K Number
K062801

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-10-04

(15 days)

Product Code
Regulation Number
878.4780
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmartSuction Solo device is an AC-powered, stand-alone device designed to be used as a suction source to replace wall suction in the operating room to remove fluids and debris from the surgical field during surgical procedures. The SmartSuction Solo device is intended for use by trained personnel under the direction of the physician.

The SmartSuction Solo device is not intended for endotracheal suction. The SmartSuction Solo device automatically adjusts suction up to 150 mmHg negative pressure when the suction tip is occluded. Do not use SmartSuction Solo device as a suction source in any procedure where 150 mmHg of suction could damage underlying tissue.

This device is intended for use by trained personnel under the direction of a physician. The attending physician is responsible for the proper use of these devices. Therefore, the device is to be used as a prescription medical device, which is indicated in the labeling as "Rx only".

Device Description

Haemonetics® SmartSuction Solo™ (HSS) is a product line extension from the SmartSuction HARMONY™ device, which was cleared by FDA under K052626 in October 2005. The SmartSuction Solo device is an AC-powered, stand-alone device designed to be used as a suction source to replace wall suction in the operating room to remove fluids and debris from the surgical field during surgical procedures. The SmartSuction Solo device is intended for use by trained personnel under the direction of the physician. The attending physician is responsible for the proper use of these devices. Therefore, both these devices are to be used as a prescription medical device, which is indicated in the labeling as "Rx only".

Both electromechanical devices are the same and consist of a vacuum pump and control circuitry.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Haemonetics® SmartSuction Solo™ device. It establishes substantial equivalence to a predicate device, the SmartSuction HARMONY™. However, the document does not contain specific acceptance criteria or a detailed study proving the device meets acceptance criteria in the format explicitly requested.

The document states: "Performance Testing - Bench Haemonetics has conducted testing to verify the electrical safety and performance characteristics as described in the Operation Manual. A detailed list of testing is provided with test protocols and reports." This indicates that performance testing was conducted, but the specifics of what those characteristics were, what the acceptance criteria for them were, and the results of those tests are not included in the provided text.

Therefore, I cannot fully complete the requested table and answer all questions directly from the provided input. However, I can extract the available information and highlight what is missing.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Hypothesized based on general device function)Reported Device Performance (as stated in the document)
Electrical Safety Standards Compliance"Haemonetics has conducted testing to verify the electrical safety..."
Performance Characteristics (e.g., suction pressure, flow rate)"...and performance characteristics as described in the Operation Manual." "The SmartSuction Solo device automatically adjusts suction up to 150 mmHg negative pressure when the suction tip is occluded."
Functional Equivalence to Predicate Device"Verification and validation testing has been completed on the SmartSuction Solo and provide valid scientific evidence to demonstrate the devices are functionally equivalent."

Missing Information:

  • Specific quantitative acceptance criteria (e.g., maximum allowable leakage current, minimum/maximum flow rates, precise suction pressure ranges beyond "up to 150 mmHg").
  • Detailed test results for each of the performance characteristics.

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of a clinical study with patients or human subjects. The performance testing mentioned is "Bench testing." Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) for a clinical study is not provided.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the document describes bench testing, not a study involving expert-established ground truth for a test set of data.


4. Adjudication method for the test set

This information is not applicable as the document describes bench testing, not a study requiring adjudication of a test set.


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is an AC-powered suction device, not an AI-assisted diagnostic tool.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical device, not an algorithm. Bench testing constitutes a "standalone" evaluation of the device's technical performance.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the bench testing, the "ground truth" would be established by engineering specifications, industry standards, and the predicate device's performance characteristics. The document states, "Haemonetics has conducted testing to verify the electrical safety and performance characteristics as described in the Operation Manual." This implies the operational manual and internal specifications serve as the "ground truth" against which the device's performance was measured.


8. The sample size for the training set

This information is not applicable. The document describes a physical medical device and bench testing, not a machine learning model requiring a training set.


9. How the ground truth for the training set was established

This information is not applicable. The document describes a physical medical device and bench testing, not a machine learning model requiring a training set.

{0}------------------------------------------------

Section 5 - 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.

Submitter

Haemonetics Corporation 400 Wood Road Braintree, MA. 02184-9114

Contact

Gabriel J. Muraca, Jr. RA Project Manager Haemonetics Corporation 355 Wood Rd. Braintree, MA. 02184-9114 T: 781-356-9553 F: 781-356-9974 Email: gmuraca@haemonetics.com

Device Name

Proprietary Name:Haemonetics® SmartSuction Solo™
Common Name:AC Powered Suction Device
Classification Name:AC Powered Suction Pump

Predicate Device

The predicate device is the SmartSuction HARMONY™ device, cleared under K052626 in October 2005.

Description

Powered suction pumps are described in FDA regulation, 21 CFR 878.4780, as:

"A powered suction pump is an AC-powered device intended to be used to remove infectious materials from wounds or fluids from patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter. The FDA classified the device as a class II medical device".

