K Number
K052830

Validate with FDA (Live)

Device Name
SPORTMESH
Manufacturer
Date Cleared
2006-01-19

(106 days)

Product Code
Regulation Number
878.3300
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sportmesh™ is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists. Sportmesh™ is also intended for reinforcement of soft tissues that are repaired by suture or suture anchors, limited to the supraspinatus, during rotator cuff repair surgery.

Sportmesh™ is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone, provide mechanical strength for the tendon repair. Sportmesh™ reinforces soft tissue and provides a degradable scaffold that is incorporated in the patient's own tissue.

Device Description

Sportmesh™is a knitted fabric made from ARTELON fibers. This construction permits the mesh to be cut into any desired shape or size without unraveling. The device is supplied in sheet form in sterile double layer peelable packaging.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Sportmesh™ device. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than proving that the device meets specific acceptance criteria through a traditional clinical study with defined endpoints and acceptance thresholds.

Therefore, the document does NOT contain information regarding:

  • A table of acceptance criteria and reported device performance in the way typically found for novel device approvals.
  • Sample sizes for a test set, data provenance, number of experts for ground truth, adjudication methods, or MRMC studies.
  • Standalone algorithm performance.
  • Training set size or how ground truth for a training set was established.

Instead, the submission relies on:

Acceptance Criteria and Device Performance (as demonstrated for 510(k) pathway):

The "acceptance criteria" in this context are primarily demonstrating equivalence to already approved predicate devices concerning safety and effectiveness. The "performance" is shown through the successful completion of specific tests and comparisons.

Acceptance Criteria (Demonstration of Equivalence)Reported Device Performance (as summarized in the 510(k))
Similar Intended Use: The device should have the same intended use as a legally marketed predicate device.Sportmesh™ is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists, and for reinforcement of soft tissues repaired by suture or suture anchors (limited to the supraspinatus) during rotator cuff repair surgery. This use is stated to be equivalent to predicate devices such as Organogenesis, Inc's FortaFlex™ Surgical Mesh, DePuy, Inc's Restore® Orthobiologic Soft Tissue Implant, Ethicon, Inc's ULTRAPRO™Mesh, and Davol's Marlex Mesh.
Similar Technological Characteristics: The device should have similar technological characteristics (e.g., material, design, operating principles) as a legally marketed predicate device, or differences in technological characteristics should not raise new questions of safety and effectiveness.Sportmesh™ is a knitted fabric made from ARTELON fibers. The construction permits it to be cut without unraveling. It is supplied in sterile sheet form. The submission claims equivalence in technological characteristics to the same predicate devices as for intended use. The document doesn't provide a detailed comparison of all characteristics, but the FDA's clearance implies these were deemed sufficiently similar or that any differences did not raise new safety/effectiveness concerns.
Performance Testing: The device should demonstrate appropriate performance through testing, including biocompatibility and mechanical properties, consistent with its intended use and recognized standards, to support that any differences in technological characteristics do not raise new questions of safety and effectiveness. (This is where "testing" comes in for the 510(k)).A collection of tests was conducted and successfully completed, including: - Biocompatibility safety studies: Conducted according to ISO 10993 standards. - Mechanical testing: Conducted in accordance with the "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh; Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance." The results "demonstrate that Sportmesh™ provides appropriate mechanical properties for its use in soft tissue repair." Based on these comparisons and performance tests, the conclusion is that Sportmesh™ is substantially equivalent and "presents no new concerns about safety and effectiveness."

Study Details (as per the provided document):

  1. Sample size used for the test set and data provenance:

    • Not applicable/Not provided. The document describes premarket notification for a Class II medical device (Sportmesh™) via the 510(k) pathway. This pathway establishes substantial equivalence to predicate devices, and typically does not require a formal "test set" with a specified sample size in the context of clinical performance evaluation in the same way a PMA (Premarket Approval) submission would. Instead, testing focused on biocompatibility and mechanical properties. The data provenance is implied to be from Artimplant AB's internal testing as part of their submission process.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. Ground truth in the sense of expert consensus on clinical outcomes for a "test set" is not detailed in this 510(k) summary. The "ground truth" for the performance tests would be the established ISO 10993 standards for biocompatibility and specific mechanical test parameters outlined in the FDA's guidance for surgical mesh, which are objectively measurable, rather than expert-derived clinical ground truth.
  3. Adjudication method for the test set:

    • Not applicable/Not provided. No clinical adjudication method is described.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a surgical mesh, not a software algorithm.
  6. The type of ground truth used:

    • The "ground truth" for the performance evaluation primarily consisted of established regulatory standards and guidance documents. For biocompatibility, this was ISO 10993 standards. For mechanical properties, it was FDA guidance for surgical mesh. These are objective, measurable criteria, not expert consensus or pathology on a clinical dataset.
  7. The sample size for the training set:

    • Not applicable/Not provided. The concept of a "training set" is not relevant for this type of device and submission, which involves physical material testing and comparison to predicates, not machine learning.
  8. How the ground truth for the training set was established:

    • Not applicable/Not provided. See point 7.

