K Number
K061588

Validate with FDA (Live)

Date Cleared
2006-07-21

(43 days)

Product Code
Regulation Number
862.1377
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only. VITROS Chemistry Products HCY Reagent is used to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma. Serum and plasma homocysteine levels can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 27 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).

For in vitro diagnostic use only. VITROS Chemistry Products HCY Performance Verifiers are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5,1 FS Chemistry Systems.

Device Description

The VITROS Chemistry Products HCY Reagent is used in conjunction with the VITROS Chemistry Products Calibrator Kit 27 and VITROS Chemistry Products FS Diluent Pack 2 (BSA/Saline) on VITROS 5,1 FS Chemistry Systems to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma.
The quantitative measurement of homocysteine (HCY) is performed using the VITROS Chemistry Products HCY Reagent in conjunction with the VITROS Chemistry Products Calibrator Kit 27 and on the VITROS 5,1 FS Chemistry Systems. The VITROS Chemistry Products HCY Reagent consists of two dual chambered reagent packs containing three ready-to- use liquid reagents. Disulfide linked homocysteine (oxidized forms) in the sample is reduced by Tris (2-Carboxyethyl) phosphine hydrochloride (TCEP) to form reduced homocysteine. Reduced homocysteine reacts with serine in the presence of cystathionine ß-synthase (CBS) to form L-cystathionine. L-cystathionine is broken down by cystathionine ß-lyase (CBL) to produce homocysteine, pyruvate and ammonia. Pyruvate is reduced to lactate by lactate dehydrogenase (LDH) using NADH as coenzyme. The concentration of HCY is directly proportional to the amount of NADH converted to NAD and is measured spectrophotometrically at 340 nm. Once a calibration has been performed, the HCY concentration in each unknown sample can be determined using the stored calibration curve and the measured absorbance obtained in the assay of the sample.

The VITROS Chemistry Products Calibrator Kit 27 are prepared from an aqueous solution containing amino acids and inorganic acid. These standards are used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).

The VITROS Chemistry Products HCY Performance Verifiers I, II and III are prepared from processed human serum to which amino acid and preservative have been added. These are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5.1 FS Chemistry Systems.

The VITROS Chemistry Products FS Diluent Pack 2 (Saline/BSA) is a common reagent that is used by multiple assays on the VITROS 5,1 FS Chemistry System. This is a dual chambered package containing two ready-to-use liquid diluents. Diluent 1 is prepared from processed water to which inorganic salt has been added. Diluent 2 is prepared from processed water to which bovine serum albumin, inorganic salts and preservatives have been added.

The VITROS 5.1 FS Chemistry System is a clinical chemistry instrument that provides automated use of the VITROS Chemistry Products MicroTip® and MicroSlides® range of products.

AI/ML Overview

The provided text describes the 510(k) summary for the VITROS Chemistry Products HCY Reagent, Calibrator Kit 27, and HCY Performance Verifiers. It states that the device is substantially equivalent to predicate devices and provides a summary of the studies performed. Here's a breakdown of the requested information based on the provided text:

1. Table of acceptance criteria and the reported device performance

The document does not explicitly present a table of "acceptance criteria" against which the device performance is directly measured in a pass/fail manner. Instead, it demonstrates substantial equivalence through correlation studies and other bench testing. The primary performance metric presented for substantial equivalence is the correlation with a predicate device.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Correlation with Predicate Device (Homocysteine Reagent)Strong correlation coefficient (likely >0.95 or similar for substantial equivalence) and a linear relationship close to y=x.y = 0.98 x + 1.0 umol/L, with a correlation coefficient = 0.97 (where y = VITROS HCY assay and x = BIO-RAD® HOMOCYSTEINE by HPLC test).
Device Similarities/Differences (HCY Reagent)Intended use, analyte, sample type, measurement type should be similar.Similarities: Intended Use, Analyte measured, Sample Type, Measurement Type (Quantitative). Differences: Reportable Range (1.0-50.0 umol/L vs 0.5-100 umol/L), Sensitivity (1.0 umol/L vs 0.5 µmol/L), Calibrator levels, format, and matrix, Instrumentation (automated clinical chemistry analyzer vs Isocratic HPLC System), Reference Interval, Method (Homogeneous Enzymatic vs Chromatography).
Device Similarities/Differences (Performance Verifiers)Intended use, matrix, product type, format, and number of levels should be similar.Similarities: Indications for Use, Matrix, Product Type (Assayed Control), Format (Liquid), Number of levels (Three). Differences: Analytes Reported (Homocysteine vs $\alpha_1$ -Antitrypsin).

