitor" used as a comparator device and interpreted by veterinary or medical professionals with expertise in cardiac electrophysiology, though their specific number and qualifications are not mentioned. For the bench testing, the ground truth would be engineering specifications and standards.
  3. Adjudication method for the test set:

    • Not applicable as the reported studies are primarily technical performance evaluations and animal comparative studies, rather than human clinical trials requiring complex adjudication of diagnostic outcomes.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This submission predates widespread AI integration in such devices and focuses on the fundamental function and safety of the recording system, not on human interpretation assistance.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The "Reveal™ Insertable Loop Recorder" acts as a recording system. Its "performance" is in accurately capturing and storing ECG data. The device itself (the implanted recorder) and its associated software for displaying waveforms do not appear to have an independent "algorithm" for automated diagnosis or interpretation. Its function is to provide raw ECG data for a clinician to analyze. Therefore, a standalone algorithm performance study, as understood in an AI context, is not applicable. The device's 'standalone' performance is its ability to record and store ECG signals reliably.
  6. The type of ground truth used:

    • In vitro/Bench Testing: Engineering specifications, industry standards, and functional requirements.
    • In vivo Canine Testing: The ground truth for the ECG information was established by simultaneously recorded data from a "surface ECG monitor" which serves as a established standard for cardiac electrical activity.
  7. The sample size for the training set:

    • Not applicable. This device is a diagnostic recording system, not a machine learning algorithm that requires a "training set" in the AI sense. Its development would involve engineering design, manufacturing standards, and rigorous testing on physical devices.
  8. How the ground truth for the training set was established:

    • Not applicable, as no training set for an AI algorithm was involved.

{0}------------------------------------------------

K972242

510(k) SUMMARY

Submitter

JAN | 6 |998

Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432

Contact: Nora K. Hadding, Product Regulation Manager Telephone: (612) 514-4189 Fax: (612) 514-6424 E-Mail: nora.hadding@medtronic.com

Date Prepared: June 13, 1997

Name of Device

Trade Name: "Reveal™ Insertable Loop Recorder" (ILR) System. The system is composed of the Model 9525 implanted recorder and the Model 6190 Activator. The Model 9802E Reveal software, Model 9790 programmer and Model 9766A telemetry head are also part of the system.

Common Name: insertable loop recorder

Classification Name: The Reveal Insertable Loop Recorder (ILR) system is a combination of existing devices. The classification names and numbers of these devices are:

  • "Telephone electrocardiograph transmitter and receiver" (870.2920) (Class II) .
  • "Implantable pacemaker pulse generator" (870.3610) (Class III) .

Predicate Devices

Table 1. Reveal ILR System - Predicate Devices

ProductFeatures Similar to the Reveal ILR System
Instromedix King of Hearts Express® 3X™ Model5328-00 (cardiac event recorder-looping memory)• Ambulatory event recorder• Continuous ECG recording• Looping memory• Pre- and post-event storage• Patient-activated
Medtronic Minix Model 8341M (single-chamberimplantable pulse generator)• Implantable• Dimensions• Subcutaneous electrode (in unipolar mode)• Programmable via a programmer unit• Transmit real time ECG waveform• Data retrieval (radio-frequency telemetry)

{1}------------------------------------------------

Device Description

System Description

The Reveal ILR system is designed to record and store subcutaneous electrocardiogram (ECG) during symptomatic events. The system consists of the Model 9525 implanted recorder and the Model 6190 Activator. A Medtronic Model 9790 programmer equipped with a Medtronic Model 9766A radio frequency telemetry head and Model 9802E software is required for programming and retrieving data from the recorder.

Model 9525 Implanted Recorder

The implanted recorder is a single-use, programmable device which is constructed of the same tissue-contacting materials used in Medtronic pacemakers. The implanted recorder is designed for long-term subcutaneous implantation. The implanted recorder is programmed using a Medtronic Model 9790 programmer. It continually records subcutaneous ECG into its looping memory and stores recorded ECG when signaled by the Activator.

Model 6190 Activator

The Activator is a hand-held, battery-operated telemetry device that communicates with the implanted recorder through the skin. It is used to signal the implanted recorder to store ECG recorded during a symptomatic event, for later retrieval and evaluation. The Activator is designed to be used during symptoms or immediately after symptoms. When the Activator is placed over the implanted recorder and the Activator button is pushed, the implanted recorder stores the ECG recorded during a pre-determined period of time before and after the Activator is used.

Stored ECG is retrieved via telemetry using a Medtronic programmer. The retrieved ECG can be viewed on the programmer screen, printed, or saved to disk. Stored ECG is erased and the recorder is restarted for continued use with a Medtronic programmer.

Programmer

Reveal Model 9802E software is loaded onto the Medtronic Model 9790 programmer. The programmer, with the Model 9766A radio frequency (RF) telemetry programming head, is used to communicate with the implanted recorder. The programmer uses radio frequency (RF) telemetry to transmit to and receive data from the implanted recorder and is used to program parameters of the implanted recorder, and to retrieve and display stored events for analysis.

Software

Reveal Model 9802E software is run from the Model 9790 programmer. It allows the clinician to program parameters of the recorder and to retrieve stored events for analysis. The clinician may select one of four modes for data storage. When the device memory is full, the software is used to erase stored events and restart recording for continued use.

When event(s) are stored in the implanted recorder's memory, the clinician may retrieve these event(s). The clinician can view the stored event on the programmer screen or print all or portions of the waveform to the programmer printer or to a full size external printer.

