K Number
K061703

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-08-02

(47 days)

Product Code
Regulation Number
862.1150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CHEM 2 CAL is an in vitro diagnostic product for the calibration of Phosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG) methods on the Dimension Vista™ System.

Device Description

CHEM 2 CAL is a liquid, multi-analyte, bovine serum albumin based product containing phosphorus, salicylate and glycerol. The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.0 mL.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Dimension Vista™ System Chemistry 2 Calibrator (CHEM 2 CAL - KC120), based on the provided text:

Acceptance Criteria and Device Performance

The primary performance characteristic detailed for the CHEM 2 CAL is its stability. The acceptance criterion for stability is a percent change over time less than or equal to 5% when comparing results of the product stored at 4°C with a control stored at -20°C.

Acceptance CriteriaReported Device Performance
Stability
Shelf-life: % change ≤ 5%Assignment at commercialization reflects real-time data on file, implying the 5% criterion is met.
On-board (punctured) stability: 7 daysNot explicitly stated if the 5% criterion was met for 7 days, but implies stability for this period.
Open vial (recapped, refrigerated) stability: 30 daysNot explicitly stated if the 5% criterion was met for 30 days, but implies stability for this period.
TraceabilityTraceable to NIST SRM 2186I (PHOS), Sodium Salicylate ACS grade (SAL), Glycerol Anhydrous ACS grade (TRIG).
Bottle Value AssignmentFinal bottle assignment for each commercial lot tested with N=45 replicates.

Study Details

  1. Sample size used for the test set and the data provenance:

    • Test set sample size: For "Bottle Value Assignment," N=45 replicates were tested for each level of the commercial lot.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data is retrospective or prospective. Given it's a submission to the FDA, the testing would likely be performed in a controlled laboratory environment, potentially in the US or at the manufacturer's facility (Dade Behring Inc., Newark, DE). The stability testing involves comparing current product results to controls, suggesting prospective testing over time.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device is a calibrator, and its "ground truth" relates to its chemical composition and the assigned values. The ground truth (traceability) for constituent concentrations is established through reference materials and primary standards, not human expert consensus.
    • PHOS: NIST SRM 2186I (National Institute of Standards and Technology - Standard Reference Material).
    • SAL: Sodium Salicylate ACS grade (American Chemical Society).
    • TRIG: Glycerol Anhydrous (ACS grade).
    • The "experts" in this context would be the organizations defining these standards (NIST, ACS) and the analytical chemists performing the value assignments and verification using primary standards.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for subjective clinical assessments where multiple human readers interpret images or data. For a chemical calibrator, validation relies on objective analytical measurements against traceable standards.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a chemical calibrator for in-vitro diagnostic instruments, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a physical calibrator material, not an algorithm. The "performance" is its ability to accurately calibrate an instrument, which is an inherent property of the material and its assigned values. The instrument itself performs the analytical measurement in a standalone manner.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the calibrator's value assignments is based on traceability to internationally recognized chemical standards (NIST SRM, ACS grade materials) and verification using primary standards.
  7. The sample size for the training set:

    • Not applicable. This is a physical calibrator, not a machine learning model that requires a training set. The "development" involves manufacturing to precise specifications and assigning values based on analytical measurements, not training an algorithm.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this device. The chemical composition and value assignments are based on the manufacturing process and analytical validation against established chemical standards. The stock solution concentrations are "verified using an instrument calibrated with primary standards," and the final bottle assignments are tested against these.

{0}------------------------------------------------

K061703

AUG 0 2 2006

510(k) Summary for the Dimension Vista™ System Chemistry 2 Calibrator (CHEM 2 CAL - KC120)

A. 510(k) Number:

B. Analytes:Phosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG)
C. Type of Test:Calibrator Material
D. Applicant:Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101Victor M. Carrio, Regulatory Affairs and Compliance ManagerOffice: (302) 631-0376 Fax: (302) 631-6299

E. Proprietary and Established Names:

Dimension Vista™ System Chemistry 2 Calibrator (CHEM 2 CAL - KC120)

F. Regulatory Information:

    1. Regulation section: 21 CFR § 862-1150 Calibrator
    1. Classification: Class II
    1. Product Code: JIX Calibrator, Multi-Analyte Mixture
    1. Panel: Clinical Chemistry
  • G. Intended Use: The CHEM 2 CAL is an in vitro diagnostic product for the calibration of Phosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG) methods on the Dimension Vista " System.

H. Device Description:

CHEM 2 CAL is a liquid, multi-analyte, bovine serum albumin based product containing phosphorus, salicylate and glycerol. The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 2.0 mL.