Other bench testing was also performed, including:

  • Assay precision
  • Linearity
  • Specificity
  • Expected values
  • Limit of detection
  • Dilution
  • Specimen matrix

However, specific acceptance criteria and detailed results for these tests are not provided in this summary.

2. Sample sized used for the test set and the data provenance

  • Test Set Sample Size: The document mentions "patient samples" were used in the correlation studies but does not specify the exact number of samples.
  • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. It only mentions "patient samples."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The ground truth for the HCY assay's performance is established by comparison to a "commercially available system BIO-RAD HOMOCYSTEINE by HPLC test," which is itself a medical device. This implies that the BIO-RAD device's results are considered the reference or "truth" for this comparison, rather than requiring expert consensus on individual patient samples.

4. Adjudication method for the test set

This information is not applicable as the comparison is against an existing commercial assay, not a subjective interpretation by experts requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is a diagnostic assay (reagent, calibrator, and verifiers) for measuring homocysteine concentration. It is not an AI-based system designed to assist human readers (e.g., radiologists interpreting images), and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant or described.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, in essence. The entire study described is a standalone performance evaluation of the VITROS Chemistry Products HCY assay. The system (analyzer + reagents) measures HCY concentration autonomously. Its performance is then compared to an existing predicate device, not to human interpretation of raw data.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for the HCY assay was established through comparison to a legally marketed predicate device, the BIO-RAD® HOMOCYSTEINE by HPLC test (K993107). The results from this predicate device served as the reference standard for evaluating the new VITROS HCY assay's performance.

8. The sample size for the training set

The document does not specify a training set sample size. This is typical for a traditional in-vitro diagnostic assay rather than a machine learning or AI algorithm that requires explicit training data. The "training" of such a system involves the development and calibration of the assay itself, for which specific data sizes are usually proprietary or part of internal validation but not explicitly called out as a "training set" in the context of this 510(k) summary.

9. How the ground truth for the training set was established

As described above, the concept of a "training set" with separate ground truth establishment is not directly applicable in the context of this traditional IVD assay's 510(k) summary. The ground truth for evaluating the device's performance (its "test set") was the predicate device's results. The development and internal validation of the assay (which could be considered analogous to "training" in a very loose sense) would have involved extensive analytical studies to ensure its accuracy, precision, and other performance characteristics, likely using certified reference materials and well-characterized samples, but this is not detailed as a "ground truth for a training set" in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Ortho-Clinical Diagnostics. The logo consists of a symbol on the left, followed by the text "Ortho-Clinical Diagnostics". The text is underlined.

Johnson-Johnson company

100 Indigo Creek Drive Rochester, New York 14626-5101

510(k) Summary

JUL 2 1 2006

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:

  • Ortho-Clinical Diagnostics, Inc. 1. Submitter 100 Indigo Creek Drive name, address, Rochester, New York 14626-5101 contact (585) 453-3143 Contact Person: Michael Byrne
  • June 1, 2006 2. Preparation Date

Trade or Proprietary Names: 3. Device name

VITROS Chemistry Products HCY Reagent VITROS Chemistry Products Calibrator Kit 27 VITROS Chemistry Products HCY Performance Verifiers I, II & III

Common Names:

HCY assay and controls

Classification Names:

Urinary homocystine (nonquantitative) test system (21 CFR 862.1377) Class II Calibrators (21 CFR 862.1150) Class II Quality Control material (assayed and unassayed) (21 CFR 862.1660) Class I (general controls). Since these devices (HCY Performance Verifiers I, II & III) are assayed controls, they meet the reserved criteria under Section 510(I) of the Food, Drug, and Cosmetic Act.

