K Number
K051991

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-10-20

(90 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PowerPICC™ catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

The Poly Per-Q-Cath PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

The Poly Per-O-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, power injection of contrast media and allow for central venous pressure monitoring. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Device Description

Subject devices:

  • Catheters range in French size from 3-5 Fr SL; 4-6 Fr DL and 6 Fr TL .
  • Catheter usable length ranges from 40 -60 cm. .
  • Catheters are open-ended catheters extruded from polyurethane material containing barium ● sulfate for radiopacity.
  • The catheter extension legs are polyurethane extrusions. Extension legs are minimum 2.2 in. . in length to promote easy application of occlusive dressings. Each extension leg has a thumb clamp.
  • The luer hub base material is Isoplast polyurethane. .
  • The catheter has a reverse taper design .
  • The user is informed of the gage size in product labeling and it is printed on the luer hub. .
  • The catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a . reference for the catheter insertion point
  • Catheters are provided sterile and are packaged with legally marketed kit components that are . preferred by clinicians
AI/ML Overview

The provided document describes a 510(k) premarket notification for Central Venous Pressure Monitoring (CVPM) capability in various catheter devices, rather than a standalone device with extensive acceptance criteria and a study demonstrating its performance in a clinical setting against a ground truth.

The document focuses on demonstrating substantial equivalence to existing predicate devices for an expanded indication for use: "allows for central venous pressure monitoring." The key point is that the catheters themselves are not new, but their indicated use now includes CVPM. Therefore, the "acceptance criteria" and "study" are geared towards proving that the existing or slightly modified catheters are suitable for this additional function, specifically focusing on the catheter lumen size for effective CVPM.

Here's the breakdown of the information requested, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
CVPM FunctionalityCatheter lumen of 20 gauge or larger for central venous pressure monitoringSubject devices, except for 21 gauge catheter lumens, met the performance criteria of design verification. As a result, the recommendation of catheter lumen of 20 gauge or larger was added to the indications for use.
Safety and EffectivenessNo change in design affecting safety or effectiveness compared to predicate devicesNo change in design that could affect the safety or effectiveness of the device.
No new types of safety and effectiveness questions raised by new characteristicsNo new types of safety and effectiveness questions.
Applicable Standards- IEC 60601-2-34:2000(E) Ed.2 (Medical electrical equipment - Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)
- AAMI TIR: 1992 (Evaluation of Clinical Systems for Invasive Blood Pressure Monitoring)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)
- ANSI/AAMI BP22:1994 (Blood Pressure Transducers)Testing was based on recognized and un-recognized standards. (Implied: device met these standards for the relevant aspects)

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a test set in the traditional sense of a clinical trial with a specific number of patients or samples. The "testing" mentioned is bench testing (laboratory studies). Therefore, there is no information on:

  • Sample size used for a clinical test set.
  • Country of origin of data for a clinical test set.
  • Retrospective or prospective nature of a clinical test set.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable as the document describes bench testing and design verification, not a clinical study involving human readers or expert consensus to establish a clinical "ground truth."

4. Adjudication Method for the Test Set

This information is not applicable for the same reasons as #3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done. The document focuses on the device's capability for CVPM, not on human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

No, a standalone algorithm performance study was not done. This submission is for physical catheter devices, not AI algorithms.

7. The Type of Ground Truth Used

The "ground truth" used for evaluating the central venous pressure monitoring capability was based on recognized and un-recognized standards for invasive blood pressure monitoring equipment and transducers, likely involving physical measurements and comparisons in a controlled bench setting. This is best described as bench test performance against established engineering or medical device performance standards. It is not clinical pathology, expert consensus on images, or patient outcomes data.

8. The Sample Size for the Training Set

This information is not applicable. The document describes physical medical devices (catheters) and their expanded indications, not a machine learning model that would require a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as #8.

{0}------------------------------------------------

Central Venous Pressure Monitoring Traditional 510(k)

ાવવા

ાજરી ર

Section 5

Central Venous Pressure Monitoring

510(k) Summary of Safety and Effectiveness 21 CFR 807.92

5.1 General Information

Submitter Name:Bard Access Systems, Inc. (BAS)[Wholly owned Subsidiary of C. R. Bard, Inc.]
Address:5425 West Amelia Earhart DriveSalt Lake City, UT 84116
Telephone Number:(801) 595-0700 ext. 7136
Fax Number:(801) 595-5425
Contact Person:Lynn M. Kirchoff
Date of Preparation:July 21, 2005
Registration Number:1720496
Additional Registration Numbers:C. R. Bard2212754

5.2 Device Information

Device Name:5 Fr SL PowerPICCTM Catheter
Trade Name:5 Fr SL PowerPICCTM Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter
Classification Panel:General Hospital
Device Name:6 Fr DL PowerPICCTM Catheter
Trade Name:6 Fr DL PowerPICCTM Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter
Classification Panel:General Hospital
Device Name:Poly Per-Q-Cath PICC (Peripherally Inserted CentralCatheter)
Trade Name:Poly Per-Q-Cath® PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter
Classification Panel:General Hospital
Device Name:Poly Per-Q-Cath®3 Triple Lumen PICC Catheter
Trade Name:Poly Per-Q-Cath®3 Triple Lumen PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class II

