Intended Use

The Digital Thermometer with Battery-Less Patch, Model KD-2100, is intended to measure and monitor human body temperature with a wireless digital LCD reader and a sensor coupled patch. The device is reusable and intended for ear and mouth measurement of adults and pediatric patients.

Device Description

The Digital Thermometer with Battery Less Patch, Model KD -2100, consists of a DC battery powered wireless LCD reader and a battery less patch. The LCD reader part consists mainly of a plastic body, four keys (ON/OFF power, memory, read and LED on), two printed circuit boards (CPU and RFID), and and CD assembly with four backlight LEDs, a buzzer, a battery compartment and four AAA type batteries. The patch part includes a plastic body, a thermister sensor, and a TAG printed circuit board.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the K-Jump Digital Thermometer with Battery-Less Patch, Model KD-2100. Based on the information provided, here's a breakdown of the acceptance criteria and the study (or lack thereof) to prove the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard Reference)Reported Device Performance
EN 60601-1-2 (2001)Complies with EN 60601-1-2 (2001)
ASTM E112-00 "Electronic Thermometer for Intermittent Determination of Patient Temperature" standardComplies with ASTM E112-00 "Electronic Thermometer for Intermittent Determination of Patient Temperature" standard
Harmonized standards for electronic safety ruled by FCC Part 15Conformity of harmonized standards for electronic safety ruled by FCC Part 15 is also included.

2. Sample size used for the test set and the data provenance

The document explicitly states "Discussion of Clinical Tests Performed: Not Applicable." This indicates that no clinical tests were performed to establish the device's performance in a real-world setting with human subjects. Therefore, there is no information regarding sample size for a test set or data provenance in the context of clinical testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no clinical tests were performed and thus no test set with human data or ground truth established by experts.

4. Adjudication method for the test set

Not applicable, as no clinical tests were performed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a digital thermometer, not an AI-powered diagnostic imaging tool. Therefore, an MRMC study or a comparison of human reader performance with or without AI assistance is irrelevant and was not conducted.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This refers to the performance of the device itself (the thermometer). The document states that the device complies with EN 60601-1-2 (2001), ASTM E112-00, and FCC Part 15. These are standards for electronic safety, and accuracy/performance of electronic thermometers, respectively. While not explicitly detailed as a "standalone study" with specific results, the compliance statements imply that the device met the technical requirements of these standards through bench testing. The submission confirms: "bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness."

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the technical standards (EN 60601-1-2, ASTM E112-00, FCC Part 15), the "ground truth" would be the established benchmarks and measurement protocols defined within those standards. For example, for temperature accuracy, the ground truth would be a highly accurate reference thermometer or temperature bath used to calibrate and verify the device's readings according to the ASTM E112-00 standard. The document doesn't provide details on how these benchmarks were applied.

8. The sample size for the training set

Not applicable. This device is a digital thermometer, not a machine learning or AI model that requires a training set of data.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Exhibit #1

1 0 2002

Barren Barren Barren

510(K) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K060240

1. Submitter's Identification:

K-Jump Health Co., Ltd.
No. 56 Wu Kung 5* Rd. Wu Ku Industrial Park Taipei Hsien, Taiwan 248 Tel: 886 2 22991378 Fax: 886 2 22991386

Contact: Mr. Danny Wang

Date Summary Prepared: January 27, 2006

2. Name of the Device:

K-Jump Digital Thermometer with Battery-Less Patch, Model KD-2100

3. Common or Usual Name:

Clinical Electronic Thermometer

4. Predicate Device Information:

K003326, Funet Amy Mama Body Thermometer, Funai Radio & Communications Corporation

5. Device Description:

The Digital Thermometer with Battery Less Patch, Model KD -2100, consists of a DC battery powered wireless LCD reader and a battery less patch. The LCD reader part consists mainly of a plastic body, four keys (ON/OFF power, memory, read and LED on), two printed circuit boards (CPU and RFID), and and CD assembly with four backlight LEDs, a buzzer, a battery compartment and four AAA type batteries. The patch part includes a plastic body, a thermister sensor, and a TAG printed circuit board.

{1}------------------------------------------------

060240

હ. Intended Use:

The Digital Thermometer with Battery-Less Patch, Model KD-2100, is intended to measure and monitor human body temperature with a wireless digital LCD reader and a sensor coupled patch. The device is reusable and intended for ear and mouth measurement of adults and pediatric patients.

7. Comparison to Predicate Devices:

The subject device is substantially equivalent to the predicate device, the Funet Amy Mama Body Thermometer, K#003326. The two devices share the same intended use and indications for use with the exception of small differences in their temperature measurement ranges, transmission frequency, minimum operational temperature and humidity, minimum storage temperature and humidity. These minor technological differences do not raise new issues of safety or effectiveness.

Discussion of Non-Clinical Tests Performed for Determination of 8. Substantial Equivalence are as follows:

The Digital Thermometer with Battery Less Patch, Model KD-2100 complies with EN 60601-1-2 (2001) and ASTM E112-00 "Electronic Thermometer for Intermittent Determination of Patient Temperature" standard. The conformity of harmonized standards for electronic safety ruled by FCC Part 15 is also included.

ರು. Discussion of Clinical Tests Performed:

Not Applicable

10. Conclusions:

The subject device and predicate devices share the same intended use and similar characteristics. Moreover, bench testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Digital Thermometer with Battery-Less Patch, Model KD-2100, is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, representing medicine and health.

Public Health Servic

Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850

MAY 19 2006

K-Jump Health Company Limited C/O Ms. Susan D. Goldstein-Falk MDI Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021

Re: K060240

Trade/Device Name: The Digital Thermometer with Battery-Less Patch, KD-2100 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: April 28, 2006 Received: May 1, 2006

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Goldstein-Falk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

Page 1 of

510(k) Number (if known): _ K (6 Ox 4 O

Device Name: The Digital Thermometer with Battery-Less Patch, Model KD-2100

Indications for Use:

The Digital Thermometer with Battery-Less Patch, Model KD-2100, is intended to measure and monitor human body temperature with a wireless digital LCD reader and a sensor coupled patch. The device is reusable and intended for ear and mouth measurement of adults and pediatric patients.

Prescription Use (Per 21 CFR 801 Subpart D)

OR

Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Euth

OF.

: Dign-Off) of Anesthesiology, General Hospital, Con Control, Dental Devices

Number. k@by44

§ 880.2910 Clinical electronic thermometer.

(a)
Identification. A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.(b)
Classification. Class II (performance standards). The device is exempt from the premarket notification procedures in part 807, subpart E of this chapter, subject to the limitations in § 880.9 and the following conditions for exemption:(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.