K Number
K060284

Validate with FDA (Live)

Date Cleared
2006-05-10

(96 days)

Product Code
Regulation Number
864.7925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

cryocheck Clot APCR is a clotting assay intended to screen for resistance to activated protein C in citrated human plasma from individuals with the factor V Leiden mutation.

Device Description

cryocheck Clot APCR consists of:

  • 5 x 2.0 mL Activator Reagent (APC-AR) .
  • 5 x 4.0 mL Russell's Viper Venom Reagent (APC-RV)
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the cryocheck™ Clot APCR™ device, based on the provided 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or correlation. Instead, it frames the acceptance criteria implicitly as "comparable precision," "identical results when challenged with interfering substances and factor deficiencies," and "identical sensitivity and specificity" when compared to the predicate device. The primary performance metric reported is a correlation coefficient.

Acceptance Criteria (Implicit)Reported Device Performance (cryocheck Clot APCR)
Comparable precision to predicate device (GradiLeiden V)Showed comparable precision to GradiLeiden V.
Identical results with interfering substances/factor deficiencies as predicateDemonstrated identical results when challenged with interfering substances and factor deficiencies.
Identical sensitivity to predicate device (GradiLeiden V)Exhibited identical sensitivity to GradiLeiden V.
Identical specificity to predicate device (GradiLeiden V)Exhibited identical specificity to GradiLeiden V.
Strong correlation with predicate deviceA correlation of R = 0.960 was obtained (with GradiLeiden V).

2. Sample Size and Data Provenance

  • Sample Size for Test Set: 207 clinical samples.
  • Data Provenance: The document states that clinical tests were performed "at Precision BioLogic and a US university hospital-based clinical coagulation laboratory." This indicates the data is prospective clinical data collected from human subjects (patients). The clinical samples were from "the target population for the assay."

3. Number of Experts and Their Qualifications for Ground Truth

The document does not specify the number of experts used to establish ground truth or their qualifications. The comparison is made against a predicate device, implies that the predicate device's results are considered the reference standard, rather than independently established expert ground truth.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method for the test set. The comparison appears to be a direct one-to-one comparison of results between the new device and the predicate device for each sample.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. The study focuses on comparing the new device's performance to a predicate device, not on assessing human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

Yes, the study is a standalone performance study. The device itself is an in vitro diagnostic (IVD) assay that provides a result (clotting time ratio). The "performance" being evaluated is the accuracy and reliability of this assay in classifying samples compared to the predicate device. There is no human-in-the-loop component described for the operation of this specific device or its interpretation in the context of the study.

7. Type of Ground Truth Used

The "ground truth" for the test set was essentially the results obtained from the predicate device (GradiLeiden V). The study compared the cryocheck Clot APCR's performance to the GradiLeiden V results across the 207 clinical samples.

8. Sample Size for the Training Set

The document does not explicitly mention a "training set" or its size. Since this is an in vitro diagnostic assay rather than an AI/machine learning algorithm, the concept of a distinct training set in the typical machine learning sense does not apply. The development of the assay would involve R&D and validation, but not a "training set" in the computational context.

9. How Ground Truth for Training Set Was Established

As noted above, the concept of a training set as it applies to AI/ML is not relevant here. The "ground truth" for the development and validation of such an assay would typically be established through robust laboratory methods, perhaps using known Factor V Leiden positive and negative samples, and comparison to established clinical diagnostic methods, but specific details of this process for assay development are not provided in the 510(k) summary.

{0}------------------------------------------------

510(K) Summary cryocheck™ Clot APCR™

MAY 10 2006

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: HOLO2 8

Part A

    1. Submitter's name, address, telephone, contact person, and the date the summary was prepared
Submitter's Name:Precision BioLogic Inc.
Submitter's Address:900 Windmill Road, Suite 100Dartmouth, Nova Scotia B3B 1P7Canada
Submitter's Telephone Number:902-468-6422
Submitter's Contact Name:Stephen L. Duff - Director of New Business DevelopmentPhone: 902-468-6422 ext. 224Fax: 902-468-6421Email: sduff@precisionbiologic.com
Date of 510(k) Preparation:February 2, 2006
    1. Name of the device, including the trade or proprietary name, the common or usual name and the classification name
Device Name & Classification:CRYOcheck Clot APCR
Common Name:Activated Protein C Resistance (APCR) Screening Assay
Classification Name:Hematology; Test, Time, Partial ThromboplastinRegulatory Class II
    1. Identification of the legally marketed device to which the submitter claims substantial equivalence
Predicate Device Name:GradiLeiden V
510(k) number: K992456
Regulation Number: 864.7925
Regulatory Class: II
Life Therapeutics Ltd.(formerly Gradipore Ltd.)
22 Rodborough Road
Frenchs Forest, NSW
2086 Australia

