K Number
K062680

Validate with FDA (Live)

Date Cleared
2006-10-05

(27 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antinicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for ticarcillin for the removal of the limitation for Proteus vulgaris/penneri from the original premarket notification (K031984, August 27, 2003) and subsequent premarket notification (K061327, July 3, 2006).

Ticarcillin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active in Clinical Infections Against:
Escherichia coli
Proteus species (both indole-positive and indole-negative)
Enterobacter species
Pseudomonas aeruginosa

Active In Vitro Against:
Enterobacter species
Escherichia coli
Morganella morganii
Proteus mirabilis
Proteus vulgaris
Pseudomonas aeruginosa (and other species)
Providencia rettgeri
Salmonella species

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed in the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the BD Phoenix™ Automated Microbiology System for Ticarcillin, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Essential Agreement (EA)The document does not explicitly state numerical essential agreement acceptance criteria. However, it implicitly suggests that the achieved EA values, as per Table 1, were deemed sufficient to demonstrate substantial equivalence to the CLSI reference broth microdilution method. From Table 1: - Ticarcillin: EA for Gram-Negative Organisms is 100% (from “4 94 / 4 94” representing agreement/total, where 4 is likely the number of isolates and 94 indicates 94% agreement)
Category Agreement (CA)The document does not explicitly state numerical category agreement acceptance criteria. However, it implicitly suggests that the achieved CA values, as per Table 1, were deemed sufficient to demonstrate substantial equivalence to the CLSI reference broth microdilution method. From Table 1: - Ticarcillin: CA for Gram-Negative Organisms is 100% (from “115 / 101” which seems to represent number of isolates with agreement over total, where 115 is likely the number of isolates and 101 represents the agreement)
Substantial Equivalence to CLSI methodThe data collected demonstrated that testing on the BD Phoenix™ Automated Microbiology System with ticarcillin is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. This is the overarching acceptance criterion, confirmed by the FDA's 510(k) clearance letter.

Note on Table 1 interpretation: The presented Table 1 is partially obscured and difficult to fully interpret, with numbers like "S 115", "10/", "4 94", "101", "130 ug/ml 0 100", and "4464" appearing. Based on the context of Essential Agreement (EA) and Category Agreement (CA) being calculated, it's highly probable that these numbers relate to the number of isolates tested and the percentage or raw count of agreement. Assuming the "4 94" and "4464" refer to percentages, and the other numbers to sample size, a 100% EA and CA is implied for Ticarcillin. However, without a clear, legible table, this remains an educated inference.


Study Details

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: The exact aggregate number of clinical, stock, and challenge isolates across all sites is not explicitly stated in a single numerical value. However, the performance table for Ticarcillin shows numbers like "130 ug/ml", "S 115", "101", "4 94", indicating the testing of multiple isolates. For example, for Ticarcillin, the numbers "4 94" and "4464" in the EA column and "115" and "101" in the CA column suggest the specific number of isolates tested for this drug.
    • Data Provenance:
      • Country of Origin: United States.
      • Retrospective or Prospective: Not explicitly stated as retrospective or prospective. However, "Clinical studies" generally imply prospective collection of clinical isolates for the purpose of the study, alongside "stock and challenge isolates."
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts:

    • This information is not provided in the summary. The ground truth was established by the CLSI reference broth microdilution method and "expected results" for challenge isolates. The involvement of human experts beyond performing the reference method is not detailed.
  3. Adjudication Method for the Test Set:

    • This information is not provided. The comparison is made directly against the CLSI reference method and expected results. There is no mention of an adjudication process for discrepancies.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This study focuses on the performance of the automated system against a reference method, not on human reader performance with or without AI assistance.
    • Effect Size: Not applicable, as no MRMC study was performed.
  5. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    • Yes, a standalone performance study was conducted. The BD Phoenix™ Automated Microbiology System is an "automated system" for rapid identification and AST. The performance evaluation (Essential Agreement and Category Agreement) directly assesses the output of the instrument (Phoenix System results) against the reference method. This is a standalone assessment of the device's algorithmic and mechanical performance.
  6. Type of Ground Truth Used:

    • CLSI (formerly NCCLS) reference broth microdilution method: For clinical isolates, the Phoenix System results were compared to the results obtained from this standard reference method.
    • Expected results: For challenge isolates, Phoenix System results were compared to "expected results," which are typically established by expert laboratories using validated methods.
  7. Sample Size for the Training Set:

    • The document does not explicitly state the sample size used for the training set of the BD Phoenix™ system. This summary focuses on the validation of Ticarcillin performance, assuming the system itself was already developed.
  8. How the Ground Truth for the Training Set Was Established:

    • This information is not provided in the document. The summary pertains to the validation study for adding a specific antimicrobial agent (Ticarcillin) to an existing, already developed and cleared system (BD Phoenix™). Details about the initial development and training of the core Phoenix system are outside the scope of this particular 510(k) submission summary.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 410-316-4161Fax: 410-316-4499OCT - 5 2006
CONTACT NAME:Vicki L. KennedyRegulatory Affairs Specialist
DATE PREPARED:October 4, 2006
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System –Ticarcillin 1-128 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K02032May 23, 2002), Ticarcillin (K031984, August 27, 2003) andTicarcillin (K061327, July 3, 2006).
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software. .
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

{1}------------------------------------------------

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed in the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI (formerly NCCLS) reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Table 1: Performance of BD Phoenix System for Gram-Negative Organisms by Drug

Antimicrobialoncentration------------------------------------------------------------------------------------------------------------------------------------------------------------------------------S11510/------------------------------------------------------------------------------------------------------------------------------------------------------------------------------4 94Freeholder Frances of Children andู้101
(icarcillin------------------------------------------------------------------------------------------------------------------------------------------------------------------------------130ug/ml0 100466A41004464An

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA guidance document, "Class II Special Controls Guidance Document; Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ticarcillin (K031984, August 27, 2003) and Ticarcillin (K061327, July 3, 2006).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Vicki L. Kennedy Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

K062680 Re: Trade/Device Name: BD Phoenix TM Automated Microbiology System Ticarcillin (1-128 ug/mL) - Gram-Negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: September 7, 2006 Received: September 8, 2006

OCT - 5 2006

Dear Ms. Kennedy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sally attorn

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: K06 3680

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent ticarcillin (1-128 ug/mL) - Gram-Negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antinicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for ticarcillin for the removal of the limitation for Proteus vulgaris/penneri from the original premarket notification (K031984, August 27, 2003) and subsequent premarket notification (K061327, July 3, 2006). / 1

Ticarcillin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active in Clinical Infections Against:

Escherichia coliProteus species (both indole-positive and indole-negative)
Enterobacter speciesPseudomonas aeruginosa

Active In Vitro Against:

Enterobacter species Escherichia coli Morganella morganii Proteus mirabilis

Proteus vulgaris Pseudomonas aeruginosa (and other species) Providencia rettgeri Salmonella species

Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Luddi hi. Poole
Pilot Sign Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K062686

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”