(374 days)
The MYKRES Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis, with objective evidence of neurological imparment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudathrosis).
The MYKRES Spinal System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients undergoing fusion by autogenous bone grafting having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.
The MYKRES Spinal System is an internal fixation device for spinal surgery consisting of rods, screws, hooks and connectors available in various lengths, diameters, and configurations to enable close conformance to patient anatomy. A series of manual instruments (not a subject of this submission) intended to assist in the insertion and placement of the implants is provided in separate trays.
The rods, screws, hooks and connectors of the MYKRES Spinal System are made of Ti-6Al-4V titanium alloy conforming to ASTM F 136.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Bench Testing per ASTM F 1717) | Reported Device Performance |
|---|---|
| Static compression performance | Substantially equivalent to legally marketed spinal fixation systems |
| Static torsion performance | Substantially equivalent to legally marketed spinal fixation systems |
| Dynamic axial compression bending performance | Substantially equivalent to legally marketed spinal fixation systems |
2. Sample Size and Data Provenance for Test Set
- Sample Size: Six (6) samples for each of the worst-case MYKRES construct for the bench testing.
- Data Provenance: The data is from laboratory bench testing performed by Showa Ika Kohgyo Co., LTD. It is not patient data with country of origin.
3. Number of Experts and Qualifications for Ground Truth (Test Set)
Not applicable. The ground truth for this device is based on objective mechanical bench testing standards (ASTM F 1717), not expert consensus on medical images or clinical outcomes.
4. Adjudication Method (Test Set)
Not applicable. This was a mechanical engineering test against established ASTM standards, not a clinical trial requiring adjudication of observations.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. The device is a spinal fixation system, and its evaluation focuses on mechanical performance and substantial equivalence to existing devices, not on human reader interpretation of images.
6. Standalone (Algorithm Only) Performance Study
Not applicable. This is a physical medical device (spinal implant), not a software algorithm orAI.
7. Type of Ground Truth Used
The ground truth used was the established performance metrics and requirements outlined in the ASTM F 1717 standard for static compression, static torsion, and dynamic axial compression bending tests of spinal implants.
8. Sample Size for Training Set
Not applicable. This submission describes the evaluation of a physical medical device (spinal implant), not an AI/ML algorithm that requires a training set.
9. How Ground Truth for Training Set Was Established
Not applicable, as there is no training set for this type of device evaluation.
{0}------------------------------------------------
510(k) Summary
. .
Showa Ika Kohgyo Co., LTD 510(k) Premarket Notification MYKRES Spinal System
JUL - 3 2006
ADMINISTRATIVE INFORMATION
:
| Manufacturer Name: | Showa Ika Kohgyo Co., LTD. |
|---|---|
| Street Address: | Shiba-koen 32 Mori-building 7F, 3-4-30 Shiba-koen, Minato-ku, Tokyo 105-0011 Japan |
| Telephone: | 81-3-5401-5340 |
| Fax: | 81-3-5401-5350 |
| Official Correspondent: | Kazuya Oribe |
| E-mail Address: | oribe@showaika.co.jp |
| Official Contact: | Kazuya Oribe |
| Representative/Consultant: | Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 |
| Telephone: | (858) 792-1235 |
| Fax: | (858) 792-1236 |
DEVICE NAME
| Classification Name: | Orthosis, Spondylolisthesis Spinal Fixation Orthosis, Spinal Pedicle Fixation Orthosis, Spinal Interlaminar Fixation |
|---|---|
| Trade/Proprietary Name: | MYKRES Spinal System |
| Common Name: | Spinal Fixation System |
ESTABLISHMENT REGISTRATION NUMBER
The establishment registration number for Showa Ika Kohgyo Co, LTD is 3004589749.
{1}------------------------------------------------
510(k) Summary
DEVICE CLASSIFICATION
Pedicle screw spinal fixation systems are classified as Class II devices (21 CFR 888.3070). The product code for Orthosis, Spondylolisthesis Spinal Fixation is MNH. The product code for Orthosis, Spinal Pedicle Fixation is MNI. Nonpedicle screw fixation systems are classified as Class II devices (21 CFR 888.3050). The product code for Spinal Interlaminar Fixation Orthosis is KWP. These device classifications are reviewed by the Orthopedic Devices Branch.
