K Number
K060162

Validate with FDA (Live)

Date Cleared
2006-02-21

(29 days)

Product Code
Regulation Number
864.5425
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACL Elite and ACL Elite Pro are fully automated, high-productivity analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Device Description

The ACL Elite and ACL Elite Pro are fully automated, high-productivity analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

The analytical modifications to the ACL 8000/9000/10000 introduced with the new family members (ACL Elite and ACL Elite Pro) include the following:

  • Modification to the test parameter (revised use of the Secondary Algorithm) for several . PT and APTT assays.
  • Addition of dedicated calibration for HemosIL Factor VIII and HemosIL Factor IX . Deficient Plasmas, when used in conjunction with HemosIL APTT-SP.
  • Introduction of automated factor parallelism, which is a manual function on the ACL . 8000/9000/10000.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ACL Elite and ACL Elite Pro as described in the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to a legally marketed predicate device (ACL 9000, ACL 8000, and ACL 10000; K000053). The goal is to demonstrate "substantial equivalence" in performance. The summary details precision and method comparison data to support this claim.

Implied Acceptance Criteria (based on predicate equivalence and good laboratory practice):

  • Precision: The device should exhibit acceptable within-run and total precision (low %CV) for various coagulation tests (FVIII, FIX, APTT, PT, Fibrinogen), comparable to or better than established laboratory standards and the predicate device.
  • Method Comparison: The device should show statistical similarity (e.g., strong correlation, slope near 1, and intercept near 0) when compared to the predicate device for various coagulation assays.

Reported Device Performance (ACL Elite Pro):

TestMaterialnMeanWithin Run % CVTotal % CV
HemosIL Factor VIII (w/ APTT-SP)Normal Control5085.34.14.5
Special Test Control Level 15072.42.93.6
Special Test Control Level 25036.34.65.4
HemosIL Factor IX (w/ APTT-SP)Normal Control50115.62.75.2
Special Test Control Level 15076.22.34.5
Special Test Control Level 25041.03.35.1
HemosIL APTT-SP (APTT)Normal Control5029.31.12.0
Low Abnormal Control5046.11.94.1
High Abnormal Control5056.92.24.7
HemosIL SynthASil (APTT)Normal Control5029.00.81.8
Low Abnormal Control5051.91.01.5
High Abnormal Control5060.51.21.2
HemosIL PT-Fibrinogen (PT)Normal Control5012.30.61.1
Low Abnormal Control5020.70.91.1
High Abnormal Control5025.50.81.4
HemosIL PT-Fibrinogen (Fibrinogen)Normal Control50277.94.44.6
Low Fibrinogen Control50132.93.65.5
HemosIL PT-Fibrinogen Recombinant (PT)Normal Control5010.71.01.0
Low Abnormal Control5035.01.11.3
High Abnormal Control5051.61.43.4
HemosIL PT-Fibrinogen Recombinant (Fibrinogen)Normal Control50297.63.86.7
Low Fibrinogen Control50113.15.36.4

Method Comparison Data (ACL Elite/ACL Elite Pro vs. ACL 9000):

TestParameternSloper
HemosIL Factor VIIIFVIII w/ APTT-SP (% Act.)470.9090.9938
HemosIL Factor IXFIX with APTT-SP (% Act.)490.9480.9914
HemosIL APTT-SPAPTT (Seconds)541.0320.9990
HemosIL SynthASilAPTT (Seconds)521.0090.9981
HemosIL PT-FibrinogenPT (sec)490.9890.9985
Fibrinogen (mg/dL)470.9960.9848
HemosIL PT-Fibrinogen RecombinantPT (sec)490.9720.9993
Fibrinogen (mg/dL)470.9740.9992

2. Sample Sizes Used for the Test Set and Data Provenance

  • Precision Test Set: For each combination of test and control material, n=50 measurements were performed. This involved control plasma samples (Normal Control, Special Test Control Levels, Low Abnormal Control, High Abnormal Control, Low Fibrinogen Control).
  • Method Comparison Test Set: Sample sizes ranged from n=47 to n=54 for each specific parameter compared between the ACL Elite/Elite Pro and the predicate ACL 9000. These were "citrated plasma samples."
  • Data Provenance: The document does not explicitly state the country of origin. It indicates the submitting company is in Lexington, MA, USA, suggesting the data was likely generated in the US or in compliance with US regulatory standards. The data presented is characteristic of a prospective study, where samples are run on the new device and compared to a predicate or against established control values.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This type of submission for an in vitro diagnostic device (IVD) for coagulation testing does not typically rely on "experts establishing ground truth" in the same way an image analysis AI might.