Continued on next page

{1}------------------------------------------------

510(k) Summary (continued)

    1. Predicate The VITROS Chemistry Products HCY assay is substantially equivalent to the previously cleared BIO-RAD® HOMOCYSTEINE by HPLC test (K993107). Devices
      The VITROS Chemistry Products HCY Performance Verifiers are substantially equivalent to the previously cleared VITROS Chemistry Products AAT Performance Verifiers (K052819).
    1. Device The VITROS Chemistry Products HCY Reagent is used in conjunction with description the VITROS Chemistry Products Calibrator Kit 27 and VITROS Chemistry Products FS Diluent Pack 2 (BSA/Saline) on VITROS 5,1 FS Chemistry Systems to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma.
      The quantitative measurement of homocysteine (HCY) is performed using the VITROS Chemistry Products HCY Reagent in conjunction with the VITROS Chemistry Products Calibrator Kit 27 and on the VITROS 5,1 FS Chemistry Systems. The VITROS Chemistry Products HCY Reagent consists of two dual chambered reagent packs containing three ready-to- use liquid reagents. Disulfide linked homocysteine (oxidized forms) in the sample is reduced by Tris (2-Carboxyethyl) phosphine hydrochloride (TCEP) to form reduced homocysteine. Reduced homocysteine reacts with serine in the presence of cystathionine ß-synthase (CBS) to form L-cystathionine. L-cystathionine is broken down by cystathionine ß-lyase (CBL) to produce homocysteine, pyruvate and ammonia. Pyruvate is reduced to lactate by lactate dehydrogenase (LDH) using NADH as coenzyme. The concentration of HCY is directly proportional to the amount of NADH converted to NAD and is measured spectrophotometrically at 340 nm. Once a calibration has been performed, the HCY concentration in each unknown sample can be determined using the stored calibration curve and the measured absorbance obtained in the assay of the sample.

The VITROS Chemistry Products Calibrator Kit 27 are prepared from an aqueous solution containing amino acids and inorganic acid. These standards are used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).

The VITROS Chemistry Products HCY Performance Verifiers I, II and III are prepared from processed human serum to which amino acid and preservative have been added. These are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5.1 FS Chemistry Systems.

Continued on next page

{2}------------------------------------------------

The VITROS Chemistry Products FS Diluent Pack 2 (Saline/BSA) is a common reagent that is used by multiple assays on the VITROS 5,1 FS Chemistry System. This is a dual chambered package containing two ready-touse liquid diluents. Diluent 1 is prepared from processed water to which inorganic salt has been added. Diluent 2 is prepared from processed water to which bovine serum albumin, inorganic salts and preservatives have been added.

The VITROS 5.1 FS Chemistry System is a clinical chemistry instrument that provides automated use of the VITROS Chemistry Products MicroTip® and MicroSlides® range of products. The VITROS 5,1 FS System was cleared for market by 510(k) premarket notification (K031924).

VITROS Chemistry Products HCY Reagent : For in vitro diagnostic use 6. Device intended only. VITROS Chemistry Products HCY Reagent is used to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma. uses Serum and plasma homocysteine levels can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.

VITROS Chemistry Products Calibrator Kit 27: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 27 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).

VITROS Chemistry Products HCY Performance Verifiers I, II & III: For in vitro diagnostic use only. VITROS Chemistry Products HCY Performance Verifiers are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5,1 FS Chemistry Systems.

The VITROS Chemistry Products HCY Reagent and VITROS Chemistry 7. Comparison to predicate Products Calibrator Kit 27 are substantially equivalent to the BIO-RAD HOMOCYSTEINE by HPLC test (K993107) (predicate device), which was devices: cleared by the FDA for IVD use.

A least squares linear regression analysis demonstrated the following relationship: y = 0.98 x + 1.0 umol/L, with a correlation coefficient = 0.97 where y = results obtained using the VITROS Chemistry Products HCY assay and x = results obtained with the commercially available system BIO-RAD HOMOCYSTEINE by HPLC test in conventional/SI units (umol/L).

The VITROS Chemistry Products HCY Performance Verifiers I, II & III are substantially equivalent to the VITROS Chemistry Products AAT Performance Verifiers I, II & III (K052819) (predicate device) which was cleared by the FDA for IVD use.

In addition to correlation studies, bench testing was performed to determine assay precision, linearity, specificity, expected values, limit of detection, dilution and specimen matrix of the VITROS HCY assay.