INTENDED USE
The MYKRES Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis, with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).
The MYKRES Spinal System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients undergoing fusion by autogenous bone grafting having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.
DEVICE DESCRIPTION
The MYKRES Spinal System is an internal fixation device for spinal surgery consisting of rods, screws, hooks and connectors available in various lengths, diameters, and configurations to enable close conformance to patient anatomy. A series of manual instruments (not a subject of this submission) intended to assist in the insertion and placement of the implants is provided in separate trays.
The rods, screws, hooks and connectors of the MYKRES Spinal System are made of Ti-6Al-4V titanium alloy conforming to ASTM F 136.
EQUIVALENCE TO MARKETED PRODUCT
Showa Ika Kohgyo has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the MYKRES Spinal System is substantially equivalent in intended use, materials, designs and operational principles to the following predicate devices:
J.B.S. Spine System with Pedicle Screws (K962757) from J.B.S. (USA), Inc. SYNERGY Spinal System - Closed VLS (K974749) from Interpore Cross International. MOSS Miami Spinal System (K955348, K983583, K011182) from DePuy AcroMed, Inc. OPTIMA Spinal System (K031585) from U&i Corp., America,
{2}------------------------------------------------
| Subject Device | Predicate Devices | ||||
|---|---|---|---|---|---|
| MYKRES Spinal System | J.B.S. SpineSystem withPedicle Screws | SYNERGY SpinalSystem - Closed VLS | MOSS MiamiSpinal System | OPTIMASpinal System | |
| J.B.S. (USA),Inc. | Interpore CrossInternational | DePuy AcroMed, Inc. | U&i Corp.,America | ||
| K962757 | K974749 | K955348, K983583,K011182 | K031585 | ||
| Intended Use | Posterior pedicle screw andnon-pedicle screw fixationindications. | Posteriorpedicle screwand non-pediclescrew fixationindications. | Posterior pediclescrew and non-pedicle screw fixationindications. | Posterior and anteriorpedicle screw andnon-pedicle screwfixation indications. | Posterior andanterior pediclescrew fixationindications. |
| Classification | MNH, MNI, KWP | MNH, KWP | MNH, KWP, KWQ | MNH, MNI, KWP,KWQ | MNH, KWQ,MNI |
| Design | Rods, polyaxial screws,monoaxial screws, hooks,connectors | Rods, polyaxialscrews,monoaxialscrews, hooks,connectors | Rods, polyaxialscrews, monoaxialscrews, hooks,connectors | Rods, polyaxialscrews, monoaxialscrews, hooks,connectors | Rods, polyaxialscrews,monoaxialscrews,connectors |
| Materials | Ti-6Al-4V | Ti-6Al-4V | Ti-6Al-4V, CP Ti,Stainless Steel | Ti-6Al-4V, StainlessSteel | Ti-6Al-4V |
Comparison of Technological Characteristics
Bench Testing Data
Showa Ika Kohgyo has submitted data from testing performed in compliance with ASTM F 1717. Static compression, static torsion, and dynamic axial compression bending tests using six samples each of a worst case MYKRES construct demonstrate that the MYKRES Spinal System is substantially equivalent to legally marketed spinal fixation systems and is therefore appropriate for use in spinal fixation as described in the indications above.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The eagle is depicted with three lines forming its body and wings, symbolizing health, human services, and well-being. The seal is in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL - 3 2006
Showa Ika Kogyo Company, Ltd. c/o Mr. Floyd Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130
Re: K051704
Trade/Device Name: Mykres Spinal System Regulation Number: 888.3070(b)(1) Regulation Name: Pedicle Screw Spinal System Regulatory Class: Class II Product Code: MNH, MNI, KWP Dated: June 15, 2006 Received: June 19. 2006
Dear Mr. Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling. and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Floyd Larson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality ovy; tem (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
barbare
tos
prekup
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K051704
Device Name: MYKRES Spinal System
Indications for Use:
The MYKRES Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis, with objective evidence of neurological imparment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudathrosis).
The MYKRES Spinal System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients undergoing fusion by autogenous bone grafting having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K051704
N/A