K Number
K061327

Validate with FDA (Live)

Date Cleared
2006-07-03

(52 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional organism groups and Amoxicillin-clavulanate (0.5/0.25 – 32/16 µg/mL), Ampicillin-sulbactam (0.5/0.25-32/16 µg/mL), and Ticarcillin (1-128 µg/mL), on the BD Phoenix Automated Microbiology System.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is assessed using Essential Agreement (EA) and Category Agreement (CA) when compared to the CLSI reference broth microdilution method. The document refers to the FDA Draft Guidance Document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, February 5, 2003" for defining acceptance criteria. While specific numerical acceptance criteria (e.g., minimum percentage for EA or CA) are not explicitly stated in the provided text, the phrase "substantially equivalent performance" implicitly indicates that the reported values must meet or exceed a predefined threshold set by the FDA guidance.

Performance MetricAcceptance Criteria (Implied by FDA Guidance)Reported Device Performance (Amoxicillin-clavulanate)Reported Device Performance (Ampicillin-sulbactam)Reported Device Performance (Ticarcillin)
Essential Agreement (EA)"Substantially equivalent performance" (as per FDA guidance)96.7%97.2%94.7%
Category Agreement (CA)"Substantially equivalent performance" (as per FDA guidance)90.9%87.5%92.7%
Intra-site ReproducibilityGreater than 90%> 90% (for each antimicrobial agent)> 90% (for each antimicrobial agent)> 90% (for each antimicrobial agent)
Inter-site ReproducibilityGreater than 95%> 95% (for each antimicrobial agent)> 95% (for each antimicrobial agent)> 95% (for each antimicrobial agent)

2. Sample Size Used for the Test Set and Data Provenance

The test set included clinical, stock, and challenge isolates.

  • Amoxicillin-clavulanate: 2249 isolates
  • Ampicillin-sulbactam: 1305 isolates
  • Ticarcillin: 2882 isolates

The data provenance is described as "across multiple geographically diverse sites across the United States." The study involved prospective collection as it describes testing isolates with the Phoenix system and comparing them to reference methods.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used or their qualifications for establishing the ground truth. It refers to the CLSI reference broth microdilution method as the ground truth, which is a standardized laboratory procedure, not an expert consensus in the human-in-the-loop sense.

4. Adjudication Method for the Test Set

No explicit adjudication method (like 2+1 or 3+1) is mentioned. The comparison is made directly between the BD Phoenix System results and the results obtained from the CLSI reference broth microdilution method (for clinical isolates) or "expected results" (for challenge isolates). This implies a direct comparison rather than an adjudicated consensus.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This device is an automated system for antimicrobial susceptibility testing, not an imaging device requiring human interpretation, so the concept of human readers improving with or without AI assistance does not apply in this context.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, a standalone performance study was done. The entire study for Essential Agreement (EA) and Category Agreement (CA) directly compares the automated Phoenix System's results against the reference method. The "Clinical Studies" section clearly states: "Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method." This is a direct comparison of the automated system's performance without human intervention in the interpretation phase for these metrics.

7. The Type of Ground Truth Used

The primary type of ground truth used was the CLSI reference broth microdilution method. For "challenge" isolates, "expected results" were used, which would also be based on established laboratory testing standards.

8. The Sample Size for the Training Set

The document does not provide information on the sample size for any training set. This is common for older 510(k) submissions, especially for systems where the "algorithm" is based on established biochemical reactions and growth curves rather than machine learning models that require explicit training datasets.

9. How the Ground Truth for the Training Set Was Established

As no training set information is provided, how its ground truth was established is not discussed in this document. The system's performance is evaluated against the CLSI reference method as the gold standard for AST.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the text "K061327" in a handwritten style, likely a code or identifier. Below this, the phrase "CONFIDENTIAL AND PROPRIETARY" is printed in a clear, block-letter font. The text suggests the document or item associated with the code is private and protected.