{1}------------------------------------------------

ItemDeviceDimension Vista™ SystemChemistry 2 CalibratorPredicatesChemistry II CalibratorK861700PredicatesSalicylate CalibratorK904307
Intended UseThe CHEM 2 CAL is an in vitro diagnostic product for the calibration of Phosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG) methods on the Dimension Vista™ System.CHEM II Calibrator is an in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the magnesium (MG), phosphorus (PHOS) and triglycerides (TRIG) methods.The Salicylate Calibrator is an in vitro diagnostic product to be used to calibrate the Dimension® clinical chemistry system for the Salicylate (SAL) method.
AnalytesPhosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG).Phosphorus (PHOS) and Triglycerides (TRIG).Salicylate (SAL).
FormLiquid.Liquid.Liquid.
TraceabilityPHOS – NIST SRM 2186I.SAL – Sodium Salicylate ACS grade.TRIG - Glycerol Anhydrous ACS grade.PHOS - NIST SRM 2186I.TRIG - Glycerol Anhydrous ACS grade.SAL - Sodium Salicylate ACS grade.
MatrixBovine serum albumin based product.Pure magnesium dissolved in a dilute solution of HCL, reagent grade potassium dihydrogen phosphate and reagent grade glycerol.Aqueous solution of ACS grade sodium salicylate.
Number ofLevelsTwo levels.Three levels.Three levels.

:

·

:

Substantial Equivalence Information 'I

{2}------------------------------------------------

J. Standard/Guidance Document Referenced:

    1. Guidance: Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final, 02/22/1999. Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, 11/30/2004.
    1. Standards: CEN 13640 Stability testing of In-Vitro Diagnostic Devices. ISO 14971:2000 Medical devices - Application of risk management to medical devices.

K. Performance Characteristics:

1. Stability:Target shelf life for the Dimension Vista™ Chemistry 2 Calibratoris 12 months. Calibrator shelf life is determined by comparingresults of the product stored at 4°C with control stored at -20°C.The method is calibrated from this stored material. The 4°Cmaterial values are recovered versus the calibration. Recoveryversus time is monitored and percent change over time isdetermined. Percent change should be less than or equal to 5%.Shelf-life stability (expiration) dating assignment atcommercialization reflects the real-time data on file at DadeBehring, Inc.A vial punctured by the instrument and stored on board is stablefor 7 days.An open vial not on instrument, but recapped and stored in arefrigerator is stable for 30 days.For testing, vials are opened /punctured on day zero. A quantitysufficient for multiple calibrations is removed and the vials arerecapped and stored at 2 – 8 °C. Opened/punctured vials are testedon days 1, 8, 15, 22, 31, and 32 versus freshly opened vials.
2. Traceability:The assigned values of the Chemistry 2 Calibrator are traceable to:
ConstituentTraceability
PHOSNISTa SRM 2186I
SALSodium Salicylate ACSb grade
TRIGGlycerol Anhydrous (ACS grade)
  • NIST-SRM: National Institute of Standards and Technology- Standard a. Reference Material.
  • b. ACS: American Chemical Society.

{3}------------------------------------------------

    1. Bottle Value Assignment:
      The new calibrator is made by adding weighed in aqueous solutions of potassium dihydrogen phosphate, sodium salicylate, and glycerol reference materials to a stock solution at target concentrations. The concentration is verified using an instrument calibrated with primary standards. The stock solutions are added to base matrix in appropriate concentrations and verified using an instrument calibrated with primary standards. The final bottle assignment for each level of the commercial lot is tested N = 45 replicates, with multiple reagent lots on multiple instruments.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circle around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Victor M. Carrio RA/QS Compliance Manager Dade Behring, Inc. 500 GBC Drive PO Box 6101, Mail Stop 514 Newark, DE 19714-6101

AUG 0 2 2006

Re: K061703

Trade/Device Name: CHEM 2 CAL Regulation Number: 21 CFR§ 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: June 15, 2006 Received: June 16, 2006

Dear Mr. Carrio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbanding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, percoits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please not the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 80.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolli-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Alberto Guti

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K061703

Device Name: CHEM 2 CAL

Indications For Use:

The CHEM 2 CAL is an in vitro diagnostic product for the calibration of Phosphorus (PHOS), Salicylate (SAL), and Triglycerides (TRIG) methods on the Dimension Vista™ System.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

CDRH, Office of I

Page 1 of __ 1

§ 862.1150 Calibrator.

(a)
Identification. A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.