  • Ground Truth for Precision: The "ground truth" for precision is derived from the mean and standard deviation of repeated measurements on stable control materials. These control materials (e.g., Normal Control, Special Test Control) themselves have established target ranges, which are typically determined by the manufacturer of the control material based on extensive testing and/or collaborative studies. There are no "experts" in the sense of human readers adjudicating results.
  • Ground Truth for Method Comparison: The "ground truth" for method comparison is the measurement obtained from the predicate device (ACL 9000). The new device's performance is compared against the predicate's measurements. Again, this does not involve human expert adjudication.

4. Adjudication Method for the Test Set

Not applicable for this type of IVD performance study. The data involves quantitative measurements from automated instruments, not qualitative assessments requiring human adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is an in vitro diagnostic device (automated analyzer) and not an AI-assisted diagnostic tool that involves human readers or interpretation of complex medical images.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, the studies reported are standalone device performance studies. The ACL Elite and ACL Elite Pro are automated analyzers. Their precision and method comparison studies evaluate the device's inherent analytical performance without any human intervention in the result generation or calculation process. The studies describe the performance of the "algorithm only" in the sense of the automated analytical process of the instrument.

7. The Type of Ground Truth Used

  • For Precision: The ground truth for precision is statistical: the mean values and the expected variability (low %CV) of the control materials, established by the control manufacturer and expected from a well-functioning analytical system.
  • For Method Comparison: The ground truth for method comparison is the measurements obtained from the predicate device (ACL 9000). The new device is considered substantially equivalent if its results correlate strongly with, and are statistically similar to, those produced by the predicate.

8. The Sample Size for the Training Set

Not applicable in the context of machine learning model training. This device is an automated analyzer, and its "algorithm" refers to its pre-programmed analytical processes and calculations, not a machine learning model that undergoes a distinct "training phase" on a dataset. The system's operational parameters and calibration procedures are established during its design and manufacturing based on general principles of coagulation testing and validation, rather than iterative training on a specific "training set" of patient data.

9. How the Ground Truth for the Training Set was Established

Not applicable for this type of device. As explained in point 8, there isn't a "training set" in the machine learning sense. The device's calibration and internal algorithms are based on established scientific principles of coagulation measurement, reagent properties, and instrument design, not on a data-driven training regimen with a "ground truth" derived from external data.

{0}------------------------------------------------

Section 3 ACL Elite and ACL Elite Pro 510(k) Summary

K060/62

Submitted by:

Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421

Contact Person:

Carol Marble, Regulatory Affairs Director 781-861-4467 Phone: 781-861-4207 Fax:

Summary Prepared:

January 20, 2006

Device Name:

ACL Elite and ACL Elite Pro

Classification Name:

Multipurpose system for in vitro coagulation studies (864.5425) Regulation Section: Classification: Class II JPA Product Code: Hematology Panel:

Legally Marketed Device:

ACL 9000 (ACL 8000 and ACL 10000) K000053

Device Description:

The ACL Elite and ACL Elite Pro are fully automated, high-productivity analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

The analytical modifications to the ACL 8000/9000/10000 introduced with the new family members (ACL Elite and ACL Elite Pro) include the following:

  • Modification to the test parameter (revised use of the Secondary Algorithm) for several . PT and APTT assays.
  • Addition of dedicated calibration for HemosIL Factor VIII and HemosIL Factor IX . Deficient Plasmas, when used in conjunction with HemosIL APTT-SP.
  • Introduction of automated factor parallelism, which is a manual function on the ACL . 8000/9000/10000.

{1}------------------------------------------------

Section 3 (Cont.) ACL Elite and ACL Elite Pro 510(k) Summary

Statement of Technological Characteristics of the Device Compared to Predicate Device:

The indications for use/intended use, operating principle and performance claims of the ACL Elite and ACL Elite Pro are substantially equivalent to the legally marketed device, the ACL 8000/9000/10000 (K000053).