Page 1 of 2

Image /page/0/Figure/17 description: The image contains the number 4 at the top, followed by the word "HAEMONETICS" in bold, uppercase letters. Below that, the phrase "COMPANY CONFIDENTIAL" is written in a smaller, non-bold font. A horizontal line is above the word "HAEMONETICS".

{1}------------------------------------------------

Haemonetics® SmartSuction Solo™ (HSS) is a product line extension from the SmartSuction HARMONY™ device, which was cleared by FDA under K052626 in October 2005. The SmartSuction Solo device is an AC-powered, stand-alone device designed to be used as a suction source to replace wall suction in the operating room to remove fluids and debris from the surgical field during surgical procedures. The SmartSuction Solo device is intended for use by trained personnel under the direction of the physician. The attending physician is responsible for the proper use of these devices. Therefore, both these devices are to be used as a prescription medical device, which is indicated in the labeling as "Rx only".

Indications for Use:

The SmartSuction Solo device is an AC-powered, stand-alone device designed to be used as a suction source to replace wall suction in the operating room to remove fluids and debris from the surgical field during surgical procedures.

The SmartSuction Solo device is not intended for endotracheal suction. The SmartSuction Solo device automatically adjusts suction up to 150 mmHg negative pressure when the suction tip is occluded. Do not use the SmartSuction Solo device as a suction source for a procedure where 150 mmHg of suction could damage underlying tissue.

Both devices are intended for use by trained operating room personnel under the direction of a physician. The attending physician is responsible for the proper use of these devices. Therefore, both these devices are to be used as a prescription medical device, which is indicated in the labeling as "Rx only".

Performance Testing - Bench

Haemonetics has conducted testing to verify the electrical safety and performance characteristics as described in the Operation Manual. A detailed list of testing is provided with test protocols and reports.

Substantial Equivalence

The substantial equivalence of the SmartSuction Solo is supported by its similarities in intended use, technological characteristics, and performance as compared to the currently marketed SmartSuction HARMONY device. Both devices have technological characteristics. They are similar in design and materials of construction. Both electromechanical devices are the same and consist of a vacuum pump and control circuitry. Verification and validation testing has been completed on the SmartSuction Solo and provide valid scientific evidence to demonstrate the devices are functionally equivalent.

Date:August 17, 2006
------------------------

Gabriel J. Muraca, Jr. Regulatory Affairs Project Manager Haemonetics Corporation

Page 2 of 2

5

HAEMONETICS

COMPANY CONFIDENTIAL

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows a stylized black and white drawing of a bird in flight. The bird's body is represented by three parallel lines that curve and taper towards the tail. The head of the bird is small and pointed, and the wings are suggested by the curved lines that extend from the body. The overall design is simple and abstract, conveying a sense of movement and freedom.

()(') = 4 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Haemonetics Corporation % Mr. Tamas Borsai TUV Rheinland of North America. Inc 12 Commerce Road Newton. Connecticut 06470

Re: K062801

Trade/Device Name: Haemonetics® SmartSuction Solo™ Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction cup Regulatory Class: Class II Product Code: BTA Dated: September 18, 2006 Received: September 19, 2006

Dear Mr. Borsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your because be device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encroours) ve regions the Medical Device Amendments, or to commerce prof to May 20, 1778, accordance with the provisions of the Federal Food, Drug, de needs may been rever do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dierelove, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de rise to such additional controls. Existing major regulations affecting your device can may be subject to back as a saterial Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I Teaso be deviced that I Draination that your device complies with other requirements of the Act that I Drimas intates and regulations administered by other Federal agencies. You must or my 1 oderal statuates and sequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Furt 007), mooning (21 CFR Part 820), and if applicable, the electronic forth in the quality byevelles (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Tamas Borsai

This letter will allow you to begin marketing your device as described in your Section 510(k) i mis letter will and in your to be a finding of substantial equivalence of your device to a legally prematted predicate device results in a classification for your device and thus, permits your device to proceed to the market.

ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n Joa the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general micronal and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Section 4 - Indications for Use Statement

Indications for Use:

510(k) Number (if known): K062801

Device Name:

Haemonetics® SmartSuction Solo™

The SmartSuction Solo device is an AC-powered, stand-alone device designed to be used as a suction source to replace wall suction in the operating room to remove fluids and debris from the surgical field during surgical procedures. The SmartSuction Solo device is intended for use by trained personnel under the direction of the physician.

The SmartSuction Solo device is not intended for endotracheal suction. The SmartSuction Solo device automatically adjusts suction up to 150 mmHg negative pressure when the suction tip is occluded. Do not use SmartSuction Solo device as a suction source in any procedure where 150 mmHg of suction could damage underlying tissue.

This device is intended for use by trained personnel under the direction of a physician. The attending physician is responsible for the proper use of these devices. Therefore, the device is to be used as a prescription medical device, which is indicated in the labeling as "Rx only".

Prescription Use (Part 21 CFR 801 Subpart D)Xand/orOver-The-Counter Use (21 CFR 807 Subpart C)
------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

Page 1 of 1

COMPANY CONFIDENTIAL

510(k) NumberK062801
------------------------

HAEMONETICS

§ 878.4780 Powered suction pump.

(a)
Identification. A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.(b)
Classification. Class II.