{0}------------------------------------------------

1/2

JAN J 9 2006

Summary of Safety and Effectiveness

In accordance with 21 CFR 807.92, the following information constitutes Artimplant AB's summary for the Sportmesh™.

SUBMITTER'S NAME:Artimplant AB
ADRESS:Hulda Mellgrens gata 5, SE-421 32 Västra Frölunda, Sweden
CONTACT PERSON:Ajrulla Zuta
TELEPHONE NUMBER:+46 31 7465600
FAX NUMBER:+46 31 7465660
DATE OF SUBMISSION:October 4, 2005

l. Identification of device

Proprietary Name: Sportmesh™ Common Name: Surgical Mesh, RC Patch, Rotator cuff patch Classification Status: Class II per regulations 878 3300 Product Code: FTM

2. Equivalent devices

Sportmesh™ is substantially equivalent to: Organogenesis, Inc's FortaFlex™ Surgical Mesh (K042809) DePuy, Inc's Restore® Orthobiologic Soft Tissue Implant (K031969, K001738) Ethicon, Inc's ULTRAPRO™Mesh (033337) Davol's Marlex Mesh (pre-Amendments)

3. Description of the Device

Sportmesh™is a knitted fabric made from ARTELON fibers. This construction permits the mesh to be cut into any desired shape or size without unraveling. The device is supplied in sheet form in sterile double layer peelable packaging.

4. Intended use

Sportmesh™ is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists. Sportmesh™ is also intended for reinforcement of soft tissues that are repaired by suture or suture anchors, limited to the supraspinatus, during rotator cuff repair surgery.

Sportnesh™ is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff. Sutures used to repair the tear, and sutures or bone anchors used to attach the tissue to the bone, provide mechanical strength for the tendon repair. Sportmesh™ reinforces soft tissue and provides a degradable scaffold that is incorporated in the patient's own tissue.

ನ. Comparison to predicate device.

Sportmesh™ is equivalent to the Organogenesis, Cuffpatch™ Surgical Mesh (K042809). DePuy, Inc's Restore® Orthobiologics Soft Tissus Implant (K03)969), Ethicon (K0428)
UI TDADBOTHM - 1, (80000), 1), 1800000), Ethicon (K031969), Ethicon, Inc's ULTRAPRO™Mesh (033337), Tissue Science's Permacol (K021969), Eldiconi, Inc S
Mesh (pre-Amendments) 337), Tissue Science's Permacol (K021056) and Davol's Marlex Mesh (pre-Amendments) with respect to intended use and technological characteristics.

{1}------------------------------------------------

Ko528 30 2/2

6. Discussion of performance testing.

A collection of tests has been conducted and successfully completed including biocompatibility safety studies (ISO 10993 standards), and mechanical testing in accordance Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh; Guidance for Industry and/or for FDA Reviewers/Staff and/or Compliance. The results demonstrate that Sportmesh™ provides appropriate mechanical properties for its use in soft tissue repair.

7. Conclusion

Based on comparison to the predicate devices, the Sportmesh™ is substantially equivalent to legally-marketed devices and presents no new concerns about safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an abstract design of an eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN ] 9 2006

Artimplant AB c/o Ms. Terry Sheridan Powell Regulatory Consultant M Squared Associates 719 A Street, NE Washington, District of Columbia 20002

Re: K052830

Trade/Device Name: Artimplant AB Sportmesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: December 9, 2005 Received: December 12, 2005

Dear Ms. Powell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Powell

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours

Sincerely yours,

Barbara Buehr

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K 0528 30

Artimplant A Device Name:

Indications for Use:

Sportmesh™ is intended for use in general surgical procedures for reinforcement of soft tissue where weakness exists. Sportmesh™ is also intended for reinforcement of soft tissues that are repaired by suture or suture anchors, limited to the supraspinatus, during rotator cuff repair surgery.

Sportmesh™ is not intended to replace normal body structure or provide the full mechanical strength to support the rotator cuff. Sutures used to repair the full
sutures or bone anchors used to ottach the i sutures or bone anchors used to attach the tissue to the bone, provide mechanical strength for the tendon repair. Sportmesh™ reinforces soft tissue and provides a degradable scaffold that is incorporated in the patient's own tissue a

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K052830 (barbarabuchbinder Page 1 of 1
(Division Sign-Off) for WM

Division of General, Re-rorative, and Neurological Devices

510(k) Number_ 1505 2430

§ 878.3300 Surgical mesh.

(a)
Identification. Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.(b)
Classification. Class II.