Summary of Performance Data:

Precision

Within run and total precision assessed over multiple runs on the ACL Elite Pro using three levels of control plasma for factors deficient plasmas, APTT and PT, and two levels of control plasmas for fibrinogen gave the following results:

TestFVIII (% Activity)
HemosILFactor VIII(K034007)w/ APTT-SPMaterialnMeanWithin Run% CVTotal%CV
Normal Control5085.34.14.5
Special Test Control Level 15072.42.93.6
Special Test Control Level 25036.34.65.4
FIX (% Activity)
HemosILFactor IX(K031829)w/ APTT-SPMaterialnMeanWithin Run% CVTotal%CV
Normal Control50115.62.75.2
Special Test Control Level 15076.22.34.5
Special Test Control Level 25041.03.35.1
APTT (Seconds)
HemosILAPTT-SP(K973306)MaterialnMeanWithin Run% CVTotal%CV
Normal Control5029.31.12.0
Low Abnormal Control5046.11.94.1
High Abnormal Control5056.92.24.7
APTT (Seconds)
HemosILSynthASil(K953981)MaterialnMeanWithin Run% CVTotal%CV
Normal Control5029.00.81.8
Low Abnormal Control5051.91.01.5
High Abnormal Control5060.51.21.2

{2}------------------------------------------------

Section 3 (Cont.) ACL Elite and ACL Elite Pro 510(k) Summary

Summary of Performance Data (Cont.):

Precision (Cont.)

TestProthrombin Time (Seconds)
MaterialnMeanWithin Run% CVTotal% CV
HemosILPT-Fibrinogen(K862301)Normal Control5012.30.61.1
Low Abnormal Control5020.70.91.1
High Abnormal Control5025.50.81.4
Fibrinogen (mg/dL)
MaterialnMeanWithin Run% CVTotal% CV
Normal Control50277.94.44.6
Low Fibrinogen Control50132.93.65.5
HemosILPT-FibrinogenRecombinant(K981479)Prothrombin Time (Seconds)
MaterialnMeanWithin Run% CV
Normal Control5010.71.0
Low Abnormal Control5035.01.1
High Abnormal Control5051.61.43.4
Fibrinogen (mg/dL)
MaterialnMeanWithin Run% CV
Normal Control50297.63.86.7
Low Fibrinogen Control50113.15.36.4

Method Comparison

In comparison studies evaluating citrated plasma samples, the ACL Elite/ACL Elite Pro versus the ACL 9000 (predicate device) was shown to be statistically similar for the tests listed below:

TestParameternSloper
HemosIL Factor VIIIFVIII w/ APTT-SP (% Act.)470.9090.9938
HemosIL Factor IXFIX with APTT-SP (% Act.)490.9480.9914
HemosIL APTT-SPAPTT (Seconds)541.0320.9990
HemosIL SynthASilAPTT (Seconds)521.0090.9981
HemosILPT (sec)490.9890.9985
PT-FibrinogenFibrinogen (mg/dL)470.9960.9848
HemosILPT (sec)490.9720.9993
PT-Fibrinogen RecombinantFibrinogen (mg/dL)470.9740.9992

Special 510(k): ACL Elite and ACL Elite Pro

Page 3 of 3

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect and promote the health and well-being of Americans. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Carol Marble Regulatory Affairs Director Instrumentation Laboratory Company 113 Hartwell Avenue Lexington, MA 02421

FEB 2 1 2006

K060162 Re: Trade/Device Name: ACL Elite and ACL Elite Pro Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: II Product Code: JPA Dated: January 20, 2006 Received: January 23, 2006

Dear Ms. Marble:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou mayy atteres, include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device n may or babyer to back of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reaso o a recer a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{4}------------------------------------------------

Page 2 -

If you desire specific information about the application of labeling requirements to your device, If you destre specific infortion and advertising of your device, please contact the Office of In of quotions on all process Evaluation and Safety at (240) 276-0484. Also, please note the vitro Diagnoutied, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Tou may oouall other generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

lobetz Beckerh

Robert L. Becker, Jr., MD, PK.D Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K060162

Device Name: ACL Elite and ACL Elite Pro

Indications for Use:

The ACL Elite and ACL Elite Pro are fully automated, high-productivity analyzers designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Jazhine Bautista

Division Sign-off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K060162

Page 1 of 1

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.