{1}------------------------------------------------

4. Description of the device

Device Description:

cryocheck Clot APCR consists of:

  • 5 x 2.0 mL Activator Reagent (APC-AR) .
  • 5 x 4.0 mL Russell's Viper Venom Reagent (APC-RV)

5. Statement of the intended use of the device

Device Intended Use:CRYOCheck Clot APCR is a clotting assay intended to screen for resistanceto activated protein C in plasma from individuals with the factor VLeiden mutation.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

6. Summary of the technological characteristics of the new device in comparison to those of the predicate device

Comparison to Predicate Device:

ParameterCRYO check Clot APCRGradiLeiden V - (K992456)
Intended UseCRYOcheck Clot APCR is a clotting assayintended to screen for resistance toactivated protein C in plasma fromindividuals with the factor VLeiden mutation.GradiLeiden V is a simple, functional testsystem intended for screening of resistance toactivated protein C in plasma from individualswith the factor VLeiden defect.
FormatFrozenLyophilized
Volume• 5 x 2.0 mL Activator Reagent (APC-AR)• 5 x 4.0 mL Russell's Viper Venom Reagent(APC-RV)• 5 x 2.0 mL Activator Reagent• 5 x 4.0 mL PR3V Reagent

Comments on Substantial Equivalence:

It is the opinion of Precision BioLogic Inc. that cxrocheck Clot APCR is substantially equivalent to GradiLeiden V, manufactured by Life Therapeutics (formerly Gradipore Ltd. Australia), and currently marketed in the United States by Rainbow Scientific. This opinion is based on the following:

  • Both products are clot-based assays, �
  • . Both products are intended to screen for resistance to activated protein C in citrated human plasma from individuals with the factor V Leiden defect.
  • Both methods rely on the activation of endogenous protein C. .
  • Both methods perform a dilute Russell's viper venom time (dRVVT) on the plasma with . saline and with activator to obtain a ratio which is used for interpretation of a diagnosis.
  • . Both products can be automated in 2 stage clotting test systems with similar activation and acquisition times.
  • . Both cryocheck Clot APCR and GradiLeiden V tests can be used with the same patient groups.
  • Both methods do not require the addition of factor V deficient plasma. .

{2}------------------------------------------------

Part B

1. Brief discussion of the non-clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence

Non-clinical tests were performed in-house to determine precision, interferences, and stability. cryocheck Clot APCR showed comparable precision to that of the predicate, GradiLeiden V. Both devices demonstrated identical results when challenged with interfering substances and factor deficiencies. Storage stability is reflected in product labeling.

2. Brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence

Clinical tests were performed at Precision BioLogic and a US university hospital-based clinical coagulation laboratory. cRYocheck Clot APCR and the predicate device exhibited identical sensitivity and specificity.

3. Conclusions drawn from the non-clinical and clinical tests that demonstrate that the device is as safe, as effective, and as well or better than the legally marketed device identified in Part A (3)

cryocheck Clot APCR and GradiLeiden V have the same intended use and can be used with the same patient groups. Precision BioLogic considers cryocheck Clot APCR to be substantially equivalent to GradiLeiden V in terms of intended use, method comparison, and overall performance characteristics. cryocheck Clot APCR was compared to GradiLeiden V using 207 clinical samples from the target population for the assay. A correlation of R = 0.960 was obtained.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the logo.

Food and Drug Administration 2098 Gaither · Road Rockville MD 20850

Mr. Stephen L. Duff Director of New Business Development Precision BioLogic Inc. 900 Windmill Road, Suite 100 Dartmouth, Nova Scotia Canada B3B 1P7

MAY 1 0 2006

Re: K060284

Trade/Device Name: CryoCheckTM Clot APCRTM Regulation Number: 21 CFR § 864.7925 Regulation Name: Partial Thromboplastin Time Tests Regulatory Class: II Product Code: GGW Dated: April 28, 2006 Received: May 2, 2006

Dear Mr. Duff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{4}------------------------------------------------

Page 2 --

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Robert L. Becker Jr.

Robert L. Becker, Jr., MD, PA. Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

SECTION 4

Indications for Use

кобраз 8

510(k) Number:

cryocheck™ Clot APCR™ Device Name:

Indications for Use:

cryocheck Clot APCR is a clotting assay intended to screen for resistance to activated protein C in citrated human plasma from individuals with the factor V Leiden mutation.

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K060284

10

§ 864.7925 Partial thromboplastin time tests.

(a)
Identification. A partial thromboplastin time test is a device used for primary screening for coagulation abnormalities, for evaluation of the effect of therapy on procoagulant disorders, and as an assay for coagulation factor deficiencies of the intrinsic coagulation pathway.(b)
Classification. Class II (performance standards).