{1}------------------------------------------------

C

Peripherally Inserted Central Catheter
Classification Panel:General Hospital
Device Name:5 Fr SL PowerHohnTM and PowerLineTM (Central VenousCatheter)
Trade Name:PowerHohnTM and PowerLineTM Catheters
Common/Usual Name:Central Venous Catheter
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class II
Classification Panel:General Hospital
Device Name:6 Fr DL PowerHohnTM and PowerLineTM (Central VenousCatheter)
Trade Name:PowerHohnTM and PowerLineTM Catheters
Common/Usual Name:Central Venous Catheter
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class II
Classification Panel:General Hospital
Subject/Predicate Device Name510(k)Clearance Date
5 Fr Single Lumen (SL) Power PICCTMK0333893/19/2004
6 Fr Dual Lumen (DL) Power PICCTMK0509316/15/2005
3, 4, 5 Fr Single Lumen (SL) and 4, 5, 6Fr Dual Lumen (DL) Poly Per-Q-Cath®K0340191/21/2004
6 Fr Triple Lumen (TL) Poly Per-Q-Cath®K0435021/14/2005
5 Fr Single Lumen (SL) PowerHohnTMand PowerLineTMK0501855/26/2005
6 Fr Dual Lumen (DL) PowerHohnTM andPowerLineTMK0514176/30/2005

5.3 Predicate Devices

The predicate devices are:

Device Name:Cook Turbo-Flo® PICC Catheter
Trade Name:Cook Turbo-Flo PICC Catheter
Common/Usual Name:Peripherally Inserted Central Catheter (PICC)
Classification Name:Long Term Intravascular Catheter (80 LJS)21 CFR 880.5970, Class IIPeripherally Inserted Central Catheter
Classification Panel:General Hospital

{2}------------------------------------------------

K051991 3 of 5

Central Venous Pressure Monit Traditional 510(k)

Predicate Device Name510(k)Clearance Date
Cook Turbo-Flo 4 Fr Single Lumen (SL)and 5 Fr Dual Lumen (DL) PICCK0215575/30/2003
Cook Triple Lumen 9 Fr Triple Lumen(TL) Central Venous Catheter (CVC)K0215575/30/2003

5.4 Device Description

Subject devices:

  • Catheters range in French size from 3-5 Fr SL; 4-6 Fr DL and 6 Fr TL .
  • Catheter usable length ranges from 40 -60 cm. .
  • Catheters are open-ended catheters extruded from polyurethane material containing barium ● sulfate for radiopacity.
  • The catheter extension legs are polyurethane extrusions. Extension legs are minimum 2.2 in. . in length to promote easy application of occlusive dressings. Each extension leg has a thumb clamp.
  • The luer hub base material is Isoplast polyurethane. .
  • The catheter has a reverse taper design .
  • The user is informed of the gage size in product labeling and it is printed on the luer hub. .
  • The catheter shaft tubing is marked with depth indicators, with "0" indicated to serve as a . reference for the catheter insertion point
  • Catheters are provided sterile and are packaged with legally marketed kit components that are . preferred by clinicians

Intended Use of Device 5.5

The intended use is the same as the predicate devices.

The Indications for Use was expanded to include allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

The Indications for Use statements are as follows:

5 Fr SL and 6 Fr DL PowerPICCTM

The PowerPICC™ catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Poly Per-Q-Cath PICC

The Poly Per-Q-Cath PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Poly Per-Q-Cath Triple Lumen PICC

The Poly Per-O-Cath® Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central

{3}------------------------------------------------

Kosteqi 4 of 5

Central Venous Pressure Monitorin Traditional 510(k)

venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

5 Fr SL and 6 Fr DL PowerHohn™ and PowerLine™

PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, power injection of contrast media and allow for central venous pressure monitoring. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Technological Comparison to Predicate Devices: ર્સ્વ

The technological characteristics of the subject devices are substantially equivalent to the predicate devices in terms of intended use, application, user population, basic design, performance, labeling, packaging, and sterilization method.

510(k) Substantial Equivalence Decision Tree: 5.7

New device compared to Marketed Device?

Yes.

Does the new device have the same indication statement as the predicates?

Yes, with expansion of indication to include allows for central venous pressure monitoring. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Does the new device have the same technological characteristics, e.g. design, material, etc.?

Yes. The principles of operation are the same as the predicate devices.

Could the new characteristics affect safety or effectiveness?

No. There is no change in design that could affect the safety or effectiveness of the device.

Do the new characteristics raise new types of safety and effectiveness questions?

No. There are no new types of safety and effectiveness questions

Do accepted scientific methods exist for assessing effects of the new characteristics?

Yes. Testing was based on recognized and un-recognized standards to evaluate the device's performance:

Are performance data available to assess effects of new characteristics? Yes. Bench testing was based on the below referenced standards.