Continued on next page

{3}------------------------------------------------

Table 1 Similarities and differences of the assays performed using the VITROS
HCY assay and the BIO-RAD HOMOCYSTEINE by HPLC test.
Device Similarities
DeviceVITROS HCY assay (New device)
CharacteristicBIO-RAD HOMOCYSTEINE by HPLC test (Predicate device)
Intended UseFor in vitro diagnostic use. Quantitative determination of total
Homocysteine in human plasma or serum.
Analyte measuredHomocysteine
Sample TypeSerum and plasma
Measurement TypeQuantitative
Device Differences
DeviceVITROS HCY assayBIO-RAD HOMOCYSTEINE
Characteristic(New device)by HPLC test
(Predicate device)
Reportable Range1.0-50.0 umol/L0.5 - 100 umol/L
Sensitivity1.0 umol/L0.5 µmol/L
Calibrator levelsTwo levels (0, 27 umol/L)Single level (15-20 umol/L)
Calibrator formatLiquidLyophilized
Calibrator matrixAqueous solution containing aminoHuman Serum
acids and inorganic acid
InstrumentationAutomated clinical chemistryIsocratic HPLC System
analyzer
Reference Interval6.6 - 14.8 umol/LMales:< 15 umol/L
4.7 - 12.6 umol/LFemales:
MethodHomogeneous EnzymaticChromatography

Table 2 Similarities and differences of the device characteristics between the VITROS HCY Performance Verifiers I, II & III with the predicate device VITROS AAT Performance Verifiers I, II & III

Device Similarities
Device CharacteristicVITROS HCY Performance Verifiers (New Device)VITROS AAT Performance Verifiers (Predicate Device)
Indications for UseFor in vitro diagnostic use. Assayed controls are used to monitor assay performance on VITROS 5,1 FS Chemistry Systems.
MatrixThe performance verifiers are prepared from processed human serum with preservatives added.
Product TypeAssayed Control
FormatLiquid
Number of levelsThree
Differences
Device CharacteristicVITROS HCY Performance Verifiers (New Device)VITROS AAT Performance Verifiers (Predicate Device)
Analytes ReportedHomocysteine (HCY)$\alpha_1$ -Antitrypsin (AAT)

{4}------------------------------------------------

    1. Conclusions The data presented in the premarket notification provide a reasonable assurance that the VITROS Chemistry Products HCY Reagent, VITROS Chemistry Products Calibrator Kit 27 and the VITROS Chemistry Products HCY Performance Verifiers are safe and effective for the stated intended uses and are substantially equivalent to the cleared predicate devices. Equivalence to the predicates was demonstrated using a commercially available assay along with patient samples.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is composed of three curved lines that resemble a stylized human figure. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JUL 2 1 2006

Mr. Michael M. Byrne Regulatory Associate Ortho-Clinical Creek Dr. 100 Indigo Creek Dr. Rochester, NY 14626

Re: K061588

Trade/Device Name: VITROS Chemistry Products HCY Regent VITROS Chemistry Products Calibrator Kit 27 VITROS Chemistry Products HCY Performance Verifiers I, II, III Regulation Number: 21 CFR8862.1377 Regulation Name: Urinary homocystine (nonquantitative) test system Regulatory Class: Class II Product Code: LPS, JIT, JJX Dated: June 7, 2006 Received: June 8, 2006

Dear Mr. Byrne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollefree number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Alberto Guts

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

Page 1 of 1

510(k) Number(if known):k 061588
Device Name:VITROS Chemistry Products HCY ReagentVITROS Chemistry Products Calibrator Kit 27VITROS Chemistry Products HCY Performance Verifiers I, II,and III
Indications forUse:For in vitro diagnostic use only. VITROS Chemistry ProductsHCY Reagent is used to quantitatively measure totalhomocysteine (HCY) concentration in human serum and plasma.Serum and plasma homocysteine levels can assist in the diagnosisand treatment of patients suspected of havinghyperhomocysteinemia and homocystinuria.
For in vitro diagnostic use only. VITROS Chemistry ProductsCalibrator Kit 27 is used to calibrate VITROS 5,1 FS ChemistrySystems for the quantitative measurement of homocysteine(HCY).
For in vitro diagnostic use only. VITROS Chemistry ProductsHCY Performance Verifiers are assayed controls used to monitorperformance of VITROS HCY Reagents on VITROS 5,1 FSChemistry Systems.
Prescription Use(Part 21 CFR 801 Subpart D)X AND/OR Over-The-Counter Use(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C. Bener
Revision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

K061583

§ 862.1377 Urinary homocystine (nonquantitative) test system.

(a)
Identification. A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.(b)
Classification. Class II.