{2}------------------------------------------------

The events can also be saved on a diskette for later use. This allows the clinician to read the events from diskette at a later time when the patient and device are not present. At that time, events can be analyzed on the programmer screen or printed.

Packaging

Two package configurations are available. The ILR system package contains the implanted recorder, the Activator, and the product information manual. A replacement Activator is available in the other package configuration. Both package configurations were fully validated.

Intended Use

The Medtronic Model 9525 Reveal Insertable Loop Recorder is an implantable, patientactivated monitoring system that records subcutaneous ECG and is designed for diagnostic evaluation of patients who experience transient symptoms that may suggest a cardiac arrhythmia.

Technological Characteristics

The technology used with the Reveal ILR system is similar to several devices currently available on the market. The table on the following page outlines the functional similarities between the Reveal ILR system and the identified predicate devices.

Device FeatureReveal™ ILRInstromedixKing of HeartsMedtronicMinix™ IPG
ECG RecorderYesYesNo
Real-timeWaveformsYesNoYes
Looping MemoryYesYesNo
ImplantableYesNoYes
Pre and Post EventStorageYesYesNo
SubcutaneousElectrodeYesNoYes(in unipolar mode)
Patient-ActivatedYesYesNo
Implantable BatteryYesNoYes
Total Storage Time42 min.*5 min.No
Maximum SingleEvent Storage Time42 min.5 min.N/A
Storage Modes4 modesProgrammableN/A
Number of Events1 or 3Up to 60N/A
Data RetrievalRadio-frequencytelemetryTranstelephonicor audio-couplingRadio-frequency telemetry
Bandwidth0.85-32 Hz0.05-30 HzN/A
Sampling Rate100 Hz218 HzN/A
Volume8 ccN/A10.4 cc
Mass17 gN/A24.8 g
Dimensions61 x 19 x 8 mmN/A50x14x6 mm

*ECG waveforms stored in modes with a total storage time to increase storage time. Data sampled at 100Hz is stored to memory at 50Hz.

{3}------------------------------------------------

Summary of Studies

The following studies were performed to ensure that the Reveal ILR system meets all of its design and performance requirements.

In Vitro/Bench Testing

To evaluate the Reveal ILR system, the following in vitro testing was completed:

  • Model 9525 implanted recorder testing (hybrid qualification testing, electromagnetic . compatibility testing, mechanical testing, and battery testing)
  • Model 6190 Activator testing .
  • . Package testing

The Reveal ILR system passed all of the in vitro requirements.

In Vivo Canine Testing

Two comparative studies of the Reveal ILR system were performed with canines. One study compared the Reveal ILR system to a surface ECG monitor. The other study compared the Reveal ILR system to a surface monitor and two competitive devices currently available: the Instromedix King of Hearts Express® 3X™ and the Instromedix HeartCard 3X™ (cardiac event recorders). The data generated by these studies demonstrate that the Reveal ILR system is safe and provides ECG information which is substantially equivalent to that provided by the surface monitor and cardiac event recorders.

Biocompatibility Information

Biocompatibility testing was not required because blood-contacting materials of the Reveal ILR implanted device are the same as the Medtronic Minix implantable pulse generator.

Sterilization Validation

The Model 9525 implanted recorder is sterilized using a 100% Ethylene Oxide (EtO) sterilization process. A process appropriate for sterilizing the Model 9525 implanted recorder was validated.

Conclusion

The testing described above provides reasonable assurance that the Reveal ILR system will perform as intended when used in accordance with its labeling. Additionally, based on similarities in design, materials, in vitro test data and canine in vivo electrical performance, Medtronic considers the Reveal ILR system to be substantially equivalent to these pre-regulation devices: the Instromedix King of Hearts Express 3X, and the Medtronic Minix implantable pulse generator.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image is a simple, high-contrast, black and white image. A thin, dark line stretches horizontally across the upper portion of the frame. The rest of the image is a blank white space.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 23 1999

Ms. Nora K. Hadding Cardiac Pacing Business Medtronic, Inc. 7000 Central Avenue, N.E. Minneapolis, MN 55432-3576

Re: K972242 The Reveal™ Insertable Loop Recorder System including the Model 9525 Implantable Recorder and the Model 6190 Patient Activator Requlatory Class: II (two) Product Code: 74 MXC Dated: October 17, 1997 Received: October 20, 1997

Dear Ms. Hadding:

This letter corrects our substantially equivalent letter of January 16, 1998, regarding the Reveal™ Insertable Loop Recorder We believe that the product code and regulatory classification System. of the device should have been 74 MXC/II(two).

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS

{5}------------------------------------------------

Page 2 - Ms. Nora K. Hadding

inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to continue marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

DA Speyh

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Indications for Use

: ・ ・

510(k) Number (if known):N/AK972242
Device Name:Reveal TM Insertable Loop Recorder System
Indications For Use:The Medtronic Model 9525 Reveal TM Insertable Loop Recorder is an implantable, patient-activated monitoring system that records subcutaneous ECG and is indicated for patients who experience transient symptoms that may suggest a cardiac arrhythmia.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)Division of Cardiovascularand Neurological Devices510(k) Number
Prescription Use ✓(Per 21 CFR 801.109)OROver-The-Counter Use
Thomas J. Callahan

(Optional Format 1-2-96)

§ 870.2800 Medical magnetic tape recorder.

(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).