{4}------------------------------------------------

  • IEC 60601-2-34:2000(E) Ed.2-Medical electrical equipment- Particular . requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment
  • AAMI TIR: 1992- Evaluation of Clinical Systems for Invasive Blood Pressure . Monitoring
  • ANSI/AAMI BP22:1994-Blood Pressure Transducers .

Do performance data demonstrate equivalence?

Yes. Performance data demonstrate that the subject devices are substantially equivalent to the predicate devices

Conclusion 5.8

Subject devices, except for 21 gauge catheter lumens, met the performance criteria of design verification as specified by applicable standards, test protocols and/or customer inputs. As a result the recommendation of catheter lumen of 20 gauge or larger was added to the indications for use. Based on FDA's decision trees, the subject devices are substantially equivalent to the legally marketed predicate devices, the Cook Turbo-Flo PICC and TL Central Venous Catheter, K021557, cleared 5/30/2003.

Subject/Predicate Device Name510(k)Clearance Date
5 Fr Single Lumen (SL) Power PICC™K0333893/19/2004
6 Fr Dual Lumen (DL) Power PICC™K0509316/15/2005
3, 4, 5 Fr Single Lumen (SL) and 4, 5, 6Fr Dual Lumen (DL) Poly Per-Q-Cath®K0340191/21/2004
6 Fr Triple Lumen (TL) Poly Per-Q-Cath®K0435021/14/2005
5 Fr Single Lumen (SL) PowerHohn™and PowerLine™K0501855/26/2005
6 Fr Dual Lumen (DL) PowerHohn™ andPowerLine™K0514176/30/2005

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the caduceus.

OCT 2 0 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Lynn M. Kirchoff Regulatory Affairs Specialist C.R. Bard, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116

Re: K051991

Trade/Device Name: POWERPICC, POLY PER-Q-CATH, 6 FR TL POLY PER-Q-CATH, POWER HOHN AND POWER LINE Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: July 21, 2005 Received: July 22, 2005

Dear Ms. Kirchoff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device microlate comments or prices that have been reclassified in accordance with the provisions of / Inchaneliers, or to are , and Cosmetic Act (Act) that do not require approval of a premarket the Federal F 600; Drag, Drag, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 WA), it may of subject in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device earred fourther announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advisou that 1 DTC issuaires on that your device complies with other requirements Incall that I DAT has mace a actes and regulations administered by other Federal agencies. of the Act of ally I edelar statues and registements, including, but not limited to: registration You must comply with an the Hee Reg (21 CFR Part 801); good manufacturing practice allu listing (21 CF R Part 067), as times (QS) regulation (21 CFR Part 820); and if requirents as set form in the quality Brovisions (Sections 531-542 of the Act); applicable, the clections product identify ou to begin marketing your device as described 21 CFK 1000-1030. This least will affication. The FDA finding of substantial equivalence III your Section 510(k) promative no or dicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), II you desire specific advise for Jour at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Suyate Y.M., Obie Omd.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section 1.2

Indications for Use

510(k) Number (if known): _________________

Device Name: Single and Dual Poly Per-Q-Cath® PICC

Indications for Use:

The Poly Per-Q-Cath PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clank le. m

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number: _ 14 15 1

Image /page/7/Picture/15 description: The image shows the number 6 repeated several times. There is a larger number 6 at the top of the image. Below the number 6, there are five smaller numbers, all of which are also the number 6. The numbers are printed in a slightly distressed font.

{8}------------------------------------------------

Section 1.2

Indications for Use

510(k) Number (if known): ___________

5 Fr Single Lumen and 6 Fr Dual Lumen Power PICC™ Device Name:

Indications for Use:

The PowerPICCTM catheter is indicated for short or long term peripheral access to the central venous system for intravenous therapy, power injection of contrast media, and allows for central venous pressure monitoring. For blood sampling, infusion or therapy, use a 4 French or larger catheter. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerPICC catheter may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

2

(Civision Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number: K025691

000007

{9}------------------------------------------------

Central Venous Pressure Monitoring Traditional 510(k)

Section 1.2

Indications for Use

510(k) Number (if known): _____________

6 Fr Triple Lumen Poly Per-Q-Cath Device Name:

Indications for Use:

The Poly Per-Q-Cath®3 Triple Lumen PICC is indicated for short or long term peripheral access to the central venous system for intravenous therapy, blood sampling and allows for central venous pressure monitoring. For blood therapy, it is recommended that a 4 French or larger catheter be used. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Cater

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

1545 49 510(k) Number _

Coopera

{10}------------------------------------------------

Section 1.2

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _ _ _ 5 Fr Single Lumen Power Hohn™ and Power Line™

Indications for Use:

PowerHohn and PowerLine Catheters are indicated for short or long term access to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, power injection of contrast media and allows for central venous pressure monitoring. The maximum recommended infusion rate is 5mL/sec. The maximum pressure of power injectors used with the PowerHohn and PowerLine catheters may not exceed 300 psi. For central venous pressure monitoring, it is recommended that catheter lumen of 20 gauge or larger be used.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Chan Un

i Tivision Sign-Off) Division of Anesthesiology, General Hospital, Intection Control, Dental Devices

510(k) Number:________________________________________________________________________________________________________________________________________________________________

00000

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”