510(K) SUMMARY

JUL - 3 2006

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410) 316-4161Fax: 410-316-4499
CONTACT NAME:Vicki L. KennedyRegulatory Affairs Specialist
DATE PREPARED:May 11, 2006
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System –Amoxicillin-clavulanate (0.5/0.25 – 32/16 µg/mL), Ampicillin-sulbactam (0.5/0.25-32/16 µg/mL), and Ticarcillin(1-128 µg/mL)
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Amoxicillin-clavulanate (K031912,July 21, 2003), Ampicillin-sulbactam (K043389,January 21, 2005), and Ticarcillin (K031984, August 27, 2003)
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .

{1}------------------------------------------------

  • BD Phoenix AST Broth used for performing AST tests only. .
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional organism groups with Amoxicillin-clavulanate (0.5/0.25 -32/16 µg/mL, Ampicillin-sulbactam (0.5/0.25-32/16 µg/mL), and Ticarcillin- 1-128 µg/mL (K031984, August 27, 2003) and the BD Phoenix™ Automated Microbiology System.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and ansociated reagents.

The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for ChalJewee wet isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Table 1:Performance of BD Phoenix System by Drug
----------------------------------------------------
AntimicrobialConcentrationEA (n)EA (%)CA (n)CA (%)
Amoxicillin-clavulanate0.5/0.25 – 32/16 µg/mL224996.7224990.9
Ampicillin-sulbactam0.5/0.25 – 32/16 µg/mL130597.2130587.5
Ticarcillin1 – 128 µg/mL288294.7288292.7

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N505 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Amoxicillin-clavulanate (K031912, July 21, 2003), Ampicillin-sulbactam (K043389, January 21, 2005), and Ticarcillin (K031984, August 27, 2003)

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Re:

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Vicki L. Kennedy Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

K061327 Trade/Device Name: BD Phoenix™ Automated Microbiology System Amoxicillin-clavulanate (0.5/0.25 - 32/16 ug/mL), Ampicillinsulbactam (0.5/0.25 - 32/16 µg/mL), and Ticarcillin (1-128 ug/mL), - Gram-negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: May 11, 2006 Received: May 12, 2006

JUL - 3 2006

Dear Ms. Kennedy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the idence) ons for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendonates, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Or ug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the I he general controls provisions of the Act include requirements for annual registration, lusting of devices, good manufacturing practice, labeling, and prohibitions against misbaranting and adulteration.

If your device is classified (sce above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition FFA may publish further announcements concerning your device in the Federal 07-2. ' w a

Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. Your must comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing pratis c requirements as set forth in the quality systems (QS) regulation (21 CFR Parts process

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of Intel Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Pan 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Sale, a 100%

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 2

510(k) Number: K061327

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Amoxicillin-clavulanate (0.5/0.25 – 32/16 µg/mL), Ampicillin-sulbactam (0.5/0.25-32/16 µg/mlgc. and Ticarcillin (1-128 µg/mL), – Gram-negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional organism groups and Amoxicillin-clavulanate (0.5/0.25 – 32/16 µg/mL), Ampicillin-sulbactam (0.5/0.25-32/16 µg/mL), and Ticarcillin (1-128 µg/mL), on the BD Phoenix Automated Microbiology System.

Amoxicillin-clavulanate has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Active In Vitro Against:

Enterobacter species Escherichia coli Klebsiella species

Proteus mirabilis

Ampicillin-sulbactam has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Beta-lactamase producing strains of: Escherichiu coli Klebsiella spp. (including K. pneumoniae)

Active In Vitro Against:

Escherichia coli (Beta-lactamase and non-beta-lactamase producing) Klebsiella species (all known strains are beta-lactamase producing) Morganella morganii Providencia rettgeri Providencia stuarti Division Sign-Off

BD Diagnostic Systems Becton. Dickinson and Company

Page S

Office of In Vitro Diagnostic Device Evaluation and Sufery

{6}------------------------------------------------

Page 2 of 2

Ticarcillin has been shown to be active in virro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agence

Active In Vitro Against:

Pseudomonas aeruginosa (and other species) Escherichia coli Proteus mirabilis Morganella morganii

Providencia rettgeri Enterobacter species Salmonella species

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Fuddi Pacci

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K061327

pg. 2 of 2

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”